Unrecognized maternal heart rate artefact in cases of perinatal mortality reported to the United States Food and Drug Administration from 2009 to 2019: a critical patient safety issue

Authors: Daniel J. Kiely, Lawrence W. Oppenheimer, James C. Dornan

Published in: BMC Pregnancy and Childbirth | Issue 1/2019

Login to get access

Abstract

Background

Maternal heart rate artefact is a signal processing error whereby the fetal heart rate is masked by the maternal pulse, potentially leading to danger by failure to recognize an abnormal fetal heart rate or a pre-existing fetal death. Maternal heart rate artefact may be exacerbated by autocorrelation algorithms in modern fetal monitors due to smooth transitions between maternal and fetal heart rates rather than breaks in the tracing. In response, manufacturers of cardiotocography monitors recommend verifying fetal life prior to monitoring and have developed safeguards including signal ambiguity detection technologies to simultaneously and continuously monitor the maternal and fetal heart rates. However, these safeguards are not emphasized in current cardiotocography clinical practice guidelines, potentially leading to a patient safety gap.

Methods

The United States Food and Drug Administration Manufacturer and User Facility Device Experience database was reviewed for records with event type “Death” for the time period March 31, 2009 to March 31, 2019, in combination with search terms selected to capture all cases reported involving cardiotocography devices. Records were reviewed to determine whether maternal heart rate artefact was probable and/or whether the report contained a recommendation from the device manufacturer regarding maternal heart rate artefact.

Results

Forty-seven cases of perinatal mortality were identified with probable maternal heart rate artefact including 14 with antepartum fetal death prior to initiation of cardiotocography, 14 with intrapartum fetal death or neonatal death after initiation of cardiotocography, and 19 where the temporal relationship between initiation of cardiotocography and death cannot be definitively established from the report. In 29 cases, there was a recommendation from the manufacturer regarding diagnosis and/or management of maternal heart rate artefact.

Conclusions

This case series indicates a recurring problem with undetected maternal heart rate artefact leading to perinatal mortality and, in cases of pre-existing fetal death, healthcare provider confusion. In response, manufacturers frequently recommend safeguards which are found in their device’s instructions for use but not in major intrapartum cardiotocography guidelines. Cardiotocography guidelines should be updated to include the latest safeguards against the risks of maternal heart rate artefact. An additional file summarizing key points for clinicians is included.
Appendix
Available only for authorised users
Literature
1.
go back to reference Miller DA. Intrapartum fetal evaluation. In: Gabbe SG, Niebyl JR, Simpson JL, Landon MB, Galan HL, Jauniaux ERM, Driscoll DA, Berghella V, Grobman WA, editors. Obstetrics: normal and problem pregnancies. 7th ed: Philadelphia: Elsevier; 2017. p. 330. Miller DA. Intrapartum fetal evaluation. In: Gabbe SG, Niebyl JR, Simpson JL, Landon MB, Galan HL, Jauniaux ERM, Driscoll DA, Berghella V, Grobman WA, editors. Obstetrics: normal and problem pregnancies. 7th ed: Philadelphia: Elsevier; 2017. p. 330.
2.
go back to reference Neilson DR Jr, Freeman RK, Mangan S. Signal ambiguity resulting in unexpected outcome with external fetal heart rate monitoring. Am J Obstet Gynecol. 2008;198(6):717–24.CrossRef Neilson DR Jr, Freeman RK, Mangan S. Signal ambiguity resulting in unexpected outcome with external fetal heart rate monitoring. Am J Obstet Gynecol. 2008;198(6):717–24.CrossRef
3.
go back to reference Freeman RK, Garite TJ, Nageotte MP, Miller LA. Fetal heart rate monitoring. 4th ed. Philadelphia: Lippincott Williams & Wilkins; 2012. Freeman RK, Garite TJ, Nageotte MP, Miller LA. Fetal heart rate monitoring. 4th ed. Philadelphia: Lippincott Williams & Wilkins; 2012.
4.
go back to reference Paquette S, Moretti F, O'Reilly K, Ferraro ZM, Oppenheimer L. The incidence of maternal artefact during intrapartum fetal heart rate monitoring. J Obstet Gynaecol Can. 2014;36(11):962–8.CrossRef Paquette S, Moretti F, O'Reilly K, Ferraro ZM, Oppenheimer L. The incidence of maternal artefact during intrapartum fetal heart rate monitoring. J Obstet Gynaecol Can. 2014;36(11):962–8.CrossRef
5.
go back to reference Fukushima T, Flores CA, Hon EH, Davidson EC Jr. Limitations of autocorrelation in fetal heart rate monitoring. Am J Obstet Gynecol. 1985;153(6):685–92.CrossRef Fukushima T, Flores CA, Hon EH, Davidson EC Jr. Limitations of autocorrelation in fetal heart rate monitoring. Am J Obstet Gynecol. 1985;153(6):685–92.CrossRef
15.
go back to reference Kiely DJ. A 2009 fetal monitor recall that arguably should have been class I. Am J Obstet Gynecol. 2018;218(3):369.CrossRef Kiely DJ. A 2009 fetal monitor recall that arguably should have been class I. Am J Obstet Gynecol. 2018;218(3):369.CrossRef
16.
go back to reference Kiely DJ. Did the maternal pulse mask the fetal heart rate of acidemic infants with no explanatory features? Am J Obstet Gynecol. 2018;219(3):314.CrossRef Kiely DJ. Did the maternal pulse mask the fetal heart rate of acidemic infants with no explanatory features? Am J Obstet Gynecol. 2018;219(3):314.CrossRef
17.
go back to reference Kiely DJ. The incidence of maternal Artefact during Intrapartum fetal heart rate monitoring. J Obstet Gynaecol Can. 2015;37(3):205–6.CrossRef Kiely DJ. The incidence of maternal Artefact during Intrapartum fetal heart rate monitoring. J Obstet Gynaecol Can. 2015;37(3):205–6.CrossRef
18.
go back to reference Kiely DJ. Management of Spontaneous Labour at term in healthy women. J Obstet Gynaecol Can. 2017;39(4):220–1.CrossRef Kiely DJ. Management of Spontaneous Labour at term in healthy women. J Obstet Gynaecol Can. 2017;39(4):220–1.CrossRef
30.
go back to reference Herman A, Ron-El R, Arieli S, Schreyer P, Caspi E. Maternal ECG recorded by internal monitoring closely mimicking fetal heart rate in a recent fetal death. Int J Gynecol Obstet. 1990;33(3):269–71.CrossRef Herman A, Ron-El R, Arieli S, Schreyer P, Caspi E. Maternal ECG recorded by internal monitoring closely mimicking fetal heart rate in a recent fetal death. Int J Gynecol Obstet. 1990;33(3):269–71.CrossRef
31.
go back to reference Ramsey PS, Johnston BW, Welter VE, Ogburn PL, Jr. Artifactual fetal electrocardiographic detection using internal monitoring following intrapartum fetal demise during VBAC trial. J Matern Fetal Med. 2000;9(6):360–1.CrossRef Ramsey PS, Johnston BW, Welter VE, Ogburn PL, Jr. Artifactual fetal electrocardiographic detection using internal monitoring following intrapartum fetal demise during VBAC trial. J Matern Fetal Med. 2000;9(6):360–1.CrossRef
32.
go back to reference Achiron R, Zakut H. Misinterpretation of fetal heart rate monitoring in case of intrauterine death. Clin Exp Obstet Gynecol. 1984;11(4):126–9.PubMed Achiron R, Zakut H. Misinterpretation of fetal heart rate monitoring in case of intrauterine death. Clin Exp Obstet Gynecol. 1984;11(4):126–9.PubMed
33.
go back to reference Barrett JM, Boehm FH. Documentation of recent fetal demise with simultaneous maternal and fetal heart rate monitoring. Obstet Gynecol. 1980;55(3 Suppl):28s–30s.CrossRef Barrett JM, Boehm FH. Documentation of recent fetal demise with simultaneous maternal and fetal heart rate monitoring. Obstet Gynecol. 1980;55(3 Suppl):28s–30s.CrossRef
34.
go back to reference Ayres-de-Campos D, Spong CY, Chandraharan E. FIGO consensus guidelines on intrapartum fetal monitoring: Cardiotocography. Int J Gynecol Obstet. 2015;131(1):13–24.CrossRef Ayres-de-Campos D, Spong CY, Chandraharan E. FIGO consensus guidelines on intrapartum fetal monitoring: Cardiotocography. Int J Gynecol Obstet. 2015;131(1):13–24.CrossRef
36.
go back to reference The American College of Obstetricians and Gynecologists. Practice bulletin no. 116: management of intrapartum fetal heart rate tracings. Obstet Gynecol. 2010;116(5):1232–40.CrossRef The American College of Obstetricians and Gynecologists. Practice bulletin no. 116: management of intrapartum fetal heart rate tracings. Obstet Gynecol. 2010;116(5):1232–40.CrossRef
38.
go back to reference Liston R, Sawchuck D, Young D. No. 197a-fetal health surveillance: antepartum consensus guideline. J Obstet Gynaecol Can. 2018;40(4):e251–e71.CrossRef Liston R, Sawchuck D, Young D. No. 197a-fetal health surveillance: antepartum consensus guideline. J Obstet Gynaecol Can. 2018;40(4):e251–e71.CrossRef
39.
go back to reference Liston R, Sawchuck D, Young D. No. 197b-fetal health surveillance: Intrapartum consensus guideline. J Obstet Gynaecol Can. 2018;40(4):e298–322.CrossRef Liston R, Sawchuck D, Young D. No. 197b-fetal health surveillance: Intrapartum consensus guideline. J Obstet Gynaecol Can. 2018;40(4):e298–322.CrossRef
40.
go back to reference Liston R, Sawchuck D, Young D. No. 197c-maintaining standards in antenatal and Intrapartum fetal surveillance: quality improvement and risk management. J Obstet Gynaecol Can. 2018;40(4):e353–e8.CrossRef Liston R, Sawchuck D, Young D. No. 197c-maintaining standards in antenatal and Intrapartum fetal surveillance: quality improvement and risk management. J Obstet Gynaecol Can. 2018;40(4):e353–e8.CrossRef
42.
go back to reference Sherman DJ, Frenkel E, Kurzweil Y, Padua A, Arieli S, Bahar M. Characteristics of maternal heart rate patterns during labor and delivery. Obstet Gynecol. 2002;99(4):542–7.PubMed Sherman DJ, Frenkel E, Kurzweil Y, Padua A, Arieli S, Bahar M. Characteristics of maternal heart rate patterns during labor and delivery. Obstet Gynecol. 2002;99(4):542–7.PubMed
43.
go back to reference Murray ML. Maternal or fetal heart rate? Avoiding intrapartum misidentification. J Obstet, Gynecol, Neonatal Nurse : JOGNN. 2004;33(1):93–104.CrossRef Murray ML. Maternal or fetal heart rate? Avoiding intrapartum misidentification. J Obstet, Gynecol, Neonatal Nurse : JOGNN. 2004;33(1):93–104.CrossRef
44.
go back to reference Odendaal HJ. False interpretation of fetal heart role monitoring in cases of intra-uterine death. S Afr Med J. 1976;50(49):1963–5.PubMed Odendaal HJ. False interpretation of fetal heart role monitoring in cases of intra-uterine death. S Afr Med J. 1976;50(49):1963–5.PubMed
45.
go back to reference Pinto P, Costa-Santos C, Gonçalves H, Ayres-De-Campos D, Bernardes J. Improvements in fetal heart rate analysis by the removal of maternal-fetal heart rate ambiguities. BMC Pregnancy Childbirth. 2015;15(1):301.CrossRef Pinto P, Costa-Santos C, Gonçalves H, Ayres-De-Campos D, Bernardes J. Improvements in fetal heart rate analysis by the removal of maternal-fetal heart rate ambiguities. BMC Pregnancy Childbirth. 2015;15(1):301.CrossRef
46.
go back to reference Pinto P, Costa-Santos C, Ayres-de-Campos D, Bernardes J. Computer analysis of maternal-fetal heart rate recordings during labor in relation with maternal-fetal attachment and prediction of newborn acidemia. J Matern Fetal Neonatal Med. 2016;29(9):1440–4.CrossRef Pinto P, Costa-Santos C, Ayres-de-Campos D, Bernardes J. Computer analysis of maternal-fetal heart rate recordings during labor in relation with maternal-fetal attachment and prediction of newborn acidemia. J Matern Fetal Neonatal Med. 2016;29(9):1440–4.CrossRef
47.
go back to reference Goncalves H, Pinto P, Silva M, Ayres-de-Campos D, Bernardes J. Toward the improvement in fetal monitoring during labor with the inclusion of maternal heart rate analysis. Med Biol Eng Comput. 2016;54(4):691–9.CrossRef Goncalves H, Pinto P, Silva M, Ayres-de-Campos D, Bernardes J. Toward the improvement in fetal monitoring during labor with the inclusion of maternal heart rate analysis. Med Biol Eng Comput. 2016;54(4):691–9.CrossRef
48.
go back to reference Wu AW. Medical error: the second victim. The doctor who makes the mistake needs help too. BMJ. 2000;320(7237):726–7.CrossRef Wu AW. Medical error: the second victim. The doctor who makes the mistake needs help too. BMJ. 2000;320(7237):726–7.CrossRef
50.
go back to reference Lee AC, Kozuki N, Blencowe H, Vos T, Bahalim A, Darmstadt GL, et al. Intrapartum-related neonatal encephalopathy incidence and impairment at regional and global levels for 2010 with trends from 1990. Pediatr Res. 2013;74(Suppl 1):50–72.CrossRef Lee AC, Kozuki N, Blencowe H, Vos T, Bahalim A, Darmstadt GL, et al. Intrapartum-related neonatal encephalopathy incidence and impairment at regional and global levels for 2010 with trends from 1990. Pediatr Res. 2013;74(Suppl 1):50–72.CrossRef
51.
go back to reference Gaba D. Improving Anesthesiologists' performance by simulating reality. Anesthesiology. 1992;76(4):491–4.CrossRef Gaba D. Improving Anesthesiologists' performance by simulating reality. Anesthesiology. 1992;76(4):491–4.CrossRef
Metadata
Title
Unrecognized maternal heart rate artefact in cases of perinatal mortality reported to the United States Food and Drug Administration from 2009 to 2019: a critical patient safety issue
Authors
Daniel J. Kiely
Lawrence W. Oppenheimer
James C. Dornan
Publication date
01-12-2019
Publisher
BioMed Central
Published in
BMC Pregnancy and Childbirth / Issue 1/2019
Electronic ISSN: 1471-2393
DOI
https://doi.org/10.1186/s12884-019-2660-5

Other articles of this Issue 1/2019

BMC Pregnancy and Childbirth 1/2019 Go to the issue