Fluoroscopic views for safe insertion of lag screws into the posterior column of the acetabulum

Authors: Wei Chen, Zekun Zhang, Yang Lu, Jia Li, Yingze Zhang, Yong Shen

Published in: BMC Musculoskeletal Disorders | Issue 1/2014

Login to get access

Abstract

Background

Percutaneous lag screw fixation is an alternative treatment for non-displaced or minimally displaced posterior column fractures. This study aims to explore new fluoroscopic views of the acetabulum for safe percutaneous insertion of posterior column lag screws.

Methods

Axial computed tomography (CT) scans were taken of sixteen embalmed adult cadavers. The axial CT images at the level of the middle height of the acetabulum were selected. The angle (angle α) between the posterior cortex of the posterior column (PCPC) and the line intersecting the axial plane and the coronal plane, and the angle (angle β) between the medial wall and the line intersecting the axial plane and the sagittal plane were identified and measured. Tangential views of the PCPC and medial wall were obtained by referencing the measured angles. A lag screw was inserted into the posterior columns of the sixteen pelvic specimens under fluoroscopic guidance using an iliac oblique view and the two tangential views. CT scans were performed to evaluate the lag screw position. Axial CT images of 52 volunteers were obtained and the angles α and β were measured following the same methods used for the cadaveric specimens.

Results

The angles α and β for the specimens were 29.3±2.8.1 and 8.1±1.4 degrees, respectively. On the tangential view of the PCPC, the posterior cortex appears as a nearly straight line between the lesser and greater sciatic notches. On the tangential view of the medial wall, the medial wall appears as a distinct straight line. Using these radiographic images, the lag screws were inserted into the posterior columns of bony pelvic specimens. Screw placement was confirmed by CT, and found to be fully intraosseous in all cases without any cortical breaches. The angles α and β were 30.4±4.1 and 9.2±1.9 degrees for male volunteers and 28.5±3.7 and 7.7±1.8 degrees for female volunteers, significant difference in these angles between cadaveric specimens and human volunteers.

Conclusion

The tangential views of both the PCPC and medial wall can be obtained following the aforementioned methods The oblique iliac view and the two tangential views enable safe insertion of posterior column lag screws.
Appendix
Available only for authorised users
Literature
1.
go back to reference Mouhsine E, Garofalo R, Borens O, Wettstein M, Blanc CH, Fischer JF, Moretti B, Leyvraz PF: Percutaneous retrograde screwing for stabilisation of acetabular fractures. Injury. 2005, 36 (11): 1330-1336.CrossRefPubMed Mouhsine E, Garofalo R, Borens O, Wettstein M, Blanc CH, Fischer JF, Moretti B, Leyvraz PF: Percutaneous retrograde screwing for stabilisation of acetabular fractures. Injury. 2005, 36 (11): 1330-1336.CrossRefPubMed
2.
go back to reference Parker PJ, Copeland C: Percutaneous fluoroscopic screw fixation of acetabular fractures. Injury. 1997, 28 (9–10): 597-600.CrossRefPubMed Parker PJ, Copeland C: Percutaneous fluoroscopic screw fixation of acetabular fractures. Injury. 1997, 28 (9–10): 597-600.CrossRefPubMed
3.
go back to reference Sanders MB, Starr AJ, Reinert C: Percutaneous screw fixation of acetabular fractures in elderly patients. Curr Opin Orthop. 2006, 17 (1): 17-24. Sanders MB, Starr AJ, Reinert C: Percutaneous screw fixation of acetabular fractures in elderly patients. Curr Opin Orthop. 2006, 17 (1): 17-24.
4.
go back to reference Starr AJ, Reinert CM, Jones AL: Percutaneous fixation of the columns of the acetabulum: a new technique. J Orthop Trauma. 1998, 12 (1): 51-58.CrossRefPubMed Starr AJ, Reinert CM, Jones AL: Percutaneous fixation of the columns of the acetabulum: a new technique. J Orthop Trauma. 1998, 12 (1): 51-58.CrossRefPubMed
5.
go back to reference Gay SB, Sistrom C, Wang GJ, Kahler DA, Boman T, McHugh N, Goitz HT: Percutaneous screw fixation of acetabular fractures with CT guidance: preliminary results of a new technique. AJR Am J Roentgenol. 1992, 158 (4): 819-822.CrossRefPubMed Gay SB, Sistrom C, Wang GJ, Kahler DA, Boman T, McHugh N, Goitz HT: Percutaneous screw fixation of acetabular fractures with CT guidance: preliminary results of a new technique. AJR Am J Roentgenol. 1992, 158 (4): 819-822.CrossRefPubMed
6.
go back to reference Starr AJ, Jones AL, Reinert CM, Borer DS: Preliminary results and complications following limited open reduction and percutaneous screw fixation of displaced fractures of the acetabulum. Injury. 2001, 32 (Suppl 1): SA45-SA50.CrossRefPubMed Starr AJ, Jones AL, Reinert CM, Borer DS: Preliminary results and complications following limited open reduction and percutaneous screw fixation of displaced fractures of the acetabulum. Injury. 2001, 32 (Suppl 1): SA45-SA50.CrossRefPubMed
7.
go back to reference Zura RD, Kahler DM: A transverse acetabular nonunion treated with computer-assisted percutaneous internal fixation. A case report. J Bone Joint Surg Am. 2000, 82 (2): 219-224.PubMed Zura RD, Kahler DM: A transverse acetabular nonunion treated with computer-assisted percutaneous internal fixation. A case report. J Bone Joint Surg Am. 2000, 82 (2): 219-224.PubMed
8.
go back to reference Brown GA, Willis MC, Firoozbakhsh K, Barmada A, Tessman CL, Montgomery A: Computed tomography image-guided surgery in complex acetabular fractures. Clin Orthop Relat Res. 2000, 370: 219-226.CrossRefPubMed Brown GA, Willis MC, Firoozbakhsh K, Barmada A, Tessman CL, Montgomery A: Computed tomography image-guided surgery in complex acetabular fractures. Clin Orthop Relat Res. 2000, 370: 219-226.CrossRefPubMed
9.
go back to reference Mears DC: Surgical treatment of acetabular fractures in elderly patients with osteoporotic bone. J Am Acad Orthop Surg. 1999, 7 (2): 128-141.CrossRefPubMed Mears DC: Surgical treatment of acetabular fractures in elderly patients with osteoporotic bone. J Am Acad Orthop Surg. 1999, 7 (2): 128-141.CrossRefPubMed
10.
go back to reference Templeman DC, Olson S, Moed BR, Duwelius P, Matta JM: Surgical treatment of acetabular fractures. Instr Course Lect. 1999, 48: 481-496.PubMed Templeman DC, Olson S, Moed BR, Duwelius P, Matta JM: Surgical treatment of acetabular fractures. Instr Course Lect. 1999, 48: 481-496.PubMed
11.
go back to reference Shahulhameed A, Roberts CS, Pomeroy CL, Acland RD, Giannoudis PV: Mapping the columns of the acetabulum–implications for percutaneous fixation. Injury. 2010, 41 (4): 339-342.CrossRefPubMed Shahulhameed A, Roberts CS, Pomeroy CL, Acland RD, Giannoudis PV: Mapping the columns of the acetabulum–implications for percutaneous fixation. Injury. 2010, 41 (4): 339-342.CrossRefPubMed
12.
go back to reference Yang JH, Chouhan DK, Oh KJ: Percutaneous screw fixation of acetabular fractures: applicability of hip arthroscopy. Arthroscopy. 2010, 26 (11): 1556-1561.CrossRefPubMed Yang JH, Chouhan DK, Oh KJ: Percutaneous screw fixation of acetabular fractures: applicability of hip arthroscopy. Arthroscopy. 2010, 26 (11): 1556-1561.CrossRefPubMed
13.
go back to reference Mosheiff R, Khoury A, Weil Y, Liebergall M: First generation computerized fluoroscopic navigation in percutaneous pelvic surgery. J Orthop Trauma. 2004, 18 (2): 106-111.CrossRefPubMed Mosheiff R, Khoury A, Weil Y, Liebergall M: First generation computerized fluoroscopic navigation in percutaneous pelvic surgery. J Orthop Trauma. 2004, 18 (2): 106-111.CrossRefPubMed
14.
go back to reference Geijer M, El-Khoury GY: Imaging of the acetabulum in the era of multidetector computed tomography. Emerg Radiol. 2007, 14 (5): 271-287.CrossRefPubMed Geijer M, El-Khoury GY: Imaging of the acetabulum in the era of multidetector computed tomography. Emerg Radiol. 2007, 14 (5): 271-287.CrossRefPubMed
15.
go back to reference Carmack DB, Moed BR, McCarroll K, Freccero D: Accuracy of detecting screw penetration of the acetabulum with intraoperative fluoroscopy and computed tomography. J Bone Joint Surg Am. 2001, 83-A (9): 1370-1375.PubMed Carmack DB, Moed BR, McCarroll K, Freccero D: Accuracy of detecting screw penetration of the acetabulum with intraoperative fluoroscopy and computed tomography. J Bone Joint Surg Am. 2001, 83-A (9): 1370-1375.PubMed
16.
go back to reference Letournel E, Judet R: Radiology of the Normal Acetabulum. Fractures of the Acetabulum. Edited by: Letournel E, Judet R, Elson RA. 1993, Berlin: Springer-Verlag, 29-61. 2CrossRef Letournel E, Judet R: Radiology of the Normal Acetabulum. Fractures of the Acetabulum. Edited by: Letournel E, Judet R, Elson RA. 1993, Berlin: Springer-Verlag, 29-61. 2CrossRef
17.
go back to reference Magid D, Fishman EK, Brooker AF, Mandelbaum BR, Siegelman SS: Multiplanar computed tomography of acetabular fractures. J Comput Assist Tomogr. 1986, 10 (5): 778-783.CrossRefPubMed Magid D, Fishman EK, Brooker AF, Mandelbaum BR, Siegelman SS: Multiplanar computed tomography of acetabular fractures. J Comput Assist Tomogr. 1986, 10 (5): 778-783.CrossRefPubMed
18.
go back to reference Mandelbaum BR, Magid D, Fishman EK, Burgess AR, Bosse M, Brooker AF: Multiplanar computed tomography: a multidimensional tool for evaluation and treatment of acetabular fractures. J Comput Tomogr. 1987, 11 (2): 167-173.CrossRefPubMed Mandelbaum BR, Magid D, Fishman EK, Burgess AR, Bosse M, Brooker AF: Multiplanar computed tomography: a multidimensional tool for evaluation and treatment of acetabular fractures. J Comput Tomogr. 1987, 11 (2): 167-173.CrossRefPubMed
19.
go back to reference Matta JM: Fractures of the acetabulum: accuracy of reduction and clinical results in patients managed operatively within three weeks after the injury. J Bone Joint Surg Am. 1996, 78 (11): 1632-1645.PubMed Matta JM: Fractures of the acetabulum: accuracy of reduction and clinical results in patients managed operatively within three weeks after the injury. J Bone Joint Surg Am. 1996, 78 (11): 1632-1645.PubMed
20.
go back to reference Olson SA, Matta JM: The computerized tomography subchondral arc: a new method of assessing acetabular articular continuity after fracture (a preliminary report). J Orthop Trauma. 1993, 7 (5): 402-413.CrossRefPubMed Olson SA, Matta JM: The computerized tomography subchondral arc: a new method of assessing acetabular articular continuity after fracture (a preliminary report). J Orthop Trauma. 1993, 7 (5): 402-413.CrossRefPubMed
21.
go back to reference White MS: Three-dimensional computed tomography in the assessment of fractures of the acetabulum. Injury. 1991, 22 (1): 13-19.CrossRefPubMed White MS: Three-dimensional computed tomography in the assessment of fractures of the acetabulum. Injury. 1991, 22 (1): 13-19.CrossRefPubMed
22.
go back to reference Wright R, Barrett K, Christie MJ, Johnson KD: Acetabular fractures: long-term follow-up of open reduction and internal fixation. J Orthop Trauma. 1994, 8 (5): 397-403.CrossRefPubMed Wright R, Barrett K, Christie MJ, Johnson KD: Acetabular fractures: long-term follow-up of open reduction and internal fixation. J Orthop Trauma. 1994, 8 (5): 397-403.CrossRefPubMed
23.
go back to reference Wang Q, Zhang Y, Pan J, Peng A, Wu X, Wang P, Li H, Zhang S, Zhang X, Zhao L: Applied anatomy of retrograde lag screw internal fixation in the posterior column of the acetabulum. Chin J Clin Anat. 2006, 24 (1): 36-38. Wang Q, Zhang Y, Pan J, Peng A, Wu X, Wang P, Li H, Zhang S, Zhang X, Zhao L: Applied anatomy of retrograde lag screw internal fixation in the posterior column of the acetabulum. Chin J Clin Anat. 2006, 24 (1): 36-38.
24.
go back to reference Kellam JF, Tile M: Surgical Techniques. Fractures of the Pelvis and Acetabulum. Edited by: Tile M. 1995, Baltimore: Williams and Wilkins, 355-396. 2 Kellam JF, Tile M: Surgical Techniques. Fractures of the Pelvis and Acetabulum. Edited by: Tile M. 1995, Baltimore: Williams and Wilkins, 355-396. 2
25.
go back to reference Ebraheim NA, Savolaine ER, Hoeflinger MJ, Jackson WT: Radiological diagnosis of screw penetration of the hip joint in acetabular fracture reconstruction. J Orthop Trauma. 1989, 3 (3): 196-201.CrossRefPubMed Ebraheim NA, Savolaine ER, Hoeflinger MJ, Jackson WT: Radiological diagnosis of screw penetration of the hip joint in acetabular fracture reconstruction. J Orthop Trauma. 1989, 3 (3): 196-201.CrossRefPubMed
26.
go back to reference Mears DC, MacLeod MD: Acetabular Fractures: Triradiate and Modified Triradiate Approaches. Master Techniques in Orthopaedic Surgery: Fractures. Edited by: Wiss DA. 1998, Philadelphia: Lippincott-Raven, 697-724. Mears DC, MacLeod MD: Acetabular Fractures: Triradiate and Modified Triradiate Approaches. Master Techniques in Orthopaedic Surgery: Fractures. Edited by: Wiss DA. 1998, Philadelphia: Lippincott-Raven, 697-724.
27.
go back to reference Helfet DL, Bartlett CS, Malkani AL: Acetabular Fractures: Extended Iliofemoral Approach. Master Techniques in Orthopaedic Surgery: Fractures. Edited by: Wiss DA. 1998, Philadelphia: Lippincott-Raven, 675-696. Helfet DL, Bartlett CS, Malkani AL: Acetabular Fractures: Extended Iliofemoral Approach. Master Techniques in Orthopaedic Surgery: Fractures. Edited by: Wiss DA. 1998, Philadelphia: Lippincott-Raven, 675-696.
28.
go back to reference Norris BL, Hahn DH, Bosse MJ, Kellam JF, Sims SH: Intraoperative fluoroscopy to evaluate fracture reduction and hardware placement during acetabular surgery. J Orthop Trauma. 1999, 13 (6): 414-417.CrossRefPubMed Norris BL, Hahn DH, Bosse MJ, Kellam JF, Sims SH: Intraoperative fluoroscopy to evaluate fracture reduction and hardware placement during acetabular surgery. J Orthop Trauma. 1999, 13 (6): 414-417.CrossRefPubMed
29.
go back to reference Kahler DM: Navigated long-bone fracture reduction. J Bone Joint Surg Am. 2009, 91 (Suppl 1): 102-107.CrossRefPubMed Kahler DM: Navigated long-bone fracture reduction. J Bone Joint Surg Am. 2009, 91 (Suppl 1): 102-107.CrossRefPubMed
30.
go back to reference Gras F, Marintschev I, Klos K, Muckley T, Hofmann GO, Kahler DM: Screw placement for acetabular fractures: which navigation modality (2-dimensional vs. 3-dimensional) should be used? An experimental study. J Orthop Trauma. 2012, 26 (8): 466-473.CrossRefPubMed Gras F, Marintschev I, Klos K, Muckley T, Hofmann GO, Kahler DM: Screw placement for acetabular fractures: which navigation modality (2-dimensional vs. 3-dimensional) should be used? An experimental study. J Orthop Trauma. 2012, 26 (8): 466-473.CrossRefPubMed
31.
go back to reference Gras F, Marintschev I, Mendler F, Wilharm A, Muckley T, Hofmann GO: 2D-fluoroscopic navigated screw osteosynthesis of acetabular fractures: a preliminary report. Z Orthop Unfall. 2008, 146 (2): 231-239.CrossRefPubMed Gras F, Marintschev I, Mendler F, Wilharm A, Muckley T, Hofmann GO: 2D-fluoroscopic navigated screw osteosynthesis of acetabular fractures: a preliminary report. Z Orthop Unfall. 2008, 146 (2): 231-239.CrossRefPubMed
Metadata
Title
Fluoroscopic views for safe insertion of lag screws into the posterior column of the acetabulum
Authors
Wei Chen
Zekun Zhang
Yang Lu
Jia Li
Yingze Zhang
Yong Shen
Publication date
01-12-2014
Publisher
BioMed Central
Published in
BMC Musculoskeletal Disorders / Issue 1/2014
Electronic ISSN: 1471-2474
DOI
https://doi.org/10.1186/1471-2474-15-303

Other articles of this Issue 1/2014

BMC Musculoskeletal Disorders 1/2014 Go to the issue