Published in: BMC Ophthalmology 1/2016

Open Access 01-12-2016 | Research article

Effects of intraocular lenses with different diopters on chromatic aberrations in human eye models

Authors: Hui Song, Xiaoyong Yuan, Xin Tang

Published in: BMC Ophthalmology | Issue 1/2016

Login to get access

Abstract

Background

In this study, the effects of intraocular lenses (IOLs) with different diopters (D) on chromatic aberration were investigated in human eye models, and the influences of the central thickness of IOLs on chromatic aberration were compared.

Methods

A Liou-Brennan-based IOL eye model was constructed using ZEMAX optical design software. Spherical IOLs with different diopters (AR40e, AMO Company, USA) were implanted; modulation transfer function (MTF) values at 3 mm of pupil diameter and from 0 to out-of-focus blur were collected and graphed.

Results

MTF values, measured at 555 nm of monochromatic light under each spatial frequency, were significantly higher than the values measured at 470 to 650 nm of polychromatic light. The influences of chromatic aberration on MTF values decreased with the increase in IOL diopter when the spatial frequency was ≤12 c/d, while increased effects were observed when the spatial frequency was ≥15 c/d. The MTF values of each IOL eye model were significantly lower than the MTF values of the Liou-Brennan eye models when measured at 555 nm of monochromatic light and at 470 to 650 nm of polychromatic light. The MTF values were also found to be increased with the increase in IOL diopter.

Conclusion

With higher diopters of IOLs, the central thickness increased accordingly, which could have created increased chromatic aberration and decreased the retinal image quality. To improve the postoperative visual quality, IOLs with lower chromatic aberration should be selected for patients with short axial lengths.
Literature
1.
go back to reference Vinas M, Dorronsoro C, Cortes D, Pascual D, Marcos S. Longitudinal chromatic aberration of the human eye in the visible and near infrared from wavefront sensing, double-pass and psychophysics. Biomed Opt Express. 2015;6(3):948–62.CrossRefPubMedPubMedCentral Vinas M, Dorronsoro C, Cortes D, Pascual D, Marcos S. Longitudinal chromatic aberration of the human eye in the visible and near infrared from wavefront sensing, double-pass and psychophysics. Biomed Opt Express. 2015;6(3):948–62.CrossRefPubMedPubMedCentral
2.
go back to reference Artal P, Manzanera S, Piers P, Weeber H. Visual effect of the combined correction of spherical and longitudinal chromatic aberrations. Opt Express. 2010;18:1637–48.CrossRefPubMed Artal P, Manzanera S, Piers P, Weeber H. Visual effect of the combined correction of spherical and longitudinal chromatic aberrations. Opt Express. 2010;18:1637–48.CrossRefPubMed
3.
go back to reference Weeber HA, Piers PA. Theoretical performance of intraocular lenses correcting both spherical and chromatic aberration. J Refract Surg. 2012;28:48–52.CrossRefPubMed Weeber HA, Piers PA. Theoretical performance of intraocular lenses correcting both spherical and chromatic aberration. J Refract Surg. 2012;28:48–52.CrossRefPubMed
4.
go back to reference Sabesan R, Zheleznyak L, Yoon G. Binocular visual performance and summation after correcting higher order aberrations. Biomed Opt Express. 2012;3:3176-89. Sabesan R, Zheleznyak L, Yoon G. Binocular visual performance and summation after correcting higher order aberrations. Biomed Opt Express. 2012;3:3176-89.
5.
go back to reference Applegate RA, Thibos LN, Hilmantel G. Optics of aberroscopy and super vision. J Cataract Refract Surg. 2001;27:1093–107.CrossRefPubMed Applegate RA, Thibos LN, Hilmantel G. Optics of aberroscopy and super vision. J Cataract Refract Surg. 2001;27:1093–107.CrossRefPubMed
6.
go back to reference Weeber HA, Pohl R, Mester U, Piers PA. Visual performance of pseudophakic eyes corrected for spherical and chromatic aberrations with an achromatic intraocular lens. Invest Ophthalmol Vis Sci. 2013;54:808.CrossRef Weeber HA, Pohl R, Mester U, Piers PA. Visual performance of pseudophakic eyes corrected for spherical and chromatic aberrations with an achromatic intraocular lens. Invest Ophthalmol Vis Sci. 2013;54:808.CrossRef
7.
go back to reference Schwarz C, Canovas C, Manzanera S, Weeber H, Prieto PM, Piers P, et al. Binocular visual acuity for the correction of spherical aberration in polychromatic and monochromatic light. J Vis. 2014;14:1–11. Schwarz C, Canovas C, Manzanera S, Weeber H, Prieto PM, Piers P, et al. Binocular visual acuity for the correction of spherical aberration in polychromatic and monochromatic light. J Vis. 2014;14:1–11.
8.
go back to reference Perez-Merino P, Dorronsoro C, Llorente L, Duran S, Jimenez-Alfaro I, Marcos S. In vivo chromatic aberration in eyes implanted with intraocular lenses. Invest Ophthalmol Vis Sci. 2013;54:2654–61.CrossRefPubMed Perez-Merino P, Dorronsoro C, Llorente L, Duran S, Jimenez-Alfaro I, Marcos S. In vivo chromatic aberration in eyes implanted with intraocular lenses. Invest Ophthalmol Vis Sci. 2013;54:2654–61.CrossRefPubMed
9.
go back to reference Siedlecki D, Ginis HS. On the longitudinal chromatic aberration of the intraocular lenses. Optom and Vis Sci. 2007;84:984–9.CrossRef Siedlecki D, Ginis HS. On the longitudinal chromatic aberration of the intraocular lenses. Optom and Vis Sci. 2007;84:984–9.CrossRef
10.
go back to reference López-Gil N, Bradley A. The potential for and challenges of spherical and chromatic aberration correction with new IOL designs. Br J Ophthalmol. 2013;97(6):677–8.CrossRefPubMed López-Gil N, Bradley A. The potential for and challenges of spherical and chromatic aberration correction with new IOL designs. Br J Ophthalmol. 2013;97(6):677–8.CrossRefPubMed
11.
go back to reference Liou HL, Brennan NA. Anatomically accurate, finite model eye for optical modeling. J Opt Soc Am A Opt Image Sci Vis. 1997;14:1684–95.CrossRefPubMed Liou HL, Brennan NA. Anatomically accurate, finite model eye for optical modeling. J Opt Soc Am A Opt Image Sci Vis. 1997;14:1684–95.CrossRefPubMed
12.
go back to reference Piers PA, Weeber HA, Artal P, Norrby S. Theoretical comparison of aberration-correcting customized and aspheric intraocular lenses. J Refract Surg. 2007;23:374–84.PubMed Piers PA, Weeber HA, Artal P, Norrby S. Theoretical comparison of aberration-correcting customized and aspheric intraocular lenses. J Refract Surg. 2007;23:374–84.PubMed
13.
go back to reference Liang J, Williams DR. Aberration and retinal image quality of the normal human eye. J Opt Soc Am A Opt Image Sci Vis. 1997;14:2873–83.CrossRefPubMed Liang J, Williams DR. Aberration and retinal image quality of the normal human eye. J Opt Soc Am A Opt Image Sci Vis. 1997;14:2873–83.CrossRefPubMed
14.
go back to reference Thibos LN, Bradley A, Still DL, Zhang X, Howarth PA. Theory and measurement of ocular chromatic aberration. Vision Res. 1990;30:33–49.CrossRefPubMed Thibos LN, Bradley A, Still DL, Zhang X, Howarth PA. Theory and measurement of ocular chromatic aberration. Vision Res. 1990;30:33–49.CrossRefPubMed
15.
go back to reference Awwad ST, El-Kateb M, McCulley JP. Comparative higher-order aberration measurement of the LADARWave and Visx WaveScan aberration at varying pupil sizes and after pharmacologic dilation and cycloplegia. J Cataract Refract Surg. 2006;32:203–14.CrossRefPubMed Awwad ST, El-Kateb M, McCulley JP. Comparative higher-order aberration measurement of the LADARWave and Visx WaveScan aberration at varying pupil sizes and after pharmacologic dilation and cycloplegia. J Cataract Refract Surg. 2006;32:203–14.CrossRefPubMed
16.
go back to reference Ravikumar S, Bradley A, Thibos LN. Chromatic aberration and polychromatic image quality with diffractive multifocal intraocular lenses. J Cataract Refract Surg. 2014;40(7):1192–204.CrossRefPubMed Ravikumar S, Bradley A, Thibos LN. Chromatic aberration and polychromatic image quality with diffractive multifocal intraocular lenses. J Cataract Refract Surg. 2014;40(7):1192–204.CrossRefPubMed
17.
go back to reference Guirao A, Redondo M, Arta LP. Optical aberrations of the human cornea as a function of age. J Opt Soc Am A Opt Image Sci Vis. 2000;17:1697–702.CrossRefPubMed Guirao A, Redondo M, Arta LP. Optical aberrations of the human cornea as a function of age. J Opt Soc Am A Opt Image Sci Vis. 2000;17:1697–702.CrossRefPubMed
19.
go back to reference Siedlecki D, Jozwik A, Zajac M, Hill-Bator A, Turno-Krecicka A. In vivo longitudinal chromatic aberration of pseudophakic eyes. Optom Vis Sci. 2014;91:240–6.PubMed Siedlecki D, Jozwik A, Zajac M, Hill-Bator A, Turno-Krecicka A. In vivo longitudinal chromatic aberration of pseudophakic eyes. Optom Vis Sci. 2014;91:240–6.PubMed
20.
go back to reference Sekiguchi N, Williams DR, Brainard DH. Efficiency in detection of isoluminant and isochromatic interference fringes. J Opt Soc Am A Opt Image Sci Vis. 1993;10:2118–33.CrossRefPubMed Sekiguchi N, Williams DR, Brainard DH. Efficiency in detection of isoluminant and isochromatic interference fringes. J Opt Soc Am A Opt Image Sci Vis. 1993;10:2118–33.CrossRefPubMed
21.
go back to reference Nio YK, Jansonius NM, Fidler V, Geraghty E, Norrby S, Kooijman AC. Age-related changes of defocus-specific contrast sensitivity in healthy subjects. Opthal Physiol Opt. 2000;20:323–34.CrossRef Nio YK, Jansonius NM, Fidler V, Geraghty E, Norrby S, Kooijman AC. Age-related changes of defocus-specific contrast sensitivity in healthy subjects. Opthal Physiol Opt. 2000;20:323–34.CrossRef
22.
go back to reference Qu J, Lv F, Mao X. Evaluation of visual correction. Chin J Ophthalmol. 2003;6:325–7. Qu J, Lv F, Mao X. Evaluation of visual correction. Chin J Ophthalmol. 2003;6:325–7.
23.
go back to reference Wang Y, Zhao K. Wavefront Aberration and Clinical Vision Correction. Beijing: People’s Medical Publishing House; 2011. Wang Y, Zhao K. Wavefront Aberration and Clinical Vision Correction. Beijing: People’s Medical Publishing House; 2011.
24.
go back to reference Kornowski JA, Petersik JT. Effect on face recognition of spatial-frequency information contained in inspection and test stimuli. J Gen Psychol. 2003;130:229–44.CrossRefPubMed Kornowski JA, Petersik JT. Effect on face recognition of spatial-frequency information contained in inspection and test stimuli. J Gen Psychol. 2003;130:229–44.CrossRefPubMed
26.
go back to reference Williams CS, Becklund OA. Introduction to the optical transfer function. Bellingham: SPIE Press; 1989. Williams CS, Becklund OA. Introduction to the optical transfer function. Bellingham: SPIE Press; 1989.
27.
go back to reference McLellan JS, Marcos S, Prieto PM, Burns SA. Imperfect optics may be the eye’s defense against chromatic blur. Nature. 2002;417:174–6.CrossRefPubMed McLellan JS, Marcos S, Prieto PM, Burns SA. Imperfect optics may be the eye’s defense against chromatic blur. Nature. 2002;417:174–6.CrossRefPubMed
28.
go back to reference Ravikumar S, Thibos LN, Bradley A. Calculation of retinal image quality for polychromatic light. J Opt Soc Am A Opt Image Sci Vis. 2008;25:2395–407.CrossRefPubMed Ravikumar S, Thibos LN, Bradley A. Calculation of retinal image quality for polychromatic light. J Opt Soc Am A Opt Image Sci Vis. 2008;25:2395–407.CrossRefPubMed
29.
go back to reference Franchini A. Compromise between spherical and chromatic aberration and depth of focus in aspheric intraocular lenses. J Cataract Refract Surg. 2007;33:497–509.CrossRefPubMed Franchini A. Compromise between spherical and chromatic aberration and depth of focus in aspheric intraocular lenses. J Cataract Refract Surg. 2007;33:497–509.CrossRefPubMed
Metadata
Title
Effects of intraocular lenses with different diopters on chromatic aberrations in human eye models
Authors
Hui Song
Xiaoyong Yuan
Xin Tang
Publication date
01-12-2016
Publisher
BioMed Central
Published in
BMC Ophthalmology / Issue 1/2016
Electronic ISSN: 1471-2415
DOI
https://doi.org/10.1186/s12886-016-0184-6

Other articles of this Issue 1/2016

BMC Ophthalmology 1/2016 Go to the issue