Open Access 01-12-2021 | Computed Tomography | Technical advance

Metal artifacts in intraoperative O-arm CBCT scans

Authors: Juha I. Peltonen, Touko Kaasalainen, Mika Kortesniemi

Published in: BMC Medical Imaging | Issue 1/2021

Login to get access

Abstract

Background

Cone-beam computed tomography (CBCT) has become an increasingly important medical imaging modality in orthopedic operating rooms. Metal implants and related image artifacts create challenges for image quality optimization in CBCT. The purpose of this study was to develop a robust and quantitative method for the comprehensive determination of metal artifacts in novel CBCT applications.

Methods

The image quality of an O-arm CBCT device was assessed with an anthropomorphic pelvis phantom in the presence of metal implants. Three different kilovoltage and two different exposure settings were used to scan the phantom both with and without the presence of metal rods.

Results

The amount of metal artifact was related to the applied CBCT imaging protocol parameters. The size of the artifact was moderate with all imaging settings. The highest applied kilovoltage and exposure level distinctly increased artifact severity.

Conclusions

The developed method offers a practical and robust way to quantify metal artifacts in CBCT. Changes in imaging parameters may have nonlinear effects on image quality which are not anticipated based on physics.
Literature
1.
go back to reference Holly LT, Foley KT. Three-dimensional fluoroscopy-guided percutaneous thoracolumbar pedicle screw placement. J Neurosurg. 2009;99:324–9. Holly LT, Foley KT. Three-dimensional fluoroscopy-guided percutaneous thoracolumbar pedicle screw placement. J Neurosurg. 2009;99:324–9.
2.
go back to reference Acosta FL Jr, Thompson TL, Campbell S, Weinstein PR, Ames CP. Use of intraoperative isocentric C-arm 3D fluoroscopy for sextant percutaneous pedicle screw placement: case report and review of the literature. Spine J. 2005;5:339–43.CrossRef Acosta FL Jr, Thompson TL, Campbell S, Weinstein PR, Ames CP. Use of intraoperative isocentric C-arm 3D fluoroscopy for sextant percutaneous pedicle screw placement: case report and review of the literature. Spine J. 2005;5:339–43.CrossRef
3.
go back to reference Bledsoe JM, Fenton D, Fogelson JL, Nottmeier EW. Accuracy of upper thoracic pedicle screw placement using three-dimensional image guidance. Spine J. 2009;9:817–21.CrossRef Bledsoe JM, Fenton D, Fogelson JL, Nottmeier EW. Accuracy of upper thoracic pedicle screw placement using three-dimensional image guidance. Spine J. 2009;9:817–21.CrossRef
4.
go back to reference Nottmeier EW, Pirris SM, Edwards S, Kimes S, Bowman C, Nelson KL. Operating room radiation exposure in cone beam computed tomography-based, image-guided spinal surgery. J Neurosurg. 2013;19:226–31. Nottmeier EW, Pirris SM, Edwards S, Kimes S, Bowman C, Nelson KL. Operating room radiation exposure in cone beam computed tomography-based, image-guided spinal surgery. J Neurosurg. 2013;19:226–31.
5.
go back to reference Mason A, Paulsen R, Babuska JM, et al. The accuracy of pedicle screw placement using intraoperative image guidance systems: a systematic review. J Neurosurg. 2014;20:196–203. Mason A, Paulsen R, Babuska JM, et al. The accuracy of pedicle screw placement using intraoperative image guidance systems: a systematic review. J Neurosurg. 2014;20:196–203.
6.
go back to reference Kleck CJ, Cullilmore I, LaFleur M, et al. A new 3-dimensional method for measuring precision in surgical navigation and methods to optimize navigation accuracy. Eur Spine J. 2016;25:1764–74.CrossRef Kleck CJ, Cullilmore I, LaFleur M, et al. A new 3-dimensional method for measuring precision in surgical navigation and methods to optimize navigation accuracy. Eur Spine J. 2016;25:1764–74.CrossRef
7.
go back to reference Dea N, Fisher CG, Batke J, et al. Economic evaluation comparing intraoperative cone beam CT-based navigation and conventional fluoroscopy for the placement of spinal pedicle screws: a patient-level data cost-effectiveness analysis. Spine J. 2016;16:23–31.CrossRef Dea N, Fisher CG, Batke J, et al. Economic evaluation comparing intraoperative cone beam CT-based navigation and conventional fluoroscopy for the placement of spinal pedicle screws: a patient-level data cost-effectiveness analysis. Spine J. 2016;16:23–31.CrossRef
8.
go back to reference Rivkin MA, Yocom SS. Thoracolumbar instrumentation with CT-guided navigation (O-arm) in 270 consecutive patients: accuracy rates and lessons learned. Neurosurg Focus. 2014;36:E7.CrossRef Rivkin MA, Yocom SS. Thoracolumbar instrumentation with CT-guided navigation (O-arm) in 270 consecutive patients: accuracy rates and lessons learned. Neurosurg Focus. 2014;36:E7.CrossRef
9.
go back to reference Ryang YM, Villard J, Obermuller T, et al. Learning curve of 3D fluoroscopy image-guided pedicle screw placement in the thoracolumbar spine. Spine J. 2015;15:467–76.CrossRef Ryang YM, Villard J, Obermuller T, et al. Learning curve of 3D fluoroscopy image-guided pedicle screw placement in the thoracolumbar spine. Spine J. 2015;15:467–76.CrossRef
10.
go back to reference Costa F, Tosi G, Attuat L, et al. Radiation exposure in spine surgery using an image-guided system based on intraoperative cone-beam computed tomography: analysis of 107 consecutive cases. J Neurosurg. 2016;25:654–9. Costa F, Tosi G, Attuat L, et al. Radiation exposure in spine surgery using an image-guided system based on intraoperative cone-beam computed tomography: analysis of 107 consecutive cases. J Neurosurg. 2016;25:654–9.
11.
go back to reference Su AW, Luo TD, McIntosh AL, et al. Switching to a pediatric dose O-arm protocol in spine surgery significantly reduced patient radiation exposure. J Pediatr Orthop. 2016;36:621–6.CrossRef Su AW, Luo TD, McIntosh AL, et al. Switching to a pediatric dose O-arm protocol in spine surgery significantly reduced patient radiation exposure. J Pediatr Orthop. 2016;36:621–6.CrossRef
12.
go back to reference Su AW, McIntosh AL, Schueler BA, et al. How does patient radiation exposure compare with low-dose o-arm versus fluoroscopy for pedicle screw placement in idiopathic scoliosis? J Pediatr Orthops. 2017;37:171–7.CrossRef Su AW, McIntosh AL, Schueler BA, et al. How does patient radiation exposure compare with low-dose o-arm versus fluoroscopy for pedicle screw placement in idiopathic scoliosis? J Pediatr Orthops. 2017;37:171–7.CrossRef
13.
go back to reference Abul-Kasim K, Söderberg M, Selariu E, Gunnarsson M, Kherad M, Ohlin A. Optimization of radiation exposure and image quality of the cone-beam O-arm intraoperative imaging system in spinal surgery. J Spinal Disord Tech. 2012;25:52–8.CrossRef Abul-Kasim K, Söderberg M, Selariu E, Gunnarsson M, Kherad M, Ohlin A. Optimization of radiation exposure and image quality of the cone-beam O-arm intraoperative imaging system in spinal surgery. J Spinal Disord Tech. 2012;25:52–8.CrossRef
14.
go back to reference Schouten R, Lee R, Boyd M, et al. Intra-operative cone-beam CT (O-arm) and stereotactic navigation in acute spinal trauma surgery. J Clin Neurosci. 2012;9:1137–43.CrossRef Schouten R, Lee R, Boyd M, et al. Intra-operative cone-beam CT (O-arm) and stereotactic navigation in acute spinal trauma surgery. J Clin Neurosci. 2012;9:1137–43.CrossRef
15.
go back to reference Qureshi S, Lu Y, McAnany S, Baird E. Three-dimensional intraoperative imaging modalities in orthopaedic surgery: a narrative review. J Am Acad Orthop Surg. 2014;22:800–9.CrossRef Qureshi S, Lu Y, McAnany S, Baird E. Three-dimensional intraoperative imaging modalities in orthopaedic surgery: a narrative review. J Am Acad Orthop Surg. 2014;22:800–9.CrossRef
16.
go back to reference Riis J, Lehman RR, Perera RA, et al. A retrospective comparison of intraoperative CT and fluoroscopy evaluating radiation exposure in posterior spinal fusions for scoliosis. Patient Saf Surg. 2017;11:32.CrossRef Riis J, Lehman RR, Perera RA, et al. A retrospective comparison of intraoperative CT and fluoroscopy evaluating radiation exposure in posterior spinal fusions for scoliosis. Patient Saf Surg. 2017;11:32.CrossRef
17.
go back to reference Boas FE, Fleischmann D. CT artifacts: causes and reduction techniques. Imaging Med. 2012;4:229–40.CrossRef Boas FE, Fleischmann D. CT artifacts: causes and reduction techniques. Imaging Med. 2012;4:229–40.CrossRef
18.
go back to reference Tang X, Krupinski EA, Xie H, Stillman AE. On the data acquisition, image reconstruction, cone beam artifacts and their suppression in axial MDCT and CBCT–A review. Med Phys. 2018;45:e761–82.CrossRef Tang X, Krupinski EA, Xie H, Stillman AE. On the data acquisition, image reconstruction, cone beam artifacts and their suppression in axial MDCT and CBCT–A review. Med Phys. 2018;45:e761–82.CrossRef
19.
go back to reference Katisko JP, Kauppinen MT, Koivukangas JP, Heikkinen ER. Stereotactic operations using the o-arm. Stereotact Funct Neurosurg. 2012;90:401–9.CrossRef Katisko JP, Kauppinen MT, Koivukangas JP, Heikkinen ER. Stereotactic operations using the o-arm. Stereotact Funct Neurosurg. 2012;90:401–9.CrossRef
20.
go back to reference Sharma M, Deogaonkar M. Accuracy and safety of targeting using intraoperative “O-arm” during placement of deep brain stimulation electrodes without electrophysiological recordings. J Clin Neurosc. 2016;27:80–6.CrossRef Sharma M, Deogaonkar M. Accuracy and safety of targeting using intraoperative “O-arm” during placement of deep brain stimulation electrodes without electrophysiological recordings. J Clin Neurosc. 2016;27:80–6.CrossRef
22.
go back to reference International Electrotechnical Commission, Medical electrical equipment - Part 2–44: Particular requirements for the basic safety and essential performance of X-ray equipment for computed tomography, IEC 60601—2-–44:2009+AMD1:2012+AMD2:2016 CSV Consolidated version, Edition 3.2, IEC, Geneva, Switzerland, 2016, 276 pages International Electrotechnical Commission, Medical electrical equipment - Part 2–44: Particular requirements for the basic safety and essential performance of X-ray equipment for computed tomography, IEC 60601—2-–44:2009+AMD1:2012+AMD2:2016 CSV Consolidated version, Edition 3.2, IEC, Geneva, Switzerland, 2016, 276 pages
23.
go back to reference Pieper S, Halle M, Kikinis R. 3D Slicer. Proceedings from the 2004 2nd IEEE International Symposium on Biomedical Imaging. Macro to. NANO. 2004;2:632–5. Pieper S, Halle M, Kikinis R. 3D Slicer. Proceedings from the 2004 2nd IEEE International Symposium on Biomedical Imaging. Macro to. NANO. 2004;2:632–5.
Metadata
Title
Metal artifacts in intraoperative O-arm CBCT scans
Authors
Juha I. Peltonen
Touko Kaasalainen
Mika Kortesniemi
Publication date
01-12-2021
Publisher
BioMed Central
Published in
BMC Medical Imaging / Issue 1/2021
Electronic ISSN: 1471-2342
DOI
https://doi.org/10.1186/s12880-020-00538-4

Other articles of this Issue 1/2021

BMC Medical Imaging 1/2021 Go to the issue