Published in: Journal of Cardiovascular Magnetic Resonance 1/2013

Open Access 01-12-2013 | Research

High spatial and temporal resolution retrospective cine cardiovascular magnetic resonance from shortened free breathing real-time acquisitions

Authors: Hui Xue, Peter Kellman, Gina LaRocca, Andrew E Arai, Michael S Hansen

Published in: Journal of Cardiovascular Magnetic Resonance | Issue 1/2013

Login to get access

Abstract

Background

Cine cardiovascular magnetic resonance (CMR) is challenging in patients who cannot perform repeated breath holds. Real-time, free-breathing acquisition is an alternative, but image quality is typically inferior. There is a clinical need for techniques that achieve similar image quality to the segmented cine using a free breathing acquisition. Previously, high quality retrospectively gated cine images have been reconstructed from real-time acquisitions using parallel imaging and motion correction. These methods had limited clinical applicability due to lengthy acquisitions and volumetric measurements obtained with such methods have not previously been evaluated systematically.

Methods

This study introduces a new retrospective reconstruction scheme for real-time cine imaging which aims to shorten the required acquisition. A real-time acquisition of 16-20s per acquired slice was inputted into a retrospective cine reconstruction algorithm, which employed non-rigid registration to remove respiratory motion and SPIRiT non-linear reconstruction with temporal regularization to fill in missing data. The algorithm was used to reconstruct cine loops with high spatial (1.3-1.8 × 1.8-2.1 mm2) and temporal resolution (retrospectively gated, 30 cardiac phases, temporal resolution 34.3 ± 9.1 ms). Validation was performed in 15 healthy volunteers using two different acquisition resolutions (256 × 144/192 × 128 matrix sizes). For each subject, 9 to 12 short axis and 3 long axis slices were imaged with both segmented and real-time acquisitions. The retrospectively reconstructed real-time cine images were compared to a traditional segmented breath-held acquisition in terms of image quality scores. Image quality scoring was performed by two experts using a scale between 1 and 5 (poor to good). For every subject, LAX and three SAX slices were selected and reviewed in the random order. The reviewers were blinded to the reconstruction approach and acquisition protocols and scores were given to segmented and retrospective cine series. Volumetric measurements of cardiac function were also compared by manually tracing the myocardium for segmented and retrospective cines.

Results

Mean image quality scores were similar for short axis and long axis views for both tested resolutions. Short axis scores were 4.52/4.31 (high/low matrix sizes) for breath-hold vs. 4.54/4.56 for real-time (paired t-test, P = 0.756/0.011). Long axis scores were 4.09/4.37 vs. 3.99/4.29 (P = 0.475/0.463). Mean ejection fraction was 60.8/61.4 for breath-held acquisitions vs. 60.3/60.3 for real-time acquisitions (P = 0.439/0.093). No significant differences were seen in end-diastolic volume (P = 0.460/0.268) but there was a trend towards a small overestimation of end-systolic volume of 2.0/2.5 ml, which did not reach statistical significance (P = 0.052/0.083).

Conclusions

Real-time free breathing CMR can be used to obtain high quality retrospectively gated cine images in 16-20s per slice. Volumetric measurements and image quality scores were similar in images from breath-held segmented and free breathing, real-time acquisitions. Further speedup of image reconstruction is still needed.
Appendix
Available only for authorised users
Literature
1.
go back to reference Bogaert J, Dymarkowski S, Taylor AM: Clinical cardiac MRI. 2012, Berlin Heidelberg New York: Springer, 2CrossRef Bogaert J, Dymarkowski S, Taylor AM: Clinical cardiac MRI. 2012, Berlin Heidelberg New York: Springer, 2CrossRef
2.
go back to reference Lee VS, Resnick D, Bundy JM, Simonetti OP, Lee P, Weinreb JC: Cardiac function: MR evaluation in One breath hold with real-time true fast imaging with steady-state Precession1. Radiology. 2002, 222: 835-42. 10.1148/radiol.2223011156.CrossRefPubMed Lee VS, Resnick D, Bundy JM, Simonetti OP, Lee P, Weinreb JC: Cardiac function: MR evaluation in One breath hold with real-time true fast imaging with steady-state Precession1. Radiology. 2002, 222: 835-42. 10.1148/radiol.2223011156.CrossRefPubMed
3.
go back to reference Kühl HP, Spuentrup E, Wall A, Franke A, Schröder J, Heussen N, Hanrath P, Günther RW, Buecker A: Assessment of myocardial function with interactive Non–breath-hold real-time MR imaging: comparison with echocardiography and breath-hold cine MR Imaging1. Radiology. 2004, 231: 198-207. 10.1148/radiol.2311021237.CrossRefPubMed Kühl HP, Spuentrup E, Wall A, Franke A, Schröder J, Heussen N, Hanrath P, Günther RW, Buecker A: Assessment of myocardial function with interactive Non–breath-hold real-time MR imaging: comparison with echocardiography and breath-hold cine MR Imaging1. Radiology. 2004, 231: 198-207. 10.1148/radiol.2311021237.CrossRefPubMed
4.
go back to reference Kaji S, Yang PC, Kerr AB, Tang WHW, Meyer CH, Macovski A, Pauly JM, Nishimura DG, Hu BS: Rapid evaluation of left ventricular volume and mass without breath-holding using real-time interactive cardiac magnetic resonance imaging system. J Am Coll Cardiol. 2001, 38: 527-33. 10.1016/S0735-1097(01)01399-7.CrossRefPubMed Kaji S, Yang PC, Kerr AB, Tang WHW, Meyer CH, Macovski A, Pauly JM, Nishimura DG, Hu BS: Rapid evaluation of left ventricular volume and mass without breath-holding using real-time interactive cardiac magnetic resonance imaging system. J Am Coll Cardiol. 2001, 38: 527-33. 10.1016/S0735-1097(01)01399-7.CrossRefPubMed
5.
go back to reference Muthurangu V, Lurz P, Critchely JD, Deanfield JE, Taylor AM, Hansen MS: Real-time assessment of right and left ventricular volumes and function in patients with congenital heart disease by using high spatiotemporal resolution radial k-t SENSE1. Radiology. 2008, 248: 782-91. 10.1148/radiol.2482071717.CrossRefPubMed Muthurangu V, Lurz P, Critchely JD, Deanfield JE, Taylor AM, Hansen MS: Real-time assessment of right and left ventricular volumes and function in patients with congenital heart disease by using high spatiotemporal resolution radial k-t SENSE1. Radiology. 2008, 248: 782-91. 10.1148/radiol.2482071717.CrossRefPubMed
6.
go back to reference Kellman P, Epstein FH, McVeigh ER: Adaptive sensitivity encoding incorporating temporal filtering (TSENSE). Magn Reson Med. 2001, 45: 846-52. 10.1002/mrm.1113.CrossRefPubMed Kellman P, Epstein FH, McVeigh ER: Adaptive sensitivity encoding incorporating temporal filtering (TSENSE). Magn Reson Med. 2001, 45: 846-52. 10.1002/mrm.1113.CrossRefPubMed
7.
go back to reference Breuer FA, Kellman P, Griswold MA, Jakob PM: Dynamic autocalibrated parallel imaging using temporal GRAPPA (TGRAPPA). Magn Reson Med. 2005, 53: 981-5. 10.1002/mrm.20430.CrossRefPubMed Breuer FA, Kellman P, Griswold MA, Jakob PM: Dynamic autocalibrated parallel imaging using temporal GRAPPA (TGRAPPA). Magn Reson Med. 2005, 53: 981-5. 10.1002/mrm.20430.CrossRefPubMed
8.
go back to reference Madore B: UNFOLD-SENSE: a parallel MRI method with self-calibration and artifact suppression. Magn Reson Med. 2004, 52: 310-20. 10.1002/mrm.20133.CrossRefPubMed Madore B: UNFOLD-SENSE: a parallel MRI method with self-calibration and artifact suppression. Magn Reson Med. 2004, 52: 310-20. 10.1002/mrm.20133.CrossRefPubMed
9.
go back to reference Tsao J, Boesiger P, Pruessmann KP: k-t BLAST and k-t SENSE: dynamic MRI with high frame rate exploiting spatiotemporal correlations. Magn Reson Med. 2003, 50: 1031-42. 10.1002/mrm.10611.CrossRefPubMed Tsao J, Boesiger P, Pruessmann KP: k-t BLAST and k-t SENSE: dynamic MRI with high frame rate exploiting spatiotemporal correlations. Magn Reson Med. 2003, 50: 1031-42. 10.1002/mrm.10611.CrossRefPubMed
10.
go back to reference Lustig M, Santos JM, Donoho DL, Pauly JM: k-t SPARSE: high frame rate dynamic MRI exploiting spatio-temporal sparsity. ISMRM proceeding; 6–12 May; seattle, washington, USA. 2006, 2420- Lustig M, Santos JM, Donoho DL, Pauly JM: k-t SPARSE: high frame rate dynamic MRI exploiting spatio-temporal sparsity. ISMRM proceeding; 6–12 May; seattle, washington, USA. 2006, 2420-
11.
go back to reference Liang D, DiBella EVR, Chen R-R, Ying L: k-t ISD: dynamic cardiac MR imaging using compressed sensing with iterative support detection. Magn Reson Med. 2011, 68: 41-53.PubMedCentralCrossRefPubMed Liang D, DiBella EVR, Chen R-R, Ying L: k-t ISD: dynamic cardiac MR imaging using compressed sensing with iterative support detection. Magn Reson Med. 2011, 68: 41-53.PubMedCentralCrossRefPubMed
12.
go back to reference Jung H, Park J, Yoo J, Ye JC: Radial k-t FOCUSS for high-resolution cardiac cine MRI. Magn Reson Med. 2010, 63: 68-78.PubMed Jung H, Park J, Yoo J, Ye JC: Radial k-t FOCUSS for high-resolution cardiac cine MRI. Magn Reson Med. 2010, 63: 68-78.PubMed
13.
go back to reference Jung H, Sung K, Nayak KS, Kim EY, Ye JC: k-t FOCUSS: a general compressed sensing framework for high resolution dynamic MRI. Magn Reson Med. 2009, 61: 103-16. 10.1002/mrm.21757.CrossRefPubMed Jung H, Sung K, Nayak KS, Kim EY, Ye JC: k-t FOCUSS: a general compressed sensing framework for high resolution dynamic MRI. Magn Reson Med. 2009, 61: 103-16. 10.1002/mrm.21757.CrossRefPubMed
14.
go back to reference Tsao J, Kozerke S: MRI temporal acceleration techniques. J Magn Reson Imaging. 2012, 36: 543-60. 10.1002/jmri.23640.CrossRefPubMed Tsao J, Kozerke S: MRI temporal acceleration techniques. J Magn Reson Imaging. 2012, 36: 543-60. 10.1002/jmri.23640.CrossRefPubMed
15.
go back to reference Kellman P, Chefd’hotel C, Lorenz CH, Mancini C, Arai AE, McVeigh ER: Fully automatic, retrospective enhancement of real-time acquired cardiac cine MR images using image-based navigators and respiratory motion-corrected averaging. Magn Reson Med. 2008, 59: 771-8. 10.1002/mrm.21509.CrossRefPubMed Kellman P, Chefd’hotel C, Lorenz CH, Mancini C, Arai AE, McVeigh ER: Fully automatic, retrospective enhancement of real-time acquired cardiac cine MR images using image-based navigators and respiratory motion-corrected averaging. Magn Reson Med. 2008, 59: 771-8. 10.1002/mrm.21509.CrossRefPubMed
16.
go back to reference Kellman P, Chefd’hotel C, Lorenz CH, Mancini C, Arai AE, McVeigh ER: High spatial and temporal resolution cardiac cine MRI from retrospective reconstruction of data acquired in real time using motion correction and resorting. Magn Reson Med. 2009, 62: 1557-64. 10.1002/mrm.22153.PubMedCentralCrossRefPubMed Kellman P, Chefd’hotel C, Lorenz CH, Mancini C, Arai AE, McVeigh ER: High spatial and temporal resolution cardiac cine MRI from retrospective reconstruction of data acquired in real time using motion correction and resorting. Magn Reson Med. 2009, 62: 1557-64. 10.1002/mrm.22153.PubMedCentralCrossRefPubMed
17.
go back to reference Hansen MS, Sørensen TS, Arai AE, Kellman P: Retrospective reconstruction of high temporal resolution cine images from real-time MRI using iterative motion correction. Magn Reson Med. 2012, 68: 741-50. 10.1002/mrm.23284.PubMedCentralCrossRefPubMed Hansen MS, Sørensen TS, Arai AE, Kellman P: Retrospective reconstruction of high temporal resolution cine images from real-time MRI using iterative motion correction. Magn Reson Med. 2012, 68: 741-50. 10.1002/mrm.23284.PubMedCentralCrossRefPubMed
18.
go back to reference Batchelor PG, Atkinson D, Irarrazaval P, Hill DLG, Hajnal J, Larkman D: Matrix description of general motion correction applied to multishot images. Magn Reson Med. 2005, 54: 1273-80. 10.1002/mrm.20656.CrossRefPubMed Batchelor PG, Atkinson D, Irarrazaval P, Hill DLG, Hajnal J, Larkman D: Matrix description of general motion correction applied to multishot images. Magn Reson Med. 2005, 54: 1273-80. 10.1002/mrm.20656.CrossRefPubMed
19.
go back to reference Inati SJ, Hansen MS, Kellman P: A solution to the phase problem in adaptive coil combination. ISMRM proceeding; 20–26 april; Salt Lake City, Utah, USA. 2013, 2672 Inati SJ, Hansen MS, Kellman P: A solution to the phase problem in adaptive coil combination. ISMRM proceeding; 20–26 april; Salt Lake City, Utah, USA. 2013, 2672
20.
go back to reference Lustig M, Pauly JM: SPIRiT: iterative self-consistent parallel imaging reconstruction from arbitrary k-space. Magn Reson Med. 2010, 64: 457-71.PubMedCentralPubMed Lustig M, Pauly JM: SPIRiT: iterative self-consistent parallel imaging reconstruction from arbitrary k-space. Magn Reson Med. 2010, 64: 457-71.PubMedCentralPubMed
21.
go back to reference Xue H, Ding Y, Chang T-C, Guetter C, Raman S, Jolly M-P, Simonetti O: Improved real-time exercise stress cardiac cine imaging using self consistent parallel imaging with temporal sensitivity estimation (TSPIRIT). J Cardiovasc Magn Reson. 2012, 14: P254-10.1186/1532-429X-14-S1-P254.PubMedCentralCrossRef Xue H, Ding Y, Chang T-C, Guetter C, Raman S, Jolly M-P, Simonetti O: Improved real-time exercise stress cardiac cine imaging using self consistent parallel imaging with temporal sensitivity estimation (TSPIRIT). J Cardiovasc Magn Reson. 2012, 14: P254-10.1186/1532-429X-14-S1-P254.PubMedCentralCrossRef
22.
go back to reference Paige CC, Saunders MA: LSQR: an algorithm for sparse linear equations and sparse least squares. ACM Trans Math Softw. 1982, 8: 43-71. 10.1145/355984.355989.CrossRef Paige CC, Saunders MA: LSQR: an algorithm for sparse linear equations and sparse least squares. ACM Trans Math Softw. 1982, 8: 43-71. 10.1145/355984.355989.CrossRef
23.
go back to reference Hansen MS, Sørensen TS: Gadgetron: an open source framework for medical image reconstruction. Magn Reson Med. 2012, 69: 1768-76.CrossRefPubMed Hansen MS, Sørensen TS: Gadgetron: an open source framework for medical image reconstruction. Magn Reson Med. 2012, 69: 1768-76.CrossRefPubMed
Metadata
Title
High spatial and temporal resolution retrospective cine cardiovascular magnetic resonance from shortened free breathing real-time acquisitions
Authors
Hui Xue
Peter Kellman
Gina LaRocca
Andrew E Arai
Michael S Hansen
Publication date
01-12-2013
Publisher
BioMed Central
Published in
Journal of Cardiovascular Magnetic Resonance / Issue 1/2013
Electronic ISSN: 1532-429X
DOI
https://doi.org/10.1186/1532-429X-15-102

Other articles of this Issue 1/2013

Journal of Cardiovascular Magnetic Resonance 1/2013 Go to the issue