Skip to main content
Top
Published in: BMC Anesthesiology 1/2017

Open Access 01-12-2017 | Research article

Role of tube size and intranasal compression of the nasotracheal tube in respiratory pressure loss during nasotracheal intubation: a laboratory study

Authors: Koichi Futagawa, Yoshihiro Takasugi, Takeharu Kobayashi, Satoshi Morishita, Takahiko Okuda

Published in: BMC Anesthesiology | Issue 1/2017

Login to get access

Abstract

Background

Small nasotracheal tubes (NTTs) and intranasal compression of the NTT in the nasal cavity may contribute to increasing airway resistance. Since the effects of size, shape, and partial compression of the NTT on airway resistance have not been investigated, values of airway resistance with partial compression of preformed NTTs of various sizes were determined.

Methods

To determine the factors affecting the respiratory pressure loss during the nasotracheal intubation, physical and fluid dynamics simulations were used. The internal minor axes of NTTs in the nasal cavity of intubated patients were measured using dial calipers. In physical and fluid dynamics simulations, pressure losses through the tubular parts, compressed parts, and slip joints of NTTs with internal diameters (IDs) of 6.0, 6.5, 7.0, 7.5, and 8.0 mm were estimated under partial compression.

Results

The median internal minor axes of the 7.0- and 7.5-mm ID NTTs in the nasal cavity were 5.2 (4.3–5.6) mm and 6.0 (4.2–7.0) mm, respectively. With a volumetric air flow rate of 30 L/min, pressure losses through uncompressed NTTs with IDs of 6.0-, 6.5-, 7.0-, 7.5- and 8.0-mm were 651.6 ± 5.7 (6.64 ± 0.06), 453.4 ± 3.9 (4.62 ± 0.04), 336.5 ± 2.2 (3.43 ± 0.02), 225.2 ± 0.2 (2.30 ± 0.00), and 179.0 ± 1.1 Pa (1.82 ± 0.01 cmH2O), respectively; the pressure losses through the slip joints were 220.3 (2.25), 131.1 (1.33), 86.8 (0.88), 57.1 (0.58), and 36.1 Pa (0.37 cmH2O), respectively; and the pressure losses through the curvature of the NTT were 71.6 (0.73), 69.0 (0.70), 64.8 (0.66), 32.5 (0.33), and 41.6 Pa (0.42 cmH2O), respectively. A maximum compression force of 34.1 N increased the pressure losses by 82.0 (0.84), 38.0 (0.39), 23.5 (0.24), 16.6 (0.17), and 9.3 Pa (0.09 cmH2O), respectively.

Conclusion

Pressure losses through NTTs are in inverse proportion to the tubes’ IDs; greater pressure losses due to slip joints, acute bending, and partial compression of the NTT were obvious in small NTTs. Pressure losses through NTTs, especially in small NTTs, could increase the work of breathing to a greater extent than that through standard tubes; intranasal compression further increases the pressure loss.
Appendix
Available only for authorised users
Literature
1.
go back to reference Shandro J. Resistance to gas flow in the "new" anaesthesia circuits: a comparative study. Can Anaesth Soc J. 1982;29:387–90.CrossRefPubMed Shandro J. Resistance to gas flow in the "new" anaesthesia circuits: a comparative study. Can Anaesth Soc J. 1982;29:387–90.CrossRefPubMed
2.
go back to reference Takasugi Y, Futagawa K, Kazuhara K, Morishita S, Okuda T. Roles of endotracheal tubes and slip joints in respiratory pressure loss: a laboratory study. J Anesth. 2016;30:789–95.CrossRefPubMed Takasugi Y, Futagawa K, Kazuhara K, Morishita S, Okuda T. Roles of endotracheal tubes and slip joints in respiratory pressure loss: a laboratory study. J Anesth. 2016;30:789–95.CrossRefPubMed
3.
go back to reference Takasugi Y, Iwamoto T, Uehara K, Shiba M, Koga Y. Compression of a nasotracheal tube in the nasal cavity of a patient with craniofacial fibrous dysplasia (in Japanese with English abstract). Masui (Jpn J Anesthesiol). 2008;57:742–4. Takasugi Y, Iwamoto T, Uehara K, Shiba M, Koga Y. Compression of a nasotracheal tube in the nasal cavity of a patient with craniofacial fibrous dysplasia (in Japanese with English abstract). Masui (Jpn J Anesthesiol). 2008;57:742–4.
4.
go back to reference Tanaka K, Konoshita M, Takahashi H. Collapse of the nasal endotracheal tube in the nasal cavity (in Japanese with English abstract). Masui (Jpn J Anesthesiol). 1997;46:1615–7. Tanaka K, Konoshita M, Takahashi H. Collapse of the nasal endotracheal tube in the nasal cavity (in Japanese with English abstract). Masui (Jpn J Anesthesiol). 1997;46:1615–7.
5.
go back to reference Maeda Y, Fujino Y, Uchiyama A, Taenaka N, Mashimo T, Nishimura M. Does the tube-compensation function of two modern mechanical ventilators provide effective work of breathing relief? Crit Care. 2003;7:R92–7.CrossRefPubMedPubMedCentral Maeda Y, Fujino Y, Uchiyama A, Taenaka N, Mashimo T, Nishimura M. Does the tube-compensation function of two modern mechanical ventilators provide effective work of breathing relief? Crit Care. 2003;7:R92–7.CrossRefPubMedPubMedCentral
6.
go back to reference Mols G, von Ungern-Sternberg B, Rohr E, Haberthür C, Geiger K, Guttmann J. Respiratory comfort and breathing pattern during volume proportional assist ventilation and pressure support ventilation: a study on volunteers with artificially reduced compliance. Crit Care Med. 2000;28:1940–6.CrossRefPubMed Mols G, von Ungern-Sternberg B, Rohr E, Haberthür C, Geiger K, Guttmann J. Respiratory comfort and breathing pattern during volume proportional assist ventilation and pressure support ventilation: a study on volunteers with artificially reduced compliance. Crit Care Med. 2000;28:1940–6.CrossRefPubMed
7.
go back to reference Brochard L, Rua F, Lorino H, Lemaire F, Harf A. Inspiratory pressure support compensates for the additional work of breathing caused by the endotracheal tube. Anesthesiology. 1991;75:739–45.CrossRefPubMed Brochard L, Rua F, Lorino H, Lemaire F, Harf A. Inspiratory pressure support compensates for the additional work of breathing caused by the endotracheal tube. Anesthesiology. 1991;75:739–45.CrossRefPubMed
8.
go back to reference Bersten AD, Rutten AJ, Vedig AE, Skowronski GA. Additional work of breathing imposed by endotracheal tubes, breathing circuits, and intensive care ventilators. Crit Care Med. 1989;17:671–7.CrossRefPubMed Bersten AD, Rutten AJ, Vedig AE, Skowronski GA. Additional work of breathing imposed by endotracheal tubes, breathing circuits, and intensive care ventilators. Crit Care Med. 1989;17:671–7.CrossRefPubMed
9.
go back to reference Hagberg CA, Artime CA. Chapter 55 airway Management in the Adult. In: Miller RD, editor. Miller's anesthesia. 8th ed. New York: Elsevier; 2014. p. P1647–83. Hagberg CA, Artime CA. Chapter 55 airway Management in the Adult. In: Miller RD, editor. Miller's anesthesia. 8th ed. New York: Elsevier; 2014. p. P1647–83.
10.
go back to reference Coe TR, Human M. The peri-operative complications of nasal intubation: a comparison of nostril side. Anaesthesia. 2001;56:447–50.CrossRefPubMed Coe TR, Human M. The peri-operative complications of nasal intubation: a comparison of nostril side. Anaesthesia. 2001;56:447–50.CrossRefPubMed
11.
go back to reference Lee JH, Kim CH, Bahk JH, Park KS. The influence of endotracheal tube tip design on nasal trauma during nasotracheal intubation: magill-tip versus murphy-tip. Anesth Analg. 2005;101:1226–9.CrossRefPubMed Lee JH, Kim CH, Bahk JH, Park KS. The influence of endotracheal tube tip design on nasal trauma during nasotracheal intubation: magill-tip versus murphy-tip. Anesth Analg. 2005;101:1226–9.CrossRefPubMed
12.
go back to reference Kim YC, Lee SH, Noh GJ, Cho SY, Yeom JH, Shin WJ, Lee DH, Ryu JS, Park YS, Cha KJ, Lee SC. Thermosoftening treatment of the nasotracheal tube before intubation can reduce epistaxis and nasal damage. Anesth Analg. 2000;91:698–701.CrossRefPubMed Kim YC, Lee SH, Noh GJ, Cho SY, Yeom JH, Shin WJ, Lee DH, Ryu JS, Park YS, Cha KJ, Lee SC. Thermosoftening treatment of the nasotracheal tube before intubation can reduce epistaxis and nasal damage. Anesth Analg. 2000;91:698–701.CrossRefPubMed
13.
go back to reference Lim CW, Min SW, Kim CS, Chang JE, Park JE, Hwang JY. The use of a nasogastric tube to facilitate nasotracheal intubation: a randomised controlled trial. Anaesthesia. 2014;69:591–7.CrossRefPubMed Lim CW, Min SW, Kim CS, Chang JE, Park JE, Hwang JY. The use of a nasogastric tube to facilitate nasotracheal intubation: a randomised controlled trial. Anaesthesia. 2014;69:591–7.CrossRefPubMed
14.
go back to reference Lu PP, Liu HP, Shyr MH, et al. Softened endothracheal tube reduces the incidence and severity of epistaxis following nasotracheal intubation. Acta Anaesthesiol Sin. 1998;36:193–7.PubMed Lu PP, Liu HP, Shyr MH, et al. Softened endothracheal tube reduces the incidence and severity of epistaxis following nasotracheal intubation. Acta Anaesthesiol Sin. 1998;36:193–7.PubMed
15.
go back to reference Hosseinzadeh H, Taheri Talesh K, Golzari SE, Gholizadeh H, Lotfi A, Hosseinzadeh P. Warming Endotracheal tube in blind Nasotracheal intubation throughout maxillofacial surgeries. J Cardiovasc Thorac Res. 2013;5:147–51.PubMedPubMedCentral Hosseinzadeh H, Taheri Talesh K, Golzari SE, Gholizadeh H, Lotfi A, Hosseinzadeh P. Warming Endotracheal tube in blind Nasotracheal intubation throughout maxillofacial surgeries. J Cardiovasc Thorac Res. 2013;5:147–51.PubMedPubMedCentral
16.
go back to reference Manczur T, Greenough A, Nicholson GP, Rafferty GF. Resistance of pediatric and neonatal endotracheal tubes: influence of flow rate, size, and shape. Crit Care Med. 2000;28:1595–8.CrossRefPubMed Manczur T, Greenough A, Nicholson GP, Rafferty GF. Resistance of pediatric and neonatal endotracheal tubes: influence of flow rate, size, and shape. Crit Care Med. 2000;28:1595–8.CrossRefPubMed
17.
go back to reference Jarreau PH, Louis B, Dassieu G, Desfrere L, Blanchard PW, Moriette G, Isabey D, Harf A. Estimation of inspiratory pressure drop in neonatal and pediatric endotracheal tubes. J Appl Physiol. 1999;87:36–46.PubMed Jarreau PH, Louis B, Dassieu G, Desfrere L, Blanchard PW, Moriette G, Isabey D, Harf A. Estimation of inspiratory pressure drop in neonatal and pediatric endotracheal tubes. J Appl Physiol. 1999;87:36–46.PubMed
18.
go back to reference Lofaso F, Louis B, Brochard L, Harf A, Isabey D. Use of the Blasius resistance formula to estimate the effective diameter of endotracheal tubes. Am Rev Respir Dis. 1992;146:974–9.CrossRefPubMed Lofaso F, Louis B, Brochard L, Harf A, Isabey D. Use of the Blasius resistance formula to estimate the effective diameter of endotracheal tubes. Am Rev Respir Dis. 1992;146:974–9.CrossRefPubMed
19.
go back to reference McGovern J. Technical Note: Friction Diagrams for Pipe Flow. Dublin Institute of Technology, 2011–10-13, 1–15. McGovern J. Technical Note: Friction Diagrams for Pipe Flow. Dublin Institute of Technology, 2011–10-13, 1–15.
20.
go back to reference Mott RL, Untener JA. Chapter 8 Reynolds number, laminar flow, turbulent flow, and energy losses due to friction. In: Mott RL, Untener JA, editors. Applied fluid mechanics. 7th ed. England: Pearson Education Limited; 2016. p. 194–220. Mott RL, Untener JA. Chapter 8 Reynolds number, laminar flow, turbulent flow, and energy losses due to friction. In: Mott RL, Untener JA, editors. Applied fluid mechanics. 7th ed. England: Pearson Education Limited; 2016. p. 194–220.
21.
go back to reference Guttmann J, Eberhard L, Fabry B, Bertschmann W, Wolff G. Continuous calculation of intratracheal pressure in tracheally intubated patients. Anesthesiology. 1993;79:503–13.CrossRefPubMed Guttmann J, Eberhard L, Fabry B, Bertschmann W, Wolff G. Continuous calculation of intratracheal pressure in tracheally intubated patients. Anesthesiology. 1993;79:503–13.CrossRefPubMed
22.
go back to reference Guttmann J, Kessler V, Mols G, Hentschel R, Haberthür C, Geiger K. Continuous calculation of intratracheal pressure in the presence of pediatric endotracheal tubes. Crit Care Med. 2000;28:1018–26.CrossRefPubMed Guttmann J, Kessler V, Mols G, Hentschel R, Haberthür C, Geiger K. Continuous calculation of intratracheal pressure in the presence of pediatric endotracheal tubes. Crit Care Med. 2000;28:1018–26.CrossRefPubMed
23.
go back to reference Hentschel R, Buntzel J, Guttmann J, Schumann S. Endotracheal tube resistance and inertance in a model of mechanical ventilation of newborns and small infants-the impact of ventilator settings on tracheal pressure swings. Physiol Meas. 2011;32:1439–51.CrossRefPubMed Hentschel R, Buntzel J, Guttmann J, Schumann S. Endotracheal tube resistance and inertance in a model of mechanical ventilation of newborns and small infants-the impact of ventilator settings on tracheal pressure swings. Physiol Meas. 2011;32:1439–51.CrossRefPubMed
24.
go back to reference Flevari AG, Maniatis N, Kremiotis TE, Siempos I, Betrosian AP, Roussos C, Douzinas E, Armaganidis A. Rohrer's constant, K2, as a factor of determining inspiratory resistance of common adult endotracheal tubes. Anaesth Intensive Care. 2011;39:410–7.PubMed Flevari AG, Maniatis N, Kremiotis TE, Siempos I, Betrosian AP, Roussos C, Douzinas E, Armaganidis A. Rohrer's constant, K2, as a factor of determining inspiratory resistance of common adult endotracheal tubes. Anaesth Intensive Care. 2011;39:410–7.PubMed
25.
go back to reference Mott RL, Untener JA. Chapter 10 minor losses. In: Mott RL, Untener JA, editors. Applied fluid mechanics. 7th ed. England: Pearson Education Limited; 2016. p. 241–79. Mott RL, Untener JA. Chapter 10 minor losses. In: Mott RL, Untener JA, editors. Applied fluid mechanics. 7th ed. England: Pearson Education Limited; 2016. p. 241–79.
26.
go back to reference Satish G, Kumar KA, Prasad VV, Pasha SM. Comparison of flow analysis of sudden and gradual change of pipe diameter using fluent software. IJRET. 2013;2:41–5. Satish G, Kumar KA, Prasad VV, Pasha SM. Comparison of flow analysis of sudden and gradual change of pipe diameter using fluent software. IJRET. 2013;2:41–5.
27.
go back to reference Lichtwarck-Aschoff M, Helmer A, Kawati R, Lattuada M, Sjöstrand UH, Zügel N, Guttman J, Hed enstierna G. Good short-term agreement between measured and calculated tracheal pressure. Brit J Anaesth. 2003;91:239–48.CrossRefPubMed Lichtwarck-Aschoff M, Helmer A, Kawati R, Lattuada M, Sjöstrand UH, Zügel N, Guttman J, Hed enstierna G. Good short-term agreement between measured and calculated tracheal pressure. Brit J Anaesth. 2003;91:239–48.CrossRefPubMed
28.
go back to reference Katz IM, Martin AR, Feng CH, Majoral C, Caillibotte G, MarxT BJE, Daviet C. Airway pressure distribution during xenon anesthesia: the insufflation phase at constant flow (volume controlled mode). Appl Cardiopulm Pathophysiol. 2012;16:5–16. Katz IM, Martin AR, Feng CH, Majoral C, Caillibotte G, MarxT BJE, Daviet C. Airway pressure distribution during xenon anesthesia: the insufflation phase at constant flow (volume controlled mode). Appl Cardiopulm Pathophysiol. 2012;16:5–16.
29.
go back to reference Mazumder QH. CFD analysis of the effect of elbow radius on pressure drop in multiphase flow. Modelling and Simulation in Engineering. 2012;2012), Article ID 125405:8.CrossRef Mazumder QH. CFD analysis of the effect of elbow radius on pressure drop in multiphase flow. Modelling and Simulation in Engineering. 2012;2012), Article ID 125405:8.CrossRef
30.
go back to reference Kamat SS, Muralidhar K. Chapter 28 volume controlled ventilation. In: Kamat SS, Muralidhar K, editors. Practical applications of mechanical ventilation. 2nd ed. New Delhi: Jaypee Brothers Medical Pub lishers; 2016. p. 372–85. Kamat SS, Muralidhar K. Chapter 28 volume controlled ventilation. In: Kamat SS, Muralidhar K, editors. Practical applications of mechanical ventilation. 2nd ed. New Delhi: Jaypee Brothers Medical Pub lishers; 2016. p. 372–85.
31.
go back to reference Brooks PJ. Back to basics: duct design fundamentals. ASHRAE J. 2010;52:30–8. Brooks PJ. Back to basics: duct design fundamentals. ASHRAE J. 2010;52:30–8.
32.
go back to reference Swamee P, Jain A. Explicit equations for pipe-flow problems. J Hydraul Div ASCE. 1976;102:657–64. Swamee P, Jain A. Explicit equations for pipe-flow problems. J Hydraul Div ASCE. 1976;102:657–64.
Metadata
Title
Role of tube size and intranasal compression of the nasotracheal tube in respiratory pressure loss during nasotracheal intubation: a laboratory study
Authors
Koichi Futagawa
Yoshihiro Takasugi
Takeharu Kobayashi
Satoshi Morishita
Takahiko Okuda
Publication date
01-12-2017
Publisher
BioMed Central
Published in
BMC Anesthesiology / Issue 1/2017
Electronic ISSN: 1471-2253
DOI
https://doi.org/10.1186/s12871-017-0432-1

Other articles of this Issue 1/2017