Published in: Radiation Oncology 1/2008

Open Access 01-12-2008 | Research

Reproducibility and geometric accuracy of the fixster system during hypofractionated stereotactic radiotherapy

Authors: Peter Lindvall, Per Bergström, Per-Olov Löfroth, Roger Henriksson, A Tommy Bergenheim

Published in: Radiation Oncology | Issue 1/2008

Login to get access

Abstract

Background

Hypofractionated radiotherapy has been used for the treatment of AVMs and brain metastases. Hypofractionation necessitates the use of a relocatable stereotactic frame that has to be applied on several occasions. The stereotactic frame needs to have a high degree of reproducibility, and patient positioning is crucial to achieve a high accuracy of the treatment.

Methods

In this study we have, by radiological means, evaluated the reproducibility of the isocenter in consecutive treatment sessions using the Fixster frame. Deviations in the X, Y and Z-axis were measured in 10 patients treated with hypofractionated radiotherapy.

Results

The mean deviation in the X-axis was 0.4 mm (range -2.1 – 2.1, median 0.7 mm) and in the Y-axis -0.3 mm (range -1.4 – 0.7, median -0.2 mm). The mean deviation in the Z-axis was -0.6 (range -1.4 – 1.4, median 0.0 mm).

Conclusion

There is a high degree of reproducibility of the isocenter during successive treatment sessions with HCSRT using the Fixster frame for stereotactic targeting. The high reducibility enables a safe treatment using hypofractionated stereotactic radiotherapy.
Appendix
Available only for authorised users
Literature
1.
go back to reference Lindvall P, Bergstrom P, Lofroth PO, Hariz MI, Henriksson R, Jonasson P, Bergenheim AT: Hypofractionated conformal stereotactic radiotherapy for arteriovenous malformations. Neurosurgery 2003,53(5):1036-42; discussion 1042-3. 10.1227/01.NEU.0000088566.82699.E6CrossRefPubMed Lindvall P, Bergstrom P, Lofroth PO, Hariz MI, Henriksson R, Jonasson P, Bergenheim AT: Hypofractionated conformal stereotactic radiotherapy for arteriovenous malformations. Neurosurgery 2003,53(5):1036-42; discussion 1042-3. 10.1227/01.NEU.0000088566.82699.E6CrossRefPubMed
2.
go back to reference Chang TC, Shirato H, Aoyama H, Ushikoshi S, Kato N, Kuroda S, Ishikawa T, Houkin K, Iwasaki Y, Miyasaka K: Stereotactic irradiation for intracranial arteriovenous malformation using stereotactic radiosurgery or hypofractionated stereotactic radiotherapy. Int J Radiat Oncol Biol Phys 2004,60(3):861-870.CrossRefPubMed Chang TC, Shirato H, Aoyama H, Ushikoshi S, Kato N, Kuroda S, Ishikawa T, Houkin K, Iwasaki Y, Miyasaka K: Stereotactic irradiation for intracranial arteriovenous malformation using stereotactic radiosurgery or hypofractionated stereotactic radiotherapy. Int J Radiat Oncol Biol Phys 2004,60(3):861-870.CrossRefPubMed
3.
go back to reference Manning MA, Cardinale RM, Benedict SH, Kavanagh BD, Zwicker RD, Amir C, Broaddus WC: Hypofractionated stereotactic radiotherapy as an alternative to radiosurgery for the treatment of patients with brain metastases. Int J Radiat Oncol Biol Phys 2000,47(3):603-608. 10.1016/S0360-3016(00)00475-2CrossRefPubMed Manning MA, Cardinale RM, Benedict SH, Kavanagh BD, Zwicker RD, Amir C, Broaddus WC: Hypofractionated stereotactic radiotherapy as an alternative to radiosurgery for the treatment of patients with brain metastases. Int J Radiat Oncol Biol Phys 2000,47(3):603-608. 10.1016/S0360-3016(00)00475-2CrossRefPubMed
4.
go back to reference Aoyama H, Shirato H, Nishioka T, Kagei K, Onimaru R, Suzuki K, Ushikoshi S, Houkin K, Kuroda S, Abe H, Miyasaka K: Treatment outcome of single or hypofractionated single-isocentric stereotactic irradiation (STI) using a linear accelerator for intracranial arteriovenous malformation. Radiother Oncol 2001,59(3):323-328. 10.1016/S0167-8140(01)00303-6CrossRefPubMed Aoyama H, Shirato H, Nishioka T, Kagei K, Onimaru R, Suzuki K, Ushikoshi S, Houkin K, Kuroda S, Abe H, Miyasaka K: Treatment outcome of single or hypofractionated single-isocentric stereotactic irradiation (STI) using a linear accelerator for intracranial arteriovenous malformation. Radiother Oncol 2001,59(3):323-328. 10.1016/S0167-8140(01)00303-6CrossRefPubMed
5.
go back to reference Aoyama H, Shirato H, Onimaru R, Kagei K, Ikeda J, Ishii N, Sawamura Y, Miyasaka K: Hypofractionated stereotactic radiotherapy alone without whole-brain irradiation for patients with solitary and oligo brain metastasis using noninvasive fixation of the skull. Int J Radiat Oncol Biol Phys 2003,56(3):793-800.CrossRefPubMed Aoyama H, Shirato H, Onimaru R, Kagei K, Ikeda J, Ishii N, Sawamura Y, Miyasaka K: Hypofractionated stereotactic radiotherapy alone without whole-brain irradiation for patients with solitary and oligo brain metastasis using noninvasive fixation of the skull. Int J Radiat Oncol Biol Phys 2003,56(3):793-800.CrossRefPubMed
6.
go back to reference Lindvall P, Bergstrom P, P-O L, Henriksson R, Bergenheim AT: Hypofractionated conformal stereotactic radiotherapy alone or in combination with whole brain radiotherapy in patients with cerebral metastases. Int J Rad Oncol Biol Phys Lindvall P, Bergstrom P, P-O L, Henriksson R, Bergenheim AT: Hypofractionated conformal stereotactic radiotherapy alone or in combination with whole brain radiotherapy in patients with cerebral metastases. Int J Rad Oncol Biol Phys
7.
go back to reference Hall EJ, Brenner DJ: The radiobiology of radiosurgery: rationale for different treatment regimes for AVMs and malignancies. Int J Radiat Oncol Biol Phys 1993,25(2):381-385.CrossRefPubMed Hall EJ, Brenner DJ: The radiobiology of radiosurgery: rationale for different treatment regimes for AVMs and malignancies. Int J Radiat Oncol Biol Phys 1993,25(2):381-385.CrossRefPubMed
8.
go back to reference Bergstrom M, Greitz T, Ribbe T: A method of stereotaxic localization adopted for conventional and digital radiography. Neuroradiology 1986,28(2):100-104. 10.1007/BF00327879CrossRefPubMed Bergstrom M, Greitz T, Ribbe T: A method of stereotaxic localization adopted for conventional and digital radiography. Neuroradiology 1986,28(2):100-104. 10.1007/BF00327879CrossRefPubMed
9.
go back to reference Greitz T, Bergstrom M, Boethius J, Kingsley D, Ribbe T: Head fixation system for integration of radiodiagnostic and therapeutic procedures. Neuroradiology 1980,19(1):1-6.PubMed Greitz T, Bergstrom M, Boethius J, Kingsley D, Ribbe T: Head fixation system for integration of radiodiagnostic and therapeutic procedures. Neuroradiology 1980,19(1):1-6.PubMed
10.
go back to reference Delannes M, Daly N, Bonnet J, Sabatier J, Tremoulet M: [Laitinen's stereo-adapter: application to the fractionated cerebral irradiation under stereotaxic conditions]. Neurochirurgie 1990,36(3):167-74; discussion 174-5.PubMed Delannes M, Daly N, Bonnet J, Sabatier J, Tremoulet M: [Laitinen's stereo-adapter: application to the fractionated cerebral irradiation under stereotaxic conditions]. Neurochirurgie 1990,36(3):167-74; discussion 174-5.PubMed
11.
go back to reference Ashamalla H, Addeo D, Ikoro NC, Ross P, Cosma M, Nasr N: Commissioning and clinical results utilizing the Gildenberg-Laitinen Adapter Device for X-ray in fractionated stereotactic radiotherapy. Int J Radiat Oncol Biol Phys 2003,56(2):592-598. 10.1016/S0360-3016(03)00119-6CrossRefPubMed Ashamalla H, Addeo D, Ikoro NC, Ross P, Cosma M, Nasr N: Commissioning and clinical results utilizing the Gildenberg-Laitinen Adapter Device for X-ray in fractionated stereotactic radiotherapy. Int J Radiat Oncol Biol Phys 2003,56(2):592-598. 10.1016/S0360-3016(03)00119-6CrossRefPubMed
12.
go back to reference Choi DR, Kim DY, Ahn YC, Huh SJ, Yeo IJ, Nam DH, Lee JI, Park K, Kim JH: Quantitative analysis of errors in fractionated stereotactic radiotherapy. Med Dosim 2001,26(4):315-318. 10.1016/S0958-3947(01)00082-6CrossRefPubMed Choi DR, Kim DY, Ahn YC, Huh SJ, Yeo IJ, Nam DH, Lee JI, Park K, Kim JH: Quantitative analysis of errors in fractionated stereotactic radiotherapy. Med Dosim 2001,26(4):315-318. 10.1016/S0958-3947(01)00082-6CrossRefPubMed
13.
go back to reference Wu A: Physics and dosimetry of the gamma knife. Neurosurg Clin N Am 1992,3(1):35-50.PubMed Wu A: Physics and dosimetry of the gamma knife. Neurosurg Clin N Am 1992,3(1):35-50.PubMed
14.
go back to reference Cernica G, de Boer SF, Diaz A, Fenstermaker RA, Podgorsak MB: Dosimetric accuracy of a staged radiosurgery treatment. Phys Med Biol 2005,50(9):1991-2002. 10.1088/0031-9155/50/9/005CrossRefPubMed Cernica G, de Boer SF, Diaz A, Fenstermaker RA, Podgorsak MB: Dosimetric accuracy of a staged radiosurgery treatment. Phys Med Biol 2005,50(9):1991-2002. 10.1088/0031-9155/50/9/005CrossRefPubMed
15.
go back to reference Hariz MI, Eriksson AT: Reproducibility of repeated mountings of a noninvasive CT/MRI stereoadapter. Appl Neurophysiol 1986,49(6):336-347. 10.1159/000100163PubMed Hariz MI, Eriksson AT: Reproducibility of repeated mountings of a noninvasive CT/MRI stereoadapter. Appl Neurophysiol 1986,49(6):336-347. 10.1159/000100163PubMed
16.
go back to reference Kumar S, Burke K, Nalder C, Jarrett P, Mubata C, A'Hern R, Humphreys M, Bidmead M, Brada M: Treatment accuracy of fractionated stereotactic radiotherapy. Radiother Oncol 2005,74(1):53-59. 10.1016/j.radonc.2004.06.008CrossRefPubMed Kumar S, Burke K, Nalder C, Jarrett P, Mubata C, A'Hern R, Humphreys M, Bidmead M, Brada M: Treatment accuracy of fractionated stereotactic radiotherapy. Radiother Oncol 2005,74(1):53-59. 10.1016/j.radonc.2004.06.008CrossRefPubMed
Metadata
Title
Reproducibility and geometric accuracy of the fixster system during hypofractionated stereotactic radiotherapy
Authors
Peter Lindvall
Per Bergström
Per-Olov Löfroth
Roger Henriksson
A Tommy Bergenheim
Publication date
01-12-2008
Publisher
BioMed Central
Published in
Radiation Oncology / Issue 1/2008
Electronic ISSN: 1748-717X
DOI
https://doi.org/10.1186/1748-717X-3-16

Other articles of this Issue 1/2008

Radiation Oncology 1/2008 Go to the issue