A quantitative non-invasive assessment of femoroacetabular impingement with CT-based dynamic simulation - cadaveric validation study

Authors: Maarten A Röling, Monique I Visser, Edwin HG Oei, Peter Pilot, Gert-Jan Kleinrensink, Rolf M Bloem

Published in: BMC Musculoskeletal Disorders | Issue 1/2015

Login to get access

Abstract

Background

Femoroacetabular impingement (FAI) is caused by an anatomic deviation of the acetabular rim or proximal femur, which causes chronic groin pain. Radiological identification of FAI can be challenging. Advances in imaging techniques with the use of computed tomography (CT) scan enable 3D simulation of FAI. We made an experimental cadaveric validation study to validate the 3D simulation imaging software.

Methods

The range of motion (ROM) of five cadaveric hips was measured using an electromagnetic tracking system (EMTS). Specific marked spots in the femur and pelvis were created as reproducible EMTS registration points. Reproducible motions were measured. Hips were subsequently imaged using high-resolution CT after introduction of artificial cam deformities. A proprietary software tool was used, Articulis (Clinical Graphics) to simulate the ROM during the presence and absence of the induced cam deformities.

Results

According to the EMTS, 13 of the 30 measured ROM end-points were restricted by > 5° due to the induced cam deformities. Using Articulis, with the same 5° threshold, we correctly detected 12 of these 13 end point limitations and detected no false positives. The median error of the measured limitations was 1.9° (interquartile range 1.1° - 4.4°). The maximum absolute error was 5.4°.

Conclusions

The use of this dynamic simulation software to determine the presence of motion limiting deformities of the femoroacetabular is validated. The simulation software is able to non-invasively detect a reduction in achievable ROM, caused by a cam type deformity.
Literature
1.
go back to reference Ganz R, Leunig M, Leunig-Ganz K, Harris WH. “The Etiology of Osteoarthritis of the Hip. An integrated mechanical concept” Symposium: papers presented at the annual closed meeting of the international hip society. Clin Orthop Relat Res. 2008;466:264–72.CrossRefPubMedPubMedCentral Ganz R, Leunig M, Leunig-Ganz K, Harris WH. “The Etiology of Osteoarthritis of the Hip. An integrated mechanical concept” Symposium: papers presented at the annual closed meeting of the international hip society. Clin Orthop Relat Res. 2008;466:264–72.CrossRefPubMedPubMedCentral
2.
go back to reference Byrd J, Jones K. Hip arthroscopy in athletes: 10-year follow-up. Am J Sports Med. 2009;11(37):2140–3.CrossRef Byrd J, Jones K. Hip arthroscopy in athletes: 10-year follow-up. Am J Sports Med. 2009;11(37):2140–3.CrossRef
3.
go back to reference Hack K, Primio G, Rakhra K. Prevalence of cam-type femoroacetabular impingement morphology in asymptomatic volunteers. J Bone Joint Surg. 2010;14(92):2436–44.CrossRef Hack K, Primio G, Rakhra K. Prevalence of cam-type femoroacetabular impingement morphology in asymptomatic volunteers. J Bone Joint Surg. 2010;14(92):2436–44.CrossRef
4.
go back to reference Aprato A.; Jayasekera N.; Viaal RN. Revision hip arthroscopic surgery: outcome at three years. Knee Surg. Sports Traumatol. Arthrosc. 2013; Jan 18, doi:10.1007/s001670-013-2373-7 Aprato A.; Jayasekera N.; Viaal RN. Revision hip arthroscopic surgery: outcome at three years. Knee Surg. Sports Traumatol. Arthrosc. 2013; Jan 18, doi:10.1007/s001670-013-2373-7
5.
go back to reference Ilizaliturri V. Complications of arthroscopic femoroacetabular impingement treatment: a review. Clin Orthop Relat Res. 2009;467(3):760–8.CrossRefPubMed Ilizaliturri V. Complications of arthroscopic femoroacetabular impingement treatment: a review. Clin Orthop Relat Res. 2009;467(3):760–8.CrossRefPubMed
6.
go back to reference Matsuda DK, Carlisle JC, Arthurs SC, Wierks CH, Philippon MJ. Comparative systematic review of the open dislocation, mini-open, and arthroscopic surgeries for femoroacetabular impingement. Arthroscopy. 2011;27(2):252–69.CrossRefPubMed Matsuda DK, Carlisle JC, Arthurs SC, Wierks CH, Philippon MJ. Comparative systematic review of the open dislocation, mini-open, and arthroscopic surgeries for femoroacetabular impingement. Arthroscopy. 2011;27(2):252–69.CrossRefPubMed
7.
go back to reference Harris JD, Erickson BJ, Bush-Joseph CA, Nho SJ. Treatment of femoroacetabular impingement: a systematic review. Curr Rev Muscoloskelet Med. 2013;6:207–18.CrossRef Harris JD, Erickson BJ, Bush-Joseph CA, Nho SJ. Treatment of femoroacetabular impingement: a systematic review. Curr Rev Muscoloskelet Med. 2013;6:207–18.CrossRef
8.
go back to reference Alradwan H, Philippon MJ, Farrokhyar F, Phil M, Chu R, Whelan D, et al. Return to Preinjury Activity Levels After Surgical Management of Femoroacetabular Impingement in Athletes. Systematic review. Arthroscopy. 2012;28(10):1567–76.CrossRefPubMed Alradwan H, Philippon MJ, Farrokhyar F, Phil M, Chu R, Whelan D, et al. Return to Preinjury Activity Levels After Surgical Management of Femoroacetabular Impingement in Athletes. Systematic review. Arthroscopy. 2012;28(10):1567–76.CrossRefPubMed
10.
go back to reference Heyworth BE, Shindle MK, Voos JE, Rukzki JR, Kelly BT. Radiologic and intraoperative findings in revision hip arthroscopy. Arthroscopy. 2007;23:1295–302.CrossRefPubMed Heyworth BE, Shindle MK, Voos JE, Rukzki JR, Kelly BT. Radiologic and intraoperative findings in revision hip arthroscopy. Arthroscopy. 2007;23:1295–302.CrossRefPubMed
11.
go back to reference Philippon MJ, Schenker ML, Briggs KK, Kuppersmith DA, Maxwell RB, Stubbs AJ. Revision hip arthroscopy. Am J Sports Med. 2007;35:1918–21.CrossRefPubMed Philippon MJ, Schenker ML, Briggs KK, Kuppersmith DA, Maxwell RB, Stubbs AJ. Revision hip arthroscopy. Am J Sports Med. 2007;35:1918–21.CrossRefPubMed
12.
go back to reference Palmer WE. Femoroacetabular impingement: caution is warranted in making imging-based assumptions and diagnosis. Radiology. 2010;257:4–7.CrossRefPubMed Palmer WE. Femoroacetabular impingement: caution is warranted in making imging-based assumptions and diagnosis. Radiology. 2010;257:4–7.CrossRefPubMed
13.
14.
go back to reference Sutter R, Dietrich TJ, Zingg PO, Pfirrmann CWA. How useful is the alpha angle for discriminationg between symptomatic patients with cam-type femoroacetabular impingement and asymptomatic volunteers? Radiology. 2012;264:514–21.CrossRefPubMed Sutter R, Dietrich TJ, Zingg PO, Pfirrmann CWA. How useful is the alpha angle for discriminationg between symptomatic patients with cam-type femoroacetabular impingement and asymptomatic volunteers? Radiology. 2012;264:514–21.CrossRefPubMed
15.
go back to reference Lepage-Saucier M, Thiéry C, Larbi A, Lecouvet FE, Vand Berg BC, Omoumi P. Femoroacetabular impingement: normal values of the quantitative morphometric parameters in asymptomatic hips. Eur Radiol. 2014;7:1707–14.CrossRef Lepage-Saucier M, Thiéry C, Larbi A, Lecouvet FE, Vand Berg BC, Omoumi P. Femoroacetabular impingement: normal values of the quantitative morphometric parameters in asymptomatic hips. Eur Radiol. 2014;7:1707–14.CrossRef
16.
go back to reference Reurink G, Jansen SPL, Bisselink JM, Vincken PWJ, Weir A, Moen MH. Reliability and validity of diagnosing acetabular labral lesions with magnetic resonance arthrography. J Bone Joint Surg Am. 2012;94:1643–8.CrossRefPubMed Reurink G, Jansen SPL, Bisselink JM, Vincken PWJ, Weir A, Moen MH. Reliability and validity of diagnosing acetabular labral lesions with magnetic resonance arthrography. J Bone Joint Surg Am. 2012;94:1643–8.CrossRefPubMed
17.
go back to reference Krekel PR, Botha CP, Vastart ER, de Bruik RW, Rozing PM, Post FH. “Interactive simulation and comparative visualisation of the bone-determinded range of motion of the human shoulder”. Proc Simulation Visualization. 2006;3:275–88. Krekel PR, Botha CP, Vastart ER, de Bruik RW, Rozing PM, Post FH. “Interactive simulation and comparative visualisation of the bone-determinded range of motion of the human shoulder”. Proc Simulation Visualization. 2006;3:275–88.
18.
go back to reference Krekel PR, Vochteloo AJ, Bloem RM, Nelissen RG. Femoroacetabular impingement and its implications on range of motion: a case report. J Med Case Rep. 2011;5:143. doi:10.1186/1752-1947-5-143.CrossRefPubMedPubMedCentral Krekel PR, Vochteloo AJ, Bloem RM, Nelissen RG. Femoroacetabular impingement and its implications on range of motion: a case report. J Med Case Rep. 2011;5:143. doi:10.1186/1752-1947-5-143.CrossRefPubMedPubMedCentral
19.
go back to reference Vochteloo AJ, Krekel PR, vd. Sande MA, Nagels J. Range of motion implications of proximal humerus fractures: a case study. Eur Orthop Traumatol. 2011;2(56):153–6.CrossRefPubMedPubMedCentral Vochteloo AJ, Krekel PR, vd. Sande MA, Nagels J. Range of motion implications of proximal humerus fractures: a case study. Eur Orthop Traumatol. 2011;2(56):153–6.CrossRefPubMedPubMedCentral
20.
go back to reference Slieker J, Theeuwes HP, Rooijen GL, Van Lange JF, Kleinrensink GJ. Training in laparoscopic colorectal surgery: a new educational model using specially embalmed human anatomical specimen. Surg Endosc. 2012;26(8):2189–94.CrossRefPubMedPubMedCentral Slieker J, Theeuwes HP, Rooijen GL, Van Lange JF, Kleinrensink GJ. Training in laparoscopic colorectal surgery: a new educational model using specially embalmed human anatomical specimen. Surg Endosc. 2012;26(8):2189–94.CrossRefPubMedPubMedCentral
21.
go back to reference Rachbauer F, Kain MS, Leunig M. The history of the anterior approach to the hip. Orthop Clin North Am. 2009;40(3):311–20.CrossRefPubMed Rachbauer F, Kain MS, Leunig M. The history of the anterior approach to the hip. Orthop Clin North Am. 2009;40(3):311–20.CrossRefPubMed
22.
go back to reference Meskers CG, Fraterman H, Vd Helm FC, Vermeulen HM, Rozing PM. Calibration of the “Flock of Birds” electromagnetic tracking device and its applications in shoulder motion studies. J Biomech. 1999;32(6):629–33.CrossRefPubMed Meskers CG, Fraterman H, Vd Helm FC, Vermeulen HM, Rozing PM. Calibration of the “Flock of Birds” electromagnetic tracking device and its applications in shoulder motion studies. J Biomech. 1999;32(6):629–33.CrossRefPubMed
23.
go back to reference Milne AD, Chess DG, Johnson JA, King GJW. Accuracy of an electromagnetic tracking device: a study of the optimal operating range and metal interference. J Biomech. 1996;29(6):791–3.CrossRefPubMed Milne AD, Chess DG, Johnson JA, King GJW. Accuracy of an electromagnetic tracking device: a study of the optimal operating range and metal interference. J Biomech. 1996;29(6):791–3.CrossRefPubMed
24.
go back to reference Beck M, Kalhor M, Leunig M, Ganz R. Hip morphology influences the pattern of damage to the acetabular cartilage: femoroacetabular impingement as a cause of early osteoarthritis of the hip. J Boint Joint Surg (Br). 2005;87:1012–8.CrossRef Beck M, Kalhor M, Leunig M, Ganz R. Hip morphology influences the pattern of damage to the acetabular cartilage: femoroacetabular impingement as a cause of early osteoarthritis of the hip. J Boint Joint Surg (Br). 2005;87:1012–8.CrossRef
25.
go back to reference Tannast M, Goricki D, Beck M, Murphy SB, Siebenrock KA. Hip damage occurs at the zone of femoroacetabular impingement. Clin Orthop Relat Res. 2008;466(2):273–80.CrossRefPubMedPubMedCentral Tannast M, Goricki D, Beck M, Murphy SB, Siebenrock KA. Hip damage occurs at the zone of femoroacetabular impingement. Clin Orthop Relat Res. 2008;466(2):273–80.CrossRefPubMedPubMedCentral
26.
go back to reference Wu G, Siegler S, Allard P, Kirtley C, Leardini A, Rosenbaum D, et al. ISB recommendation on definitions of joint coordinate system of various joints for the reporting of human joint motion - part I: ankle, hip, and spine. J Biomech. 2002;35:543–8.CrossRefPubMed Wu G, Siegler S, Allard P, Kirtley C, Leardini A, Rosenbaum D, et al. ISB recommendation on definitions of joint coordinate system of various joints for the reporting of human joint motion - part I: ankle, hip, and spine. J Biomech. 2002;35:543–8.CrossRefPubMed
27.
go back to reference Puls M, Echer TM, Tannast M, Steppacker SD, Siebenrock KA, Kowal JH. The equidistant method – a novel hip joint simulation algorithm for detection of femoroacetabular impingement. Comput Aided Surg. 2010;15(4–6):75–82.CrossRefPubMed Puls M, Echer TM, Tannast M, Steppacker SD, Siebenrock KA, Kowal JH. The equidistant method – a novel hip joint simulation algorithm for detection of femoroacetabular impingement. Comput Aided Surg. 2010;15(4–6):75–82.CrossRefPubMed
28.
go back to reference Barton C, Salineros MJ, Rakhra KS, Beaulé PE. Validity of the alpha angle measurement on plain radiographs in the evaluation of cam-type femoroacetabular impingement. Clin Orthop Relat Res. 2011;469(2):464–9.CrossRefPubMed Barton C, Salineros MJ, Rakhra KS, Beaulé PE. Validity of the alpha angle measurement on plain radiographs in the evaluation of cam-type femoroacetabular impingement. Clin Orthop Relat Res. 2011;469(2):464–9.CrossRefPubMed
29.
go back to reference Nepple JJ, Martel JM, Kim YJ, Zaltz I, Clohisy JC. Anchor study group. Do plain radiographs correlate with CT for imaging of cam-type femoroacetabular impingement? Clin Orthop Relat Res. 2012;470(12):3313–20.CrossRefPubMedPubMedCentral Nepple JJ, Martel JM, Kim YJ, Zaltz I, Clohisy JC. Anchor study group. Do plain radiographs correlate with CT for imaging of cam-type femoroacetabular impingement? Clin Orthop Relat Res. 2012;470(12):3313–20.CrossRefPubMedPubMedCentral
30.
go back to reference de Bruin F, Reijnierse M, Farhang-Razi V, Bloem JL. Radiographic signs associated with femoroacetabular impingement occur with high prevalence at all ages in a hospital population. Eur Radiol. 2013;23(11):3131–9.CrossRefPubMed de Bruin F, Reijnierse M, Farhang-Razi V, Bloem JL. Radiographic signs associated with femoroacetabular impingement occur with high prevalence at all ages in a hospital population. Eur Radiol. 2013;23(11):3131–9.CrossRefPubMed
31.
go back to reference Gervaise A, Osemont B, Louis M, Lecocq S, Teixeira P, Blum A. Standard dose versus low-dose abdominal and pelvic CT: comparison between filtered back projection versus adaptive iterative dose reduction 3D. Diagn Interv Imaging. 2014;95(1):47–53.CrossRefPubMed Gervaise A, Osemont B, Louis M, Lecocq S, Teixeira P, Blum A. Standard dose versus low-dose abdominal and pelvic CT: comparison between filtered back projection versus adaptive iterative dose reduction 3D. Diagn Interv Imaging. 2014;95(1):47–53.CrossRefPubMed
32.
go back to reference Becce F, Ben Salah Y, Verdun FR, Vande Berg BC, Lecouvet FE, Leuli R, et al. Computed tomography of the cervical spine: comparison of image quality between a standard-dose and a low-dose protocol using filtered back-projection and iterative reconstruction. Skeletal Radiol. 2013;42(7):937–45.CrossRefPubMed Becce F, Ben Salah Y, Verdun FR, Vande Berg BC, Lecouvet FE, Leuli R, et al. Computed tomography of the cervical spine: comparison of image quality between a standard-dose and a low-dose protocol using filtered back-projection and iterative reconstruction. Skeletal Radiol. 2013;42(7):937–45.CrossRefPubMed
Metadata
Title
A quantitative non-invasive assessment of femoroacetabular impingement with CT-based dynamic simulation - cadaveric validation study
Authors
Maarten A Röling
Monique I Visser
Edwin HG Oei
Peter Pilot
Gert-Jan Kleinrensink
Rolf M Bloem
Publication date
01-12-2015
Publisher
BioMed Central
Published in
BMC Musculoskeletal Disorders / Issue 1/2015
Electronic ISSN: 1471-2474
DOI
https://doi.org/10.1186/s12891-015-0504-7

Other articles of this Issue 1/2015

BMC Musculoskeletal Disorders 1/2015 Go to the issue