Distal radius plate of CFR-PEEK has minimal effect compared to titanium plates on bone parameters in high-resolution peripheral quantitative computed tomography: a pilot study

Authors: Joost J. A. de Jong, Arno Lataster, Bert van Rietbergen, Jacobus J. Arts, Piet P. Geusens, Joop P. W. van den Bergh, Paul C. Willems

Published in: BMC Medical Imaging | Issue 1/2017

Login to get access

Abstract

Background

Carbon-fiber-reinforced poly-ether-ether-ketone (CFR-PEEK) has superior radiolucency compared to other orthopedic implant materials, e.g. titanium or stainless steel, thus allowing metal-artifact-free postoperative monitoring by computed tomography (CT). Recently, high-resolution peripheral quantitative CT (HRpQCT) proved to be a promising technique to monitor the recovery of volumetric bone mineral density (vBMD), micro-architecture and biomechanical parameters in stable conservatively treated distal radius fractures. When using HRpQCT to monitor unstable distal radius fractures that require volar distal radius plating for fixation, radiolucent CFR-PEEK plates may be a better alternative to currently used titanium plates to allow for reliable assessment. In this pilot study, we assessed the effect of a volar distal radius plate made from CFR-PEEK on bone parameters obtained from HRpQCT in comparison to two titanium plates.

Methods

Plates were instrumented in separate cadaveric human fore-arms (n = 3). After instrumentation and after removal of the plates duplicate HRpQCT scans were made of the region covered by the plate. HRpQCT images were visually checked for artifacts. vBMD, micro-architectural and biomechanical parameters were calculated, and compared between the uninstrumented and instrumented radii.

Results

No visible image artifacts were observed in the CFR-PEEK plate instrumented radius, and errors in bone parameters ranged from −3.2 to 2.6%. In the radii instrumented with the titanium plates, severe image artifacts were observed and errors in bone parameters ranged between −30.2 and 67.0%.

Conclusions

We recommend using CFR-PEEK plates in longitudinal in vivo studies that monitor the healing process of unstable distal radius fractures treated operatively by plating or bone graft ingrowth.
Literature
1.
go back to reference Zimel MN, Hwang S, Riedel ER, Healey JH. Carbon fiber intramedullary nails reduce artifact in postoperative advanced imaging. Skeletal Radiol. 2015;44:1317–25.CrossRefPubMed Zimel MN, Hwang S, Riedel ER, Healey JH. Carbon fiber intramedullary nails reduce artifact in postoperative advanced imaging. Skeletal Radiol. 2015;44:1317–25.CrossRefPubMed
2.
go back to reference Heary RF, Kheterpal A, Mammis A, Kumar S. Stackable carbon fiber cages for thoracolumbar interbody fusion after corpectomy: long-term outcome analysis. Neurosurgery. 2011;68:810–8.CrossRefPubMed Heary RF, Kheterpal A, Mammis A, Kumar S. Stackable carbon fiber cages for thoracolumbar interbody fusion after corpectomy: long-term outcome analysis. Neurosurgery. 2011;68:810–8.CrossRefPubMed
3.
go back to reference Pace N, Marinelli M, Spurio S. Technical and histologic analysis of a retrieved carbon fiber-reinforced poly-ether-ether-ketone composite alumina-bearing liner 28 months after implantation. J Arthroplasty. 2008;23:151–5.CrossRefPubMed Pace N, Marinelli M, Spurio S. Technical and histologic analysis of a retrieved carbon fiber-reinforced poly-ether-ether-ketone composite alumina-bearing liner 28 months after implantation. J Arthroplasty. 2008;23:151–5.CrossRefPubMed
4.
go back to reference Nakahara I, Takao M, Bandoh S, Bertollo N, Walsh WR, Sugano N. In vivo implant fixation of carbon fiber-reinforced PEEK hip prostheses in an ovine model. J Orthop Res. 2013;31:485–92.CrossRefPubMed Nakahara I, Takao M, Bandoh S, Bertollo N, Walsh WR, Sugano N. In vivo implant fixation of carbon fiber-reinforced PEEK hip prostheses in an ovine model. J Orthop Res. 2013;31:485–92.CrossRefPubMed
5.
go back to reference Zimel MN, Farfalli GL, Zindman AM, Riedel ER, Morris CD, Boland PJ, et al. Revision distal femoral arthroplasty with the compress((R)) prosthesis has a low rate of mechanical failure at 10 years. Clin Orthop Relat Res. 2016;474:528–36.CrossRefPubMed Zimel MN, Farfalli GL, Zindman AM, Riedel ER, Morris CD, Boland PJ, et al. Revision distal femoral arthroplasty with the compress((R)) prosthesis has a low rate of mechanical failure at 10 years. Clin Orthop Relat Res. 2016;474:528–36.CrossRefPubMed
6.
go back to reference de Jong JJ, Heyer FL, Arts JJ, Poeze M, Keszei AP, Willems PC, et al. Fracture repair in the distal radius in post-menopausal women: a follow-up two years post-fracture using HRpQCT. J Bone Miner Res. 2016;31:1114–22.CrossRefPubMed de Jong JJ, Heyer FL, Arts JJ, Poeze M, Keszei AP, Willems PC, et al. Fracture repair in the distal radius in post-menopausal women: a follow-up two years post-fracture using HRpQCT. J Bone Miner Res. 2016;31:1114–22.CrossRefPubMed
7.
go back to reference de Jong JJ, Willems PC, Arts JJ, Bours SG, Brink PR, van Geel TA, et al. Assessment of the healing process in distal radius fractures by high resolution peripheral quantitative computed tomography. Bone. 2014;64C:65–74.CrossRef de Jong JJ, Willems PC, Arts JJ, Bours SG, Brink PR, van Geel TA, et al. Assessment of the healing process in distal radius fractures by high resolution peripheral quantitative computed tomography. Bone. 2014;64C:65–74.CrossRef
8.
go back to reference de Jong JJ, Arts JJ, Meyer U, Willems PC, Geusens PP, van den Bergh JP, et al. Effect of a cast on short-term reproducibility and bone parameters obtained from HR-pQCT measurements at the distal end of the radius. J Bone Joint Surg Am. 2016;98:356–62.CrossRefPubMed de Jong JJ, Arts JJ, Meyer U, Willems PC, Geusens PP, van den Bergh JP, et al. Effect of a cast on short-term reproducibility and bone parameters obtained from HR-pQCT measurements at the distal end of the radius. J Bone Joint Surg Am. 2016;98:356–62.CrossRefPubMed
9.
go back to reference Hammert WC, Kramer RC, Graham B, Keith MW. AAOS appropriate use criteria: treatment of distal radius fractures. J Am Acad Orthop Surg. 2013;21:506–9.CrossRefPubMed Hammert WC, Kramer RC, Graham B, Keith MW. AAOS appropriate use criteria: treatment of distal radius fractures. J Am Acad Orthop Surg. 2013;21:506–9.CrossRefPubMed
10.
go back to reference Mueller TL, Wirth AJ, van Lenthe GH, Goldhahn J, Schense J, Jamieson V, et al. Mechanical stability in a human radius fracture treated with a novel tissue-engineered bone substitute: a non-invasive, longitudinal assessment using high-resolution pQCT in combination with finite element analysis. J Tissue Eng Regen Med. 2011;5:415–20.CrossRefPubMed Mueller TL, Wirth AJ, van Lenthe GH, Goldhahn J, Schense J, Jamieson V, et al. Mechanical stability in a human radius fracture treated with a novel tissue-engineered bone substitute: a non-invasive, longitudinal assessment using high-resolution pQCT in combination with finite element analysis. J Tissue Eng Regen Med. 2011;5:415–20.CrossRefPubMed
11.
go back to reference Boas FE, Fleischmann D. Evaluation of two iterative techniques for reducing metal artifacts in computed tomography. Radiology. 2011;259:894–902.CrossRefPubMed Boas FE, Fleischmann D. Evaluation of two iterative techniques for reducing metal artifacts in computed tomography. Radiology. 2011;259:894–902.CrossRefPubMed
12.
go back to reference Boutroy S, Bouxsein ML, Munoz F, Delmas PD. In vivo assessment of trabecular bone microarchitecture by high-resolution peripheral quantitative computed tomography. J Clin Endocrinol Metab. 2005;90:6508–15.CrossRefPubMed Boutroy S, Bouxsein ML, Munoz F, Delmas PD. In vivo assessment of trabecular bone microarchitecture by high-resolution peripheral quantitative computed tomography. J Clin Endocrinol Metab. 2005;90:6508–15.CrossRefPubMed
13.
go back to reference Pistoia W, van Rietbergen B, Lochmuller EM, Lill CA, Eckstein F, Ruegsegger P. Estimation of distal radius failure load with micro-finite element analysis models based on three-dimensional peripheral quantitative computed tomography images. Bone. 2002;30:842–8.CrossRefPubMed Pistoia W, van Rietbergen B, Lochmuller EM, Lill CA, Eckstein F, Ruegsegger P. Estimation of distal radius failure load with micro-finite element analysis models based on three-dimensional peripheral quantitative computed tomography images. Bone. 2002;30:842–8.CrossRefPubMed
14.
go back to reference Pistoia W, van Rietbergen B, Lochmuller EM, Lill CA, Eckstein F, Ruegsegger P. Image-based micro-finite-element modeling for improved distal radius strength diagnosis: moving from bench to bedside. J Clin Densitom. 2004;7:153–60.CrossRefPubMed Pistoia W, van Rietbergen B, Lochmuller EM, Lill CA, Eckstein F, Ruegsegger P. Image-based micro-finite-element modeling for improved distal radius strength diagnosis: moving from bench to bedside. J Clin Densitom. 2004;7:153–60.CrossRefPubMed
15.
go back to reference Gluer CC, Blake G, Lu Y, Blunt BA, Jergas M, Genant HK. Accurate assessment of precision errors: how to measure the reproducibility of bone densitometry techniques. Osteoporos Int. 1995;5:262–70.CrossRefPubMed Gluer CC, Blake G, Lu Y, Blunt BA, Jergas M, Genant HK. Accurate assessment of precision errors: how to measure the reproducibility of bone densitometry techniques. Osteoporos Int. 1995;5:262–70.CrossRefPubMed
16.
go back to reference MacNeil JA, Boyd SK. Improved reproducibility of high-resolution peripheral quantitative computed tomography for measurement of bone quality. Med Eng Phys. 2008;30:792–9.CrossRefPubMed MacNeil JA, Boyd SK. Improved reproducibility of high-resolution peripheral quantitative computed tomography for measurement of bone quality. Med Eng Phys. 2008;30:792–9.CrossRefPubMed
17.
go back to reference Knott PT, Mardjetko SM, Kim RH, Cotter TM, Dunn MM, Patel ST, et al. A comparison of magnetic and radiographic imaging artifact after using three types of metal rods: stainless steel, titanium, and vitallium. Spine J. 2010;10:789–94.CrossRefPubMed Knott PT, Mardjetko SM, Kim RH, Cotter TM, Dunn MM, Patel ST, et al. A comparison of magnetic and radiographic imaging artifact after using three types of metal rods: stainless steel, titanium, and vitallium. Spine J. 2010;10:789–94.CrossRefPubMed
Metadata
Title
Distal radius plate of CFR-PEEK has minimal effect compared to titanium plates on bone parameters in high-resolution peripheral quantitative computed tomography: a pilot study
Authors
Joost J. A. de Jong
Arno Lataster
Bert van Rietbergen
Jacobus J. Arts
Piet P. Geusens
Joop P. W. van den Bergh
Paul C. Willems
Publication date
01-12-2017
Publisher
BioMed Central
Published in
BMC Medical Imaging / Issue 1/2017
Electronic ISSN: 1471-2342
DOI
https://doi.org/10.1186/s12880-017-0190-z

Other articles of this Issue 1/2017

BMC Medical Imaging 1/2017 Go to the issue