Published analysis of contraceptive effectiveness of Daysy and DaysyView app is fatally flawed

Author: Chelsea B. Polis

Published in: Reproductive Health | Issue 1/2018

Login to get access

Abstract

Background

In March 2018, Dr. Martin C. Koch and colleagues published an analysis purporting to measure the effectiveness of the Daysy device and DaysyView app for the prevention of unintended pregnancy. Unfortunately, the analysis was flawed in multiple ways which render the estimates unreliable. Unreliable estimates of contraceptive effectiveness can endanger public health.

Main body

This commentary details multiple concerns pertaining to the collection and analysis of data in Koch et al. 2018. A key concern pertains to the inappropriate exclusion of all women with fewer than 13 cycles of use from the Pearl Index calculations, which has no basis in standard effectiveness calculations. Multiple additional methodological concerns, as well as prior attempts to directly convey concerns to the manufacturer regarding marketing materials based on prior inaccurate analyses, are also discussed.

Conclusion

The Koch et al. 2018 publication produced unreliable estimates of contraceptive effectiveness for the Daysy device and DaysyView app, which are likely substantially higher than the actual contraceptive effectiveness of the device and app. Those estimates are being used in marketing materials which may inappropriately inflate consumer confidence and leave consumers more vulnerable than expected to the risk of unintended pregnancy. Prior attempts to directly convey concerns to the manufacturer of this device were unsuccessful in preventing publication of subsequent inaccurate analyses. To protect public health, concerns with this analysis should be documented in the published literature, the Koch et al. 2018 analysis should be retracted, and marketing materials on contraceptive effectiveness should be subjected to appropriate oversight.
Footnotes
1
For example, in addition to multiple concerns pertaining to the methods of pregnancy ascertainment in Freundl 1998 (collected via a mailed, retrospective questionnaire), the calculation which generated a 0.7 perfect use Pearl Index [10] used an incorrect denominator which included all cycles, rather than only cycles of perfect use, which would inflate perfect use effectiveness rates [3].
 
Literature
1.
go back to reference Koch MC, Lermann J, van de Roemer N, Renner SK, Burghaus S, Hackl J, et al. Improving usability and pregnancy rates of a fertility monitor by an additional mobile application: results of a retrospective efficacy study of Daysy and DaysyView app. Reprod Health. 2018;15:37.CrossRef Koch MC, Lermann J, van de Roemer N, Renner SK, Burghaus S, Hackl J, et al. Improving usability and pregnancy rates of a fertility monitor by an additional mobile application: results of a retrospective efficacy study of Daysy and DaysyView app. Reprod Health. 2018;15:37.CrossRef
3.
go back to reference Trussell J. Methodological pitfalls in the analysis of contraceptive failure. Stat Med. 1991;10:201–20.CrossRef Trussell J. Methodological pitfalls in the analysis of contraceptive failure. Stat Med. 1991;10:201–20.CrossRef
4.
go back to reference Trussell J, Portman D. The creeping pearl: why has the rate of contraceptive failure increased in clinical trials of combined hormonal contraceptive pills? Contraception. 2013;88:604–10.CrossRef Trussell J, Portman D. The creeping pearl: why has the rate of contraceptive failure increased in clinical trials of combined hormonal contraceptive pills? Contraception. 2013;88:604–10.CrossRef
5.
go back to reference Dominik R, Trussell J, Walsh T. Failure rates among perfect users and during perfect use: a distinction that matters. Contraception. 1999;60:315–20.CrossRef Dominik R, Trussell J, Walsh T. Failure rates among perfect users and during perfect use: a distinction that matters. Contraception. 1999;60:315–20.CrossRef
6.
go back to reference Jones RK, Kost K. Underreporting of induced and spontaneous abortion in the United States: an analysis of the 2002 National Survey of family growth. Stud Fam Plan. 2007;38:187–97.CrossRef Jones RK, Kost K. Underreporting of induced and spontaneous abortion in the United States: an analysis of the 2002 National Survey of family growth. Stud Fam Plan. 2007;38:187–97.CrossRef
7.
go back to reference Jagannathan R. Relying on surveys to understand abortion behavior: some cautionary evidence. Am J Public Health. 2001;91:1825–31.CrossRef Jagannathan R. Relying on surveys to understand abortion behavior: some cautionary evidence. Am J Public Health. 2001;91:1825–31.CrossRef
8.
go back to reference Klerman LV. The intendedness of pregnancy: a concept in transition. Matern Child Health J. 2000;4:155–62.CrossRef Klerman LV. The intendedness of pregnancy: a concept in transition. Matern Child Health J. 2000;4:155–62.CrossRef
9.
go back to reference Peragallo Urrutia R, Polis CB, Jensen ET, Greene ME, Kennedy E, Stanford JB. Effectiveness of fertility awareness-based methods for pregnancy prevention: a systematic review. Obstet Gynecol. In press Peragallo Urrutia R, Polis CB, Jensen ET, Greene ME, Kennedy E, Stanford JB. Effectiveness of fertility awareness-based methods for pregnancy prevention: a systematic review. Obstet Gynecol. In press
10.
go back to reference Freundl G, Frank-Herrmann P, Godehardt E, Klemm R, Bachhofer M. Retrospective clinical trial of contraceptive effectiveness of the electronic fertility indicator Ladycomp/Babycomp. Adv Contracept. 1998;14:97–108.CrossRef Freundl G, Frank-Herrmann P, Godehardt E, Klemm R, Bachhofer M. Retrospective clinical trial of contraceptive effectiveness of the electronic fertility indicator Ladycomp/Babycomp. Adv Contracept. 1998;14:97–108.CrossRef
11.
go back to reference Demianczyk A, Michaluk K. Evaluation of the effectiveness of selected natural fertility symptoms used for contraception: estimation of the pearl index of lady-comp, pearly and Daysy cycle computers based on 10 years of observation in the polish market. Ginekol Pol. 2016;87:793–7.CrossRef Demianczyk A, Michaluk K. Evaluation of the effectiveness of selected natural fertility symptoms used for contraception: estimation of the pearl index of lady-comp, pearly and Daysy cycle computers based on 10 years of observation in the polish market. Ginekol Pol. 2016;87:793–7.CrossRef
12.
go back to reference Polis CB. Email from Chelsea B. Polis to Niels van de Roemer, director. Medical Affairs. 2017; Polis CB. Email from Chelsea B. Polis to Niels van de Roemer, director. Medical Affairs. 2017;
13.
go back to reference Marcus A, Oransky I. Science publishing: The paper is not sacred. Nature. 2011;480:449–50.CrossRef Marcus A, Oransky I. Science publishing: The paper is not sacred. Nature. 2011;480:449–50.CrossRef
Metadata
Title
Published analysis of contraceptive effectiveness of Daysy and DaysyView app is fatally flawed
Author
Chelsea B. Polis
Publication date
01-12-2018
Publisher
BioMed Central
Published in
Reproductive Health / Issue 1/2018
Electronic ISSN: 1742-4755
DOI
https://doi.org/10.1186/s12978-018-0560-1

Other articles of this Issue 1/2018

Reproductive Health 1/2018 Go to the issue