Uncalibrated pulse power analysis fails to reliably measure cardiac output in patients undergoing coronary artery bypass surgery

Authors: Ole Broch, Jochen Renner, Jan Höcker, Matthias Gruenewald, Patrick Meybohm, Jan Schöttler, Markus Steinfath, Berthold Bein

Published in: Critical Care | Issue 1/2011

Login to get access

Abstract

Introduction

Uncalibrated arterial pulse power analysis has been recently introduced for continuous monitoring of cardiac index (CI). The aim of the present study was to compare the accuracy of arterial pulse power analysis with intermittent transpulmonary thermodilution (TPTD) before and after cardiopulmonary bypass (CPB).

Methods

Forty-two patients scheduled for elective coronary surgery were studied after induction of anaesthesia, before and after CPB respectively. Each patient was monitored with the pulse contour cardiac output (PiCCO) system, a central venous line and the recently introduced LiDCO monitoring system. Haemodynamic variables included measurement of CI derived by transpulmonary thermodilution (CITPTD) or CI derived by pulse power analysis (CIPP), before and after calibration (CIPPnon-cal., CIPPcal.). Percentage changes of CI (ΔCITPTD, ΔCIPPnon-cal./PPcal.) were calculated to analyse directional changes.

Results

Before CPB there was no significant correlation between CIPPnon-cal. and CITPTD (r2 = 0.04, P = 0.08) with a percentage error (PE) of 86%. Higher mean arterial pressure (MAP) values were significantly correlated with higher CIPPnon-cal. (r2 = 0.26, P < 0.0001). After CPB, CIPPcal. revealed a significant correlation compared with CITPTD (r2 = 0.77, P < 0.0001) with PE of 28%. Changes in CIPPcal. (ΔCIPPcal.) showed a correlation with changes in CITPTD (ΔCITPTD) only after CPB (r2 = 0.52, P = 0.005).

Conclusions

Uncalibrated pulse power analysis was significantly influenced by MAP and was not able to reliably measure CI compared with TPTD. Calibration improved accuracy, but pulse power analysis was still not consistently interchangeable with TPTD. Only calibrated pulse power analysis was able to reliably track haemodynamic changes and trends.
Appendix
Available only for authorised users
Literature
1.
go back to reference Grocott MP, Mythen MG, Gan TJ: Perioperative fluid management and clinical outcomes in adults. Anesth Analg. 2005, 100: 1093-1106. 10.1213/01.ANE.0000148691.33690.AC.PubMedCrossRef Grocott MP, Mythen MG, Gan TJ: Perioperative fluid management and clinical outcomes in adults. Anesth Analg. 2005, 100: 1093-1106. 10.1213/01.ANE.0000148691.33690.AC.PubMedCrossRef
2.
go back to reference Jans O, Tollund C, Bundgaard-Nielsen M, Selmer C, Warberg J, Secher NH: Goal-directed fluid therapy: stroke volume optimisation and cardiac dimensions in supine healthy humans. Acta Anaesthesiol Scand. 2008, 52: 536-540. 10.1111/j.1399-6576.2008.01585.x.PubMedCrossRef Jans O, Tollund C, Bundgaard-Nielsen M, Selmer C, Warberg J, Secher NH: Goal-directed fluid therapy: stroke volume optimisation and cardiac dimensions in supine healthy humans. Acta Anaesthesiol Scand. 2008, 52: 536-540. 10.1111/j.1399-6576.2008.01585.x.PubMedCrossRef
3.
go back to reference Pearse R, Dawson D, Fawcett J, Rhodes A, Grounds RM, Bennett ED: Early goal-directed therapy after major surgery reduces complications and duration of hospital stay. A randomised, controlled trial (ISRCTN38797445). Crit Care. 2005, 9: R687-693. 10.1186/cc3887.PubMedPubMedCentralCrossRef Pearse R, Dawson D, Fawcett J, Rhodes A, Grounds RM, Bennett ED: Early goal-directed therapy after major surgery reduces complications and duration of hospital stay. A randomised, controlled trial (ISRCTN38797445). Crit Care. 2005, 9: R687-693. 10.1186/cc3887.PubMedPubMedCentralCrossRef
4.
go back to reference Rivers E, Nguyen B, Havstad S, Ressler J, Muzzin A, Knoblich B, Peterson E, Tomlanovich M: Early goal-directed therapy in the treatment of severe sepsis and septic shock. N Engl J Med. 2001, 345: 1368-1377. 10.1056/NEJMoa010307.PubMedCrossRef Rivers E, Nguyen B, Havstad S, Ressler J, Muzzin A, Knoblich B, Peterson E, Tomlanovich M: Early goal-directed therapy in the treatment of severe sepsis and septic shock. N Engl J Med. 2001, 345: 1368-1377. 10.1056/NEJMoa010307.PubMedCrossRef
5.
go back to reference Breukers RM, Sepehrkhouy S, Spiegelenberg SR, Groeneveld AB: Cardiac output measured by a new arterial pressure waveform analysis method without calibration compared with thermodilution after cardiac surgery. J Cardiothorac Vasc Anesth. 2007, 21: 632-635. 10.1053/j.jvca.2007.01.001.PubMedCrossRef Breukers RM, Sepehrkhouy S, Spiegelenberg SR, Groeneveld AB: Cardiac output measured by a new arterial pressure waveform analysis method without calibration compared with thermodilution after cardiac surgery. J Cardiothorac Vasc Anesth. 2007, 21: 632-635. 10.1053/j.jvca.2007.01.001.PubMedCrossRef
6.
go back to reference Friesecke S, Heinrich A, Abel P, Felix SB: Comparison of pulmonary artery and aortic transpulmonary thermodilution for monitoring of cardiac output in patients with severe heart failure: validation of a novel method. Crit Care Med. 2009, 37: 119-123. 10.1097/CCM.0b013e31819290d5.PubMedCrossRef Friesecke S, Heinrich A, Abel P, Felix SB: Comparison of pulmonary artery and aortic transpulmonary thermodilution for monitoring of cardiac output in patients with severe heart failure: validation of a novel method. Crit Care Med. 2009, 37: 119-123. 10.1097/CCM.0b013e31819290d5.PubMedCrossRef
7.
go back to reference Reinke RT, Higgins CB: Pulmonary infarction complicating the use of Swan-Ganz catheters. Br J Radiol. 1975, 48: 885-888. 10.1259/0007-1285-48-575-885.PubMedCrossRef Reinke RT, Higgins CB: Pulmonary infarction complicating the use of Swan-Ganz catheters. Br J Radiol. 1975, 48: 885-888. 10.1259/0007-1285-48-575-885.PubMedCrossRef
8.
go back to reference Connors AF, Speroff T, Dawson NV, Thomas C, Harrell FE, Wagner D, Desbiens N, Goldman L, Wu AW, Califf RM, Fulkerson WJ, Vidaillet H, Broste S, Bellamy P, Lynn J, Knaus WA: The effectiveness of right heart catheterization in the initial care of critically ill patients. SUPPORT Investigators. JAMA. 1996, 276: 889-897. 10.1001/jama.276.11.889.PubMedCrossRef Connors AF, Speroff T, Dawson NV, Thomas C, Harrell FE, Wagner D, Desbiens N, Goldman L, Wu AW, Califf RM, Fulkerson WJ, Vidaillet H, Broste S, Bellamy P, Lynn J, Knaus WA: The effectiveness of right heart catheterization in the initial care of critically ill patients. SUPPORT Investigators. JAMA. 1996, 276: 889-897. 10.1001/jama.276.11.889.PubMedCrossRef
9.
go back to reference Richard C, Warszawski J, Anguel N, Deye N, Combes A, Barnoud D, Boulain T, Lefort Y, Fartoukh M, Baud F, Boyer A, Brochard L, Teboul JL: Early use of the pulmonary artery catheter and outcomes in patients with shock and acute respiratory distress syndrome: a randomized controlled trial. JAMA. 2003, 290: 2713-2720. 10.1001/jama.290.20.2713.PubMedCrossRef Richard C, Warszawski J, Anguel N, Deye N, Combes A, Barnoud D, Boulain T, Lefort Y, Fartoukh M, Baud F, Boyer A, Brochard L, Teboul JL: Early use of the pulmonary artery catheter and outcomes in patients with shock and acute respiratory distress syndrome: a randomized controlled trial. JAMA. 2003, 290: 2713-2720. 10.1001/jama.290.20.2713.PubMedCrossRef
10.
go back to reference Sander M, von Heymann C, Foer A, von Dossow V, Grosse J, Dushe S, Konertz WF, Spies CD: Pulse contour analysis after normothermic cardiopulmonary bypass in cardiac surgery patients. Crit Care. 2005, 9: R729-734. 10.1186/cc3903.PubMedPubMedCentralCrossRef Sander M, von Heymann C, Foer A, von Dossow V, Grosse J, Dushe S, Konertz WF, Spies CD: Pulse contour analysis after normothermic cardiopulmonary bypass in cardiac surgery patients. Crit Care. 2005, 9: R729-734. 10.1186/cc3903.PubMedPubMedCentralCrossRef
11.
go back to reference Ritter S, Rudiger A, Maggiorini M: Transpulmonary thermodilution-derived cardiac function index identifies cardiac dysfunction in acute heart failure and septic patients: an observational study. Crit Care. 2009, 13: R133-10.1186/cc7994.PubMedPubMedCentralCrossRef Ritter S, Rudiger A, Maggiorini M: Transpulmonary thermodilution-derived cardiac function index identifies cardiac dysfunction in acute heart failure and septic patients: an observational study. Crit Care. 2009, 13: R133-10.1186/cc7994.PubMedPubMedCentralCrossRef
12.
go back to reference Hillis LD, Firth BG, Winniford MD: Comparison of thermodilution and indocyanine green dye in low cardiac output or left-sided regurgitation. Am J Cardiol. 1986, 57: 1201-1202. 10.1016/0002-9149(86)90704-6.PubMedCrossRef Hillis LD, Firth BG, Winniford MD: Comparison of thermodilution and indocyanine green dye in low cardiac output or left-sided regurgitation. Am J Cardiol. 1986, 57: 1201-1202. 10.1016/0002-9149(86)90704-6.PubMedCrossRef
13.
go back to reference Sakka SG, Reinhart K, Meier-Hellmann A: Comparison of pulmonary artery and arterial thermodilution cardiac output in critically ill patients. Intensive Care Med. 1999, 25: 843-846. 10.1007/s001340050962.PubMedCrossRef Sakka SG, Reinhart K, Meier-Hellmann A: Comparison of pulmonary artery and arterial thermodilution cardiac output in critically ill patients. Intensive Care Med. 1999, 25: 843-846. 10.1007/s001340050962.PubMedCrossRef
14.
go back to reference Breukers RM, Groeneveld AB, de Wilde RB, Jansen JR: Transpulmonary versus continuous thermodilution cardiac output after valvular and coronary artery surgery. Interact Cardiovasc Thorac Surg. 2009, 9: 4-8. 10.1510/icvts.2009.204545.PubMedCrossRef Breukers RM, Groeneveld AB, de Wilde RB, Jansen JR: Transpulmonary versus continuous thermodilution cardiac output after valvular and coronary artery surgery. Interact Cardiovasc Thorac Surg. 2009, 9: 4-8. 10.1510/icvts.2009.204545.PubMedCrossRef
15.
go back to reference Sander M, Spies CD, Grubitzsch H, Foer A, Muller M, von Heymann C: Comparison of uncalibrated arterial waveform analysis in cardiac surgery patients with thermodilution cardiac output measurements. Crit Care. 2006, 10: R164-10.1186/cc5103.PubMedPubMedCentralCrossRef Sander M, Spies CD, Grubitzsch H, Foer A, Muller M, von Heymann C: Comparison of uncalibrated arterial waveform analysis in cardiac surgery patients with thermodilution cardiac output measurements. Crit Care. 2006, 10: R164-10.1186/cc5103.PubMedPubMedCentralCrossRef
16.
go back to reference Hamilton TT, Huber LM, Jessen ME: PulseCO: a less-invasive method to monitor cardiac output from arterial pressure after cardiac surgery. Ann Thorac Surg. 2002, 74: S1408-1412. 10.1016/S0003-4975(02)04059-6.PubMedCrossRef Hamilton TT, Huber LM, Jessen ME: PulseCO: a less-invasive method to monitor cardiac output from arterial pressure after cardiac surgery. Ann Thorac Surg. 2002, 74: S1408-1412. 10.1016/S0003-4975(02)04059-6.PubMedCrossRef
17.
go back to reference Belloni L, Pisano A, Natale A, Piccirillo MR, Piazza L, Ismeno G, De Martino G: Assessment of fluid-responsiveness parameters for off-pump coronary artery bypass surgery: a comparison among LiDCO, transesophageal echochardiography, and pulmonary artery catheter. J Cardiothorac Vasc Anesth. 2008, 22: 243-248. 10.1053/j.jvca.2007.07.007.PubMedCrossRef Belloni L, Pisano A, Natale A, Piccirillo MR, Piazza L, Ismeno G, De Martino G: Assessment of fluid-responsiveness parameters for off-pump coronary artery bypass surgery: a comparison among LiDCO, transesophageal echochardiography, and pulmonary artery catheter. J Cardiothorac Vasc Anesth. 2008, 22: 243-248. 10.1053/j.jvca.2007.07.007.PubMedCrossRef
18.
go back to reference Bein B, Meybohm P, Cavus E, Renner J, Tonner PH, Steinfath M, Scholz J, Doerges V: The reliability of pulse contour-derived cardiac output during hemorrhage and after vasopressor administration. Anesth Analg. 2007, 105: 107-113. 10.1213/01.ane.0000268140.02147.ed.PubMedCrossRef Bein B, Meybohm P, Cavus E, Renner J, Tonner PH, Steinfath M, Scholz J, Doerges V: The reliability of pulse contour-derived cardiac output during hemorrhage and after vasopressor administration. Anesth Analg. 2007, 105: 107-113. 10.1213/01.ane.0000268140.02147.ed.PubMedCrossRef
19.
go back to reference Yamashita K, Nishiyama T, Yokoyama T, Abe H, Manabe M: Effects of vasodilation on cardiac output measured by PulseCO. J Clin Monit Comput. 2007, 21: 335-339. 10.1007/s10877-007-9093-9.PubMedCrossRef Yamashita K, Nishiyama T, Yokoyama T, Abe H, Manabe M: Effects of vasodilation on cardiac output measured by PulseCO. J Clin Monit Comput. 2007, 21: 335-339. 10.1007/s10877-007-9093-9.PubMedCrossRef
20.
go back to reference Critchley LA, Lee A, Ho AM: A critical review of the ability of continuous cardiac output monitors to measure trends in cardiac output. Anesth Analg. 2010, 111: 1180-1192. 10.1213/ANE.0b013e3181f08a5b.PubMedCrossRef Critchley LA, Lee A, Ho AM: A critical review of the ability of continuous cardiac output monitors to measure trends in cardiac output. Anesth Analg. 2010, 111: 1180-1192. 10.1213/ANE.0b013e3181f08a5b.PubMedCrossRef
21.
go back to reference Critchley LA, Critchley JA: A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques. J Clin Monit Comput. 1999, 15: 85-91. 10.1023/A:1009982611386.PubMedCrossRef Critchley LA, Critchley JA: A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques. J Clin Monit Comput. 1999, 15: 85-91. 10.1023/A:1009982611386.PubMedCrossRef
22.
go back to reference Halvorsen PS, Espinoza A, Lundblad R, Cvancarova M, Hol PK, Fosse E, Tonnessen TI: Agreement between PiCCO pulse-contour analysis, pulmonal artery thermodilution and transthoracic thermodilution during off-pump coronary artery by-pass surgery. Acta Anaesthesiol Scand. 2006, 50: 1050-1057. 10.1111/j.1399-6576.2006.01118.x.PubMedCrossRef Halvorsen PS, Espinoza A, Lundblad R, Cvancarova M, Hol PK, Fosse E, Tonnessen TI: Agreement between PiCCO pulse-contour analysis, pulmonal artery thermodilution and transthoracic thermodilution during off-pump coronary artery by-pass surgery. Acta Anaesthesiol Scand. 2006, 50: 1050-1057. 10.1111/j.1399-6576.2006.01118.x.PubMedCrossRef
23.
go back to reference Yamashita K, Nishiyama T, Yokoyama T, Abe H, Manabe M: The effects of vasodilation on cardiac output measured by PiCCO. J Cardiothorac Vasc Anesth. 2008, 22: 688-692. 10.1053/j.jvca.2008.04.007.PubMedCrossRef Yamashita K, Nishiyama T, Yokoyama T, Abe H, Manabe M: The effects of vasodilation on cardiac output measured by PiCCO. J Cardiothorac Vasc Anesth. 2008, 22: 688-692. 10.1053/j.jvca.2008.04.007.PubMedCrossRef
24.
go back to reference Squara P, Cecconi M, Rhodes A, Singer M, Chiche JD: Tracking changes in cardiac output: methodological considerations for the validation of monitoring devices. Intensive Care Med. 2009, 35: 1801-1808. 10.1007/s00134-009-1570-9.PubMedCrossRef Squara P, Cecconi M, Rhodes A, Singer M, Chiche JD: Tracking changes in cardiac output: methodological considerations for the validation of monitoring devices. Intensive Care Med. 2009, 35: 1801-1808. 10.1007/s00134-009-1570-9.PubMedCrossRef
25.
go back to reference Cooper ES, Muir WW: Continuous cardiac output monitoring via arterial pressure waveform analysis following severe hemorrhagic shock in dogs. Crit Care Med. 2007, 35: 1724-1729. 10.1097/01.CCM.0000266590.25109.F2.PubMedCrossRef Cooper ES, Muir WW: Continuous cardiac output monitoring via arterial pressure waveform analysis following severe hemorrhagic shock in dogs. Crit Care Med. 2007, 35: 1724-1729. 10.1097/01.CCM.0000266590.25109.F2.PubMedCrossRef
26.
go back to reference Boulain T, Achard JM, Teboul JL, Richard C, Perrotin D, Ginies G: Changes in BP induced by passive leg raising predict response to fluid loading in critically ill patients. Chest. 2002, 121: 1245-1252. 10.1378/chest.121.4.1245.PubMedCrossRef Boulain T, Achard JM, Teboul JL, Richard C, Perrotin D, Ginies G: Changes in BP induced by passive leg raising predict response to fluid loading in critically ill patients. Chest. 2002, 121: 1245-1252. 10.1378/chest.121.4.1245.PubMedCrossRef
27.
go back to reference Sagawa K, Lie RK, Schaefer J: Translation of Otto Frank's paper "Die Grundform des Arteriellen Pulses" Zeitschrift fur Biologie 37: 483-526 (1899). J Mol Cell Cardiol. 1990, 22: 253-277. 10.1016/0022-2828(90)91459-K.PubMedCrossRef Sagawa K, Lie RK, Schaefer J: Translation of Otto Frank's paper "Die Grundform des Arteriellen Pulses" Zeitschrift fur Biologie 37: 483-526 (1899). J Mol Cell Cardiol. 1990, 22: 253-277. 10.1016/0022-2828(90)91459-K.PubMedCrossRef
28.
go back to reference Mayer J, Boldt J, Schollhorn T, Rohm KD, Mengistu AM, Suttner S: Semi-invasive monitoring of cardiac output by a new device using arterial pressure waveform analysis: a comparison with intermittent pulmonary artery thermodilution in patients undergoing cardiac surgery. Br J Anaesth. 2007, 98: 176-182. 10.1093/bja/ael341.PubMedCrossRef Mayer J, Boldt J, Schollhorn T, Rohm KD, Mengistu AM, Suttner S: Semi-invasive monitoring of cardiac output by a new device using arterial pressure waveform analysis: a comparison with intermittent pulmonary artery thermodilution in patients undergoing cardiac surgery. Br J Anaesth. 2007, 98: 176-182. 10.1093/bja/ael341.PubMedCrossRef
29.
go back to reference Winer N, Sowers JR, Weber MA: Gender differences in vascular compliance in young, healthy subjects assessed by pulse contour analysis. J Clin Hypertens (Greenwich). 2001, 3: 145-152. 10.1111/j.1524-6175.2001.00704.x.CrossRef Winer N, Sowers JR, Weber MA: Gender differences in vascular compliance in young, healthy subjects assessed by pulse contour analysis. J Clin Hypertens (Greenwich). 2001, 3: 145-152. 10.1111/j.1524-6175.2001.00704.x.CrossRef
30.
go back to reference Covic A, Haydar AA, Bhamra-Ariza P, Gusbeth-Tatomir P, Goldsmith DJ: Aortic pulse wave velocity and arterial wave reflections predict the extent and severity of coronary artery disease in chronic kidney disease patients. J Nephrol. 2005, 18: 388-396.PubMed Covic A, Haydar AA, Bhamra-Ariza P, Gusbeth-Tatomir P, Goldsmith DJ: Aortic pulse wave velocity and arterial wave reflections predict the extent and severity of coronary artery disease in chronic kidney disease patients. J Nephrol. 2005, 18: 388-396.PubMed
31.
go back to reference Eleftheriadis S, Galatoudis Z, Didilis V, Bougioukas I, Schon J, Heinze H, Berger KU, Heringlake M: Variations in arterial blood pressure are associated with parallel changes in FlowTrac/Vigileo-derived cardiac output measurements: a prospective comparison study. Crit Care. 2009, 13: R179-10.1186/cc8161.PubMedPubMedCentralCrossRef Eleftheriadis S, Galatoudis Z, Didilis V, Bougioukas I, Schon J, Heinze H, Berger KU, Heringlake M: Variations in arterial blood pressure are associated with parallel changes in FlowTrac/Vigileo-derived cardiac output measurements: a prospective comparison study. Crit Care. 2009, 13: R179-10.1186/cc8161.PubMedPubMedCentralCrossRef
32.
go back to reference Manecke GR: Cardiac output from the arterial catheter: deceptively simple. J Cardiothorac Vasc Anesth. 2007, 21: 629-631. 10.1053/j.jvca.2007.07.001.PubMedCrossRef Manecke GR: Cardiac output from the arterial catheter: deceptively simple. J Cardiothorac Vasc Anesth. 2007, 21: 629-631. 10.1053/j.jvca.2007.07.001.PubMedCrossRef
33.
go back to reference Sakka SG, Kozieras J, Thuemer O, van Hout N: Measurement of cardiac output: a comparison between transpulmonary thermodilution and uncalibrated pulse contour analysis. Br J Anaesth. 2007, 99: 337-342. 10.1093/bja/aem177.PubMedCrossRef Sakka SG, Kozieras J, Thuemer O, van Hout N: Measurement of cardiac output: a comparison between transpulmonary thermodilution and uncalibrated pulse contour analysis. Br J Anaesth. 2007, 99: 337-342. 10.1093/bja/aem177.PubMedCrossRef
34.
go back to reference Monnet X, Anguel N, Naudin B, Jabot J, Richard C, Teboul JL: Arterial pressure-based cardiac output in septic patients: different accuracy of pulse contour and uncalibrated pressure waveform devices. Crit Care. 2010, 14: R109-10.1186/cc9058.PubMedPubMedCentralCrossRef Monnet X, Anguel N, Naudin B, Jabot J, Richard C, Teboul JL: Arterial pressure-based cardiac output in septic patients: different accuracy of pulse contour and uncalibrated pressure waveform devices. Crit Care. 2010, 14: R109-10.1186/cc9058.PubMedPubMedCentralCrossRef
35.
go back to reference Gruenewald M, Renner J, Meybohm P, Hocker J, Scholz J, Bein B: Reliability of continuous cardiac output measurement during intra-abdominal hypertension relies on repeated calibrations: an experimental animal study. Crit Care. 2008, 12: R132-10.1186/cc7102.PubMedPubMedCentralCrossRef Gruenewald M, Renner J, Meybohm P, Hocker J, Scholz J, Bein B: Reliability of continuous cardiac output measurement during intra-abdominal hypertension relies on repeated calibrations: an experimental animal study. Crit Care. 2008, 12: R132-10.1186/cc7102.PubMedPubMedCentralCrossRef
36.
go back to reference Linton NW, Linton RA: Is comparison of changes in cardiac output, assessed by different methods, better than only comparing cardiac output to the reference method?. Br J Anaesth. 2002, 89: 336-337. 10.1093/bja/aef530. author reply 337-339PubMedCrossRef Linton NW, Linton RA: Is comparison of changes in cardiac output, assessed by different methods, better than only comparing cardiac output to the reference method?. Br J Anaesth. 2002, 89: 336-337. 10.1093/bja/aef530. author reply 337-339PubMedCrossRef
Metadata
Title
Uncalibrated pulse power analysis fails to reliably measure cardiac output in patients undergoing coronary artery bypass surgery
Authors
Ole Broch
Jochen Renner
Jan Höcker
Matthias Gruenewald
Patrick Meybohm
Jan Schöttler
Markus Steinfath
Berthold Bein
Publication date
01-02-2011
Publisher
BioMed Central
Published in
Critical Care / Issue 1/2011
Electronic ISSN: 1364-8535
DOI
https://doi.org/10.1186/cc10065

Other articles of this Issue 1/2011

Critical Care 1/2011 Go to the issue