Open Access 01-12-2019 | Case report

Albumin losses during hemodiafiltration: all dialyzers are not created equal - a case report

Authors: Charles Cuvelier, Michel Tintillier, Gabriela Migali, Charlotte Van Ende, Jean-Michel Pochet

Published in: BMC Nephrology | Issue 1/2019

Login to get access

Abstract

Background

Online hemodiafiltration (OL-HDF) is associated with better removal of both small and middle molecules and might improve survival compared to conventional hemodialysis (HD). Nevertheless, hemodiafiltration (HDF) can lead to an increase in albumin loss across the dialyzer, especially with high permeability membrane and high convective volume (CV). We present the case of a patient treated by OL-HDF who developed severe hypoalbuminemia resulting from massive albumin loss into dialysate.

Case presentation

A 71-year-old woman with ESRD started renal replacement therapy in December 2016. She was treated by high volume post-dilution OL-HDF, 4 h, 3 times per week. The dialyzer was the Phylther HF20SD (a 2.0m2 heat sterilized high flux (HF) polyphenylene membrane from Bellco). At the initiation of dialysis, the serum albumin was 4.0 g/dl. During the following months, the patient developed severe hypoalbuminemia. The lowest value observed was 2.26 g/dl in July 2017. Diagnostic workup excluded nephrotic syndrome, hepatic failure and malabsorption. The patient was shifted from OL-HDF to standard HF HD, keeping the same dialyzer and dialysis schedule. During the following months, we observed a progressive correction of the hypoalbuminemia (3.82 g/dl at last follow-up). To precise the impact of the epuration technique on the albumin losses in this patient, we measured the amount of albumin in dialysate during one session with the Phylther HF20SD on OL-HDF and one session with the same filter but on standard HD. The CV was 29.0 l for the HDF session. The total albumin losses were 23.6 g on OL-HDF and 4.6 g on HD.

Conclusion

OL-HDF can lead to significant albumin loss into the dialysate, especially with high permeability membrane and high CV. When prescribing post-dilutional OL-HDF, the choice of the dialyzer membrane should be made with caution. Users of the steam sterilized polyphenylene membrane, the Phylther SD, should be informed of the risk of large albumin loss with this membrane during post-dilution OL-HDF.
Literature
5.
go back to reference Ahrenholz PG, Winkler RE, Michelsen A, Lang DA, Bowry SK. Dialysis membrane-dependent removal of middle molecules during hemodiafiltration: the beta2-microglobulin/albumin relationship. Clin Nephrol. 2004;62(1):21–8.CrossRefPubMed Ahrenholz PG, Winkler RE, Michelsen A, Lang DA, Bowry SK. Dialysis membrane-dependent removal of middle molecules during hemodiafiltration: the beta2-microglobulin/albumin relationship. Clin Nephrol. 2004;62(1):21–8.CrossRefPubMed
25.
go back to reference Bove S, D’Avanzo L, Terlizi V et al. FP458 Comparison of high flux hemodialyzers with medium cut-off dialyzer on the removal of middle high size uremic toxins. Nephrol Dial Transplant. 2018;33, Issue suppl_1, Pages i190. doi.10.1093, gfy104.FP458. Bove S, D’Avanzo L, Terlizi V et al. FP458 Comparison of high flux hemodialyzers with medium cut-off dialyzer on the removal of middle high size uremic toxins. Nephrol Dial Transplant. 2018;33, Issue suppl_1, Pages i190. doi.10.​1093, gfy104.​FP458.
Metadata
Title
Albumin losses during hemodiafiltration: all dialyzers are not created equal - a case report
Authors
Charles Cuvelier
Michel Tintillier
Gabriela Migali
Charlotte Van Ende
Jean-Michel Pochet
Publication date
01-12-2019
Publisher
BioMed Central
Published in
BMC Nephrology / Issue 1/2019
Electronic ISSN: 1471-2369
DOI
https://doi.org/10.1186/s12882-019-1567-8

Other articles of this Issue 1/2019

BMC Nephrology 1/2019 Go to the issue