Realization of fully automated quantification of left ventricular volumes and systolic function using transthoracic 3D echocardiography

Authors: Lina Sun, Haiyan Feng, Lujia Ni, Hui Wang, Dongmei Gao

Published in: Cardiovascular Ultrasound | Issue 1/2018

Login to get access

Abstract

Background

Study on automated three-dimensional (3D) quantification of left heart parameters by using Heartmodel software is still in the early stage and fully automatic analysis was not clearly achieved. The aim of our study was to evaluate the performance of this new technology in measuring left ventricular (LV) volume and ejection fraction (EF) in patients with a variety of heart diseases on the basis of rationally determining the default endocardial border values.

Methods

Subjects with a variety of heart diseases were included prospectively. High quality Heartmodel images were selected to determine the end-diastolic and end-systolic default values of endocardial border. The accuracy and reproducibility of automated three-dimensional echocardiography (3DE) for measuring LV end-diastolic volume (EDV), end-systolic volume (ESV) and EF were evaluated with the traditional manual 3DE as the relative standard.

Results

Ninety seven subjects were enrolled in the study. The default endocardial border values were determined as 66% and 40% for end-diastole (ED) and end-systole (ES), respectively. Most of the subjects (84/97) were automatically analyzed by Heartmodel software without manual adjustment, revealing a close correlation of automated 3DE with manual 3DE in measuring EDV, ESV and EF (r-values: EDV: 0.96, ESV: 0.97, EF: 0.96). The EDV and ESV values obtained by automated 3DE were higher than those measured by manual 3DE (biases: EDV: 16 ± 18 ml, ESV: 11 ± 12 ml). The intra- and inter-observer reproducibility of automated 3DE was better than that of manual 3DE. Automated 3DE with manual adjustment showed good consistency with manual 3DE in assessing the impairment degree of systolic function in patients with wall motion abnormalities (n = 58), (Kappa = 0.74, P = 0.00).

Conclusion

Fully automated 3DE quantification of LV volume and EF could be achieved in most patients. Since automated 3DE was accurate and more reproducible, it could replace the existing manual 3DE technology and be routinely used in clinical practice.
Literature
1.
go back to reference MossAJ ZW, Hall WJ, Klein H, Wilber DJ, Cannom DS, Daubert JP, Higgins SL, Brown MW, Andrews ML. Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction. N Engl J Med. 2002;346:877–83. [PubMed]CrossRef MossAJ ZW, Hall WJ, Klein H, Wilber DJ, Cannom DS, Daubert JP, Higgins SL, Brown MW, Andrews ML. Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction. N Engl J Med. 2002;346:877–83. [PubMed]CrossRef
2.
go back to reference Packer M, Fowler MB, Roecker EB, Coats AJ, Katus HA, Krum H, Mohacsi P, Rouleau JL, Tendera M, Staiger C, et al. Effect of carvedilol on the morbidity of patients with severe chronic heart failure: results of the carvedilol prospective randomized cumulative survival (COPERNICUS) study. Circulation. 2002;106:2194–9. [PubMed]CrossRefPubMed Packer M, Fowler MB, Roecker EB, Coats AJ, Katus HA, Krum H, Mohacsi P, Rouleau JL, Tendera M, Staiger C, et al. Effect of carvedilol on the morbidity of patients with severe chronic heart failure: results of the carvedilol prospective randomized cumulative survival (COPERNICUS) study. Circulation. 2002;106:2194–9. [PubMed]CrossRefPubMed
3.
go back to reference Dorosz JL, Lezotte DC, Weitzenkamp DA, Allen LA, Salcedo EE. Performance of 3-dimensional echocardiography in measuring left ventricular volumes and ejection fraction: a systematic review and meta-analysis. J Am Coll Cardiol. 2012;59:1799–808. [PubMed]CrossRefPubMedPubMedCentral Dorosz JL, Lezotte DC, Weitzenkamp DA, Allen LA, Salcedo EE. Performance of 3-dimensional echocardiography in measuring left ventricular volumes and ejection fraction: a systematic review and meta-analysis. J Am Coll Cardiol. 2012;59:1799–808. [PubMed]CrossRefPubMedPubMedCentral
4.
go back to reference Jenkins C, Chan J, Hanekom L, Marwick TH. Accuracy and feasibility of online 3-dimensional echocardiography for measurement of left ventricular parameters. J Am Soc Echocardiogr. 2006;19:1119–28. [PubMed]CrossRefPubMed Jenkins C, Chan J, Hanekom L, Marwick TH. Accuracy and feasibility of online 3-dimensional echocardiography for measurement of left ventricular parameters. J Am Soc Echocardiogr. 2006;19:1119–28. [PubMed]CrossRefPubMed
5.
go back to reference Spitzer E, Ren B, Soliman OI, Zijlstra F, Van Mieghem NM, Geleijnse ML. Accuracy of an automated transthoracic echocardiographic tool for 3D assessment of left heart chamber volumes. Echocardiography. 2017;34:199–209. [PubMed]CrossRefPubMed Spitzer E, Ren B, Soliman OI, Zijlstra F, Van Mieghem NM, Geleijnse ML. Accuracy of an automated transthoracic echocardiographic tool for 3D assessment of left heart chamber volumes. Echocardiography. 2017;34:199–209. [PubMed]CrossRefPubMed
6.
go back to reference Medvedofsky D, Mor-Avi V, Amzulescu M, Fernández-Golfín C, Hinojar R, Monaghan MJ, Otani K, Reiken J, Takeuchi M, Tsang W, et al. Three-dimensional echocardiographic quantification of the left-heart chambers using an automated adaptive analytics algorithm: multicentre validation study. Eur Heart J Cardiovasc Imaging 2017; doi: https://doi.org/10.1093/ehjci/jew328. [PubMed]. Medvedofsky D, Mor-Avi V, Amzulescu M, Fernández-Golfín C, Hinojar R, Monaghan MJ, Otani K, Reiken J, Takeuchi M, Tsang W, et al. Three-dimensional echocardiographic quantification of the left-heart chambers using an automated adaptive analytics algorithm: multicentre validation study. Eur Heart J Cardiovasc Imaging 2017; doi: https://​doi.​org/​10.​1093/​ehjci/​jew328. [PubMed].
7.
go back to reference Levy F, Schouver ED, Lacuzio L, Civaia F, Rusek S, Dommerc C, Marechaux S, Dor V, Tribouilloy C, Dreyfus G. Performance of new automated transthoracic three-dimensional echocardiographic software for left ventricular volumes and function assessment in routine clinical practice: Comparison with 3 Tesla cardiac magnetic resonance. Arch Cardiovasc Dis 2017; doi:https://doi.org/10.1016/j.acvd.2016.12.015. [PubMed]. Levy F, Schouver ED, Lacuzio L, Civaia F, Rusek S, Dommerc C, Marechaux S, Dor V, Tribouilloy C, Dreyfus G. Performance of new automated transthoracic three-dimensional echocardiographic software for left ventricular volumes and function assessment in routine clinical practice: Comparison with 3 Tesla cardiac magnetic resonance. Arch Cardiovasc Dis 2017; doi:https://​doi.​org/​10.​1016/​j.​acvd.​2016.​12.​015. [PubMed].
8.
go back to reference Tsang W, Salgo IS, Medvedofsky D, Takeuchi M, Prater D, Weinert L, Yamat M, Mor-Avi V, Patel AR, Lang RM. Transthoracic 3D Echocardiographic left heart chamber quantification using an automated adaptive analytics algorithm. JACC Cardiovasc Imaging. 2016;9:769–82. [PubMed]CrossRefPubMed Tsang W, Salgo IS, Medvedofsky D, Takeuchi M, Prater D, Weinert L, Yamat M, Mor-Avi V, Patel AR, Lang RM. Transthoracic 3D Echocardiographic left heart chamber quantification using an automated adaptive analytics algorithm. JACC Cardiovasc Imaging. 2016;9:769–82. [PubMed]CrossRefPubMed
9.
go back to reference Otani K, Nakazono A, Salgo IS, Lang RM, Takeuchi M. Three-dimensional Echocardiographic assessment of left heart chamber size and function with fully automated quantification software in patients with Atrial fibrillation. J Am Soc Echocardiogr. 2016;29:955–65. [PubMed]CrossRefPubMed Otani K, Nakazono A, Salgo IS, Lang RM, Takeuchi M. Three-dimensional Echocardiographic assessment of left heart chamber size and function with fully automated quantification software in patients with Atrial fibrillation. J Am Soc Echocardiogr. 2016;29:955–65. [PubMed]CrossRefPubMed
10.
go back to reference Feng C, Chen L, Li J, Wang J, Dong F, Xu J. Three-dimensional echocardiographic measurements using automated quantification software for big data processing. J Xray Sci Technol. 2017;25:313–21. [PubMed]PubMed Feng C, Chen L, Li J, Wang J, Dong F, Xu J. Three-dimensional echocardiographic measurements using automated quantification software for big data processing. J Xray Sci Technol. 2017;25:313–21. [PubMed]PubMed
11.
go back to reference Yang LT, Nagata Y, Otani K, Kado Y, Otsuji Y, Takeuchi M. Feasibility of one-beat real-time full-volume three-dimensional echocardiography for assessing left ventricular volumes and deformation parameters. J Am Soc Echocardiogr. 2016;29:853–60. [PubMed]CrossRefPubMed Yang LT, Nagata Y, Otani K, Kado Y, Otsuji Y, Takeuchi M. Feasibility of one-beat real-time full-volume three-dimensional echocardiography for assessing left ventricular volumes and deformation parameters. J Am Soc Echocardiogr. 2016;29:853–60. [PubMed]CrossRefPubMed
12.
go back to reference Lang RM, Badano LP, Mor-Avi V, Afilalo J, Armstrong A, Ernande L, Flachskampf FA, Foster E, Goldstein SA, Kuznetsova T, et al. Recommendations for cardiac chamber quantification by echocardiography in adults: an update from the American Society of Echocardiography and the European Association of Cardiovascular Imaging. J Am Soc Echocardiogr. 2015;28:1–39. [PubMed]CrossRefPubMed Lang RM, Badano LP, Mor-Avi V, Afilalo J, Armstrong A, Ernande L, Flachskampf FA, Foster E, Goldstein SA, Kuznetsova T, et al. Recommendations for cardiac chamber quantification by echocardiography in adults: an update from the American Society of Echocardiography and the European Association of Cardiovascular Imaging. J Am Soc Echocardiogr. 2015;28:1–39. [PubMed]CrossRefPubMed
13.
go back to reference Thavendiranathan P, Liu S, Verhaert D, Calleja A, Nitinunu A, Van Houten T, De Michelis N, Simonetti O, Rajagopalan S, Ryan T, et al. Feasibility, accuracy, and reproducibility of real-time full-volume 3D transthoracic echocardiography to measure LV volumes and systolic function: a fully automated endocardial contouring algorithm in sinus rhythm and atrial fibrillation. JACC Cardiovasc Imaging. 2012;5:239–51. [PubMed]CrossRefPubMed Thavendiranathan P, Liu S, Verhaert D, Calleja A, Nitinunu A, Van Houten T, De Michelis N, Simonetti O, Rajagopalan S, Ryan T, et al. Feasibility, accuracy, and reproducibility of real-time full-volume 3D transthoracic echocardiography to measure LV volumes and systolic function: a fully automated endocardial contouring algorithm in sinus rhythm and atrial fibrillation. JACC Cardiovasc Imaging. 2012;5:239–51. [PubMed]CrossRefPubMed
14.
go back to reference Mor-Avi V, Jenkins C, Kühl HP, Nesser HJ, Marwick T, Franke A, Ebner C, Freed BH, Steringer-Mascherbauer R, Pollard H, et al. Real-time 3-dimensional echocardiographic quantification of left ventricular volumes: multicenter study for validation with magnetic resonance imaging and investigation of sources of error. JACC Cardiovasc Imaging. 2008;1:413–23. [PubMed]CrossRefPubMed Mor-Avi V, Jenkins C, Kühl HP, Nesser HJ, Marwick T, Franke A, Ebner C, Freed BH, Steringer-Mascherbauer R, Pollard H, et al. Real-time 3-dimensional echocardiographic quantification of left ventricular volumes: multicenter study for validation with magnetic resonance imaging and investigation of sources of error. JACC Cardiovasc Imaging. 2008;1:413–23. [PubMed]CrossRefPubMed
15.
go back to reference Medvedofsky D, Mor-Avi V, Byku I, Singh A, Weinert L, Yamat M, Kruse E, Ciszek B, Nelson A, Otani K, et al. Three-dimensional Echocardiographic automated quantification of left heart chamber volumes using an adaptive analytics algorithm: feasibility and impact of image quality in nonselected patients. J Am Soc Echocardiogr. 2017;30:879–85. [PubMed]CrossRefPubMedPubMedCentral Medvedofsky D, Mor-Avi V, Byku I, Singh A, Weinert L, Yamat M, Kruse E, Ciszek B, Nelson A, Otani K, et al. Three-dimensional Echocardiographic automated quantification of left heart chamber volumes using an adaptive analytics algorithm: feasibility and impact of image quality in nonselected patients. J Am Soc Echocardiogr. 2017;30:879–85. [PubMed]CrossRefPubMedPubMedCentral
16.
go back to reference Thavendiranathan P, Grant AD, Negishi T, Plana JC, Popović ZB, Marwick TH. Reproducibility of echocardiographic techniques for sequential assessment of left ventricular ejection fraction and volumes: application to patients undergoing cancer chemotherapy. J Am CollCardiol. 2013;61:77–84. [PubMed]CrossRef Thavendiranathan P, Grant AD, Negishi T, Plana JC, Popović ZB, Marwick TH. Reproducibility of echocardiographic techniques for sequential assessment of left ventricular ejection fraction and volumes: application to patients undergoing cancer chemotherapy. J Am CollCardiol. 2013;61:77–84. [PubMed]CrossRef
Metadata
Title
Realization of fully automated quantification of left ventricular volumes and systolic function using transthoracic 3D echocardiography
Authors
Lina Sun
Haiyan Feng
Lujia Ni
Hui Wang
Dongmei Gao
Publication date
01-12-2018
Publisher
BioMed Central
Published in
Cardiovascular Ultrasound / Issue 1/2018
Electronic ISSN: 1476-7120
DOI
https://doi.org/10.1186/s12947-017-0121-8

Other articles of this Issue 1/2018

Cardiovascular Ultrasound 1/2018 Go to the issue