Published in: Cardiovascular Diabetology 1/2012

Open Access 01-12-2012 | Original investigation

Increased blood glycohemoglobin A1c levels lead to overestimation of arterial oxygen saturation by pulse oximetry in patients with type 2 diabetes

Authors: Li Jin Pu, Ying Shen, Lin Lu, Rui Yan Zhang, Qi Zhang, Wei Feng Shen

Published in: Cardiovascular Diabetology | Issue 1/2012

Login to get access

Abstract

Background

Non-enzymatic glycation increases hemoglobin-oxygen affinity and reduces oxygen delivery to tissues by altering the structure and function of hemoglobin.

Objectives

We investigated whether an elevated blood concentration of glycosylated hemoglobin (HbA1c) could induce falsely high pulse oximeter oxygen saturation (SpO2) in type 2 diabetic patients during mechanical ventilation or oxygen therapy.

Methods

Arterial oxygen saturation (SaO2) and partial pressure of oxygen (PO2) were determined with simultaneous monitoring of SpO2 in 261 type 2 diabetic patients during ventilation or oxygen inhalation.

Results

Blood concentration of HbA1c was >7% in 114 patients and ≤ 7% in 147 patients. Both SaO2 (96.2 ± 2.9%, 95% confidence interval [CI] 95.7-96.7% vs. 95.1 ± 2.8%, 95% CI 94.7-95.6%) and SpO2 (98.0 ± 2.6%, 95% CI 97.6-98.5% vs. 95.3 ± 2.8%, 95% CI 94.9-95.8%) were significantly higher in patients with HbA1c >7% than in those with HbA1c ≤ 7% (Data are mean ± SD, all p < 0.01), but PO2 did not significantly differ between the two groups. Bland-Altman analysis demonstrated a significant bias between SpO2 and SaO2 (1.83 ±0.55%, 95% CI 1.73% -1.94%) and limits of agreement (0.76% and 2.92%) in patients with HbA1c >7%. The differences between SpO2 and SaO2 correlated closely with blood HbA1c levels (Pearson’s r = 0.307, p < 0.01).

Conclusions

Elevated blood HbA1c levels lead to an overestimation of SaO2 by SpO2, suggesting that arterial blood gas analysis may be needed for type 2 diabetic patients with poor glycemic control during the treatment of hypoxemia.
Appendix
Available only for authorised users
Literature
1.
go back to reference Miedema K: Electrospray mass spectrometry for measurement of glycohemoglobin. Clin Chem. 1997, 43: 705-707.PubMed Miedema K: Electrospray mass spectrometry for measurement of glycohemoglobin. Clin Chem. 1997, 43: 705-707.PubMed
2.
go back to reference American Diabetes Association: Clinical practice recommendations. Diabetes Care. 2000, 23: S32-42. American Diabetes Association: Clinical practice recommendations. Diabetes Care. 2000, 23: S32-42.
3.
go back to reference Daida H, Takayama T, Hiro T, Yamagishi M, Hirayama A, Saito S, Yamaguchi T, Matsuzaki M: High HbA1c levels correlate with reduced plaque regression during statin treatment in patients with stable coronary artery disease: Results of the coronary atherosclerosis study measuring effects of rosuvastatin using intravascular ultrasound in Japanese subjects (COSMOS). Cardiovasc Diabetol. 2012, 11: 87-10.1186/1475-2840-11-87.PubMedCentralCrossRefPubMed Daida H, Takayama T, Hiro T, Yamagishi M, Hirayama A, Saito S, Yamaguchi T, Matsuzaki M: High HbA1c levels correlate with reduced plaque regression during statin treatment in patients with stable coronary artery disease: Results of the coronary atherosclerosis study measuring effects of rosuvastatin using intravascular ultrasound in Japanese subjects (COSMOS). Cardiovasc Diabetol. 2012, 11: 87-10.1186/1475-2840-11-87.PubMedCentralCrossRefPubMed
4.
go back to reference Nishimura R, Nakagami T, Sone H, Ohashi Y, Tajima N: Relationship between hemoglobin A1c and cardiovascular disease in mild-to-moderate hypercholesterolemic Japanese individuals: subanalysis of a large-scale randomized controlled trial. Cardiovasc Diabetol. 2011, 10: 58-10.1186/1475-2840-10-58.PubMedCentralCrossRefPubMed Nishimura R, Nakagami T, Sone H, Ohashi Y, Tajima N: Relationship between hemoglobin A1c and cardiovascular disease in mild-to-moderate hypercholesterolemic Japanese individuals: subanalysis of a large-scale randomized controlled trial. Cardiovasc Diabetol. 2011, 10: 58-10.1186/1475-2840-10-58.PubMedCentralCrossRefPubMed
5.
go back to reference Ikeda N, Iijima R, Hara H, Moroi M, Nakamura M, Sugi K: Glycated hemoglobin is associated with the complexity of coronary artery disease, even in non-diabetic adults. J Atheroscler Thromb. 2012, 19: [Epub ahead of print] Ikeda N, Iijima R, Hara H, Moroi M, Nakamura M, Sugi K: Glycated hemoglobin is associated with the complexity of coronary artery disease, even in non-diabetic adults. J Atheroscler Thromb. 2012, 19: [Epub ahead of print]
6.
go back to reference De Rosa MC, Sanna MT, Messana I, Castagnola M, Galtieri A, Tellone E, Scatena R, Botta B, Botta M, Giardina B: Glycated human hemoglobin (HbA1c): functional characteristics and molecular modeling studies. Biophys Chem. 1998, 72: 323-335. 10.1016/S0301-4622(98)00117-3.CrossRefPubMed De Rosa MC, Sanna MT, Messana I, Castagnola M, Galtieri A, Tellone E, Scatena R, Botta B, Botta M, Giardina B: Glycated human hemoglobin (HbA1c): functional characteristics and molecular modeling studies. Biophys Chem. 1998, 72: 323-335. 10.1016/S0301-4622(98)00117-3.CrossRefPubMed
7.
go back to reference Samaja M, Melotti D, Carenini A, Pozza G: Glycosylated haemoglobins and the oxygen affinity of whole blood. Diabetologia. 1982, 23: 399-402.6.CrossRefPubMed Samaja M, Melotti D, Carenini A, Pozza G: Glycosylated haemoglobins and the oxygen affinity of whole blood. Diabetologia. 1982, 23: 399-402.6.CrossRefPubMed
8.
go back to reference Bunn HF: The glycosylation of hemoglobin: relevance to diabetes mellitus. Science. 1978, 200: 21-27. 10.1126/science.635569.CrossRefPubMed Bunn HF: The glycosylation of hemoglobin: relevance to diabetes mellitus. Science. 1978, 200: 21-27. 10.1126/science.635569.CrossRefPubMed
9.
go back to reference Ditzel J: Affinity hypoxia as a pathogenic factor of microangiopathy with particular reference to diabetic retinopathy. Acta Endocrinologica Supplementum. 1980, 238: 39-55.PubMed Ditzel J: Affinity hypoxia as a pathogenic factor of microangiopathy with particular reference to diabetic retinopathy. Acta Endocrinologica Supplementum. 1980, 238: 39-55.PubMed
10.
go back to reference Jubran A, Tobin MJ: Reliability of pulse oximetry in titrating supplemental oxygen therapy in ventilator-dependent patients. Chest. 1990, 97: 1420-1425. 10.1378/chest.97.6.1420.CrossRefPubMed Jubran A, Tobin MJ: Reliability of pulse oximetry in titrating supplemental oxygen therapy in ventilator-dependent patients. Chest. 1990, 97: 1420-1425. 10.1378/chest.97.6.1420.CrossRefPubMed
11.
go back to reference Moloney ED, Kiely JL, McNicholas WT: Controlled oxygen therapy and carbon dioxide retention during exacerbations of chronic obstructive pulmonary disease. Lancet. 2001, 357: 526-528. 10.1016/S0140-6736(00)04049-6.CrossRefPubMed Moloney ED, Kiely JL, McNicholas WT: Controlled oxygen therapy and carbon dioxide retention during exacerbations of chronic obstructive pulmonary disease. Lancet. 2001, 357: 526-528. 10.1016/S0140-6736(00)04049-6.CrossRefPubMed
12.
go back to reference Kress JP, Pohlman AS, Hall JB: Determination of hemoglobin saturation in patients with acute sickle chest syndrome: a comparison of arterial blood gases and pulse oximetry. Chest. 1999, 115: 1316-1320. 10.1378/chest.115.5.1316.CrossRefPubMed Kress JP, Pohlman AS, Hall JB: Determination of hemoglobin saturation in patients with acute sickle chest syndrome: a comparison of arterial blood gases and pulse oximetry. Chest. 1999, 115: 1316-1320. 10.1378/chest.115.5.1316.CrossRefPubMed
13.
go back to reference Wilson BJ, Cowan HJ, Lord JA, Zuege DJ, Zygun DA: The accuracy of pulse oximetry in emergency department patients with severe sepsis and septic shock: a retrospective cohort study. BMC Emerg Med. 2010, 10: 9-10.1186/1471-227X-10-9.PubMedCentralCrossRefPubMed Wilson BJ, Cowan HJ, Lord JA, Zuege DJ, Zygun DA: The accuracy of pulse oximetry in emergency department patients with severe sepsis and septic shock: a retrospective cohort study. BMC Emerg Med. 2010, 10: 9-10.1186/1471-227X-10-9.PubMedCentralCrossRefPubMed
14.
go back to reference Dyer DG, Dunn JA, Thorpe SR, Bailie KE, Lyons TJ, McCance DR, Baynes JW: Accumulation of Maillard reaction products in skin collagen in diabetes and aging. J Clin Invest. 1993, 91: 2463-2469. 10.1172/JCI116481.PubMedCentralCrossRefPubMed Dyer DG, Dunn JA, Thorpe SR, Bailie KE, Lyons TJ, McCance DR, Baynes JW: Accumulation of Maillard reaction products in skin collagen in diabetes and aging. J Clin Invest. 1993, 91: 2463-2469. 10.1172/JCI116481.PubMedCentralCrossRefPubMed
15.
go back to reference Bos DC, de Ranitz-Greven WL, de Valk HW: Advanced glycation end products, measured as skin autofluorescence and diabetes complications: a systematic review. Diabetes Technol Ther. 2011, 13: 773-779. 10.1089/dia.2011.0034.CrossRefPubMed Bos DC, de Ranitz-Greven WL, de Valk HW: Advanced glycation end products, measured as skin autofluorescence and diabetes complications: a systematic review. Diabetes Technol Ther. 2011, 13: 773-779. 10.1089/dia.2011.0034.CrossRefPubMed
16.
go back to reference Expert Committee on the Diagnosis and Classification of Diabetes Mellitus: Report of the expert committee on the diagnosis and classification of diabetes mellitus. Diabetes Care. 2003, 26 (Suppl 1): S5-20. Expert Committee on the Diagnosis and Classification of Diabetes Mellitus: Report of the expert committee on the diagnosis and classification of diabetes mellitus. Diabetes Care. 2003, 26 (Suppl 1): S5-20.
17.
go back to reference Kassaian SE, Goodarzynejad H, Boroumand MA, Salarifar M, Masoudkabir F, Mohajeri-Tehrani MR, Pourhoseini H, Sadeghian S, Ramezanpour N, Alidoosti M, Hakki E, Saadat S, Nematipour E: Glycosylated hemoglobin (HbA1c) levels and clinical outcomes in diabetic patients following coronary artery stenting. Cardiovasc Diabetol. 2012, 11: 82-10.1186/1475-2840-11-82.PubMedCentralCrossRefPubMed Kassaian SE, Goodarzynejad H, Boroumand MA, Salarifar M, Masoudkabir F, Mohajeri-Tehrani MR, Pourhoseini H, Sadeghian S, Ramezanpour N, Alidoosti M, Hakki E, Saadat S, Nematipour E: Glycosylated hemoglobin (HbA1c) levels and clinical outcomes in diabetic patients following coronary artery stenting. Cardiovasc Diabetol. 2012, 11: 82-10.1186/1475-2840-11-82.PubMedCentralCrossRefPubMed
18.
19.
go back to reference Bland JM, Altman DG: Measuring agreement in method comparison studies. Stat Methods Med Res. 1999, 8: 161-179. 10.1191/096228099668524590.CrossRef Bland JM, Altman DG: Measuring agreement in method comparison studies. Stat Methods Med Res. 1999, 8: 161-179. 10.1191/096228099668524590.CrossRef
20.
go back to reference de Jonge E, Peelen L, Keijzers PJ, Joore H, de Lange D, van der Voort PH, Bosman RJ, de Waal RA, Wesselink R, de Keizer NF: Association between administered oxygen, arterial partial oxygen pressure and mortality in mechanically ventilated intensive care unit patients. Crit Care. 2008, 12: R156-10.1186/cc7150.PubMedCentralCrossRefPubMed de Jonge E, Peelen L, Keijzers PJ, Joore H, de Lange D, van der Voort PH, Bosman RJ, de Waal RA, Wesselink R, de Keizer NF: Association between administered oxygen, arterial partial oxygen pressure and mortality in mechanically ventilated intensive care unit patients. Crit Care. 2008, 12: R156-10.1186/cc7150.PubMedCentralCrossRefPubMed
21.
go back to reference Birnbaum S: Pulse oximetry: identifying its applications, coding, and reimbursement. Chest. 2009, 135: 838-841. 10.1378/chest.07-3127.CrossRefPubMed Birnbaum S: Pulse oximetry: identifying its applications, coding, and reimbursement. Chest. 2009, 135: 838-841. 10.1378/chest.07-3127.CrossRefPubMed
22.
go back to reference Van de Louw A, Cracco C, Cerf C, Harf A, Duvaldestin P, Lemaire F, Brochard L: Accuracy of pulse oximetry in the intensive care unit. Intensive Care Med. 2001, 27: 1606-1613. 10.1007/s001340101064.CrossRefPubMed Van de Louw A, Cracco C, Cerf C, Harf A, Duvaldestin P, Lemaire F, Brochard L: Accuracy of pulse oximetry in the intensive care unit. Intensive Care Med. 2001, 27: 1606-1613. 10.1007/s001340101064.CrossRefPubMed
23.
go back to reference Lee WW, Mayberry K, Crapo R, Jensen RL: The accuracy of pulse oximetry in the emergency department. Am J Emerg Med. 2000, 18: 427-431.CrossRefPubMed Lee WW, Mayberry K, Crapo R, Jensen RL: The accuracy of pulse oximetry in the emergency department. Am J Emerg Med. 2000, 18: 427-431.CrossRefPubMed
24.
go back to reference Seguin P, Le Rouzo A, Tanguy M, Guillou YM, Feuillu A, Malledant Y: Evidence for the need of bedside accuracy of pulse oximetry in an intensive care unit. Crit Care Med. 2000, 28: 703-706. 10.1097/00003246-200003000-00017.CrossRefPubMed Seguin P, Le Rouzo A, Tanguy M, Guillou YM, Feuillu A, Malledant Y: Evidence for the need of bedside accuracy of pulse oximetry in an intensive care unit. Crit Care Med. 2000, 28: 703-706. 10.1097/00003246-200003000-00017.CrossRefPubMed
26.
go back to reference Mihm FG, Halperin BD: Noninvasive detection of profound arterial desaturations using a pulse oximetry device. Anesthesiology. 1985, 62: 85-87. 10.1097/00000542-198501000-00020.CrossRefPubMed Mihm FG, Halperin BD: Noninvasive detection of profound arterial desaturations using a pulse oximetry device. Anesthesiology. 1985, 62: 85-87. 10.1097/00000542-198501000-00020.CrossRefPubMed
27.
go back to reference Marschner JP, Seidlitz T, Rietbrock N: Effect of 2,3-diphosphoglycerate on O2-dissociation kinetics of hemoglobin and glycosylated hemoglobin using the stopped flow technique and an improved in vitro method for hemoglobin glycosylation. Int J Clin Pharmacol Ther. 1994, 32: 116-121.PubMed Marschner JP, Seidlitz T, Rietbrock N: Effect of 2,3-diphosphoglycerate on O2-dissociation kinetics of hemoglobin and glycosylated hemoglobin using the stopped flow technique and an improved in vitro method for hemoglobin glycosylation. Int J Clin Pharmacol Ther. 1994, 32: 116-121.PubMed
28.
go back to reference Peterson KP, Pavlovich JG, Goldstein D, Little R, England J, Peterson CM: What is hemoglobin A1c? An analysis of glycated hemoglobins by electrospray ionization mass spectrometry. Clin Chem. 1998, 44: 1951-1958.PubMed Peterson KP, Pavlovich JG, Goldstein D, Little R, England J, Peterson CM: What is hemoglobin A1c? An analysis of glycated hemoglobins by electrospray ionization mass spectrometry. Clin Chem. 1998, 44: 1951-1958.PubMed
29.
go back to reference Roberts NB, Amara AB, Morris M, Green BN: Long-term evaluation of electrospray ionization mass spectrometric analysis of glycated hemoglobin. Clin Chem. 2001, 47: 316-321.PubMed Roberts NB, Amara AB, Morris M, Green BN: Long-term evaluation of electrospray ionization mass spectrometric analysis of glycated hemoglobin. Clin Chem. 2001, 47: 316-321.PubMed
30.
go back to reference Martins RA, Jones JG, Cumming SP, Coelho e Silva MJ, Teixeira AM, Veríssimo MT: Glycated hemoglobin and associated risk factors in older adults. Cardiovasc Diabetol. 2012, 11: 13-10.1186/1475-2840-11-13.PubMedCentralCrossRefPubMed Martins RA, Jones JG, Cumming SP, Coelho e Silva MJ, Teixeira AM, Veríssimo MT: Glycated hemoglobin and associated risk factors in older adults. Cardiovasc Diabetol. 2012, 11: 13-10.1186/1475-2840-11-13.PubMedCentralCrossRefPubMed
31.
go back to reference Dunn JA, McCance DR, Thorpe SR, Lyons TJ, Baynes JW: Age-dependent accumulation of Nε-(carboxymethyl)lysine and Nε-(carboxymethyl)hydroxylysine in human skin collagen. Biochemistry. 1991, 30: 1205-1210. 10.1021/bi00219a007.CrossRefPubMed Dunn JA, McCance DR, Thorpe SR, Lyons TJ, Baynes JW: Age-dependent accumulation of Nε-(carboxymethyl)lysine and Nε-(carboxymethyl)hydroxylysine in human skin collagen. Biochemistry. 1991, 30: 1205-1210. 10.1021/bi00219a007.CrossRefPubMed
32.
go back to reference Gerrits EG, Lutgers HL, Kleefstra N, Graaff R, Groenier KH, Smit AJ, Gans RO, Bilo HJ: Skin autofluorescence: a tool to identify type 2 diabetic patients at risk for developing microvascular complications. Diabetes Care. 2008, 31: 517-521.CrossRefPubMed Gerrits EG, Lutgers HL, Kleefstra N, Graaff R, Groenier KH, Smit AJ, Gans RO, Bilo HJ: Skin autofluorescence: a tool to identify type 2 diabetic patients at risk for developing microvascular complications. Diabetes Care. 2008, 31: 517-521.CrossRefPubMed
33.
go back to reference Bickler PE, Feiner JR, Severinghaus JW: Effects of skin pigmentation on pulse oximeter accuracy at low saturation. Anesthesiology. 2005, 102: 715-719. 10.1097/00000542-200504000-00004.CrossRefPubMed Bickler PE, Feiner JR, Severinghaus JW: Effects of skin pigmentation on pulse oximeter accuracy at low saturation. Anesthesiology. 2005, 102: 715-719. 10.1097/00000542-200504000-00004.CrossRefPubMed
34.
go back to reference Feiner JR, Severinghaus JW, Bickler PE: Dark skin decreases the accuracy of pulse oximeters at low oxygen saturation: the effects of oximeter probe type and gender. Anesth Analg. 2007, 105: S18-23. 10.1213/01.ane.0000285988.35174.d9.CrossRefPubMed Feiner JR, Severinghaus JW, Bickler PE: Dark skin decreases the accuracy of pulse oximeters at low oxygen saturation: the effects of oximeter probe type and gender. Anesth Analg. 2007, 105: S18-23. 10.1213/01.ane.0000285988.35174.d9.CrossRefPubMed
35.
go back to reference Severinghaus JW: Simple, accurate equations for human blood O2 dissociation computations. J Appl Physiol. 1979, 46: 599-602.PubMed Severinghaus JW: Simple, accurate equations for human blood O2 dissociation computations. J Appl Physiol. 1979, 46: 599-602.PubMed
Metadata
Title
Increased blood glycohemoglobin A1c levels lead to overestimation of arterial oxygen saturation by pulse oximetry in patients with type 2 diabetes
Authors
Li Jin Pu
Ying Shen
Lin Lu
Rui Yan Zhang
Qi Zhang
Wei Feng Shen
Publication date
01-12-2012
Publisher
BioMed Central
Published in
Cardiovascular Diabetology / Issue 1/2012
Electronic ISSN: 1475-2840
DOI
https://doi.org/10.1186/1475-2840-11-110

Other articles of this Issue 1/2012

Cardiovascular Diabetology 1/2012 Go to the issue