Monte Carlo dose verification of prostate patients treated with simultaneous integrated boost intensity modulated radiation therapy

Authors: Nesrin Dogan, Ivaylo Mihaylov, Yan Wu, Paul J Keall, Jeffrey V Siebers, Michael P Hagan

Published in: Radiation Oncology | Issue 1/2009

Login to get access

Abstract

Background

To evaluate the dosimetric differences between Superposition/Convolution (SC) and Monte Carlo (MC) calculated dose distributions for simultaneous integrated boost (SIB) prostate cancer intensity modulated radiotherapy (IMRT) compared to experimental (film) measurements and the implications for clinical treatments.

Methods

Twenty-two prostate patients treated with an in-house SIB-IMRT protocol were selected. SC-based plans used for treatment were re-evaluated with EGS4-based MC calculations for treatment verification. Accuracy was evaluated with-respect-to film-based dosimetry. Comparisons used gamma (γ)-index, distance-to-agreement (DTA), and superimposed dose distributions. The treatment plans were also compared based on dose-volume indices and 3-D γ index for targets and critical structures.

Results

Flat-phantom comparisons demonstrated that the MC algorithm predicted measurements better than the SC algorithm. The average PTVprostate D98 agreement between SC and MC was 1.2% ± 1.1. For rectum, the average differences in SC and MC calculated D50 ranged from -3.6% to 3.4%. For small bowel, there were up to 30.2% ± 40.7 (range: 0.2%, 115%) differences between SC and MC calculated average D50 index. For femurs, the differences in average D50 reached up to 8.6% ± 3.6 (range: 1.2%, 14.5%). For PTVprostate and PTVnodes, the average gamma scores were >95.0%.

Conclusion

MC agrees better with film measurements than SC. Although, on average, SC-calculated doses agreed with MC calculations within the targets within 2%, there were deviations up to 5% for some patient's treatment plans. For some patients, the magnitude of such deviations might decrease the intended target dose levels that are required for the treatment protocol, placing the patients in different dose levels that do not satisfy the protocol dose requirements.
Appendix
Available only for authorised users
Literature
1.
go back to reference Wu Q, Manning M, Schmidt-Ullrich R, Mohan R: The potential for sparing of parotids and escalation of biologically effective dose with intensity-modulated radiation treatments of head and neck cancers: a treatment design study. Int J Radiat Oncol Biol Phys 2000, 46: 195-205. 10.1016/S0360-3016(99)00304-1CrossRefPubMed Wu Q, Manning M, Schmidt-Ullrich R, Mohan R: The potential for sparing of parotids and escalation of biologically effective dose with intensity-modulated radiation treatments of head and neck cancers: a treatment design study. Int J Radiat Oncol Biol Phys 2000, 46: 195-205. 10.1016/S0360-3016(99)00304-1CrossRefPubMed
2.
go back to reference Cozzi L, Fogliata A: IMRT in the treatment of head and neck cancer: is the present already the future? Expert Rev Anticancer Ther 2002, 2: 297-308. 10.1586/14737140.2.3.297CrossRefPubMed Cozzi L, Fogliata A: IMRT in the treatment of head and neck cancer: is the present already the future? Expert Rev Anticancer Ther 2002, 2: 297-308. 10.1586/14737140.2.3.297CrossRefPubMed
3.
go back to reference Eisbruch A, Dawson LA, Kim HM, Bradford CR, Terrell JE, Chepeha DB, Teknos TN, Anzai Y, Marsh LH, Martel MK, Ten Haken RK, Wolf GT, Ship JA: Conformal and intensity modulated irradiation of head and neck cancer: the potential for improved target irradiation, salivary gland function, and quality of life. Acta Otorhinolaryngol Belg 1999, 53: 271-275.PubMed Eisbruch A, Dawson LA, Kim HM, Bradford CR, Terrell JE, Chepeha DB, Teknos TN, Anzai Y, Marsh LH, Martel MK, Ten Haken RK, Wolf GT, Ship JA: Conformal and intensity modulated irradiation of head and neck cancer: the potential for improved target irradiation, salivary gland function, and quality of life. Acta Otorhinolaryngol Belg 1999, 53: 271-275.PubMed
4.
go back to reference Mackie TR, Scrimger JW, Battista JJ: A convolution method of calculating dose for 15-MV x rays. Med Phys 1985, 12: 188-196. 10.1118/1.595774CrossRefPubMed Mackie TR, Scrimger JW, Battista JJ: A convolution method of calculating dose for 15-MV x rays. Med Phys 1985, 12: 188-196. 10.1118/1.595774CrossRefPubMed
5.
go back to reference Yu CX, Mackie TR, Wong JW: Photon dose calculation incorporating explicit electron transport. Med Phys 1995, 22: 1157-1165. 10.1118/1.597611CrossRefPubMed Yu CX, Mackie TR, Wong JW: Photon dose calculation incorporating explicit electron transport. Med Phys 1995, 22: 1157-1165. 10.1118/1.597611CrossRefPubMed
6.
go back to reference Arnfield MR, Siantar CH, Siebers J, Garmon P, Cox L, Mohan R: The impact of electron transport on the accuracy of computed dose. Med Phys 2000, 27: 1266-1274. 10.1118/1.599004CrossRefPubMed Arnfield MR, Siantar CH, Siebers J, Garmon P, Cox L, Mohan R: The impact of electron transport on the accuracy of computed dose. Med Phys 2000, 27: 1266-1274. 10.1118/1.599004CrossRefPubMed
7.
go back to reference Mohan R, Arnfield M, Tong S, Wu Q, Siebers J: The impact of fluctuations in intensity patterns on the number of monitor units and the quality and accuracy of intensity modulated radiotherapy. Med Phys 2000, 27: 1226-1237. 10.1118/1.599000CrossRefPubMed Mohan R, Arnfield M, Tong S, Wu Q, Siebers J: The impact of fluctuations in intensity patterns on the number of monitor units and the quality and accuracy of intensity modulated radiotherapy. Med Phys 2000, 27: 1226-1237. 10.1118/1.599000CrossRefPubMed
8.
go back to reference Keall PJ, Siebers JV, Jeraj R, Mohan R: The effect of dose calculation uncertainty on the evaluation of radiotherapy plans. Med Phys 2000, 27: 478-484. 10.1118/1.598916CrossRefPubMed Keall PJ, Siebers JV, Jeraj R, Mohan R: The effect of dose calculation uncertainty on the evaluation of radiotherapy plans. Med Phys 2000, 27: 478-484. 10.1118/1.598916CrossRefPubMed
9.
go back to reference Jeraj R, Keall PJ, Siebers JV: The effect of dose calculation accuracy on inverse treatment planning. Phys Med Biol 2002, 47: 391-407. 10.1088/0031-9155/47/3/303CrossRefPubMed Jeraj R, Keall PJ, Siebers JV: The effect of dose calculation accuracy on inverse treatment planning. Phys Med Biol 2002, 47: 391-407. 10.1088/0031-9155/47/3/303CrossRefPubMed
10.
go back to reference Siebers JV, Keall PJ, Kim JO, Mohan R: A method for photon beam Monte Carlo multileaf collimator particle transport. Phys Med Biol 2002, 47: 3225-3249. 10.1088/0031-9155/47/17/312CrossRefPubMed Siebers JV, Keall PJ, Kim JO, Mohan R: A method for photon beam Monte Carlo multileaf collimator particle transport. Phys Med Biol 2002, 47: 3225-3249. 10.1088/0031-9155/47/17/312CrossRefPubMed
11.
go back to reference Kim JO, Siebers JV, Keall PJ, Arnfield MR, Mohan R: A Monte Carlo study of radiation transport through multileaf collimators. Med Phys 2001, 28: 2497-2506. 10.1118/1.1420734CrossRefPubMed Kim JO, Siebers JV, Keall PJ, Arnfield MR, Mohan R: A Monte Carlo study of radiation transport through multileaf collimators. Med Phys 2001, 28: 2497-2506. 10.1118/1.1420734CrossRefPubMed
12.
go back to reference Keall PJ, Siebers JV, Arnfield M, Kim JO, Mohan R: Monte Carlo dose calculations for dynamic IMRT treatments. Phys Med Biol 2001, 46: 929-941. 10.1088/0031-9155/46/4/302CrossRefPubMed Keall PJ, Siebers JV, Arnfield M, Kim JO, Mohan R: Monte Carlo dose calculations for dynamic IMRT treatments. Phys Med Biol 2001, 46: 929-941. 10.1088/0031-9155/46/4/302CrossRefPubMed
13.
go back to reference Yang J, Li JS, Qin L, Xiong W, Ma CM: Modelling of electron contamination in clinical photon beams for Monte Carlo dose calculation. Phys Med Biol 2004, 49: 2657-2673. 10.1088/0031-9155/49/12/013CrossRefPubMed Yang J, Li JS, Qin L, Xiong W, Ma CM: Modelling of electron contamination in clinical photon beams for Monte Carlo dose calculation. Phys Med Biol 2004, 49: 2657-2673. 10.1088/0031-9155/49/12/013CrossRefPubMed
14.
go back to reference Yang J, Li J, Chen L, Price R, McNeeley S, Qin L, Wang L, Xiong W, Ma CM: Dosimetric verification of IMRT treatment planning using Monte Carlo simulations for prostate cancer. Phys Med Biol 2005, 50: 869-878. 10.1088/0031-9155/50/5/011CrossRefPubMed Yang J, Li J, Chen L, Price R, McNeeley S, Qin L, Wang L, Xiong W, Ma CM: Dosimetric verification of IMRT treatment planning using Monte Carlo simulations for prostate cancer. Phys Med Biol 2005, 50: 869-878. 10.1088/0031-9155/50/5/011CrossRefPubMed
15.
go back to reference Wang L, Yorke E, Chui CS: Monte Carlo evaluation of 6 MV intensity modulated radiotherapy plans for head and neck and lung treatments. Med Phys 2002, 29: 2705-2717. 10.1118/1.1517291CrossRefPubMed Wang L, Yorke E, Chui CS: Monte Carlo evaluation of 6 MV intensity modulated radiotherapy plans for head and neck and lung treatments. Med Phys 2002, 29: 2705-2717. 10.1118/1.1517291CrossRefPubMed
16.
go back to reference Sakthi N, Keall P, Mihaylov I, Wu Q, Wu Y, Williamson JF, Schmidt-Ullrich R, Siebers JV: Monte Carlo-based dosimetry of head-and-neck patients treated with SIB-IMRT. Int J Radiat Oncol Biol Phys 2006, 64: 968-977.CrossRefPubMed Sakthi N, Keall P, Mihaylov I, Wu Q, Wu Y, Williamson JF, Schmidt-Ullrich R, Siebers JV: Monte Carlo-based dosimetry of head-and-neck patients treated with SIB-IMRT. Int J Radiat Oncol Biol Phys 2006, 64: 968-977.CrossRefPubMed
17.
go back to reference Rogers DWO, Faddegon BA, Ding GX, Ma CM, We J, Mackie TR: BEAM: a Monte Carlo code to simulate radiotherapy treatment units. Med Phys 1995, 22: 503-524. 10.1118/1.597552CrossRefPubMed Rogers DWO, Faddegon BA, Ding GX, Ma CM, We J, Mackie TR: BEAM: a Monte Carlo code to simulate radiotherapy treatment units. Med Phys 1995, 22: 503-524. 10.1118/1.597552CrossRefPubMed
18.
go back to reference Ma CM, Pawlicki T, Jiang SB, Li JS, Deng J, Mok E, Kapur A, Xing L, Ma L, Boyer AL: Monte Carlo verification of IMRT dose distributions from a commercial treatment planning optimization system. Phys Med Biol 2000, 45: 2483-2495. 10.1088/0031-9155/45/9/303CrossRefPubMed Ma CM, Pawlicki T, Jiang SB, Li JS, Deng J, Mok E, Kapur A, Xing L, Ma L, Boyer AL: Monte Carlo verification of IMRT dose distributions from a commercial treatment planning optimization system. Phys Med Biol 2000, 45: 2483-2495. 10.1088/0031-9155/45/9/303CrossRefPubMed
19.
go back to reference Ma CM, Mok E, Kapur A, Pawlicki T, Findley D, Brain S, Forster K, Boyer AL: Clinical implementation of a Monte Carlo treatment planning system. Med Phys 1999, 26: 2133-2143. 10.1118/1.598729CrossRefPubMed Ma CM, Mok E, Kapur A, Pawlicki T, Findley D, Brain S, Forster K, Boyer AL: Clinical implementation of a Monte Carlo treatment planning system. Med Phys 1999, 26: 2133-2143. 10.1118/1.598729CrossRefPubMed
20.
go back to reference Ma CM, Li JS, Pawlicki T, Jiang SB, Deng J, Lee MC, Koumrian T, Luxton M, Brain S: A Monte Carlo dose calculation tool for radiotherapy treatment planning. Phys Med Biol 2002, 47: 1671-1689. 10.1088/0031-9155/47/10/305CrossRefPubMed Ma CM, Li JS, Pawlicki T, Jiang SB, Deng J, Lee MC, Koumrian T, Luxton M, Brain S: A Monte Carlo dose calculation tool for radiotherapy treatment planning. Phys Med Biol 2002, 47: 1671-1689. 10.1088/0031-9155/47/10/305CrossRefPubMed
21.
go back to reference Leal A, Sanchez-Doblado F, Arrans R, Rosello J, Pavon EC, Lagares JI: Routine IMRT verification by means of an automated Monte Carlo simulation system. Int J Radiat Oncol Biol Phys 2003, 56: 58-68.CrossRefPubMed Leal A, Sanchez-Doblado F, Arrans R, Rosello J, Pavon EC, Lagares JI: Routine IMRT verification by means of an automated Monte Carlo simulation system. Int J Radiat Oncol Biol Phys 2003, 56: 58-68.CrossRefPubMed
22.
go back to reference Francescon P, Cora S, Chiovati P: Dose verification of an IMRT treatment planning system with the BEAM EGS4-based Monte Carlo code. Med Phys 2003, 30: 144-157. 10.1118/1.1538236CrossRefPubMed Francescon P, Cora S, Chiovati P: Dose verification of an IMRT treatment planning system with the BEAM EGS4-based Monte Carlo code. Med Phys 2003, 30: 144-157. 10.1118/1.1538236CrossRefPubMed
23.
go back to reference Fan J, Li J, Chen L, Stathakis S, Luo W, Du Plessis F, Xiong W, Yang J, Ma CM: A practical Monte Carlo MU verification tool for IMRT quality assurance. Phys Med Biol 2006, 51: 2503-2515. 10.1088/0031-9155/51/10/010CrossRefPubMed Fan J, Li J, Chen L, Stathakis S, Luo W, Du Plessis F, Xiong W, Yang J, Ma CM: A practical Monte Carlo MU verification tool for IMRT quality assurance. Phys Med Biol 2006, 51: 2503-2515. 10.1088/0031-9155/51/10/010CrossRefPubMed
24.
go back to reference Boudreau C, Heath E, Seuntjens J, Ballivy O, Parker W: IMRT head and neck treatment planning with a commercially available Monte Carlo based planning system. Phys Med Biol 2005, 50: 879-890. 10.1088/0031-9155/50/5/012CrossRefPubMed Boudreau C, Heath E, Seuntjens J, Ballivy O, Parker W: IMRT head and neck treatment planning with a commercially available Monte Carlo based planning system. Phys Med Biol 2005, 50: 879-890. 10.1088/0031-9155/50/5/012CrossRefPubMed
25.
go back to reference Dogan N, Siebers JV, Keall PJ, Lerma F, Wu Y, Fatyga M, Williamson JF, Schmidt-Ullrich RK: Improving IMRT dose accuracy via deliverable Monte Carlo optimization for the treatment of head and neck cancer patients. Med Phys 2006, 33: 4033-4043. 10.1118/1.2357835CrossRefPubMed Dogan N, Siebers JV, Keall PJ, Lerma F, Wu Y, Fatyga M, Williamson JF, Schmidt-Ullrich RK: Improving IMRT dose accuracy via deliverable Monte Carlo optimization for the treatment of head and neck cancer patients. Med Phys 2006, 33: 4033-4043. 10.1118/1.2357835CrossRefPubMed
26.
go back to reference Spirou SV, Chui CS: Generation of arbitrary intensity profiles by combining the scanning beam with dynamic multileaf collimation. Med Phys 1996, 23: 1-8. 10.1118/1.597672CrossRefPubMed Spirou SV, Chui CS: Generation of arbitrary intensity profiles by combining the scanning beam with dynamic multileaf collimation. Med Phys 1996, 23: 1-8. 10.1118/1.597672CrossRefPubMed
27.
go back to reference Spirou SV, Chui CS: Generation of arbitrary intensity profiles by dynamic jaws or multileaf collimators. Med Phys 1994, 21: 1031-1041. 10.1118/1.597345CrossRefPubMed Spirou SV, Chui CS: Generation of arbitrary intensity profiles by dynamic jaws or multileaf collimators. Med Phys 1994, 21: 1031-1041. 10.1118/1.597345CrossRefPubMed
28.
go back to reference Mihaylov IB, Lerma FA, Wu Y, Siebers JV: Analytic IMRT dose calculations utilizing Monte Carlo to predict MLC fluence modulation. Med Phys 2006, 33: 828-839. 10.1118/1.2178449PubMedCentralCrossRefPubMed Mihaylov IB, Lerma FA, Wu Y, Siebers JV: Analytic IMRT dose calculations utilizing Monte Carlo to predict MLC fluence modulation. Med Phys 2006, 33: 828-839. 10.1118/1.2178449PubMedCentralCrossRefPubMed
29.
go back to reference Wu Q, Mohan R: Algorithms and functionality of an intensity modulated radiotherapy optimization system. Med Phys 2000, 27: 701-711. 10.1118/1.598932CrossRefPubMed Wu Q, Mohan R: Algorithms and functionality of an intensity modulated radiotherapy optimization system. Med Phys 2000, 27: 701-711. 10.1118/1.598932CrossRefPubMed
30.
go back to reference An ADAC Pinnacle3 Physics Guide ADAC Laboratories, Milpitas, California; 2001. An ADAC Pinnacle3 Physics Guide ADAC Laboratories, Milpitas, California; 2001.
31.
go back to reference Nelson WR, Hirayama H, Rogers DWO: The EGS4 Code System. SLAC-265, Stanford Linear Accelerator Center 1985. Nelson WR, Hirayama H, Rogers DWO: The EGS4 Code System. SLAC-265, Stanford Linear Accelerator Center 1985.
32.
go back to reference Ma CM, Reckwerdt PJ, Holmes M, Roger DWO, Geiser B: DOSXYZ user's manual. PIRS-0509b. NRCC, Ottawa, Canada 1995. Ma CM, Reckwerdt PJ, Holmes M, Roger DWO, Geiser B: DOSXYZ user's manual. PIRS-0509b. NRCC, Ottawa, Canada 1995.
33.
go back to reference Miften M, Wiesmeyer M, Kapur A, Ma CM: Comparison of RTP dose distributions in heterogeneous phantoms with the BEAM Monte Carlo simulation system. J Appl Clin Med Phys 2001, 2: 21-31. 10.1120/1.1328616CrossRefPubMed Miften M, Wiesmeyer M, Kapur A, Ma CM: Comparison of RTP dose distributions in heterogeneous phantoms with the BEAM Monte Carlo simulation system. J Appl Clin Med Phys 2001, 2: 21-31. 10.1120/1.1328616CrossRefPubMed
34.
go back to reference Jeraj R, Keall P: The effect of statistical uncertainty on inverse treatment planning based on Monte Carlo dose calculation. Phys Med Biol 2000, 45: 3601-3613. 10.1088/0031-9155/45/12/307CrossRefPubMed Jeraj R, Keall P: The effect of statistical uncertainty on inverse treatment planning based on Monte Carlo dose calculation. Phys Med Biol 2000, 45: 3601-3613. 10.1088/0031-9155/45/12/307CrossRefPubMed
35.
go back to reference Dogan N, Siebers JV, Keall PJ: Clinical comparison of head and neck and prostate IMRT plans using absorbed dose to medium and absorbed dose to water. Phys Med Biol 2006, 51: 4967-4980. 10.1088/0031-9155/51/19/015CrossRefPubMed Dogan N, Siebers JV, Keall PJ: Clinical comparison of head and neck and prostate IMRT plans using absorbed dose to medium and absorbed dose to water. Phys Med Biol 2006, 51: 4967-4980. 10.1088/0031-9155/51/19/015CrossRefPubMed
36.
go back to reference Siebers JV, Keall PJ, Nahum AE, Mohan R: Converting absorbed dose to medium to absorbed dose to water for Monte Carlo based photon beam dose calculations. Phys Med Biol 2000, 45: 983-995. 10.1088/0031-9155/45/4/313CrossRefPubMed Siebers JV, Keall PJ, Nahum AE, Mohan R: Converting absorbed dose to medium to absorbed dose to water for Monte Carlo based photon beam dose calculations. Phys Med Biol 2000, 45: 983-995. 10.1088/0031-9155/45/4/313CrossRefPubMed
37.
go back to reference Siebers JV, Keall PJ, Libby B, Mohan R: Comparison of EGS4 and MCNP4b Monte Carlo codes for generation of photon phase space distributions for a Varian 2100C. Phys Med Biol 1999, 44: 3009-3026. 10.1088/0031-9155/44/12/311CrossRefPubMed Siebers JV, Keall PJ, Libby B, Mohan R: Comparison of EGS4 and MCNP4b Monte Carlo codes for generation of photon phase space distributions for a Varian 2100C. Phys Med Biol 1999, 44: 3009-3026. 10.1088/0031-9155/44/12/311CrossRefPubMed
38.
go back to reference Libby B, Siebers J, Mohan R: Validation of Monte Carlo generated phase-space descriptions of medical linear accelerators. Med Phys 1999, 26: 1476-1483. 10.1118/1.598643CrossRefPubMed Libby B, Siebers J, Mohan R: Validation of Monte Carlo generated phase-space descriptions of medical linear accelerators. Med Phys 1999, 26: 1476-1483. 10.1118/1.598643CrossRefPubMed
39.
go back to reference Keall PJ, Siebers JV, Libby B, Mohan R: Determining the incident electron fluence for Monte Carlo-based photon treatment planning using a standard measured data set. Med Phys 2003, 30: 574-582. 10.1118/1.1561623CrossRefPubMed Keall PJ, Siebers JV, Libby B, Mohan R: Determining the incident electron fluence for Monte Carlo-based photon treatment planning using a standard measured data set. Med Phys 2003, 30: 574-582. 10.1118/1.1561623CrossRefPubMed
40.
go back to reference Low DA, Harms WB, Mutic S, Purdy JA: A technique for the quantitative evaluation of dose distributions. Med Phys 1998, 25: 656-661. 10.1118/1.598248CrossRefPubMed Low DA, Harms WB, Mutic S, Purdy JA: A technique for the quantitative evaluation of dose distributions. Med Phys 1998, 25: 656-661. 10.1118/1.598248CrossRefPubMed
41.
go back to reference Harms WB Sr, Low DA, Wong JW, Purdy JA: A software tool for the quantitative evaluation of 3D dose calculation algorithms. Med Phys 1998, 25: 1830-1836. 10.1118/1.598363CrossRefPubMed Harms WB Sr, Low DA, Wong JW, Purdy JA: A software tool for the quantitative evaluation of 3D dose calculation algorithms. Med Phys 1998, 25: 1830-1836. 10.1118/1.598363CrossRefPubMed
42.
go back to reference Wendling M, Zijp LJ, McDermott LN, Smit EJ, Sonke JJ, Mijnheer BJ, van Herk M: A fast algorithm for gamma evaluation in 3D. Med Phys 2007, 34: 1647-1654. 10.1118/1.2721657CrossRefPubMed Wendling M, Zijp LJ, McDermott LN, Smit EJ, Sonke JJ, Mijnheer BJ, van Herk M: A fast algorithm for gamma evaluation in 3D. Med Phys 2007, 34: 1647-1654. 10.1118/1.2721657CrossRefPubMed
43.
go back to reference Mihaylov IB, Lerma FA, Fatyga M, Sibers JV: Quantification of the impact of MLC modeling and tissue heterogeneities on dynamic IMRT dose calculations. Med Phys 2007, 34: 1244-1252. 10.1118/1.2712413CrossRefPubMed Mihaylov IB, Lerma FA, Fatyga M, Sibers JV: Quantification of the impact of MLC modeling and tissue heterogeneities on dynamic IMRT dose calculations. Med Phys 2007, 34: 1244-1252. 10.1118/1.2712413CrossRefPubMed
Metadata
Title
Monte Carlo dose verification of prostate patients treated with simultaneous integrated boost intensity modulated radiation therapy
Authors
Nesrin Dogan
Ivaylo Mihaylov
Yan Wu
Paul J Keall
Jeffrey V Siebers
Michael P Hagan
Publication date
01-12-2009
Publisher
BioMed Central
Published in
Radiation Oncology / Issue 1/2009
Electronic ISSN: 1748-717X
DOI
https://doi.org/10.1186/1748-717X-4-18

Other articles of this Issue 1/2009

Radiation Oncology 1/2009 Go to the issue