Open Access 01-12-2020 | Digital Volume Tomography | Research article

Dental implant location via surface scanner: a pilot study

Authors: Miao Zhou, Hui Zhou, Shu-yi Li, Yuan-ming Geng

Published in: BMC Oral Health | Issue 1/2020

Login to get access

Abstract

Purpose

Implant location is performed after placement to verify that the safety of neighboring anatomic structure and the realizability of prosthetic plan. Routine postoperative location is based on radiological scanning and raises the concerns on radiation exposure and inconveniency in practice. In the present study a location method based on surface scanning was introduced and the accuracy of this method was assessed in vitro.

Material and methods

A total of 40 implants were placed in 10 resin mandible models. The models were scanned with intraoral scanner (IS group) and extraoral scanner (ES group). The implant position was located with fusing the images of surface scanning and cone beam computerized tomography (CBCT) after implant placement. Deviations were measured between positions located by surface scanner and postoperative CBCT with the parameters: central deviation at apex (cda), central deviation at hex (cdh), horizontal deviation at apex (hda), horizontal deviation at hex (hdh), vertical deviation at apex (vda), vertical deviation at hex (vdh) and angular deviation (ad).

Results

In IS group, the mean value of cda, cdh, hda, hdh, vda, vdh and ad was 0.27 mm, 0.23 mm, 0.12 mm, 0.10 mm, 0.21 mm, 0.19 mm and 0.72°, respectively. In ES group, the mean value of cda, cdh, hda, hdh, vda, vdh and ad was 0.28 mm, 0.25 mm 0.14 mm, 0.11 mm, 0.22 mm, 0.20 mm and 0.68°, respectively. The implant deviations in IS and ES groups were of no significant difference for any of the measurements.

Conclusions

Dental implant can be located via surface scanner with acceptable accuracy for postoperative verification. Further clinical investigation is needed to assess the feasibility of the method.
Literature
1.
go back to reference Raico Gallardo YN, da Silva-Olivio IRT, Mukai E, Morimoto S, Sesma N, Cordaro L. Accuracy comparison of guided surgery for dental implants according to the tissue of support: a systematic review and meta-analysis. Clin Oral Implant Res. 2017;28(5):602–12.CrossRef Raico Gallardo YN, da Silva-Olivio IRT, Mukai E, Morimoto S, Sesma N, Cordaro L. Accuracy comparison of guided surgery for dental implants according to the tissue of support: a systematic review and meta-analysis. Clin Oral Implant Res. 2017;28(5):602–12.CrossRef
2.
go back to reference D’Haese J, Ackhurst J, Wismeijer D, De Bruyn H, Tahmaseb A. Current state of the art of computer-guided implant surgery. Periodontol 2000. 2017;73(1):121–33.CrossRef D’Haese J, Ackhurst J, Wismeijer D, De Bruyn H, Tahmaseb A. Current state of the art of computer-guided implant surgery. Periodontol 2000. 2017;73(1):121–33.CrossRef
3.
go back to reference Jacobs R, Salmon B, Codari M, Hassan B, Bornstein MM. Cone beam computed tomography in implant dentistry: recommendations for clinical use. BMC Oral Health. 2018;18(1):88.CrossRef Jacobs R, Salmon B, Codari M, Hassan B, Bornstein MM. Cone beam computed tomography in implant dentistry: recommendations for clinical use. BMC Oral Health. 2018;18(1):88.CrossRef
4.
go back to reference Stratis A, Zhang G, Jacobs R, Bogaerts R, Bosmans H. The growing concern of radiation dose in paediatric dental and maxillofacial CBCT: an easy guide for daily practice. Eur Radiol. 2019;29:7009–18.CrossRef Stratis A, Zhang G, Jacobs R, Bogaerts R, Bosmans H. The growing concern of radiation dose in paediatric dental and maxillofacial CBCT: an easy guide for daily practice. Eur Radiol. 2019;29:7009–18.CrossRef
5.
go back to reference Naitoh M, Ariji E, Okumura S, Ohsaki C, Kurita K, Ishigami T. Can implants be correctly angulated based on surgical templates used for osseointegrated dental impla nts? Clin Oral Implant Res. 2000;11(5):409–14.CrossRef Naitoh M, Ariji E, Okumura S, Ohsaki C, Kurita K, Ishigami T. Can implants be correctly angulated based on surgical templates used for osseointegrated dental impla nts? Clin Oral Implant Res. 2000;11(5):409–14.CrossRef
6.
go back to reference Nickenig HJ, Eitner S. An alternative method to match planned and achieved positions of implants, after virtual planning using cone-beam CT data and surgical guide templates–a method reducing patient radiation exposure (par t I). J Craniomaxillofac Surg. 2010;38(6):436–40.CrossRef Nickenig HJ, Eitner S. An alternative method to match planned and achieved positions of implants, after virtual planning using cone-beam CT data and surgical guide templates–a method reducing patient radiation exposure (par t I). J Craniomaxillofac Surg. 2010;38(6):436–40.CrossRef
7.
go back to reference Rutkūnas V, Gečiauskaitė A, Jegelevičius D, Vaitiekūnas M. Accuracy of digital implant impressions with intraoral scanners. A systematic review. Eur J Oral Implantol. 2017;10(Suppl 1):101–20. Rutkūnas V, Gečiauskaitė A, Jegelevičius D, Vaitiekūnas M. Accuracy of digital implant impressions with intraoral scanners. A systematic review. Eur J Oral Implantol. 2017;10(Suppl 1):101–20.
8.
go back to reference Van Assche N, Vercruyssen M, Coucke W, Teughels W, Jacobs R, Quirynen M. Accuracy of computer-aided implant placement. Clin Oral Implant Res. 2012;23(Suppl 6):112–23.CrossRef Van Assche N, Vercruyssen M, Coucke W, Teughels W, Jacobs R, Quirynen M. Accuracy of computer-aided implant placement. Clin Oral Implant Res. 2012;23(Suppl 6):112–23.CrossRef
9.
go back to reference Mangano FG, Hauschild U, Veronesi G, Imburgia M, Mangano C, Admakin O. Trueness and precision of 5 intraoral scanners in the impressions of single and multiple implants: a comparative in vitro study. BMC Oral Health. 2019;19(1):101.CrossRef Mangano FG, Hauschild U, Veronesi G, Imburgia M, Mangano C, Admakin O. Trueness and precision of 5 intraoral scanners in the impressions of single and multiple implants: a comparative in vitro study. BMC Oral Health. 2019;19(1):101.CrossRef
10.
go back to reference Runkel C, Güth JF, Erdelt K, Keul C. Digital impressions in dentistry-accuracy of impression digitalisation by desktop scanners. Clin Oral Investig. 2020;24(3):1249–57.CrossRef Runkel C, Güth JF, Erdelt K, Keul C. Digital impressions in dentistry-accuracy of impression digitalisation by desktop scanners. Clin Oral Investig. 2020;24(3):1249–57.CrossRef
11.
go back to reference Flügge T, van der Meer WJ, Gonzalez BG, Vach K, Wismeijer D, Wang P. The accuracy of different dental impression techniques for implant-supported dental prostheses: a systematic review and meta-analysis. Clin Oral Implant Res. 2018;29(Suppl 16):374–92.CrossRef Flügge T, van der Meer WJ, Gonzalez BG, Vach K, Wismeijer D, Wang P. The accuracy of different dental impression techniques for implant-supported dental prostheses: a systematic review and meta-analysis. Clin Oral Implant Res. 2018;29(Suppl 16):374–92.CrossRef
12.
go back to reference Bover-Ramos F, Viña-Almunia J, Cervera-Ballester J, Peñarrocha-Diago M, García-Mira B. Accuracy of implant placement with computer-guided surgery: a systematic review and meta-analysis comparing cadaver, clinical, and in vitro studies. Int J Oral Maxillofac Implants. 2018;33(1):101–15.CrossRef Bover-Ramos F, Viña-Almunia J, Cervera-Ballester J, Peñarrocha-Diago M, García-Mira B. Accuracy of implant placement with computer-guided surgery: a systematic review and meta-analysis comparing cadaver, clinical, and in vitro studies. Int J Oral Maxillofac Implants. 2018;33(1):101–15.CrossRef
13.
go back to reference Sicilia A, Botticelli D, Working G. Computer-guided implant therapy and soft- and hard-tissue aspects. The third EAO consensus conference 2012. Clin Oral Implants Research. 2012;23(Suppl 6):157–61.CrossRef Sicilia A, Botticelli D, Working G. Computer-guided implant therapy and soft- and hard-tissue aspects. The third EAO consensus conference 2012. Clin Oral Implants Research. 2012;23(Suppl 6):157–61.CrossRef
14.
go back to reference Bohner LOL, De Luca CG, Marció BS, Laganá DC, Sesma N, Tortamano Neto P. Computer-aided analysis of digital dental impressions obtained from intraoral and extraoral scanners. J Prosthet Dent. 2017;118(5):617–23. Bohner LOL, De Luca CG, Marció BS, Laganá DC, Sesma N, Tortamano Neto P. Computer-aided analysis of digital dental impressions obtained from intraoral and extraoral scanners. J Prosthet Dent. 2017;118(5):617–23.
15.
go back to reference Keul C, Güth JF. Accuracy of full-arch digital impressions: an in vitro and in vivo comparison. Clin Oral Investig. 2020;24(2):735–45.CrossRef Keul C, Güth JF. Accuracy of full-arch digital impressions: an in vitro and in vivo comparison. Clin Oral Investig. 2020;24(2):735–45.CrossRef
16.
go back to reference Giménez B, Özcan M, Martínez-Rus F, Pradíes G. Accuracy of a digital impression system based on active wavefront sampling technology for implants considering operator experience, implant angulation, and depth. Clin Implant Dent Relat Res. 2015;17(Suppl 1):e54-64.CrossRef Giménez B, Özcan M, Martínez-Rus F, Pradíes G. Accuracy of a digital impression system based on active wavefront sampling technology for implants considering operator experience, implant angulation, and depth. Clin Implant Dent Relat Res. 2015;17(Suppl 1):e54-64.CrossRef
17.
go back to reference Lim JH, Park JM, Kim M, Heo SJ, Myung JY. Comparison of digital intraoral scanner reproducibility and image trueness considering repetitive exp erience. J Prosthet Dent. 2018;119(2):225–32.CrossRef Lim JH, Park JM, Kim M, Heo SJ, Myung JY. Comparison of digital intraoral scanner reproducibility and image trueness considering repetitive exp erience. J Prosthet Dent. 2018;119(2):225–32.CrossRef
18.
go back to reference Rangel FA, Maal TJJ, de Koning MJJ, Bronkhorst EM, Bergé SJ, Kuijpers-Jagtman AM. Integration of digital dental casts in cone beam computed tomography scans-a clinical validation study. Clin Oral Investig. 2018;22(3):1215–22.CrossRef Rangel FA, Maal TJJ, de Koning MJJ, Bronkhorst EM, Bergé SJ, Kuijpers-Jagtman AM. Integration of digital dental casts in cone beam computed tomography scans-a clinical validation study. Clin Oral Investig. 2018;22(3):1215–22.CrossRef
19.
go back to reference Woo S, Lee S, Chae J, Rim J, Lee J, Seo J, Lee C. Automatic matching of computed tomography and stereolithography data. Comput Methods Programs Biomed. 2019;175:215–22.CrossRef Woo S, Lee S, Chae J, Rim J, Lee J, Seo J, Lee C. Automatic matching of computed tomography and stereolithography data. Comput Methods Programs Biomed. 2019;175:215–22.CrossRef
Metadata
Title
Dental implant location via surface scanner: a pilot study
Authors
Miao Zhou
Hui Zhou
Shu-yi Li
Yuan-ming Geng
Publication date
01-12-2020
Publisher
BioMed Central
Published in
BMC Oral Health / Issue 1/2020
Electronic ISSN: 1472-6831
DOI
https://doi.org/10.1186/s12903-020-01297-y

Other articles of this Issue 1/2020

BMC Oral Health 1/2020 Go to the issue