Abstract

Background

Quantification of the longitudinal- and transverse relaxation time in the myocardium has shown to provide important information in cardiac diagnostics. Methods for cardiac relaxation time mapping generally demand a long breath hold to measure either T1 or T2 in a single 2D slice. In this paper we present and evaluate a novel method for 3D interleaved T1 and T2 mapping of the whole left ventricular myocardium within a single breath hold of 15 heartbeats.

Methods

The 3D-QALAS (3D-quantification using an interleaved Look-Locker acquisition sequence with T2 preparation pulse) is based on a 3D spoiled Turbo Field Echo sequence using inversion recovery with interleaved T2 preparation. Quantification of both T1 and T2 in a volume of 13 slices with a resolution of 2.0x2.0x6.0 mm is obtained from five measurements by using simulations of the longitudinal magnetizations Mz. This acquisition scheme is repeated three times to sample k-space. The method was evaluated both in-vitro (validated against Inversion Recovery and Multi Echo) and in-vivo (validated against MOLLI and Dual Echo).

Results

In-vitro, a strong relation was found between 3D-QALAS and Inversion Recovery (R = 0.998; N = 10; p < 0.01) and between 3D-QALAS and Multi Echo (R = 0.996; N = 10; p < 0.01). The 3D-QALAS method showed no dependence on e.g. heart rate in the interval of 40–120 bpm. In healthy myocardium, the mean T1 value was 1083 ± 43 ms (mean ± SD) for 3D-QALAS and 1089 ± 54 ms for MOLLI, while the mean T2 value was 50.4 ± 3.6 ms 3D-QALAS and 50.3 ± 3.5 ms for Dual Echo. No significant difference in in-vivo relaxation times was found between 3D-QALAS and MOLLI (N = 10; p = 0.65) respectively 3D-QALAS and Dual Echo (N = 10; p = 0.925) for the ten healthy volunteers.

Conclusions

The 3D-QALAS method has demonstrated good accuracy and intra-scan variability both in-vitro and in-vivo. It allows rapid acquisition and provides quantitative information of both T1 and T2 relaxation times in the same scan with full coverage of the left ventricle, enabling clinical application in a broader spectrum of cardiac disorders.
Appendix
Available only for authorised users
Literature
1.
go back to reference Chang KJ, Jara H. Applications of quantitative T1, T2, and proton density to diagnosis.Appl Radiol. 2005; 34–42., Chang KJ, Jara H. Applications of quantitative T1, T2, and proton density to diagnosis.Appl Radiol. 2005; 34–42.,
2.
go back to reference Riederer S, Lee J, Farzaneh F, Wang H, Wright R: Magnetic resonance image synthesis. clinical implementation. Acta Radiol Suppl. 1986, 369: 466-PubMed Riederer S, Lee J, Farzaneh F, Wang H, Wright R: Magnetic resonance image synthesis. clinical implementation. Acta Radiol Suppl. 1986, 369: 466-PubMed
3.
go back to reference Messroghli DR, Walters K, Plein S, Sparrow P, Friedrich MG, Ridgway JP, Sivananthan MU: Myocardial T1 mapping: application to patients with acute and chronic myocardial infarction. Magn Reson Med. 2007, 58: 34-40. 10.1002/mrm.21272.CrossRefPubMed Messroghli DR, Walters K, Plein S, Sparrow P, Friedrich MG, Ridgway JP, Sivananthan MU: Myocardial T1 mapping: application to patients with acute and chronic myocardial infarction. Magn Reson Med. 2007, 58: 34-40. 10.1002/mrm.21272.CrossRefPubMed
4.
go back to reference Iles L, Pfluger H, Phrommintikul A, Cherayath J, Aksit P, Gupta SN, Kaye DM, Taylor AJ: Evaluation of diffuse myocardial fibrosis in heart failure with cardiac magnetic resonance contrast-enhanced T1 mapping. J Am Coll Cardiol. 2008, 52: 1574-1580. 10.1016/j.jacc.2008.06.049.CrossRefPubMed Iles L, Pfluger H, Phrommintikul A, Cherayath J, Aksit P, Gupta SN, Kaye DM, Taylor AJ: Evaluation of diffuse myocardial fibrosis in heart failure with cardiac magnetic resonance contrast-enhanced T1 mapping. J Am Coll Cardiol. 2008, 52: 1574-1580. 10.1016/j.jacc.2008.06.049.CrossRefPubMed
5.
go back to reference Sparrow P, Messroghli DR, Reid S, Ridgway JP, Bainbridge G, Sivananthan MU: Myocardial T1 mapping for detection of left ventricular myocardial fibrosis in chronic aortic regurgitation: pilot study. Am J Roentgenol. 2006, 187 (6): W630-W635. 10.2214/AJR.05.1264.CrossRef Sparrow P, Messroghli DR, Reid S, Ridgway JP, Bainbridge G, Sivananthan MU: Myocardial T1 mapping for detection of left ventricular myocardial fibrosis in chronic aortic regurgitation: pilot study. Am J Roentgenol. 2006, 187 (6): W630-W635. 10.2214/AJR.05.1264.CrossRef
6.
go back to reference Flett AS, Hayward MP, Ashworth MT, Hansen MS, Taylor AM, Elliott PM, McGregor C, Moon JC: Equilibrium contrast cardiovascular magnetic resonance for the measurement of diffuse myocardial fibrosis preliminary validation in humans. Circulation. 2010, 122: 138-144. 10.1161/CIRCULATIONAHA.109.930636.CrossRefPubMed Flett AS, Hayward MP, Ashworth MT, Hansen MS, Taylor AM, Elliott PM, McGregor C, Moon JC: Equilibrium contrast cardiovascular magnetic resonance for the measurement of diffuse myocardial fibrosis preliminary validation in humans. Circulation. 2010, 122: 138-144. 10.1161/CIRCULATIONAHA.109.930636.CrossRefPubMed
7.
go back to reference Verhaert D, Thavendiranathan P, Giri S, Mihai G, Rajagopalan S, Simonetti OP, Raman SV: Direct T2 quantification of myocardial edema in acute ischemic injury. JACC Cardiovasc Imaging. 2011, 4: 269-278. 10.1016/j.jcmg.2010.09.023.PubMedCentralCrossRefPubMed Verhaert D, Thavendiranathan P, Giri S, Mihai G, Rajagopalan S, Simonetti OP, Raman SV: Direct T2 quantification of myocardial edema in acute ischemic injury. JACC Cardiovasc Imaging. 2011, 4: 269-278. 10.1016/j.jcmg.2010.09.023.PubMedCentralCrossRefPubMed
8.
go back to reference Giri S, Chung Y-C, Merchant A, Mihai G, Rajagopalan S, Raman S, Simonetti O: T2 quantification for improved detection of myocardial edema. J Cardiovasc Magn Reson. 2009, 11: 56-10.1186/1532-429X-11-56.PubMedCentralCrossRefPubMed Giri S, Chung Y-C, Merchant A, Mihai G, Rajagopalan S, Raman S, Simonetti O: T2 quantification for improved detection of myocardial edema. J Cardiovasc Magn Reson. 2009, 11: 56-10.1186/1532-429X-11-56.PubMedCentralCrossRefPubMed
9.
go back to reference Ferreira VM, Piechnik SK, Dall’Armellina E, Karamitsos TD, Francis JM, Choudhury RP, Friedrich MG, Robson MD, Neubauer S: Non-contrast T1-mapping detects acute myocardial edema with high diagnostic accuracy: a comparison to T2-weighted cardiovascular magnetic resonance. J Cardiovasc Magn Reson. 2012, 14: 42-10.1186/1532-429X-14-42.PubMedCentralCrossRefPubMed Ferreira VM, Piechnik SK, Dall’Armellina E, Karamitsos TD, Francis JM, Choudhury RP, Friedrich MG, Robson MD, Neubauer S: Non-contrast T1-mapping detects acute myocardial edema with high diagnostic accuracy: a comparison to T2-weighted cardiovascular magnetic resonance. J Cardiovasc Magn Reson. 2012, 14: 42-10.1186/1532-429X-14-42.PubMedCentralCrossRefPubMed
10.
go back to reference Ugander M, Bagi PS, Oki AJ, Chen B, Hsu L-Y, Aletras AH, Shah S, Greiser A, Kellman P, Arai AE: Myocardial edema as detected by pre-contrast T1 and T2 CMR delineates area at risk associated with acute myocardial infarction. JACC Cardiovasc Imaging. 2012, 5: 596-603. 10.1016/j.jcmg.2012.01.016.PubMedCentralCrossRefPubMed Ugander M, Bagi PS, Oki AJ, Chen B, Hsu L-Y, Aletras AH, Shah S, Greiser A, Kellman P, Arai AE: Myocardial edema as detected by pre-contrast T1 and T2 CMR delineates area at risk associated with acute myocardial infarction. JACC Cardiovasc Imaging. 2012, 5: 596-603. 10.1016/j.jcmg.2012.01.016.PubMedCentralCrossRefPubMed
11.
go back to reference Look DC, Locker DR: Time saving in measurement of NMR and EPR relaxation times. Rev Sci Instrum. 1970, 41: 250-251. 10.1063/1.1684482.CrossRef Look DC, Locker DR: Time saving in measurement of NMR and EPR relaxation times. Rev Sci Instrum. 1970, 41: 250-251. 10.1063/1.1684482.CrossRef
12.
go back to reference Messroghli DR, Radjenovic A, Kozerke S, Higgins DM, Sivananthan MU, Ridgway JP: Modified look-locker inversion recovery (MOLLI) for high-resolution T 1 mapping of the heart. Magn Reson Med. 2004, 52: 141-146. 10.1002/mrm.20110.CrossRefPubMed Messroghli DR, Radjenovic A, Kozerke S, Higgins DM, Sivananthan MU, Ridgway JP: Modified look-locker inversion recovery (MOLLI) for high-resolution T 1 mapping of the heart. Magn Reson Med. 2004, 52: 141-146. 10.1002/mrm.20110.CrossRefPubMed
13.
go back to reference Piechnik SK, Ferreira VM, Dall’Armellina E, Cochlin LE, Greiser A, Neubauer S, Robson MD: Shortened Modified Look-Locker Inversion recovery (ShMOLLI) for clinical myocardial T1-mapping at 1.5 and 3 T within a 9 heartbeat breathhold. J Cardiovasc Magn Reson. 2010, 12: 69-10.1186/1532-429X-12-69.PubMedCentralCrossRefPubMed Piechnik SK, Ferreira VM, Dall’Armellina E, Cochlin LE, Greiser A, Neubauer S, Robson MD: Shortened Modified Look-Locker Inversion recovery (ShMOLLI) for clinical myocardial T1-mapping at 1.5 and 3 T within a 9 heartbeat breathhold. J Cardiovasc Magn Reson. 2010, 12: 69-10.1186/1532-429X-12-69.PubMedCentralCrossRefPubMed
14.
go back to reference Coniglio A, Di Renzi P, Vilches Freixas G, Della Longa G, Santarelli A, Capparella R, Nardiello B, Loiudice C, Bianchi S, D’Arienzo M, Begnozzi L: Multiple 3D inversion recovery imaging for volume T1 mapping of the heart. Magn Reson Med. 2013, 69: 163-170. 10.1002/mrm.24248.CrossRefPubMed Coniglio A, Di Renzi P, Vilches Freixas G, Della Longa G, Santarelli A, Capparella R, Nardiello B, Loiudice C, Bianchi S, D’Arienzo M, Begnozzi L: Multiple 3D inversion recovery imaging for volume T1 mapping of the heart. Magn Reson Med. 2013, 69: 163-170. 10.1002/mrm.24248.CrossRefPubMed
15.
go back to reference Warntjes MJ, Kihlberg J, Engvall J: Rapid T1 quantification based on 3D phase sensitive inversion recovery. BMC Med Imaging. 2010, 10: 19-10.1186/1471-2342-10-19.PubMedCentralCrossRefPubMed Warntjes MJ, Kihlberg J, Engvall J: Rapid T1 quantification based on 3D phase sensitive inversion recovery. BMC Med Imaging. 2010, 10: 19-10.1186/1471-2342-10-19.PubMedCentralCrossRefPubMed
16.
go back to reference Kellman P, Aletras AH, Mancini C, McVeigh ER, Arai AE: T2-prepared SSFP improves diagnostic confidence in edema imaging in acute myocardial infarction compared to turbo spin echo. Magn Reson Med. 2007, 57: 891-897. 10.1002/mrm.21215.PubMedCentralCrossRefPubMed Kellman P, Aletras AH, Mancini C, McVeigh ER, Arai AE: T2-prepared SSFP improves diagnostic confidence in edema imaging in acute myocardial infarction compared to turbo spin echo. Magn Reson Med. 2007, 57: 891-897. 10.1002/mrm.21215.PubMedCentralCrossRefPubMed
17.
go back to reference Moon JC, Messroghli DR, Kellman P, Piechnik SK, Robson MD, Ugander M, Gatehouse PD, Arai AE, Friedrich MG, Neubauer S: Myocardial T1 mapping and extracellular volume quantification: a Society for Cardiovascular Magnetic Resonance (SCMR) and CMR working group of the european society of cardiology consensus statement. J Cardiovasc Magn Reson. 2013, 15: 92-10.1186/1532-429X-15-92.PubMedCentralCrossRefPubMed Moon JC, Messroghli DR, Kellman P, Piechnik SK, Robson MD, Ugander M, Gatehouse PD, Arai AE, Friedrich MG, Neubauer S: Myocardial T1 mapping and extracellular volume quantification: a Society for Cardiovascular Magnetic Resonance (SCMR) and CMR working group of the european society of cardiology consensus statement. J Cardiovasc Magn Reson. 2013, 15: 92-10.1186/1532-429X-15-92.PubMedCentralCrossRefPubMed
18.
go back to reference Huang T-Y, Liu Y-J, Stemmer A, Poncelet BP: T2 measurement of the human myocardium using a T2-prepared transient-state trueFISP sequence. Magn Reson Med. 2007, 57: 960-966. 10.1002/mrm.21208.CrossRefPubMed Huang T-Y, Liu Y-J, Stemmer A, Poncelet BP: T2 measurement of the human myocardium using a T2-prepared transient-state trueFISP sequence. Magn Reson Med. 2007, 57: 960-966. 10.1002/mrm.21208.CrossRefPubMed
19.
go back to reference Messroghli DR, Plein S, Jones TR, Sivananthan MU, Higgins DM, Ridgway JP, Walters K: Human myocardium: single-breath-hold MR T1 mapping with high spatial resolution - reproducibility study. Radiology. 2006, 238: 1004-1012. 10.1148/radiol.2382041903.CrossRefPubMed Messroghli DR, Plein S, Jones TR, Sivananthan MU, Higgins DM, Ridgway JP, Walters K: Human myocardium: single-breath-hold MR T1 mapping with high spatial resolution - reproducibility study. Radiology. 2006, 238: 1004-1012. 10.1148/radiol.2382041903.CrossRefPubMed
20.
go back to reference von Knobelsdorff-Brenkenhoff F, Prothmann M, Dieringer M, Wassmuth R, Greiser A, Schwenke C, Niendorf T, Schulz-Menger J: Myocardial T1 and T2 mapping at 3 T: reference values, influencing factors and implications. J Cardiovasc Magn Reson. 2013, 15: 53-10.1186/1532-429X-15-53.PubMedCentralCrossRefPubMed von Knobelsdorff-Brenkenhoff F, Prothmann M, Dieringer M, Wassmuth R, Greiser A, Schwenke C, Niendorf T, Schulz-Menger J: Myocardial T1 and T2 mapping at 3 T: reference values, influencing factors and implications. J Cardiovasc Magn Reson. 2013, 15: 53-10.1186/1532-429X-15-53.PubMedCentralCrossRefPubMed
22.
go back to reference Dabir D, Rogers T, Voigt T, Schaeffter T, Nagel E, Puntmann VO: Age-gender reference values of native myocardial T1 at 1.5 T and 3T: comparison of MOLLI and shMOLLI. J Cardiovasc Magn Reson. 2013, 15: P91-10.1186/1532-429X-15-S1-P91.PubMedCentralCrossRef Dabir D, Rogers T, Voigt T, Schaeffter T, Nagel E, Puntmann VO: Age-gender reference values of native myocardial T1 at 1.5 T and 3T: comparison of MOLLI and shMOLLI. J Cardiovasc Magn Reson. 2013, 15: P91-10.1186/1532-429X-15-S1-P91.PubMedCentralCrossRef
23.
go back to reference Lee JJ, Liu S, Nacif MS, Ugander M, Han J, Kawel N, Sibley CT, Kellman P, Arai AE, Bluemke DA: Myocardial T1 and extracellular volume fraction mapping at 3 Tesla. J Cardiovasc Magn Reson. 2011, 13: 75-10.1186/1532-429X-13-75.PubMedCentralCrossRefPubMed Lee JJ, Liu S, Nacif MS, Ugander M, Han J, Kawel N, Sibley CT, Kellman P, Arai AE, Bluemke DA: Myocardial T1 and extracellular volume fraction mapping at 3 Tesla. J Cardiovasc Magn Reson. 2011, 13: 75-10.1186/1532-429X-13-75.PubMedCentralCrossRefPubMed
Metadata
Title
Simultaneous three-dimensional myocardial T1 and T2 mapping in one breath hold with 3D-QALAS
Authors
Sofia Kvernby
Marcel Jan Bertus Warntjes
Henrik Haraldsson
Carl-Johan Carlhäll
Jan Engvall
Tino Ebbers
Publication date
01-12-2014
Publisher
BioMed Central
Published in
Journal of Cardiovascular Magnetic Resonance / Issue 1/2014
Electronic ISSN: 1532-429X
DOI
https://doi.org/10.1186/s12968-014-0102-0

Other articles of this Issue 1/2014

Journal of Cardiovascular Magnetic Resonance 1/2014 Go to the issue