Open Access 01-12-2011 | Original research

SUVref: reducing reconstruction-dependent variation in PET SUV

Authors: Matthew D Kelly, Jerome M Declerck

Published in: EJNMMI Research | Issue 1/2011

Login to get access

Abstract

Background

We propose a new methodology, reference Standardised Uptake Value (SUVref), for reducing the quantitative variation resulting from differences in reconstruction protocol. Such variation that is not directly addressed by the use of SUV or the recently proposed PERCIST can impede comparability between positron emission tomography (PET)/CT scans.

Methods

SUVref applies a reconstruction-protocol-specific phantom-optimised filter to clinical PET scans for the purpose of improving comparability of quantification. The ability of this filter to reduce variability due to differences in reconstruction protocol was assessed using both phantom and clinical data.

Results

SUVref reduced the variability between recovery coefficients measured with the NEMA image quality phantom across a range of reconstruction protocols to below that measured for a single reconstruction protocol. In addition, it enabled quantitative conformance to the recently proposed EANM guidelines. For the clinical data, a significant reduction in bias and variance in the distribution of differences in SUV, resulting from differences in reconstruction protocol, greatly reduced the number of hot spots that would be misclassified as undergoing a clinically significant change in SUV.

Conclusions

SUVref significantly reduces reconstruction-dependent variation in SUV measurements, enabling increased confidence in quantitative comparison of clinical images for monitoring treatment response or disease progression. This new methodology could be similarly applied to reduce variability from scanner hardware.
Appendix
Available only for authorised users
Literature
1.
go back to reference Wahl RL, Jacene H, Kasamon Y, Lodge MA, From RECIST to PERCIST: Evolving Considerations for PET Response Criteria in Solid Tumors. J Nucl Med 2009, 50: 122S-150S. 10.2967/jnumed.108.057307PubMedCentralPubMedCrossRef Wahl RL, Jacene H, Kasamon Y, Lodge MA, From RECIST to PERCIST: Evolving Considerations for PET Response Criteria in Solid Tumors. J Nucl Med 2009, 50: 122S-150S. 10.2967/jnumed.108.057307PubMedCentralPubMedCrossRef
2.
go back to reference Huang H: Anatomy of SUV. Nucl Med and Biol 2000, 27: 643–646. 10.1016/S0969-8051(00)00155-4CrossRef Huang H: Anatomy of SUV. Nucl Med and Biol 2000, 27: 643–646. 10.1016/S0969-8051(00)00155-4CrossRef
3.
go back to reference Jaskowiak CJ, Bianco JA, Perlman SB, Fine JP: Influence of Reconstruction Iterations on 18 F-FDG PET/CT Standardized Uptake Values. J Nucl Med 2005, 46: 424–428.PubMed Jaskowiak CJ, Bianco JA, Perlman SB, Fine JP: Influence of Reconstruction Iterations on 18 F-FDG PET/CT Standardized Uptake Values. J Nucl Med 2005, 46: 424–428.PubMed
4.
go back to reference Boellaard R, O'Doherty MJ, Weber WA, Mottaghy FM, Lonsdale MN, Stroobants SG, Oyen WJG, Kotzerke J, Hoekstra OS, Pruim J, Marsden PK, Tatsch K, Hoekstra CK, Visser EP, Arends B, Verzijlbergen FJ, Zijlstra JM, Comans EFI, Lammertsma AA, Paans AM, Willemsen AT, Beyer T, Bockisch A, Schaefer-Prokop C, Delbeke D, Baum RP, Chiti A, Krause BJ: FDG PET and PET/CT: EANM procedure guidelines for tumour PET imaging: version 1.0. Eur J Nucl Med Mol Imaging 2010, 37: 181–200. 10.1007/s00259-009-1297-4PubMedCentralPubMedCrossRef Boellaard R, O'Doherty MJ, Weber WA, Mottaghy FM, Lonsdale MN, Stroobants SG, Oyen WJG, Kotzerke J, Hoekstra OS, Pruim J, Marsden PK, Tatsch K, Hoekstra CK, Visser EP, Arends B, Verzijlbergen FJ, Zijlstra JM, Comans EFI, Lammertsma AA, Paans AM, Willemsen AT, Beyer T, Bockisch A, Schaefer-Prokop C, Delbeke D, Baum RP, Chiti A, Krause BJ: FDG PET and PET/CT: EANM procedure guidelines for tumour PET imaging: version 1.0. Eur J Nucl Med Mol Imaging 2010, 37: 181–200. 10.1007/s00259-009-1297-4PubMedCentralPubMedCrossRef
5.
go back to reference Boellaard R, Oyen WJG, Hoekstra CJ, Hoekstra OS, Visser EP, Willemsen AT, Arends B, Verzijlbergen FJ, Zijlstra J, Paans AM, Comans EF, Pruim J: The Netherlands protocol for standardisation and quantification of FDG whole body PET studies in multi-centre trials. Eur J Nucl Med Mol Imaging 2008, 35: 2320–2333. 10.1007/s00259-008-0874-2PubMedCrossRef Boellaard R, Oyen WJG, Hoekstra CJ, Hoekstra OS, Visser EP, Willemsen AT, Arends B, Verzijlbergen FJ, Zijlstra J, Paans AM, Comans EF, Pruim J: The Netherlands protocol for standardisation and quantification of FDG whole body PET studies in multi-centre trials. Eur J Nucl Med Mol Imaging 2008, 35: 2320–2333. 10.1007/s00259-008-0874-2PubMedCrossRef
6.
go back to reference National Electrical Manufacturers Association. NEMA Standards Publication NU 2–2007. Performance Measurements of Positron Emission Tomographs NEMA 2007. National Electrical Manufacturers Association. NEMA Standards Publication NU 2–2007. Performance Measurements of Positron Emission Tomographs NEMA 2007.
7.
go back to reference Weber WA, Figlin R: Monitoring cancer treatment with PET/CT: Does it make a difference? J Nucl Med 2007, 48: 36S-44S.PubMedCrossRef Weber WA, Figlin R: Monitoring cancer treatment with PET/CT: Does it make a difference? J Nucl Med 2007, 48: 36S-44S.PubMedCrossRef
8.
go back to reference Joshi A, Koeppe RA, Fessler JA: Reducing between scanner differences in multi-centre PET studies. NeuroImage 2009, 49: 154–159.CrossRef Joshi A, Koeppe RA, Fessler JA: Reducing between scanner differences in multi-centre PET studies. NeuroImage 2009, 49: 154–159.CrossRef
9.
go back to reference Kadrmas DJ, Casey ME, Conti M, Jakoby BW, Lois C, Townsend DW: Impact of Time-of-Flight on PET Tumor Detection. J Nucl Med 2009, 50: 1315–1323. 10.2967/jnumed.109.063016PubMedCentralPubMedCrossRef Kadrmas DJ, Casey ME, Conti M, Jakoby BW, Lois C, Townsend DW: Impact of Time-of-Flight on PET Tumor Detection. J Nucl Med 2009, 50: 1315–1323. 10.2967/jnumed.109.063016PubMedCentralPubMedCrossRef
10.
go back to reference Panin V, Kehren F, Michel C, Casey M: Fully 3-D PET Reconstruction with system matrix derived from point source measurements. IEEE Trans Med Imaging 2007, 25: 907–921.CrossRef Panin V, Kehren F, Michel C, Casey M: Fully 3-D PET Reconstruction with system matrix derived from point source measurements. IEEE Trans Med Imaging 2007, 25: 907–921.CrossRef
11.
go back to reference Young H, Baum R, Cremerius U, Herholz K, Hoekstra O, Lammertsma AA, Pruim J, Price P: Measurement of Clinical and Subclinical Tumour Response using [ 18 F]-fluorodeoxyglucose and Positron Emission Tomography: Review and 1999 EORTC Recommendations. Eur J Cancer 1999, 35: 1773–1782. 10.1016/S0959-8049(99)00229-4PubMedCrossRef Young H, Baum R, Cremerius U, Herholz K, Hoekstra O, Lammertsma AA, Pruim J, Price P: Measurement of Clinical and Subclinical Tumour Response using [ 18 F]-fluorodeoxyglucose and Positron Emission Tomography: Review and 1999 EORTC Recommendations. Eur J Cancer 1999, 35: 1773–1782. 10.1016/S0959-8049(99)00229-4PubMedCrossRef
12.
13.
go back to reference Boellaard R, Krak N, Hoekstra OS, Lammertsma AA: Effects of Noise, Image Resolution, and ROI Definition on the Accuracy of Standard Uptake Values: A Simulation Study. J Nucl Med 2004, 45: 1519–1527.PubMed Boellaard R, Krak N, Hoekstra OS, Lammertsma AA: Effects of Noise, Image Resolution, and ROI Definition on the Accuracy of Standard Uptake Values: A Simulation Study. J Nucl Med 2004, 45: 1519–1527.PubMed
Metadata
Title
SUVref: reducing reconstruction-dependent variation in PET SUV
Authors
Matthew D Kelly
Jerome M Declerck
Publication date
01-12-2011
Publisher
Springer Berlin Heidelberg
Published in
EJNMMI Research / Issue 1/2011
Electronic ISSN: 2191-219X
DOI
https://doi.org/10.1186/2191-219X-1-16

Other articles of this Issue 1/2011

EJNMMI Research 1/2011 Go to the issue