Published in: BMC Musculoskeletal Disorders 1/2012

Open Access 01-12-2012 | Research article

Prosthesis alignment affects axial rotation motion after total knee replacement: a prospective in vivo study combining computed tomography and fluoroscopic evaluations

Authors: Melinda K Harman, Scott A Banks, Stephan Kirschner, Jörg Lützner

Published in: BMC Musculoskeletal Disorders | Issue 1/2012

Login to get access

Abstract

Background

Clinical consequences of alignment errors in total knee replacement (TKR) have led to the rigorous evaluation of surgical alignment techniques. Rotational alignment in the transverse plane has proven particularly problematic, with errors due to component malalignment relative to bone anatomic landmarks and an overall mismatch between the femoral and tibial components’ relative positions. Ranges of nominal rotational alignment are not well defined, especially for the tibial component and for relative rotational mismatch, and some studies advocate the use of mobile-bearing TKR to accommodate the resulting small rotation errors. However, the relationships between prosthesis rotational alignment and mobile-bearing polyethylene insert motion are poorly understood. This prospective, in vivo study evaluates whether component malalignment and mismatch affect axial rotation motions during passive knee flexion after TKR.

Methods

Eighty patients were implanted with mobile-bearing TKR. Rotational alignment of the femoral and tibial components was measured from postoperative CT scans. All TKR were categorized into nominal or outlier groups based on defined norms for surgical rotational alignment relative to bone anatomic landmarks and relative rotational mismatch between the femoral and tibial components. Axial rotation motion of the femoral, tibial and polyethylene bearing components was measured from fluoroscopic images acquired during passive knee flexion.

Results

Axial rotation motion was generally accomplished in two phases, dominated by polyethylene bearing rotation on the tibial component in early to mid-flexion and then femoral component rotation on the polyethylene articular surface in later flexion. Opposite rotations of the femur-bearing and bearing-baseplate articulations were evident at flexion greater than 80°. Knees with outlier alignment had lower magnitudes of axial rotation and distinct transitions from external to internal rotation during mid-flexion. Knees with femoral-tibial rotational mismatch had significantly lower total axial rotation compared to knees with nominal alignment.

Conclusions

Maintaining relative rotational mismatch within ±5° during TKR provided for controlled knee axial rotation during flexion. TKR with rotational alignment outside of defined surgical norms, with either positive or negative mismatch, experienced measurable kinematic differences and presented different patterns of axial rotation motions during passive knee flexion compared to TKR with nominal mismatch. These findings support previous studies linking prosthesis rotational alignment with inferior clinical and functional outcomes.

Trial Registration

Clinical Trials NCT01022099
Appendix
Available only for authorised users
Literature
1.
go back to reference Anderson KC, Buehler KC, Markel DC: Computer assisted navigation in total knee arthroplasty: Comparison with conventional methods. J Arthroplasty. 2005, 20 (7 Suppl 3): 132-138.CrossRefPubMed Anderson KC, Buehler KC, Markel DC: Computer assisted navigation in total knee arthroplasty: Comparison with conventional methods. J Arthroplasty. 2005, 20 (7 Suppl 3): 132-138.CrossRefPubMed
2.
go back to reference Berend ME, Ritter MA, Meding JB, Faris PM, Keating EM, Redelman R, Faris GW, Davis KE: Tibial component failure mechanisms in total knee arthroplasty. Clin Orthop Relat Res. 2004, 428: 26-34.CrossRefPubMed Berend ME, Ritter MA, Meding JB, Faris PM, Keating EM, Redelman R, Faris GW, Davis KE: Tibial component failure mechanisms in total knee arthroplasty. Clin Orthop Relat Res. 2004, 428: 26-34.CrossRefPubMed
3.
go back to reference Daubresse F, Vajeu C, Loquet J: Total knee arthroplasty with conventional or navigated technique: Comparison of the learning curves in a community hospital. Acta Orthop Belg. 2005, 71 (6): 710-713.PubMed Daubresse F, Vajeu C, Loquet J: Total knee arthroplasty with conventional or navigated technique: Comparison of the learning curves in a community hospital. Acta Orthop Belg. 2005, 71 (6): 710-713.PubMed
4.
go back to reference Ensini A, Catani F, Leardini A, Romagnoli M, Giannini S: Alignments and clinical results in conventional and navigated total knee arthroplasty. Clin Orthop Relat Res. 2007, 457: 156-162.PubMed Ensini A, Catani F, Leardini A, Romagnoli M, Giannini S: Alignments and clinical results in conventional and navigated total knee arthroplasty. Clin Orthop Relat Res. 2007, 457: 156-162.PubMed
5.
go back to reference Lotke PA, Ecker ML: Influence of positioning of prosthesis in total knee replacement. J Bone Joint Surg Am. 1977, 59 (1): 77-79.PubMed Lotke PA, Ecker ML: Influence of positioning of prosthesis in total knee replacement. J Bone Joint Surg Am. 1977, 59 (1): 77-79.PubMed
6.
go back to reference Chauhan SK, Scott RG, Breidahl W, Beaver RJ: Computer-assisted knee arthroplasty versus a conventional jig-based technique: A randomised prospective trial. J Bone Joint Surg Br. 2004, 86 (3): 372-377. 10.1302/0301-620X.86B3.14643.CrossRefPubMed Chauhan SK, Scott RG, Breidahl W, Beaver RJ: Computer-assisted knee arthroplasty versus a conventional jig-based technique: A randomised prospective trial. J Bone Joint Surg Br. 2004, 86 (3): 372-377. 10.1302/0301-620X.86B3.14643.CrossRefPubMed
7.
go back to reference Conditt MA, Noble PC, Thompson MT, Ismaily SK, Moy GJ, Mathis KB: A computerized bioskills system for surgical skills training in total knee replacement. Proc Inst Mech Eng H. 2007, 221 (1): 61-69.CrossRefPubMed Conditt MA, Noble PC, Thompson MT, Ismaily SK, Moy GJ, Mathis KB: A computerized bioskills system for surgical skills training in total knee replacement. Proc Inst Mech Eng H. 2007, 221 (1): 61-69.CrossRefPubMed
8.
go back to reference Eckhoff DG, Metzger RG, Vandewalle MV: Malrotation associated with implant alignment technique in total knee arthroplasty. Clin Orthop Relat Res. 1995, 321: 28-31.PubMed Eckhoff DG, Metzger RG, Vandewalle MV: Malrotation associated with implant alignment technique in total knee arthroplasty. Clin Orthop Relat Res. 1995, 321: 28-31.PubMed
9.
go back to reference Ikeuchi M, Yamanaka N, Okanoue Y, Ueta E, Tani T: Determining the rotational alignment of the tibial component at total knee replacement: A comparison of two techniques. J Bone Joint Surg Br. 2007, 89 (1): 45-49. 10.1302/0301-620X.89B1.17728.CrossRefPubMed Ikeuchi M, Yamanaka N, Okanoue Y, Ueta E, Tani T: Determining the rotational alignment of the tibial component at total knee replacement: A comparison of two techniques. J Bone Joint Surg Br. 2007, 89 (1): 45-49. 10.1302/0301-620X.89B1.17728.CrossRefPubMed
10.
go back to reference Mizu-uchi H, Matsuda S, Miura H, Okazaki K, Akasaki Y, Iwamoto Y: The evaluation of post-operative alignment in total knee replacement using a CT-based navigation system. J Bone Joint Surg Br. 2008, 90 (8): 1025-1031. 10.1302/0301-620X.90B8.20265.CrossRefPubMed Mizu-uchi H, Matsuda S, Miura H, Okazaki K, Akasaki Y, Iwamoto Y: The evaluation of post-operative alignment in total knee replacement using a CT-based navigation system. J Bone Joint Surg Br. 2008, 90 (8): 1025-1031. 10.1302/0301-620X.90B8.20265.CrossRefPubMed
11.
go back to reference Spencer JM, Chauhan SK, Sloan K, Taylor A, Beaver RJ: Computer navigation versus conventional total knee replacement: No difference in functional results at two years. J Bone Joint Surg Br. 2007, 89 (4): 477-480. 10.1302/0301-620X.89B4.18094.CrossRefPubMed Spencer JM, Chauhan SK, Sloan K, Taylor A, Beaver RJ: Computer navigation versus conventional total knee replacement: No difference in functional results at two years. J Bone Joint Surg Br. 2007, 89 (4): 477-480. 10.1302/0301-620X.89B4.18094.CrossRefPubMed
12.
go back to reference Stöckl B, Nogler M, Rosiek R, Fischer M, Krismer M, Kessler O: Navigation improves accuracy of rotational alignment in total knee arthroplasty. Clin Orthop Relat Res. 2004, 426: 180-186.CrossRefPubMed Stöckl B, Nogler M, Rosiek R, Fischer M, Krismer M, Kessler O: Navigation improves accuracy of rotational alignment in total knee arthroplasty. Clin Orthop Relat Res. 2004, 426: 180-186.CrossRefPubMed
13.
go back to reference van der Linden HM, van der Zwaag HM, Valstar ER, van der Molen AJ, Nelissen RG: Transepicondylar axis accuracy in computer assisted knee surgery: A comparison of the CT-based measured axis versus the CAS-determined axis. Comput Aided Surg. 2008, 13 (4): 200-206. 10.3109/10929080802240134.CrossRef van der Linden HM, van der Zwaag HM, Valstar ER, van der Molen AJ, Nelissen RG: Transepicondylar axis accuracy in computer assisted knee surgery: A comparison of the CT-based measured axis versus the CAS-determined axis. Comput Aided Surg. 2008, 13 (4): 200-206. 10.3109/10929080802240134.CrossRef
14.
go back to reference Barrack RL, Schrader T, Bertot AJ, Wolfe MW, Myers L: Component rotation and anterior knee pain after total knee arthroplasty. Clin Orthop Relat Res. 2001, 392: 46-55.CrossRefPubMed Barrack RL, Schrader T, Bertot AJ, Wolfe MW, Myers L: Component rotation and anterior knee pain after total knee arthroplasty. Clin Orthop Relat Res. 2001, 392: 46-55.CrossRefPubMed
15.
go back to reference Lützner J, Krummenauer F, Wolf C, Günther KP, Kirschner S: Computer-assisted and conventional total knee replacement: A comparative prospective randomised study with radiological and CT evaluation. J Bone Joint Surg Br. 2008, 90 (8): 1039-1044. 10.1302/0301-620X.90B8.20553.CrossRefPubMed Lützner J, Krummenauer F, Wolf C, Günther KP, Kirschner S: Computer-assisted and conventional total knee replacement: A comparative prospective randomised study with radiological and CT evaluation. J Bone Joint Surg Br. 2008, 90 (8): 1039-1044. 10.1302/0301-620X.90B8.20553.CrossRefPubMed
16.
go back to reference Matziolis G, Krocker D, Weiss U, Tohtz S, Perka C: A prospective randomized study of computer-assisted and conventional total knee arthroplasty: Three-dimensional evaluation of implant alignment and rotation. J Bone Joint Surg Am. 2007, 89 (2): 236-243.CrossRefPubMed Matziolis G, Krocker D, Weiss U, Tohtz S, Perka C: A prospective randomized study of computer-assisted and conventional total knee arthroplasty: Three-dimensional evaluation of implant alignment and rotation. J Bone Joint Surg Am. 2007, 89 (2): 236-243.CrossRefPubMed
17.
go back to reference Siston RA, Goodman SB, Patel JJ, Delp SL, Giori NJ: The high variability of tibial rotational alignment in total knee arthroplasty. Clin Orthop Rel Res. 2006, 452: 65-69.CrossRef Siston RA, Goodman SB, Patel JJ, Delp SL, Giori NJ: The high variability of tibial rotational alignment in total knee arthroplasty. Clin Orthop Rel Res. 2006, 452: 65-69.CrossRef
18.
go back to reference Berger RA, Crossett LS, Jacobs JJ, Rubash HE: Malrotation causing patellofemoral complications after total knee arthroplasty. Clin Orthop Relat Res. 1998, 356: 144-153.CrossRefPubMed Berger RA, Crossett LS, Jacobs JJ, Rubash HE: Malrotation causing patellofemoral complications after total knee arthroplasty. Clin Orthop Relat Res. 1998, 356: 144-153.CrossRefPubMed
19.
go back to reference Uehara K, Kadoya Y, Kobayashi A, Ohashi H, Yamano Y: Bone anatomy and rotational alignment in total knee arthroplasty. Clin Orthop Relat Res. 2002, 402: 196-201.CrossRefPubMed Uehara K, Kadoya Y, Kobayashi A, Ohashi H, Yamano Y: Bone anatomy and rotational alignment in total knee arthroplasty. Clin Orthop Relat Res. 2002, 402: 196-201.CrossRefPubMed
20.
go back to reference Wasielewski RC, Galante JO, Leighty RM, Natarajan RN, Rosenberg AG: Wear patterns on retrieved polyethylene tibial inserts and their relationship to technical considerations during total knee arthroplasty. Clin Orthop Relat Res. 1994, 299: 31-43.PubMed Wasielewski RC, Galante JO, Leighty RM, Natarajan RN, Rosenberg AG: Wear patterns on retrieved polyethylene tibial inserts and their relationship to technical considerations during total knee arthroplasty. Clin Orthop Relat Res. 1994, 299: 31-43.PubMed
21.
go back to reference Breugem SJ, Sierevelt IN, Schafroth MU, Blankevoort L, Schaap GR, van Dijk CN: Less anterior knee pain with mobile-bearing prosthesis compared with a fixed-bearing prosthesis. Clin Orthop Relat Res. 2008, 466: 1959-1965. 10.1007/s11999-008-0320-6.CrossRefPubMedPubMedCentral Breugem SJ, Sierevelt IN, Schafroth MU, Blankevoort L, Schaap GR, van Dijk CN: Less anterior knee pain with mobile-bearing prosthesis compared with a fixed-bearing prosthesis. Clin Orthop Relat Res. 2008, 466: 1959-1965. 10.1007/s11999-008-0320-6.CrossRefPubMedPubMedCentral
22.
go back to reference Huddleston JI, Scott RD, Wimberley DW: Determination of neutral tibial rotational alignment in rotating platform TKA. Clin Orthop Relat Res. 2005, 440: 101-106. 10.1097/01.blo.0000185448.43622.77.CrossRefPubMed Huddleston JI, Scott RD, Wimberley DW: Determination of neutral tibial rotational alignment in rotating platform TKA. Clin Orthop Relat Res. 2005, 440: 101-106. 10.1097/01.blo.0000185448.43622.77.CrossRefPubMed
23.
go back to reference Kaper BP, Smith PN, Bourne RB, Rorabeck CH, Robertson D: Medium-term results of a mobile bearing total knee replacement. Clin Orthop Relat Res. 1999, 367: 201-209.CrossRefPubMed Kaper BP, Smith PN, Bourne RB, Rorabeck CH, Robertson D: Medium-term results of a mobile bearing total knee replacement. Clin Orthop Relat Res. 1999, 367: 201-209.CrossRefPubMed
24.
go back to reference Walker PS, Komistek RD, Barrett DS, Anderson D, Dennis DA, Sampson M: Motion of a mobile bearing knee allowing translation and rotation. J Arthroplasty. 2002, 17 (1): 11-19. 10.1054/arth.2002.28731.CrossRefPubMed Walker PS, Komistek RD, Barrett DS, Anderson D, Dennis DA, Sampson M: Motion of a mobile bearing knee allowing translation and rotation. J Arthroplasty. 2002, 17 (1): 11-19. 10.1054/arth.2002.28731.CrossRefPubMed
25.
go back to reference Fantozzi S, Leardini A, Banks SA, Marcacci M, Giannini S, Catani F: Dynamic in-vivo tibio-femoral and bearing motions in mobile bearing knee arthroplasty. Knee Surg Sports Traumatol Arthrosc. 2004, 12 (2): 144-151. 10.1007/s00167-003-0384-5.CrossRefPubMed Fantozzi S, Leardini A, Banks SA, Marcacci M, Giannini S, Catani F: Dynamic in-vivo tibio-femoral and bearing motions in mobile bearing knee arthroplasty. Knee Surg Sports Traumatol Arthrosc. 2004, 12 (2): 144-151. 10.1007/s00167-003-0384-5.CrossRefPubMed
26.
go back to reference Haas BD, Komistek RD, Dennis DA: In vivo kinematics of the Low Contract Stress rotating platform total knee. Orthopedics. 2002, 25 (2 Supplement): 219-226. Haas BD, Komistek RD, Dennis DA: In vivo kinematics of the Low Contract Stress rotating platform total knee. Orthopedics. 2002, 25 (2 Supplement): 219-226.
27.
go back to reference Stiehl JB, Dennis DA, Komistek RD, Keblish PA: In vivo kinematic analysis of a mobile bearing total knee prosthesis. Clin Orthop Relat Res. 1997, 345: 60-66.CrossRefPubMed Stiehl JB, Dennis DA, Komistek RD, Keblish PA: In vivo kinematic analysis of a mobile bearing total knee prosthesis. Clin Orthop Relat Res. 1997, 345: 60-66.CrossRefPubMed
28.
go back to reference Stukenborg-Colsman C, Ostermeier S, Wenger KH, Wirth CJ: Relative motion of a mobile bearing inlay after total knee arthroplasty: A dynamic in vitro study. Clin Biomech. 2002, 17: 49-55. 10.1016/S0268-0033(01)00103-6.CrossRef Stukenborg-Colsman C, Ostermeier S, Wenger KH, Wirth CJ: Relative motion of a mobile bearing inlay after total knee arthroplasty: A dynamic in vitro study. Clin Biomech. 2002, 17: 49-55. 10.1016/S0268-0033(01)00103-6.CrossRef
29.
go back to reference der Zwaag HMJ VdL-v, Bos J, van der Heide HJL, Nelissen RGHH: A computed tomography based study on rotational alignment accuracy of the femoral component in total knee arthroplasty using computer-assisted orthopaedic surgery. Int Orthop. 2011, 35: 845-850. 10.1007/s00264-010-1082-9.CrossRef der Zwaag HMJ VdL-v, Bos J, van der Heide HJL, Nelissen RGHH: A computed tomography based study on rotational alignment accuracy of the femoral component in total knee arthroplasty using computer-assisted orthopaedic surgery. Int Orthop. 2011, 35: 845-850. 10.1007/s00264-010-1082-9.CrossRef
30.
go back to reference Jazrawi LM, Birdzell L, Kummer FJ, Di Cesare PE: The accuracy of computed tomography for determining femoral and tibial total knee arthroplasty component rotation. J Arthrop. 2000, 15 (6): 761-766. 10.1054/arth.2000.8193.CrossRef Jazrawi LM, Birdzell L, Kummer FJ, Di Cesare PE: The accuracy of computed tomography for determining femoral and tibial total knee arthroplasty component rotation. J Arthrop. 2000, 15 (6): 761-766. 10.1054/arth.2000.8193.CrossRef
31.
go back to reference Merican AM, Ghosh KM, Iranpour F, Deehan DJ, Amis AA: The effect of femoral component rotation on the kinematics of the tibiofemoral and patellofemoral joints after total knee arthroplasty. Knee Surg Sports Traumatol Arthrosc. 2011, 19: 1479-1487. 10.1007/s00167-011-1499-8.CrossRefPubMed Merican AM, Ghosh KM, Iranpour F, Deehan DJ, Amis AA: The effect of femoral component rotation on the kinematics of the tibiofemoral and patellofemoral joints after total knee arthroplasty. Knee Surg Sports Traumatol Arthrosc. 2011, 19: 1479-1487. 10.1007/s00167-011-1499-8.CrossRefPubMed
32.
go back to reference Nicoll D, Rowley DI: Internal rotational error of the tibial component is a major cause of pain after total knee replacement. J Bone Joint Surg. 2010, 92B: 1238-1244.CrossRef Nicoll D, Rowley DI: Internal rotational error of the tibial component is a major cause of pain after total knee replacement. J Bone Joint Surg. 2010, 92B: 1238-1244.CrossRef
33.
go back to reference Zihlmann MS, Stacoff A, Romero J, Quervain IK, Stüssi E: Biomechanical background and clinical observations of rotational malalignment in TKA: Literature review and consequences. Clin Biomech. 2005, 20: 661-668. 10.1016/j.clinbiomech.2005.03.014.CrossRef Zihlmann MS, Stacoff A, Romero J, Quervain IK, Stüssi E: Biomechanical background and clinical observations of rotational malalignment in TKA: Literature review and consequences. Clin Biomech. 2005, 20: 661-668. 10.1016/j.clinbiomech.2005.03.014.CrossRef
34.
go back to reference Rhoads DD, Noble PC, Reuben JD, Tullos HS: The effect of femoral component position on the kinematics of total knee arthroplasty. Clin Orthop Relat Res. 1993, 286: 122-129.PubMed Rhoads DD, Noble PC, Reuben JD, Tullos HS: The effect of femoral component position on the kinematics of total knee arthroplasty. Clin Orthop Relat Res. 1993, 286: 122-129.PubMed
35.
go back to reference Banks SA, Hodge WA: Accurate measurement of three-dimensional knee replacement kinematics using single-plane fluoroscopy. IEEE Trans Biomed Eng. 1996, 43: 638-649. 10.1109/10.495283.CrossRefPubMed Banks SA, Hodge WA: Accurate measurement of three-dimensional knee replacement kinematics using single-plane fluoroscopy. IEEE Trans Biomed Eng. 1996, 43: 638-649. 10.1109/10.495283.CrossRefPubMed
36.
go back to reference Chouteau J, Lerat JL, Testa R, Moyen B, Fessy MH, Banks SA: Mobile-bearing insert translational and rotational kinematics in a PCL-retaining total knee arthroplasty. Ortho Traum Surg Res. 2009, 95: 254-259. 10.1016/j.otsr.2009.03.012.CrossRef Chouteau J, Lerat JL, Testa R, Moyen B, Fessy MH, Banks SA: Mobile-bearing insert translational and rotational kinematics in a PCL-retaining total knee arthroplasty. Ortho Traum Surg Res. 2009, 95: 254-259. 10.1016/j.otsr.2009.03.012.CrossRef
37.
go back to reference Delport HP, Banks SA, De Schepper J, Bellemans J: A kinematic comparison of fixed- and mobile-bearing knee replacements. J Bone Joint Surg Br. 2006, 88 (8): 1016-1021. 10.1302/0301-620X.88B8.17529.CrossRefPubMed Delport HP, Banks SA, De Schepper J, Bellemans J: A kinematic comparison of fixed- and mobile-bearing knee replacements. J Bone Joint Surg Br. 2006, 88 (8): 1016-1021. 10.1302/0301-620X.88B8.17529.CrossRefPubMed
38.
go back to reference Gamada K, Jayasekera N, Kashif F, Fennema P, Schmotzer H, Banks SA: Does ligament balancing technique affect kinematics in rotating platform PCL retaining knee arthroplasties? A prospective randomized study. Knee Surg Sports Traumatol Arthrosc. 2008, 16 (2): 160-166. 10.1007/s00167-007-0447-0.CrossRefPubMed Gamada K, Jayasekera N, Kashif F, Fennema P, Schmotzer H, Banks SA: Does ligament balancing technique affect kinematics in rotating platform PCL retaining knee arthroplasties? A prospective randomized study. Knee Surg Sports Traumatol Arthrosc. 2008, 16 (2): 160-166. 10.1007/s00167-007-0447-0.CrossRefPubMed
39.
go back to reference Garling EH, Kaptein BL, Nelissen RGHH, Valstar ER: Limited rotation of the mobile-bearing in a rotating platform total knee prosthesis. J Biomech. 2007, 40 (Suppl 1): S25-S30.CrossRefPubMed Garling EH, Kaptein BL, Nelissen RGHH, Valstar ER: Limited rotation of the mobile-bearing in a rotating platform total knee prosthesis. J Biomech. 2007, 40 (Suppl 1): S25-S30.CrossRefPubMed
40.
go back to reference Colwell CW, Chen PC, D’Lima D: Extensor malalignment arising from femoral component malrotation in knee arthroplasty: Effect of rotating-bearing. Clin Biomech. 2011, 26: 52-57. 10.1016/j.clinbiomech.2010.08.009.CrossRef Colwell CW, Chen PC, D’Lima D: Extensor malalignment arising from femoral component malrotation in knee arthroplasty: Effect of rotating-bearing. Clin Biomech. 2011, 26: 52-57. 10.1016/j.clinbiomech.2010.08.009.CrossRef
41.
go back to reference Lützner J, Günther KP, Kirschner S: Functional outcome after computer-assisted versus conventional total knee arthroplasty: A randomized controlled study. Knee Surg Sports Traumatol Arthrosc. 2010, 18: 1339-1344. 10.1007/s00167-010-1153-x.CrossRefPubMed Lützner J, Günther KP, Kirschner S: Functional outcome after computer-assisted versus conventional total knee arthroplasty: A randomized controlled study. Knee Surg Sports Traumatol Arthrosc. 2010, 18: 1339-1344. 10.1007/s00167-010-1153-x.CrossRefPubMed
Metadata
Title
Prosthesis alignment affects axial rotation motion after total knee replacement: a prospective in vivo study combining computed tomography and fluoroscopic evaluations
Authors
Melinda K Harman
Scott A Banks
Stephan Kirschner
Jörg Lützner
Publication date
01-12-2012
Publisher
BioMed Central
Published in
BMC Musculoskeletal Disorders / Issue 1/2012
Electronic ISSN: 1471-2474
DOI
https://doi.org/10.1186/1471-2474-13-206