Published in: BMC Anesthesiology 1/2017

Open Access 01-12-2017 | Research article

Pediatric supraglottic airway devices in clinical practice: A prospective observational study

Authors: Maren Kleine-Brueggeney, Anne Gottfried, Sabine Nabecker, Robert Greif, Malte Book, Lorenz Theiler

Published in: BMC Anesthesiology | Issue 1/2017

Login to get access

Abstract

Background

Supraglottic airway devices (SGA) are commonly used in pediatric anesthesia and serve as primary or back-up devices for difficult airway management. Most SGA are marketed without proper clinical evaluation. The purpose of this study was to evaluate the performance of the pediatric LMA Supreme™, Air-Q® and Ambu® Aura-i™.

Methods

This prospective observational study was performed at Bern University Hospital, Switzerland. With ethics committee approval and a waiver for written informed consent 240 children undergoing elective surgery with an ASA class I-III and a weight of 5-30 kg were included. Three different pediatric supraglottic airway devices were assessed: The LMA Supreme™, Air-Q® and Ambu® Aura-i™. Primary outcome parameter was airway leak pressure. Secondary outcome parameters included first attempt and overall success rate, insertion time, fiberoptic view through the SGA, and adverse events. The primary hypothesis was that the mean airway leak pressure of each tested SGA was 20 cmH2O ± 10%.

Results

None of the SGA showed a mean airway leak pressure of 20 cmH2O ± 10%, but mean airway leak pressures differed significantly between devices [LMA Supreme™ 18.0 (3.4) cmH2O, Air-Q® 15.9 (3.2) cmH2O, Ambu® Aura-i™ 17.3 (3.7) cmH2O, p < 0.001]. First attempt success rates (LMA Supreme™ 100%, Air-Q® 90%, Ambu® Aura-i™ 91%, p = 0.02) and overall success rates (LMA Supreme™ 100%, Air-Q® 91%, Ambu® Aura-i™ 95%, p = 0.02) also differed significantly. Insertion times ranged from 20 (7) seconds (Air-Q®) to 24 (6) seconds (LMA Supreme™, <p = 0.005). Insertion was rated easiest with the LMA Supreme™ (very easy in 97% vs. Air-Q® 70%, Ambu® Aura-i™ 72%, p < 0.001). Fiberoptic view was similar between the SGA. Adverse events were rare.

Conclusions

Airway leak pressures ranged from 16 to 18 cmH2O, enabling positive pressure ventilation with all successful SGA. The highest success rates were achieved by the LMA Supreme™, which was also rated easiest to insert.

Trials Registration

ClinicalTrials.​gov, identifier NCT01625858. Registered 31 May 2012.
Literature
1.
go back to reference Bradley AE, White MC, Engelhardt T, Bayley G, Beringer RM. Current UK practice of pediatric supraglottic airway devices - a survey of members of the Association of Paediatric Anaesthetists of Great Britain and Ireland. Paediatr Anaesth. 2013;23:1006–9.CrossRefPubMed Bradley AE, White MC, Engelhardt T, Bayley G, Beringer RM. Current UK practice of pediatric supraglottic airway devices - a survey of members of the Association of Paediatric Anaesthetists of Great Britain and Ireland. Paediatr Anaesth. 2013;23:1006–9.CrossRefPubMed
2.
go back to reference Luce V, Harkouk H, Brasher C, Michelet D, Hilly J, Maesani M, et al. Supraglottic airway devices vs tracheal intubation in children: a quantitative meta-analysis of respiratory complications. Paediatr Anaesth. 2014;24:1088–98.CrossRefPubMed Luce V, Harkouk H, Brasher C, Michelet D, Hilly J, Maesani M, et al. Supraglottic airway devices vs tracheal intubation in children: a quantitative meta-analysis of respiratory complications. Paediatr Anaesth. 2014;24:1088–98.CrossRefPubMed
3.
go back to reference Acquaviva MA, Horn ND, Gupta SK. Endotracheal Intubation Versus Laryngeal Mask Airway for Esophagogastroduodenoscopy in Children. J Pediatr Gastroenterol Nutr. 2014;59:54–6.CrossRefPubMed Acquaviva MA, Horn ND, Gupta SK. Endotracheal Intubation Versus Laryngeal Mask Airway for Esophagogastroduodenoscopy in Children. J Pediatr Gastroenterol Nutr. 2014;59:54–6.CrossRefPubMed
4.
go back to reference Zhao N, Deng F, Yu C. Anesthesia for pediatric day-case dental surgery: a study comparing the classic laryngeal mask airway with nasal trachea intubation. J Craniofac Surg. 2014;25:e245–8.CrossRefPubMed Zhao N, Deng F, Yu C. Anesthesia for pediatric day-case dental surgery: a study comparing the classic laryngeal mask airway with nasal trachea intubation. J Craniofac Surg. 2014;25:e245–8.CrossRefPubMed
5.
go back to reference Jagannathan N, Sohn LE, Sawardekar A, Chang E, Langen KE, Anderson K. A randomised trial comparing the laryngeal mask airway Supreme with the laryngeal mask airway Unique in children. Anaesthesia. 2012;67:139–44.CrossRefPubMed Jagannathan N, Sohn LE, Sawardekar A, Chang E, Langen KE, Anderson K. A randomised trial comparing the laryngeal mask airway Supreme with the laryngeal mask airway Unique in children. Anaesthesia. 2012;67:139–44.CrossRefPubMed
6.
go back to reference Kim MS, Lee JH, Han SW, Im YJ, Kang HJ, Lee JR. A randomized comparison of the i-gel with the self-pressurized air-Q intubating laryngeal airway in children. Paediatr Anaesth. 2015;25:405–12.CrossRefPubMed Kim MS, Lee JH, Han SW, Im YJ, Kang HJ, Lee JR. A randomized comparison of the i-gel with the self-pressurized air-Q intubating laryngeal airway in children. Paediatr Anaesth. 2015;25:405–12.CrossRefPubMed
7.
go back to reference Jagannathan N, Sohn LE, Sawardekar A, Gordon J, Shah RD, Mukherji II, et al. A randomized trial comparing the Ambu (R) Aura-i with the air-Q intubating laryngeal airway as conduits for tracheal intubation in children. Paediatr Anaesth. 2012;22:1197–204.CrossRefPubMed Jagannathan N, Sohn LE, Sawardekar A, Gordon J, Shah RD, Mukherji II, et al. A randomized trial comparing the Ambu (R) Aura-i with the air-Q intubating laryngeal airway as conduits for tracheal intubation in children. Paediatr Anaesth. 2012;22:1197–204.CrossRefPubMed
8.
go back to reference Theiler LG, Kleine-Brueggeney M, Kaiser D, Urwyler N, Luyet C, Vogt A, et al. Crossover Comparison of the Laryngeal Mask Supreme and the i-gel in Simulated Difficult Airway Scenario in Anesthetized Patients. Anesthesiology. 2009;111:55–62.CrossRefPubMed Theiler LG, Kleine-Brueggeney M, Kaiser D, Urwyler N, Luyet C, Vogt A, et al. Crossover Comparison of the Laryngeal Mask Supreme and the i-gel in Simulated Difficult Airway Scenario in Anesthetized Patients. Anesthesiology. 2009;111:55–62.CrossRefPubMed
9.
go back to reference Sudhir G, Redfern D, Hall JE, Wilkes AR, Cann C. A comparison of the disposable Ambu AuraOnce Laryngeal Mask with the reusable LMA Classic laryngeal mask airway. Anaesthesia. 2007;62:719–22.CrossRefPubMed Sudhir G, Redfern D, Hall JE, Wilkes AR, Cann C. A comparison of the disposable Ambu AuraOnce Laryngeal Mask with the reusable LMA Classic laryngeal mask airway. Anaesthesia. 2007;62:719–22.CrossRefPubMed
10.
go back to reference Theiler LG, Kleine-Brueggeney M, Luepold B, Stucki F, Seiler S, Urwyler N, et al. Performance of the pediatric-sized i-gel compared with the Ambu AuraOnce laryngeal mask in anesthetized and ventilated children. Anesthesiology. 2011;115:102–10.CrossRefPubMed Theiler LG, Kleine-Brueggeney M, Luepold B, Stucki F, Seiler S, Urwyler N, et al. Performance of the pediatric-sized i-gel compared with the Ambu AuraOnce laryngeal mask in anesthetized and ventilated children. Anesthesiology. 2011;115:102–10.CrossRefPubMed
11.
go back to reference Keller C, Brimacombe JR, Keller K, Morris R. Comparison of four methods for assessing airway sealing pressure with the laryngeal mask airway in adult patients. Br J Anaesth. 1999;82:286–7.CrossRefPubMed Keller C, Brimacombe JR, Keller K, Morris R. Comparison of four methods for assessing airway sealing pressure with the laryngeal mask airway in adult patients. Br J Anaesth. 1999;82:286–7.CrossRefPubMed
12.
go back to reference Kelly F, Sale S, Bayley G, Cook T, Stoddart P, White M. A cohort evaluation of the pediatric ProSeal laryngeal mask airway in 100 children. Paediatr Anaesth. 2008;18:947–51.CrossRefPubMed Kelly F, Sale S, Bayley G, Cook T, Stoddart P, White M. A cohort evaluation of the pediatric ProSeal laryngeal mask airway in 100 children. Paediatr Anaesth. 2008;18:947–51.CrossRefPubMed
13.
go back to reference Lopez-Gil M, Brimacombe J, Garcia G. A randomized non-crossover study comparing the ProSeal and Classic laryngeal mask airway in anaesthetized children. Br J Anaesth. 2005;95:827–30.CrossRefPubMed Lopez-Gil M, Brimacombe J, Garcia G. A randomized non-crossover study comparing the ProSeal and Classic laryngeal mask airway in anaesthetized children. Br J Anaesth. 2005;95:827–30.CrossRefPubMed
14.
go back to reference Brimacombe J, Keller C, Brimacombe L. A comparison of the laryngeal mask airway ProSeal and the laryngeal tube airway in paralyzed anesthetized adult patients undergoing pressure-controlled ventilation. Anesth Analg. 2002;95:770–6.CrossRefPubMed Brimacombe J, Keller C, Brimacombe L. A comparison of the laryngeal mask airway ProSeal and the laryngeal tube airway in paralyzed anesthetized adult patients undergoing pressure-controlled ventilation. Anesth Analg. 2002;95:770–6.CrossRefPubMed
15.
go back to reference Francksen H, Bein B, Cavus E, Renner J, Scholz J, Steinfath M, et al. Comparison of LMA Unique, Ambu laryngeal mask and Soft Seal laryngeal mask during routine surgical procedures. Eur J Anaesthesiol. 2007;24:134–40.CrossRefPubMed Francksen H, Bein B, Cavus E, Renner J, Scholz J, Steinfath M, et al. Comparison of LMA Unique, Ambu laryngeal mask and Soft Seal laryngeal mask during routine surgical procedures. Eur J Anaesthesiol. 2007;24:134–40.CrossRefPubMed
16.
go back to reference Theiler L, Gutzmann M, Kleine-Brueggeney M, Urwyler N, Kaempfen B, Greif R. i-gel supraglottic airway in clinical practice: a prospective observational multicentre study. Br J Anaesth. 2012;109:990–5.CrossRefPubMed Theiler L, Gutzmann M, Kleine-Brueggeney M, Urwyler N, Kaempfen B, Greif R. i-gel supraglottic airway in clinical practice: a prospective observational multicentre study. Br J Anaesth. 2012;109:990–5.CrossRefPubMed
17.
go back to reference Cook TM, Cranshaw J. Randomized crossover comparison of ProSeal Laryngeal Mask Airway with Laryngeal Tube Sonda during anaesthesia with controlled ventilation. Br J Anaesth. 2005;95:261–6.CrossRefPubMed Cook TM, Cranshaw J. Randomized crossover comparison of ProSeal Laryngeal Mask Airway with Laryngeal Tube Sonda during anaesthesia with controlled ventilation. Br J Anaesth. 2005;95:261–6.CrossRefPubMed
18.
go back to reference Kapila A, Addy EV, Verghese C, Brain AI. The intubating laryngeal mask airway: an initial assessment of performance. Br J Anaesth. 1997;79:710–3.CrossRefPubMed Kapila A, Addy EV, Verghese C, Brain AI. The intubating laryngeal mask airway: an initial assessment of performance. Br J Anaesth. 1997;79:710–3.CrossRefPubMed
19.
go back to reference Verghese C, Berlet J, Kapila A, Pollard R. Clinical assessment of the single use laryngeal mask airway--the LMA-unique. Br J Anaesth. 1998;80:677–9.CrossRefPubMed Verghese C, Berlet J, Kapila A, Pollard R. Clinical assessment of the single use laryngeal mask airway--the LMA-unique. Br J Anaesth. 1998;80:677–9.CrossRefPubMed
20.
go back to reference Jagannathan N, Sohn LE, Mankoo R, Langen KE, Roth AG, Hall SC. Prospective evaluation of the self-pressurized air-Q intubating laryngeal airway in children. Paediatr Anaesth. 2011;21:673–80.CrossRefPubMed Jagannathan N, Sohn LE, Mankoo R, Langen KE, Roth AG, Hall SC. Prospective evaluation of the self-pressurized air-Q intubating laryngeal airway in children. Paediatr Anaesth. 2011;21:673–80.CrossRefPubMed
21.
go back to reference Richez B, Saltel L, Banchereau F, Torrielli R, Cros AM. A new single use supraglottic airway device with a noninflatable cuff and an esophageal vent: an observational study of the i-gel. Anesth Analg. 2008;106:1137–9.CrossRefPubMed Richez B, Saltel L, Banchereau F, Torrielli R, Cros AM. A new single use supraglottic airway device with a noninflatable cuff and an esophageal vent: an observational study of the i-gel. Anesth Analg. 2008;106:1137–9.CrossRefPubMed
22.
go back to reference Jagannathan N, Sommers K, Sohn LE, Sawardekar A, Shah RD, Mukherji II, et al. A randomized equivalence trial comparing the i-gel and laryngeal mask airway Supreme in children. Paediatr Anaesth. 2013;23:127–33.CrossRefPubMed Jagannathan N, Sommers K, Sohn LE, Sawardekar A, Shah RD, Mukherji II, et al. A randomized equivalence trial comparing the i-gel and laryngeal mask airway Supreme in children. Paediatr Anaesth. 2013;23:127–33.CrossRefPubMed
23.
go back to reference Kus A, Gok CN, Hosten T, Gurkan Y, Solak M, Toker K. The LMA-Supreme versus the I-gel in simulated difficult airway in children: a randomised study. Eur J Anaesthesiol. 2014;31:280–4.CrossRefPubMed Kus A, Gok CN, Hosten T, Gurkan Y, Solak M, Toker K. The LMA-Supreme versus the I-gel in simulated difficult airway in children: a randomised study. Eur J Anaesthesiol. 2014;31:280–4.CrossRefPubMed
24.
go back to reference Theiler L, Hermann K, Schoettker P, Savoldelli G, Urwyler N, Kleine-Brueggeney M, et al. SWIVIT--Swiss video-intubation trial evaluating video-laryngoscopes in a simulated difficult airway scenario: study protocol for a multicenter prospective randomized controlled trial in Switzerland. Trials. 2013;14:94.CrossRefPubMedPubMedCentral Theiler L, Hermann K, Schoettker P, Savoldelli G, Urwyler N, Kleine-Brueggeney M, et al. SWIVIT--Swiss video-intubation trial evaluating video-laryngoscopes in a simulated difficult airway scenario: study protocol for a multicenter prospective randomized controlled trial in Switzerland. Trials. 2013;14:94.CrossRefPubMedPubMedCentral
25.
go back to reference Kim MS, Oh JT, Min JY, Lee KH, Lee JR. A randomised comparison of the i-gel and the Laryngeal Mask Airway Classic in infants. Anaesthesia. 2014;69:362–7.CrossRefPubMed Kim MS, Oh JT, Min JY, Lee KH, Lee JR. A randomised comparison of the i-gel and the Laryngeal Mask Airway Classic in infants. Anaesthesia. 2014;69:362–7.CrossRefPubMed
26.
go back to reference Sinha R, Chandralekha RBR. Evaluation of air-Q intubating laryngeal airway as a conduit for tracheal intubation in infants--a pilot study. Paediatr Anaesth. 2012;22:156–60.CrossRefPubMed Sinha R, Chandralekha RBR. Evaluation of air-Q intubating laryngeal airway as a conduit for tracheal intubation in infants--a pilot study. Paediatr Anaesth. 2012;22:156–60.CrossRefPubMed
27.
go back to reference Kleine-Brueggeney M, Nicolet A, Nabecker S, Seiler S, Stucki F, Greif R, et al. Blind intubation of anaesthetised children with supraglottic airway devices AmbuAura-i and Air-Q cannot be recommended: A randomised controlled trial. Eur J Anaesthesiol. 2015;32:631–9.CrossRefPubMed Kleine-Brueggeney M, Nicolet A, Nabecker S, Seiler S, Stucki F, Greif R, et al. Blind intubation of anaesthetised children with supraglottic airway devices AmbuAura-i and Air-Q cannot be recommended: A randomised controlled trial. Eur J Anaesthesiol. 2015;32:631–9.CrossRefPubMed
Metadata
Title
Pediatric supraglottic airway devices in clinical practice: A prospective observational study
Authors
Maren Kleine-Brueggeney
Anne Gottfried
Sabine Nabecker
Robert Greif
Malte Book
Lorenz Theiler
Publication date
01-12-2017
Publisher
BioMed Central
Published in
BMC Anesthesiology / Issue 1/2017
Electronic ISSN: 1471-2253
DOI
https://doi.org/10.1186/s12871-017-0403-6