Published in: BMC Nephrology 1/2015

Open Access 01-12-2015 | Research article

Is it useful to increase dialysate flow rate to improve the delivered Kt?

Authors: Marta Albalate, Rafael Pérez-García, Patricia de Sequera, Elena Corchete, Roberto Alcazar, Mayra Ortega, Marta Puerta

Published in: BMC Nephrology | Issue 1/2015

Login to get access

Abstract

Background

Increasing dialysate flow rates (Qd) from 500 to 800 ml/min has been recommended to increase dialysis efficiency. A few publications show that increasing Qd no longer led to an increase in mass transfer area coefficient (KoA) or Kt/V measurement.
Our objectives were: 1) Studying the effect in Kt of using a Qd of 400, 500, 700 ml/min and autoflow (AF) with different modern dialysers. 2) Comparing the effect on Kt of water consumption vs. dialysis time to obtain an individual objective of Kt (Ktobj) adjusted to body surface.

Methods

This is a prospective single-centre study with crossover design. Thirty-one patients were studied and six sessions with each Qd were performed. HD parameters were acquired directly from the monitor display: effective blood flow rate (Qbe), Qd, effective dialysis time (Te) and measured by conductivity monitoring, final Kt.

Results

We studied a total of 637 sessions: 178 with 500 ml/min, 173 with 700 ml/min, 160 with AF and 126 with 400 ml/min. Kt rose a 4% comparing 400 with 500 ml/min, and 3% comparing 500 with 700 ml/min. Ktobj was reached in 82.4, 88.2, 88.2 and 94.1% of patients with 400, AF, 500 and 700 ml/min, respectively. We did not find statistical differences between dialysers.
The difference between programmed time and Te was 8′ when Qd was 400 and 500 ml/min and 8.8′ with Qd = 700 ml/min.
Calculating an average time loss of eight minutes/session, we can say that a patient loses 24′ weekly, 312′ monthly and 62.4 hours yearly. Identical Kt could be obtained with Qd of 400 and 500 ml/min, increasing dialysis time 9.1′ and saving 20% of dialysate.

Conclusions

Our data suggest that increasing Qd over 400 ml/min for these dialysers offers a limited benefit. Increasing time is a better alternative with demonstrated benefits to the patient and also less water consumption.
Appendix
Available only for authorised users
Literature
1.
go back to reference Manzoni C, Di Filippo S, Corti M, Locatelli F. Ionic dialysance as a method for the on-line monitoring of delivered dialysis without blood sampling. Nephrol Dial Transplant. 1996;11:2023–30.CrossRefPubMed Manzoni C, Di Filippo S, Corti M, Locatelli F. Ionic dialysance as a method for the on-line monitoring of delivered dialysis without blood sampling. Nephrol Dial Transplant. 1996;11:2023–30.CrossRefPubMed
2.
go back to reference Lowrie EG, Chertow GM, Lew NL, Lazarus JM, Owen WF. The urea {clearance x dialysis time} product (Kt) as an outcome-based measure of hemodialysis dose. Kidney Int. 1999;56:729–37.CrossRefPubMed Lowrie EG, Chertow GM, Lew NL, Lazarus JM, Owen WF. The urea {clearance x dialysis time} product (Kt) as an outcome-based measure of hemodialysis dose. Kidney Int. 1999;56:729–37.CrossRefPubMed
3.
go back to reference Lowrie EG, Li Z, Ofsthun NJ, Lazarus JM. Evaluating a new method to judge dialysis treatment using online measurements of ionic clearance. Kidney Int. 2006;70:211–7.CrossRefPubMed Lowrie EG, Li Z, Ofsthun NJ, Lazarus JM. Evaluating a new method to judge dialysis treatment using online measurements of ionic clearance. Kidney Int. 2006;70:211–7.CrossRefPubMed
4.
go back to reference Maduell F, Ramos R, Palomares I, Martín-Malo A, Molina M, Bustamante J, et al. Impact of targeting Kt instead of Kt/V. Nephrol Dial Transplant. 2013;28:2595–603.CrossRefPubMed Maduell F, Ramos R, Palomares I, Martín-Malo A, Molina M, Bustamante J, et al. Impact of targeting Kt instead of Kt/V. Nephrol Dial Transplant. 2013;28:2595–603.CrossRefPubMed
5.
go back to reference Leypoldt JK, Cheung AK, Agodoa LY, Daugirdas JT, Greene T, Keshaviah PR. Hemodialyzer mass transfer-area coefficients for urea increase at high dialysate flow rates. Kidney Int. 1997;51:2013–7.CrossRefPubMed Leypoldt JK, Cheung AK, Agodoa LY, Daugirdas JT, Greene T, Keshaviah PR. Hemodialyzer mass transfer-area coefficients for urea increase at high dialysate flow rates. Kidney Int. 1997;51:2013–7.CrossRefPubMed
6.
go back to reference Ouseph R, Ward RA. Increasing dialysate flow rate increases dialyzer urea mass transfer-area coefficients during clinical use. Am J Kidney Dis. 2001;37:316–20.CrossRefPubMed Ouseph R, Ward RA. Increasing dialysate flow rate increases dialyzer urea mass transfer-area coefficients during clinical use. Am J Kidney Dis. 2001;37:316–20.CrossRefPubMed
7.
go back to reference Hirano A, Kida S, Yamamoto K, Sakai K. Experimental evaluation of flow and dialysis performance of hollow-fiber dialyzers with different packing densities. J Int Organs. 2012;15:168–75.CrossRef Hirano A, Kida S, Yamamoto K, Sakai K. Experimental evaluation of flow and dialysis performance of hollow-fiber dialyzers with different packing densities. J Int Organs. 2012;15:168–75.CrossRef
8.
go back to reference Hirano A, Yamamoto K, Matsuda M, Ogawa T, Yakushiji T, Miyasaka T, et al. Evaluation of dialyzer jacket structure and hollow-fiber dialysis membranes to achieve high dialysis performance. Ther Apher Dial. 2011;15:66–74.CrossRefPubMed Hirano A, Yamamoto K, Matsuda M, Ogawa T, Yakushiji T, Miyasaka T, et al. Evaluation of dialyzer jacket structure and hollow-fiber dialysis membranes to achieve high dialysis performance. Ther Apher Dial. 2011;15:66–74.CrossRefPubMed
9.
go back to reference Ronco C, Scabardi M, Goldoni M, Brendolan A, Crepaldi C, La Greca G. Impact of spacing filaments external to hollow fibers on dialysate flow distribution and dialyzer performance. Int J Artif Org. 1997;20:261–6. Ronco C, Scabardi M, Goldoni M, Brendolan A, Crepaldi C, La Greca G. Impact of spacing filaments external to hollow fibers on dialysate flow distribution and dialyzer performance. Int J Artif Org. 1997;20:261–6.
10.
go back to reference Ronco C, Brendolan A, Crepaldi C, Rodighiero M, Scabardi M. Blood and dialysate flow distributions in hollow-fiber hemodialyzers analyzed by computerized helical scanning technique. J Am Soc Nephrol. 2002;13:S53–61.PubMed Ronco C, Brendolan A, Crepaldi C, Rodighiero M, Scabardi M. Blood and dialysate flow distributions in hollow-fiber hemodialyzers analyzed by computerized helical scanning technique. J Am Soc Nephrol. 2002;13:S53–61.PubMed
11.
go back to reference Bhimani J, Ouseph R, Ward R. Effect of increasing dialysate flow rate on diffusive mass transfer of urea, phosphate and β2-microglobulin during clinical haemodialysis. Nephrol Dial Transplant. 2010;25:3990–5.CrossRefPubMedPubMedCentral Bhimani J, Ouseph R, Ward R. Effect of increasing dialysate flow rate on diffusive mass transfer of urea, phosphate and β2-microglobulin during clinical haemodialysis. Nephrol Dial Transplant. 2010;25:3990–5.CrossRefPubMedPubMedCentral
12.
go back to reference Ward RA, Idoux JW, Hamdan H, Ouseph R, Depner TA, Golper TA. Dialysate flow rate and delivered Kt/Vurea for dialyzers with enhanced dialysate flow distribution. Clin J Am Soc Nephrol. 2011;6:2235–9.CrossRefPubMedPubMedCentral Ward RA, Idoux JW, Hamdan H, Ouseph R, Depner TA, Golper TA. Dialysate flow rate and delivered Kt/Vurea for dialyzers with enhanced dialysate flow distribution. Clin J Am Soc Nephrol. 2011;6:2235–9.CrossRefPubMedPubMedCentral
13.
go back to reference Alayoud A, Benyahia M, Montassir D, Hamzi A, Zajjari Y, Bahadi A, et al. A model to predict optimal dialysate flow. Ther Apher Dial. 2012;16:152–8.CrossRefPubMed Alayoud A, Benyahia M, Montassir D, Hamzi A, Zajjari Y, Bahadi A, et al. A model to predict optimal dialysate flow. Ther Apher Dial. 2012;16:152–8.CrossRefPubMed
14.
go back to reference Singh S, Choi P, Power A, Ashby D, Cairns T, Griffith M, et al. Ten-year patient survival on maintenance haemodialysis: association with treatment time and dialysis dose. J Nephrol. 2013;26:763–70.CrossRefPubMed Singh S, Choi P, Power A, Ashby D, Cairns T, Griffith M, et al. Ten-year patient survival on maintenance haemodialysis: association with treatment time and dialysis dose. J Nephrol. 2013;26:763–70.CrossRefPubMed
15.
go back to reference Tentori F, Zhang J, Li Y, Karaboyas A, Kerr P, Saran R, et al. Longer dialysis session length is associated with better intermediate outcomes and survival among patients on in-center three times per week hemodialysis: results from the Dialysis Outcomes and Practice Patterns Study (DOPPS). Nephrol Dial Transplant. 2012;27:4180–8.CrossRefPubMedPubMedCentral Tentori F, Zhang J, Li Y, Karaboyas A, Kerr P, Saran R, et al. Longer dialysis session length is associated with better intermediate outcomes and survival among patients on in-center three times per week hemodialysis: results from the Dialysis Outcomes and Practice Patterns Study (DOPPS). Nephrol Dial Transplant. 2012;27:4180–8.CrossRefPubMedPubMedCentral
16.
go back to reference Flythe J, Curhan G, Brunelli S. Shorter length dialysis sessions are associated with increased mortality, independent of body weight. Kidney Int. 2013;83:104–13.CrossRefPubMed Flythe J, Curhan G, Brunelli S. Shorter length dialysis sessions are associated with increased mortality, independent of body weight. Kidney Int. 2013;83:104–13.CrossRefPubMed
17.
go back to reference Kim JC, Kim JH, Kimg HC, Kang E, Kim KG, Kim HC, et al. Effect of fiber structure on dialysate flow profile and hollow-fiber hemodialyzer reliability: CT perfusion study. Int J of Art Organs. 2008;31:944–50. Kim JC, Kim JH, Kimg HC, Kang E, Kim KG, Kim HC, et al. Effect of fiber structure on dialysate flow profile and hollow-fiber hemodialyzer reliability: CT perfusion study. Int J of Art Organs. 2008;31:944–50.
18.
go back to reference Yamamoto K, Matsuda M, Hirano A, Takizawa N, Iwashima S, Yakushiji T, et al. Computational evaluation of dialysis fluid flow in dialyzers with variously designed jackets. Artif Organs. 2009;33:481–6.CrossRefPubMed Yamamoto K, Matsuda M, Hirano A, Takizawa N, Iwashima S, Yakushiji T, et al. Computational evaluation of dialysis fluid flow in dialyzers with variously designed jackets. Artif Organs. 2009;33:481–6.CrossRefPubMed
19.
go back to reference Kahiwagi T, Sato K, Kawakami S, Kiyomoto M, Enomoto M, Suzuki T, et al. Effects of reduced dialysis fluid flow in hemodialysis. J Nippon Med Sch. 2013;80:119–30.CrossRef Kahiwagi T, Sato K, Kawakami S, Kiyomoto M, Enomoto M, Suzuki T, et al. Effects of reduced dialysis fluid flow in hemodialysis. J Nippon Med Sch. 2013;80:119–30.CrossRef
20.
go back to reference Chazot C, Jean G. The advantages and challenges of increasing the duration and frequency of maintenance dialysis sessions. Nat Clin Pract Nephrol. 2008;5:34–44.CrossRefPubMed Chazot C, Jean G. The advantages and challenges of increasing the duration and frequency of maintenance dialysis sessions. Nat Clin Pract Nephrol. 2008;5:34–44.CrossRefPubMed
Metadata
Title
Is it useful to increase dialysate flow rate to improve the delivered Kt?
Authors
Marta Albalate
Rafael Pérez-García
Patricia de Sequera
Elena Corchete
Roberto Alcazar
Mayra Ortega
Marta Puerta
Publication date
01-12-2015
Publisher
BioMed Central
Published in
BMC Nephrology / Issue 1/2015
Electronic ISSN: 1471-2369
DOI
https://doi.org/10.1186/s12882-015-0013-9

Other articles of this Issue 1/2015

BMC Nephrology 1/2015 Go to the issue