Published in: Radiation Oncology 1/2008

Open Access 01-12-2008 | Research

Investigation of the usability of conebeam CT data sets for dose calculation

Authors: Anne Richter, Qiaoqiao Hu, Doreen Steglich, Kurt Baier, Jürgen Wilbert, Matthias Guckenberger, Michael Flentje

Published in: Radiation Oncology | Issue 1/2008

Login to get access

Abstract

Background

To investigate the feasibility and accuracy of dose calculation in cone beam CT (CBCT) data sets.

Methods

Kilovoltage CBCT images were acquired with the Elekta XVI system, CT studies generated with a conventional multi-slice CT scanner (Siemens Somatom Sensation Open) served as reference images. Material specific volumes of interest (VOI) were defined for commercial CT Phantoms (CATPhan® and Gammex RMI®) and CT values were evaluated in CT and CBCT images. For CBCT imaging, the influence of image acquisition parameters such as tube voltage, with or without filter (F1 or F0) and collimation on the CT values was investigated. CBCT images of 33 patients (pelvis n = 11, thorax n = 11, head n = 11) were compared with corresponding planning CT studies. Dose distributions for three different treatment plans were calculated in CT and CBCT images and differences were evaluated. Four different correction strategies to match CT values (HU) and density (D) in CBCT images were analysed: standard CT HU-D table without adjustment for CBCT; phantom based HU-D tables; patient group based HU-D tables (pelvis, thorax, head); and patient specific HU-D tables.

Results

CT values in the CBCT images of the CATPhan® were highly variable depending on the image acquisition parameters: a mean difference of 564 HU ± 377 HU was calculated between CT values determined from the planning CT and CBCT images. Hence, two protocols were selected for CBCT imaging in the further part of the study and HU-D tables were always specific for these protocols (pelvis and thorax with M20F1 filter, 120 kV; head S10F0 no filter, 100 kV). For dose calculation in real patient CBCT images, the largest differences between CT and CBCT were observed for the standard CT HU-D table: differences were 8.0% ± 5.7%, 10.9% ± 6.8% and 14.5% ± 10.4% respectively for pelvis, thorax and head patients using clinical treatment plans. The use of patient and group based HU-D tables resulted in small dose differences between planning CT and CBCT: 0.9% ± 0.9%, 1.8% ± 1.6%, 1.5% ± 2.5% for pelvis, thorax and head patients, respectively. The application of the phantom based HU-D table was acceptable for the head patients but larger deviations were determined for the pelvis and thorax patient populations.

Conclusion

The generation of three HU-D tables specific for the anatomical regions pelvis, thorax and head and specific for the corresponding CBCT image acquisition parameters resulted in accurate dose calculation in CBCT images. Once these HU-D tables are created, direct dose calculation on CBCT datasets is possible without the need of a reference CT images for pixel value calibration.
Appendix
Available only for authorised users
Literature
1.
go back to reference Jaffray DA, Siewerdsen JH, Wong JW, et al.: Flat-panel cone-beam computed tomography for image-guided radiation therapy. Int J Radiat Oncol Biol Phys 2002, 53: 1337-1349. 10.1016/S0360-3016(02)02884-5CrossRefPubMed Jaffray DA, Siewerdsen JH, Wong JW, et al.: Flat-panel cone-beam computed tomography for image-guided radiation therapy. Int J Radiat Oncol Biol Phys 2002, 53: 1337-1349. 10.1016/S0360-3016(02)02884-5CrossRefPubMed
2.
go back to reference Gong Y, Wang J, Bai S, et al.: Conventionally-fractionated image-guided intensity modulated radiotherapy (IG-IMRT): a safe and effective treatment for cancer spinal metastasis. Radiat Oncol 2008, 3: 11. 10.1186/1748-717X-3-11PubMedCentralCrossRefPubMed Gong Y, Wang J, Bai S, et al.: Conventionally-fractionated image-guided intensity modulated radiotherapy (IG-IMRT): a safe and effective treatment for cancer spinal metastasis. Radiat Oncol 2008, 3: 11. 10.1186/1748-717X-3-11PubMedCentralCrossRefPubMed
3.
go back to reference Meyer J, Wilbert J, Baier K, et al.: Positioning accuracy of cone-beam computed tomography in combination with a HexaPOD robot treatment table. Int J Radiat Oncol Biol Phys 2007, 67: 1220-1228.CrossRefPubMed Meyer J, Wilbert J, Baier K, et al.: Positioning accuracy of cone-beam computed tomography in combination with a HexaPOD robot treatment table. Int J Radiat Oncol Biol Phys 2007, 67: 1220-1228.CrossRefPubMed
4.
go back to reference Oldham M, Letourneau D, Watt L, et al.: Cone-beam-CT guided radiation therapy: A model for on-line application. Radiother Oncol 2005, 75: 271-278. 10.1016/j.radonc.2005.03.026CrossRefPubMed Oldham M, Letourneau D, Watt L, et al.: Cone-beam-CT guided radiation therapy: A model for on-line application. Radiother Oncol 2005, 75: 271-278. 10.1016/j.radonc.2005.03.026CrossRefPubMed
5.
go back to reference Paskalev K, Ma CM, Jacob R, et al.: Daily target localization for prostate patients based on 3D image correlation. Phys Med Biol 2004, 49: 931-939. 10.1088/0031-9155/49/6/005CrossRefPubMed Paskalev K, Ma CM, Jacob R, et al.: Daily target localization for prostate patients based on 3D image correlation. Phys Med Biol 2004, 49: 931-939. 10.1088/0031-9155/49/6/005CrossRefPubMed
6.
go back to reference Polat B, Wilbert J, Baier K, et al.: Nonrigid patient setup errors in the head-and-neck region. Strahlenther Onkol 2007, 183: 506-511. 10.1007/s00066-007-1747-5CrossRefPubMed Polat B, Wilbert J, Baier K, et al.: Nonrigid patient setup errors in the head-and-neck region. Strahlenther Onkol 2007, 183: 506-511. 10.1007/s00066-007-1747-5CrossRefPubMed
7.
go back to reference Jensen HR, Hansen O, Hjelm-Hansen M, et al.: Inter- and intrafractional movement of the tumour in extracranial stereotactic radiotherapy of NSCLC. Acta Oncol 2008, 47: 1432-1437. 10.1080/02841860802251567CrossRefPubMed Jensen HR, Hansen O, Hjelm-Hansen M, et al.: Inter- and intrafractional movement of the tumour in extracranial stereotactic radiotherapy of NSCLC. Acta Oncol 2008, 47: 1432-1437. 10.1080/02841860802251567CrossRefPubMed
8.
go back to reference Verellen D, De Ridder M, Storme G: A (short) history of image-guided radiotherapy. Radiother Oncol 2008, 86: 4-13. 10.1016/j.radonc.2007.11.023CrossRefPubMed Verellen D, De Ridder M, Storme G: A (short) history of image-guided radiotherapy. Radiother Oncol 2008, 86: 4-13. 10.1016/j.radonc.2007.11.023CrossRefPubMed
9.
go back to reference Verellen D, Ridder MD, Linthout N, et al.: Innovations in image-guided radiotherapy. Nat Rev Cancer 2007, 7: 949-960. 10.1038/nrc2288CrossRefPubMed Verellen D, Ridder MD, Linthout N, et al.: Innovations in image-guided radiotherapy. Nat Rev Cancer 2007, 7: 949-960. 10.1038/nrc2288CrossRefPubMed
10.
go back to reference Guckenberger M, Meyer J, Wilbert J, et al.: Cone-beam CT based image-guidance for extracranial stereotactic radiotherapy of intrapulmonary tumors. Acta Oncol 2006, 45: 897-906. 10.1080/02841860600904839CrossRefPubMed Guckenberger M, Meyer J, Wilbert J, et al.: Cone-beam CT based image-guidance for extracranial stereotactic radiotherapy of intrapulmonary tumors. Acta Oncol 2006, 45: 897-906. 10.1080/02841860600904839CrossRefPubMed
11.
go back to reference Morin O, Chen J, Aubin M, et al.: Dose calculation using megavoltage cone-beam CT. Int J Radiat Oncol Biol Phys 2007, 67: 1201-1210.CrossRefPubMed Morin O, Chen J, Aubin M, et al.: Dose calculation using megavoltage cone-beam CT. Int J Radiat Oncol Biol Phys 2007, 67: 1201-1210.CrossRefPubMed
12.
go back to reference Tanyi JA, Fuss MH: Volumetric image-guidance: Does routine usage prompt adaptive re-planning? An institutional review. Acta Oncol 2008, 47: 1444-1453. 10.1080/02841860802279741CrossRefPubMed Tanyi JA, Fuss MH: Volumetric image-guidance: Does routine usage prompt adaptive re-planning? An institutional review. Acta Oncol 2008, 47: 1444-1453. 10.1080/02841860802279741CrossRefPubMed
13.
go back to reference Letourneau D, Wong R, Moseley D, et al.: Online planning and delivery technique for radiotherapy of spinal metastases using cone-beam CT: image quality and system performance. Int J Radiat Oncol Biol Phys 2007, 67: 1229-1237.CrossRefPubMed Letourneau D, Wong R, Moseley D, et al.: Online planning and delivery technique for radiotherapy of spinal metastases using cone-beam CT: image quality and system performance. Int J Radiat Oncol Biol Phys 2007, 67: 1229-1237.CrossRefPubMed
14.
go back to reference Chi Y, Wu Q, Yan D: Dose calculation on cone beam CT (CBCT). Medical Physics 2007, 34: 2438. 10.1118/1.2760822CrossRef Chi Y, Wu Q, Yan D: Dose calculation on cone beam CT (CBCT). Medical Physics 2007, 34: 2438. 10.1118/1.2760822CrossRef
15.
go back to reference Depuydt T, Hrbacek J, Slagmolen P, et al.: Cone-beam CT Hounsfield unit correction method and application on images of the pelvic region. Radiotherapy and oncology 2006,81(Supplement 1):29. Depuydt T, Hrbacek J, Slagmolen P, et al.: Cone-beam CT Hounsfield unit correction method and application on images of the pelvic region. Radiotherapy and oncology 2006,81(Supplement 1):29.
16.
go back to reference Houser C, Nawaz AO, Galvin J, et al.: Quantitive Evaluation of Cone Beam CT Data Used for Treatment Planning. Medical Physics 2006, 33: 2285-2286. 10.1118/1.2241926CrossRef Houser C, Nawaz AO, Galvin J, et al.: Quantitive Evaluation of Cone Beam CT Data Used for Treatment Planning. Medical Physics 2006, 33: 2285-2286. 10.1118/1.2241926CrossRef
17.
go back to reference Lee L, Le QT, Xing L: Retrospective IMRT Dose Reconstruction Based on Cone-Beam CT and MLC Log-File. Int J Radiat Oncol Biol Phys 2008, 70: 634-644.CrossRefPubMed Lee L, Le QT, Xing L: Retrospective IMRT Dose Reconstruction Based on Cone-Beam CT and MLC Log-File. Int J Radiat Oncol Biol Phys 2008, 70: 634-644.CrossRefPubMed
18.
go back to reference Marchant TE, Moore CJ, Rowbottom CG, et al.: Shading correction algorithm for improvement of cone-beam CT images in radiotherapy. Phys Med Biol 2008, 53: 5719-5733. 10.1088/0031-9155/53/20/010CrossRefPubMed Marchant TE, Moore CJ, Rowbottom CG, et al.: Shading correction algorithm for improvement of cone-beam CT images in radiotherapy. Phys Med Biol 2008, 53: 5719-5733. 10.1088/0031-9155/53/20/010CrossRefPubMed
19.
go back to reference Paliwal B, Tewatia D, Orton N, et al.: Cone beam CT based treatment planning. Medical Physics 2006, 33: 2029. 10.1118/1.2240820CrossRef Paliwal B, Tewatia D, Orton N, et al.: Cone beam CT based treatment planning. Medical Physics 2006, 33: 2029. 10.1118/1.2240820CrossRef
20.
go back to reference Tucking T, Nill S, Oelfke U: Dose calculation on linac integrated KV-cone beam CT. Radiotherapy and oncology 2006,81(Supplement 1):26. Tucking T, Nill S, Oelfke U: Dose calculation on linac integrated KV-cone beam CT. Radiotherapy and oncology 2006,81(Supplement 1):26.
21.
go back to reference van Zijtveld M, Dirkx M, Heijmen B: Correction of conebeam CT values using a planning CT for derivation of the "dose of the day". Radiother Oncol 2007, 85: 195-200. 10.1016/j.radonc.2007.08.010CrossRefPubMed van Zijtveld M, Dirkx M, Heijmen B: Correction of conebeam CT values using a planning CT for derivation of the "dose of the day". Radiother Oncol 2007, 85: 195-200. 10.1016/j.radonc.2007.08.010CrossRefPubMed
22.
go back to reference Yang Y, Schreibmann E, Li T, et al.: Evaluation of on-board kV cone beam CT (CBCT)-based dose calculation. Phys Med Biol 2007, 52: 685-705. 10.1088/0031-9155/52/3/011CrossRefPubMed Yang Y, Schreibmann E, Li T, et al.: Evaluation of on-board kV cone beam CT (CBCT)-based dose calculation. Phys Med Biol 2007, 52: 685-705. 10.1088/0031-9155/52/3/011CrossRefPubMed
23.
go back to reference Yoo S, Yin FF: Dosimetric feasibility of cone-beam CT-based treatment planning compared to CT-based treatment planning. Int J Radiat Oncol Biol Phys 2006, 66: 1553-1561.CrossRefPubMed Yoo S, Yin FF: Dosimetric feasibility of cone-beam CT-based treatment planning compared to CT-based treatment planning. Int J Radiat Oncol Biol Phys 2006, 66: 1553-1561.CrossRefPubMed
24.
go back to reference Siewerdsen JH, Jaffray DA: Cone-beam computed tomography with a flat-panel imager: magnitude and effects of x-ray scatter. Med Phys 2001, 28: 220-231. 10.1118/1.1339879CrossRefPubMed Siewerdsen JH, Jaffray DA: Cone-beam computed tomography with a flat-panel imager: magnitude and effects of x-ray scatter. Med Phys 2001, 28: 220-231. 10.1118/1.1339879CrossRefPubMed
25.
go back to reference Parker RP, Hobday PA, Cassell KJ: The direct use of CT numbers in radiotherapy dosage calculations for inhomogeneous media. Phys Med Biol 1979, 24: 802-809. 10.1088/0031-9155/24/4/011CrossRefPubMed Parker RP, Hobday PA, Cassell KJ: The direct use of CT numbers in radiotherapy dosage calculations for inhomogeneous media. Phys Med Biol 1979, 24: 802-809. 10.1088/0031-9155/24/4/011CrossRefPubMed
26.
go back to reference Saw CB, Loper A, Komanduri K, et al.: Determination of CT-to-density conversion relationship for image-based treatment planning systems. Med Dosim 2005, 30: 145-148. 10.1016/j.meddos.2005.05.001CrossRefPubMed Saw CB, Loper A, Komanduri K, et al.: Determination of CT-to-density conversion relationship for image-based treatment planning systems. Med Dosim 2005, 30: 145-148. 10.1016/j.meddos.2005.05.001CrossRefPubMed
27.
go back to reference Constantinou C, Harrington JC, DeWerd LA: An electron density calibration phantom for CT-based treatment planning computers. Med Phys 1992, 19: 325-327. 10.1118/1.596862CrossRefPubMed Constantinou C, Harrington JC, DeWerd LA: An electron density calibration phantom for CT-based treatment planning computers. Med Phys 1992, 19: 325-327. 10.1118/1.596862CrossRefPubMed
28.
go back to reference Elekta: Clinical User Manual for XVI R3.5 General Release. 2006. Elekta: Clinical User Manual for XVI R3.5 General Release. 2006.
29.
go back to reference Schneider U, Pedroni E, Lomax A: The calibration of CT Hounsfield units for radiotherapy treatment planning. Phys Med Biol 1996, 41: 111-124. 10.1088/0031-9155/41/1/009CrossRefPubMed Schneider U, Pedroni E, Lomax A: The calibration of CT Hounsfield units for radiotherapy treatment planning. Phys Med Biol 1996, 41: 111-124. 10.1088/0031-9155/41/1/009CrossRefPubMed
Metadata
Title
Investigation of the usability of conebeam CT data sets for dose calculation
Authors
Anne Richter
Qiaoqiao Hu
Doreen Steglich
Kurt Baier
Jürgen Wilbert
Matthias Guckenberger
Michael Flentje
Publication date
01-12-2008
Publisher
BioMed Central
Published in
Radiation Oncology / Issue 1/2008
Electronic ISSN: 1748-717X
DOI
https://doi.org/10.1186/1748-717X-3-42

Other articles of this Issue 1/2008

Radiation Oncology 1/2008 Go to the issue