Published in: Journal of Cardiovascular Magnetic Resonance 1/2013

Open Access 01-12-2013 | Research

Modified look-locker inversion recovery T1 mapping indices: assessment of accuracy and reproducibility between magnetic resonance scanners

Authors: Fabio S Raman, Nadine Kawel-Boehm, Neville Gai, Melanie Freed, Jing Han, Chia-Ying Liu, Joao AC Lima, David A Bluemke, Songtao Liu

Published in: Journal of Cardiovascular Magnetic Resonance | Issue 1/2013

Login to get access

Abstract

Background

Cardiovascular magnetic resonance (CMR) T1 mapping indices, such as T1 time and partition coefficient (λ), have shown potential to assess diffuse myocardial fibrosis. The purpose of this study was to investigate how scanner and field strength variation affect the accuracy and precision/reproducibility of T1 mapping indices.

Methods

CMR studies were performed on two 1.5T and three 3T scanners. Eight phantoms were made to mimic the T1/T2 of pre- and post-contrast myocardium and blood at 1.5T and 3T. T1 mapping using MOLLI was performed with simulated heart rate of 40-100 bpm. Inversion recovery spin echo (IR-SE) was the reference standard for T1 determination. Accuracy was defined as the percent error between MOLLI and IR-SE, and scan/re-scan reproducibility was defined as the relative percent mean difference between repeat MOLLI scans. Partition coefficient was estimated by ΔR1myocardium phantom/ΔR1blood phantom. Generalized linear mixed model was used to compare the accuracy and precision/reproducibility of T1 and λ across field strength, scanners, and protocols.

Results

Field strength significantly affected MOLLI T1 accuracy (6.3% error for 1.5T vs. 10.8% error for 3T, p<0.001) but not λ accuracy (8.8% error for 1.5T vs. 8.0% error for 3T, p=0.11). Partition coefficients of MOLLI were not different between two 1.5T scanners (47.2% vs. 47.9%, p=0.13), and showed only slight variation across three 3T scanners (49.2% vs. 49.8% vs. 49.9%, p=0.016). Partition coefficient also had significantly lower percent error for precision (better scan/re-scan reproducibility) than measurement of individual T1 values (3.6% for λ vs. 4.3%-4.8% for T1 values, approximately, for pre/post blood and myocardium values).

Conclusion

Based on phantom studies, T1 errors using MOLLI ranged from 6-14% across various MR scanners while errors for partition coefficient were less (6-10%). Compared with absolute T1 times, partition coefficient showed less variability across platforms and field strengths as well as higher precision.
Appendix
Available only for authorised users
Literature
1.
go back to reference Ugander M, Oki AJ, Hsu LY, Kellman P, Greiser A, Aletras AH, Sibley CT, Chen MY, Bandettini WP, Arai AE: Extracellular volume imaging by magnetic resonance imaging provides insights into overt and sub-clinical myocardial pathology. Eur Heart J. 2012, 33 (10): 1268-1278. 10.1093/eurheartj/ehr481.PubMedCentralCrossRefPubMed Ugander M, Oki AJ, Hsu LY, Kellman P, Greiser A, Aletras AH, Sibley CT, Chen MY, Bandettini WP, Arai AE: Extracellular volume imaging by magnetic resonance imaging provides insights into overt and sub-clinical myocardial pathology. Eur Heart J. 2012, 33 (10): 1268-1278. 10.1093/eurheartj/ehr481.PubMedCentralCrossRefPubMed
2.
go back to reference Arheden H, Saeed M, Higgins CB, Gao DW, Bremerich J, Wyttenbach R, Dae MW, Wendland MF: Measurement of the distribution volume of gadopentetate dimeglumine at echo-planar MR imaging to quantify myocardial infarction: comparison with 99mTc-DTPA autoradiography in rats. Radiology. 1999, 211 (3): 698-708.CrossRefPubMed Arheden H, Saeed M, Higgins CB, Gao DW, Bremerich J, Wyttenbach R, Dae MW, Wendland MF: Measurement of the distribution volume of gadopentetate dimeglumine at echo-planar MR imaging to quantify myocardial infarction: comparison with 99mTc-DTPA autoradiography in rats. Radiology. 1999, 211 (3): 698-708.CrossRefPubMed
3.
go back to reference Jerosch-Herold M, Kwong RY: Magnetic resonance imaging in the assessment of ventricular remodeling and viability. Curr Heart Fail Rep. 2008, 5 (1): 5-10. 10.1007/s11897-008-0002-4.PubMedCentralCrossRefPubMed Jerosch-Herold M, Kwong RY: Magnetic resonance imaging in the assessment of ventricular remodeling and viability. Curr Heart Fail Rep. 2008, 5 (1): 5-10. 10.1007/s11897-008-0002-4.PubMedCentralCrossRefPubMed
4.
go back to reference Sparrow P, Messroghli DR, Reid S, Ridgway JP, Bainbridge G, Sivananthan MU: Myocardial T1 mapping for detection of left ventricular myocardial fibrosis in chronic aortic regurgitation: pilot study. AJR Am J Roentgenol. 2006, 187 (6): W630-635. 10.2214/AJR.05.1264.CrossRefPubMed Sparrow P, Messroghli DR, Reid S, Ridgway JP, Bainbridge G, Sivananthan MU: Myocardial T1 mapping for detection of left ventricular myocardial fibrosis in chronic aortic regurgitation: pilot study. AJR Am J Roentgenol. 2006, 187 (6): W630-635. 10.2214/AJR.05.1264.CrossRefPubMed
5.
go back to reference Iles L, Pfluger H, Phrommintikul A, Cherayath J, Aksit P, Gupta SN, Kaye DM, Taylor AJ: Evaluation of diffuse myocardial fibrosis in heart failure with cardiac magnetic resonance contrast-enhanced T1 mapping. J Am Coll Cardiol. 2008, 52 (19): 1574-1580. 10.1016/j.jacc.2008.06.049.CrossRefPubMed Iles L, Pfluger H, Phrommintikul A, Cherayath J, Aksit P, Gupta SN, Kaye DM, Taylor AJ: Evaluation of diffuse myocardial fibrosis in heart failure with cardiac magnetic resonance contrast-enhanced T1 mapping. J Am Coll Cardiol. 2008, 52 (19): 1574-1580. 10.1016/j.jacc.2008.06.049.CrossRefPubMed
6.
go back to reference Flett AS, Hayward MP, Ashworth MT, Hansen MS, Taylor AM, Elliott PM, McGregor C, Moon JC: Equilibrium contrast cardiovascular magnetic resonance for the measurement of diffuse myocardial fibrosis. Preliminary Validation in Humans. Circulation. 2010, 122 (2): 138-144. 10.1161/CIRCULATIONAHA.109.930636.CrossRefPubMed Flett AS, Hayward MP, Ashworth MT, Hansen MS, Taylor AM, Elliott PM, McGregor C, Moon JC: Equilibrium contrast cardiovascular magnetic resonance for the measurement of diffuse myocardial fibrosis. Preliminary Validation in Humans. Circulation. 2010, 122 (2): 138-144. 10.1161/CIRCULATIONAHA.109.930636.CrossRefPubMed
7.
go back to reference Broberg CS, Chugh S, Conklin C, Sahn DJ, Jerosch-Herold M: Quantification of diffuse myocardial fibrosis and its association with myocardial dysfunction in congenital heart disease. Circ Cardiovasc Imaging. 2010, 3 (6): 727-734. 10.1161/CIRCIMAGING.108.842096.PubMedCentralCrossRefPubMed Broberg CS, Chugh S, Conklin C, Sahn DJ, Jerosch-Herold M: Quantification of diffuse myocardial fibrosis and its association with myocardial dysfunction in congenital heart disease. Circ Cardiovasc Imaging. 2010, 3 (6): 727-734. 10.1161/CIRCIMAGING.108.842096.PubMedCentralCrossRefPubMed
8.
go back to reference Lee JJ, Liu S, Nacif MS, Ugander M, Han J, Kawel N, Sibley CT, Kellman P, Arai AE, Bluemke DA: Myocardial T1 and extracellular volume fraction mapping at 3 tesla. J Cardiovasc Magn Reson. 2011, 13: 75-10.1186/1532-429X-13-75.PubMedCentralCrossRefPubMed Lee JJ, Liu S, Nacif MS, Ugander M, Han J, Kawel N, Sibley CT, Kellman P, Arai AE, Bluemke DA: Myocardial T1 and extracellular volume fraction mapping at 3 tesla. J Cardiovasc Magn Reson. 2011, 13: 75-10.1186/1532-429X-13-75.PubMedCentralCrossRefPubMed
9.
go back to reference Schelbert EB, Testa SM, Meier CG, Ceyrolles WJ, Levenson JE, Blair AJ, Kellman P, Jones BL, Ludwig DR, Schwartzman D,et al: Myocardial extravascular extracellular volume fraction measurement by gadolinium cardiovascular magnetic resonance in humans: slow infusion versus bolus. J Cardiovasc Magn Reson. 2011, 13 (1): 16-10.1186/1532-429X-13-16.PubMedCentralCrossRefPubMed Schelbert EB, Testa SM, Meier CG, Ceyrolles WJ, Levenson JE, Blair AJ, Kellman P, Jones BL, Ludwig DR, Schwartzman D,et al: Myocardial extravascular extracellular volume fraction measurement by gadolinium cardiovascular magnetic resonance in humans: slow infusion versus bolus. J Cardiovasc Magn Reson. 2011, 13 (1): 16-10.1186/1532-429X-13-16.PubMedCentralCrossRefPubMed
10.
go back to reference Kawel N, Nacif M, Zavodni A, Jones J, Liu S, Sibley CT, Bluemke DA: T1 Mapping of the myocardium: intra-individual assessment of the effect of field strength, cardiac cycle and variation by myocardial region. J Cardiovasc Magn Reson. 2012, 14 (1): 27-10.1186/1532-429X-14-27.PubMedCentralCrossRefPubMed Kawel N, Nacif M, Zavodni A, Jones J, Liu S, Sibley CT, Bluemke DA: T1 Mapping of the myocardium: intra-individual assessment of the effect of field strength, cardiac cycle and variation by myocardial region. J Cardiovasc Magn Reson. 2012, 14 (1): 27-10.1186/1532-429X-14-27.PubMedCentralCrossRefPubMed
11.
go back to reference Kawel N, Nacif M, Zavodni A, Jones J, Liu S, Sibley CT, Bluemke DA: T1 Mapping of the myocardium: intra-individual assessment of post-contrast T1 time evolution and extracellular volume fraction at 3T for Gd-DTPA and Gd-BOPTA. J Cardiovasc Magn Reson. 2012, 14 (1): 26-10.1186/1532-429X-14-26.PubMedCentralCrossRefPubMed Kawel N, Nacif M, Zavodni A, Jones J, Liu S, Sibley CT, Bluemke DA: T1 Mapping of the myocardium: intra-individual assessment of post-contrast T1 time evolution and extracellular volume fraction at 3T for Gd-DTPA and Gd-BOPTA. J Cardiovasc Magn Reson. 2012, 14 (1): 26-10.1186/1532-429X-14-26.PubMedCentralCrossRefPubMed
12.
go back to reference Sado DM, Flett AS, Banypersad SM, White SK, Maestrini V, Quarta G, Lachmann RH, Murphy E, Mehta A, Hughes DA,et al: Cardiovascular magnetic resonance measurement of myocardial extracellular volume in health and disease. Heart. 2012, 98 (19): 1436-1441. 10.1136/heartjnl-2012-302346.CrossRefPubMed Sado DM, Flett AS, Banypersad SM, White SK, Maestrini V, Quarta G, Lachmann RH, Murphy E, Mehta A, Hughes DA,et al: Cardiovascular magnetic resonance measurement of myocardial extracellular volume in health and disease. Heart. 2012, 98 (19): 1436-1441. 10.1136/heartjnl-2012-302346.CrossRefPubMed
13.
go back to reference Fontana M, White SK, Banypersad SM, Sado DM, Maestrini V, Flett AS, Piechnik SK, Neubauer S, Roberts N, Moon JC: Comparison of T1 mapping techniques for ECV quantification. Histological validation and reproducibility of ShMOLLI versus multibreath-hold T1 quantification equilibrium contrast CMR. J Cardiovasc Magn Reson. 2012, 14 (1): 88-10.1186/1532-429X-14-88.PubMedCentralCrossRefPubMed Fontana M, White SK, Banypersad SM, Sado DM, Maestrini V, Flett AS, Piechnik SK, Neubauer S, Roberts N, Moon JC: Comparison of T1 mapping techniques for ECV quantification. Histological validation and reproducibility of ShMOLLI versus multibreath-hold T1 quantification equilibrium contrast CMR. J Cardiovasc Magn Reson. 2012, 14 (1): 88-10.1186/1532-429X-14-88.PubMedCentralCrossRefPubMed
14.
go back to reference Nacif MS, Turkbey EB, Gai N, Nazarian S, van der Geest RJ, Noureldin RA, Sibley CT, Ugander M, Liu S, Arai AE,et al: Myocardial T1 mapping with MRI: comparison of look-locker and MOLLI sequences. J Magn Reson Imaging. 2011, 34 (6): 1367-1373. 10.1002/jmri.22753.PubMedCentralCrossRefPubMed Nacif MS, Turkbey EB, Gai N, Nazarian S, van der Geest RJ, Noureldin RA, Sibley CT, Ugander M, Liu S, Arai AE,et al: Myocardial T1 mapping with MRI: comparison of look-locker and MOLLI sequences. J Magn Reson Imaging. 2011, 34 (6): 1367-1373. 10.1002/jmri.22753.PubMedCentralCrossRefPubMed
15.
go back to reference Tofts PS, Shuter B, Pope JM: Ni-DTPA doped agarose gel–a phantom material for Gd-DTPA enhancement measurements. Magnetic resonance imaging. 1993, 11 (1): 125-133. 10.1016/0730-725X(93)90420-I.CrossRefPubMed Tofts PS, Shuter B, Pope JM: Ni-DTPA doped agarose gel–a phantom material for Gd-DTPA enhancement measurements. Magnetic resonance imaging. 1993, 11 (1): 125-133. 10.1016/0730-725X(93)90420-I.CrossRefPubMed
16.
go back to reference Stanisz GJ, Odrobina EE, Pun J, Escaravage M, Graham SJ, Bronskill MJ, Henkelman RM: T1, T2 Relaxation and magnetization transfer in tissue at 3T. Magn Reson Med. 2005, 54 (3): 507-512. 10.1002/mrm.20605.CrossRefPubMed Stanisz GJ, Odrobina EE, Pun J, Escaravage M, Graham SJ, Bronskill MJ, Henkelman RM: T1, T2 Relaxation and magnetization transfer in tissue at 3T. Magn Reson Med. 2005, 54 (3): 507-512. 10.1002/mrm.20605.CrossRefPubMed
17.
go back to reference Kellman P, Wilson JR, Xue H, Ugander M, Arai AE: Extracellular volume fraction mapping in the myocardium, part 1: evaluation of an automated method. J Cardiovasc Magn Reson. 2012, 14: 63-10.1186/1532-429X-14-63.PubMedCentralCrossRefPubMed Kellman P, Wilson JR, Xue H, Ugander M, Arai AE: Extracellular volume fraction mapping in the myocardium, part 1: evaluation of an automated method. J Cardiovasc Magn Reson. 2012, 14: 63-10.1186/1532-429X-14-63.PubMedCentralCrossRefPubMed
18.
go back to reference Schar M, Kozerke S, Fischer SE, Boesiger P: Cardiac SSFP imaging at 3 tesla. Magn Reson Med. 2004, 51 (4): 799-806. 10.1002/mrm.20024.CrossRefPubMed Schar M, Kozerke S, Fischer SE, Boesiger P: Cardiac SSFP imaging at 3 tesla. Magn Reson Med. 2004, 51 (4): 799-806. 10.1002/mrm.20024.CrossRefPubMed
19.
go back to reference Messroghli DR, Plein S, Higgins DM, Walters K, Jones TR, Ridgway JP, Sivananthan MU: Human myocardium: single-breath-hold MR T1 mapping with high spatial resolution–reproducibility study. Radiology. 2006, 238 (3): 1004-1012. 10.1148/radiol.2382041903.CrossRefPubMed Messroghli DR, Plein S, Higgins DM, Walters K, Jones TR, Ridgway JP, Sivananthan MU: Human myocardium: single-breath-hold MR T1 mapping with high spatial resolution–reproducibility study. Radiology. 2006, 238 (3): 1004-1012. 10.1148/radiol.2382041903.CrossRefPubMed
20.
go back to reference Liu S, Han J, Nacif MS, Jones J, Kawel N, Kellman P, Sibley CT, Bluemke DA: Diffuse myocardial fibrosis evaluation using cardiac magnetic resonance T1 mapping: sample size considerations for clinical trials. J Cardiovasc Magn Reson. 2012, 14: 90-10.1186/1532-429X-14-90.PubMedCentralCrossRefPubMed Liu S, Han J, Nacif MS, Jones J, Kawel N, Kellman P, Sibley CT, Bluemke DA: Diffuse myocardial fibrosis evaluation using cardiac magnetic resonance T1 mapping: sample size considerations for clinical trials. J Cardiovasc Magn Reson. 2012, 14: 90-10.1186/1532-429X-14-90.PubMedCentralCrossRefPubMed
21.
go back to reference Messroghli DR, Rudolph A, Abdel-Aty H, Wassmuth R, Kuhne T, Dietz R, Schulz-Menger J: An open-source software tool for the generation of relaxation time maps in magnetic resonance imaging. BMC Med Imaging. 2010, 10 (1): 16-10.1186/1471-2342-10-16.PubMedCentralCrossRefPubMed Messroghli DR, Rudolph A, Abdel-Aty H, Wassmuth R, Kuhne T, Dietz R, Schulz-Menger J: An open-source software tool for the generation of relaxation time maps in magnetic resonance imaging. BMC Med Imaging. 2010, 10 (1): 16-10.1186/1471-2342-10-16.PubMedCentralCrossRefPubMed
22.
go back to reference White SK, Sado DM, Flett AS, Moon JC: Characterising the myocardial interstitial space: the clinical relevance of non-invasive imaging. Heart. 2012, 98 (10): 773-779. 10.1136/heartjnl-2011-301515.CrossRefPubMed White SK, Sado DM, Flett AS, Moon JC: Characterising the myocardial interstitial space: the clinical relevance of non-invasive imaging. Heart. 2012, 98 (10): 773-779. 10.1136/heartjnl-2011-301515.CrossRefPubMed
23.
go back to reference Judd RM, Kim RJ: Extracellular space measurements with CMR imaging. JACC Cardiovasc Imaging. 2012, 5 (9): 908-910. 10.1016/j.jcmg.2012.04.007.CrossRefPubMed Judd RM, Kim RJ: Extracellular space measurements with CMR imaging. JACC Cardiovasc Imaging. 2012, 5 (9): 908-910. 10.1016/j.jcmg.2012.04.007.CrossRefPubMed
24.
go back to reference Plein S, Kidambi A: Understanding LV remodeling following myocardial infarction: Are T1 maps by CMR the New guide?. JACC Cardiovasc Imaging. 2012, 5 (9): 894-896. 10.1016/j.jcmg.2012.07.006.CrossRefPubMed Plein S, Kidambi A: Understanding LV remodeling following myocardial infarction: Are T1 maps by CMR the New guide?. JACC Cardiovasc Imaging. 2012, 5 (9): 894-896. 10.1016/j.jcmg.2012.07.006.CrossRefPubMed
25.
go back to reference Liu CY, Liu YC, Wu C, Armstrong A, Volpe GJ, van der Geest R, Liu Y, Hundley WG, Gomes AS, Liu S, Nacif MS, Bluemke DA, Lima JAC: Evaluation of age related interstitial myocardial fibrosis with cardiac magnetic resonance contrast-enhanced T1 Mapping in the Multi-ethnic Study of Atherosclerosis (MESA). J Am Coll Cardiol. 2013, 10.1016/j.jacc.2013.05.078. Liu CY, Liu YC, Wu C, Armstrong A, Volpe GJ, van der Geest R, Liu Y, Hundley WG, Gomes AS, Liu S, Nacif MS, Bluemke DA, Lima JAC: Evaluation of age related interstitial myocardial fibrosis with cardiac magnetic resonance contrast-enhanced T1 Mapping in the Multi-ethnic Study of Atherosclerosis (MESA). J Am Coll Cardiol. 2013, 10.1016/j.jacc.2013.05.078.
26.
go back to reference Wong TC, Piehler KM, Kang IA, Kadakkal A, Kellman P, Schwartzman DS, Mulukutla SR, Simon MA, Shroff SG, Kuller LH,et al: Myocardial extracellular volume fraction quantified by cardiovascular magnetic resonance is increased in diabetes and associated with mortality and incident heart failure admission. Eur Heart J. 2013, [Epub ahead of print] Wong TC, Piehler KM, Kang IA, Kadakkal A, Kellman P, Schwartzman DS, Mulukutla SR, Simon MA, Shroff SG, Kuller LH,et al: Myocardial extracellular volume fraction quantified by cardiovascular magnetic resonance is increased in diabetes and associated with mortality and incident heart failure admission. Eur Heart J. 2013, [Epub ahead of print]
27.
go back to reference Karamitsos TD, Piechnik SK, Banypersad SM, Fontana M, Ntusi NB, Ferreira VM, Whelan CJ, Myerson SG, Robson MD, Hawkins PN,et al: Noncontrast T1 mapping for the diagnosis of cardiac amyloidosis. JACC Cardiovasc Imaging. 2013, 6 (4): 488-497. 10.1016/j.jcmg.2012.11.013.CrossRefPubMed Karamitsos TD, Piechnik SK, Banypersad SM, Fontana M, Ntusi NB, Ferreira VM, Whelan CJ, Myerson SG, Robson MD, Hawkins PN,et al: Noncontrast T1 mapping for the diagnosis of cardiac amyloidosis. JACC Cardiovasc Imaging. 2013, 6 (4): 488-497. 10.1016/j.jcmg.2012.11.013.CrossRefPubMed
28.
go back to reference Salerno M, Janardhanan R, Jiji RS, Brooks J, Adenaw N, Mehta B, Yang Y, Antkowiak P, Kramer CM, Epstein FH: Comparison of methods for determining the partition coefficient of gadolinium in the myocardium using T(1) mapping. J Magn Reson Imaging. 2012, 38 (1): 217-224.PubMedCentralCrossRefPubMed Salerno M, Janardhanan R, Jiji RS, Brooks J, Adenaw N, Mehta B, Yang Y, Antkowiak P, Kramer CM, Epstein FH: Comparison of methods for determining the partition coefficient of gadolinium in the myocardium using T(1) mapping. J Magn Reson Imaging. 2012, 38 (1): 217-224.PubMedCentralCrossRefPubMed
29.
go back to reference Piechnik SK, Ferreira VM, Dall’Armellina E, Cochlin LE, Greiser A, Neubauer S, Robson MD: Shortened modified look-locker inversion recovery (ShMOLLI) for clinical myocardial T1-mapping at 1.5 And 3 T within a 9 heartbeat breathhold. J Cardiovasc Magn Reson. 2010, 12: 69-10.1186/1532-429X-12-69.PubMedCentralCrossRefPubMed Piechnik SK, Ferreira VM, Dall’Armellina E, Cochlin LE, Greiser A, Neubauer S, Robson MD: Shortened modified look-locker inversion recovery (ShMOLLI) for clinical myocardial T1-mapping at 1.5 And 3 T within a 9 heartbeat breathhold. J Cardiovasc Magn Reson. 2010, 12: 69-10.1186/1532-429X-12-69.PubMedCentralCrossRefPubMed
30.
go back to reference Flewitt JA, Chow K, Pagano JJ, Green JD, Friedrich MG, Thompson RB: Effect of systematic T1 errors on lambda caclulations: comparison of different T1 mapping techniques. ISMRM 20th Annual Meeting. 2012 Flewitt JA, Chow K, Pagano JJ, Green JD, Friedrich MG, Thompson RB: Effect of systematic T1 errors on lambda caclulations: comparison of different T1 mapping techniques. ISMRM 20th Annual Meeting. 2012
31.
go back to reference Fitts M, Breton E, Kholmovski EG, Dosdall DJ, Vijayakumar S, Hong KP, Ranjan R, Marrouche NF, Axel L, Kim D: Arrhythmia insensitive rapid cardiac T(1) mapping pulse sequence. Magn Reson Med. 2012, [Epub ahead of print]. Fitts M, Breton E, Kholmovski EG, Dosdall DJ, Vijayakumar S, Hong KP, Ranjan R, Marrouche NF, Axel L, Kim D: Arrhythmia insensitive rapid cardiac T(1) mapping pulse sequence. Magn Reson Med. 2012, [Epub ahead of print].
32.
go back to reference Messroghli DR, Greiser A, Frohlich M, Dietz R, Schulz-Menger J: Optimization and validation of a fully-integrated pulse sequence for modified look-locker inversion-recovery (MOLLI) T1 mapping of the heart. J Magn Reson Imaging. 2007, 26 (4): 1081-1086. 10.1002/jmri.21119.CrossRefPubMed Messroghli DR, Greiser A, Frohlich M, Dietz R, Schulz-Menger J: Optimization and validation of a fully-integrated pulse sequence for modified look-locker inversion-recovery (MOLLI) T1 mapping of the heart. J Magn Reson Imaging. 2007, 26 (4): 1081-1086. 10.1002/jmri.21119.CrossRefPubMed
33.
go back to reference Wong TC, Piehler K, Meier CG, Testa SM, Klock AM, Aneizi AA, Shakesprere J, Kellman P, Shroff SG, Schwartzman DS,et al: Association between extracellular matrix expansion quantified by cardiovascular magnetic resonance and short-term mortality. Circulation. 2012, 126 (10): 1206-1216. 10.1161/CIRCULATIONAHA.111.089409.PubMedCentralCrossRefPubMed Wong TC, Piehler K, Meier CG, Testa SM, Klock AM, Aneizi AA, Shakesprere J, Kellman P, Shroff SG, Schwartzman DS,et al: Association between extracellular matrix expansion quantified by cardiovascular magnetic resonance and short-term mortality. Circulation. 2012, 126 (10): 1206-1216. 10.1161/CIRCULATIONAHA.111.089409.PubMedCentralCrossRefPubMed
34.
go back to reference Kellman P, Herzka DA, Hansen MS: Adiabatic inversion pulses for myocardial T1 mapping. Magn Reson Med. 2013, [Epub ahead of print] Kellman P, Herzka DA, Hansen MS: Adiabatic inversion pulses for myocardial T1 mapping. Magn Reson Med. 2013, [Epub ahead of print]
35.
go back to reference Bottomley PA, Foster TH, Argersinger RE, Pfeifer LM: A review of normal tissue hydrogen NMR relaxation times and relaxation mechanisms from 1–100 MHz: dependence on tissue type, NMR frequency, temperature, species, excision, and age. Med Phys. 1984, 11 (4): 425-448. 10.1118/1.595535.CrossRefPubMed Bottomley PA, Foster TH, Argersinger RE, Pfeifer LM: A review of normal tissue hydrogen NMR relaxation times and relaxation mechanisms from 1–100 MHz: dependence on tissue type, NMR frequency, temperature, species, excision, and age. Med Phys. 1984, 11 (4): 425-448. 10.1118/1.595535.CrossRefPubMed
36.
go back to reference Sasaki M, Yamada K, Watanabe Y, Matsui M, Ida M, Fujiwara S, Shibata E, Acute Stroke Imaging Standardization Group-Japan I: Variability in absolute apparent diffusion coefficient values across different platforms may be substantial: a multivendor, multi-institutional comparison study. Radiology. 2008, 249 (2): 624-630. 10.1148/radiol.2492071681.CrossRefPubMed Sasaki M, Yamada K, Watanabe Y, Matsui M, Ida M, Fujiwara S, Shibata E, Acute Stroke Imaging Standardization Group-Japan I: Variability in absolute apparent diffusion coefficient values across different platforms may be substantial: a multivendor, multi-institutional comparison study. Radiology. 2008, 249 (2): 624-630. 10.1148/radiol.2492071681.CrossRefPubMed
37.
go back to reference Takao H, Hayashi N, Ohtomo K: Effect of scanner in asymmetry studies using diffusion tensor imaging. NeuroImage. 2011, 54 (2): 1053-1062. 10.1016/j.neuroimage.2010.09.023.CrossRefPubMed Takao H, Hayashi N, Ohtomo K: Effect of scanner in asymmetry studies using diffusion tensor imaging. NeuroImage. 2011, 54 (2): 1053-1062. 10.1016/j.neuroimage.2010.09.023.CrossRefPubMed
38.
go back to reference Focke NK, Helms G, Kaspar S, Diederich C, Toth V, Dechent P, Mohr A, Paulus W: Multi-site voxel-based morphometry–not quite there yet. NeuroImage. 2011, 56 (3): 1164-1170. 10.1016/j.neuroimage.2011.02.029.CrossRefPubMed Focke NK, Helms G, Kaspar S, Diederich C, Toth V, Dechent P, Mohr A, Paulus W: Multi-site voxel-based morphometry–not quite there yet. NeuroImage. 2011, 56 (3): 1164-1170. 10.1016/j.neuroimage.2011.02.029.CrossRefPubMed
Metadata
Title
Modified look-locker inversion recovery T1 mapping indices: assessment of accuracy and reproducibility between magnetic resonance scanners
Authors
Fabio S Raman
Nadine Kawel-Boehm
Neville Gai
Melanie Freed
Jing Han
Chia-Ying Liu
Joao AC Lima
David A Bluemke
Songtao Liu
Publication date
01-12-2013
Publisher
BioMed Central
Published in
Journal of Cardiovascular Magnetic Resonance / Issue 1/2013
Electronic ISSN: 1532-429X
DOI
https://doi.org/10.1186/1532-429X-15-64

Other articles of this Issue 1/2013

Journal of Cardiovascular Magnetic Resonance 1/2013 Go to the issue