Published in: BMC Ophthalmology 1/2014

Open Access 01-12-2014 | Research article

Optical coherence tomography for an in-vivo study of posterior-capsule-opacification types and their influence on the total-pulse energy required for Nd:YAG capsulotomy: a case series

Authors: Gregor Hawlina, Darko Perovšek, Brigita Drnovšek-Olup, Janez Možina, Peter Gregorčič

Published in: BMC Ophthalmology | Issue 1/2014

Login to get access

Abstract

Background

Posterior capsule opacification (PCO) is the most common post-operative complication associated with cataract surgery and is mostly treated with Nd:YAG laser capsulotomy. Here, we demonstrate the use of high-resolution spectral-domain optical coherence tomography (OCT) as a technique for PCO analysis. Additionally, we evaluate the influence of PCO types and the distance between the intraocular lens (IOL) and the posterior capsule (PC), i.e., the IOL/PC distance, on the total-pulse energy required for the Nd:YAG laser posterior capsulotomy.

Methods

47 eyes with PCO scheduled for the Nd:YAG procedure were examined and divided into four categories: fibrosis, pearl, mixed type and late-postoperative capsular bag distension syndrome. Using custom-made computer software for OCT image analysis, the IOL/PC distances in two dimensions were measured. The IOL/PC distances were compared with those of a control group of 15 eyes without PCO. The influence of the different PCO types and the IOL/PC distance on the total-pulse energy required for the Nd:YAG procedure was analyzed.

Results

The total-pulse energy required for a laser capsulotomy differs significantly between PCO types (p = 0.005, Kruskal-Wallis test). The highest energy was required for the fibrosis PCO type, followed by mixed, pearl and late-postoperative capsular bag distension syndrome. The IOL/PC distance also significantly influenced the total-pulse energy required for laser capsulotomy (p = 0.028, linear regression). Lower total-pulse energy was expected for a larger IOL/PC distance.

Conclusions

Our study indicates that the PCO types and the IOL/PC distance influence the total-pulse energy required for Nd:YAG capsulotomy. The presented OCT method has the potential to become an additional tool for PCO characterization. Our results are important for a better understanding of the photodisruptive mechanisms in Nd:YAG capsulotomy.
Appendix
Available only for authorised users
Literature
1.
go back to reference Hollick EJ, Spalton DJ, Ursell PG, Pande MV, Barman SA, Boyce JF, Tilling K: The effect of polymethylmethacrylate, silicone, and polyacrylic intraocular lenses on posterior capsular opacification 3 years after cataract surgery. Ophthalmology. 1999, 106 (1): 49-54. 10.1016/S0161-6420(99)90047-7. discussion 54–45CrossRefPubMed Hollick EJ, Spalton DJ, Ursell PG, Pande MV, Barman SA, Boyce JF, Tilling K: The effect of polymethylmethacrylate, silicone, and polyacrylic intraocular lenses on posterior capsular opacification 3 years after cataract surgery. Ophthalmology. 1999, 106 (1): 49-54. 10.1016/S0161-6420(99)90047-7. discussion 54–45CrossRefPubMed
2.
go back to reference Schaumberg DA, Dana MR, Christen WG, Glynn RJ: A systematic overview of the incidence of posterior capsule opacification. Ophthalmology. 1998, 105 (7): 1213-1221. 10.1016/S0161-6420(98)97023-3.CrossRefPubMed Schaumberg DA, Dana MR, Christen WG, Glynn RJ: A systematic overview of the incidence of posterior capsule opacification. Ophthalmology. 1998, 105 (7): 1213-1221. 10.1016/S0161-6420(98)97023-3.CrossRefPubMed
3.
go back to reference Apple DJ, Solomon KD, Tetz MR, Assia EI, Holland EY, Legler UF, Tsai JC, Castaneda VE, Hoggatt JP, Kostick AM: Posterior capsule opacification. Surv Ophthalmol. 1992, 37 (2): 73-116. 10.1016/0039-6257(92)90073-3.CrossRefPubMed Apple DJ, Solomon KD, Tetz MR, Assia EI, Holland EY, Legler UF, Tsai JC, Castaneda VE, Hoggatt JP, Kostick AM: Posterior capsule opacification. Surv Ophthalmol. 1992, 37 (2): 73-116. 10.1016/0039-6257(92)90073-3.CrossRefPubMed
4.
go back to reference Hiles DA, Johnson BL: The role of the crystalline lens epithelium in postpseudophakos membrane formation. J Am Intraocul Implant Soc. 1980, 6 (2): 141-147. 10.1016/S0146-2776(80)80007-3.CrossRefPubMed Hiles DA, Johnson BL: The role of the crystalline lens epithelium in postpseudophakos membrane formation. J Am Intraocul Implant Soc. 1980, 6 (2): 141-147. 10.1016/S0146-2776(80)80007-3.CrossRefPubMed
5.
go back to reference McDonnell PJ, Green WR, Maumenee AE, Iliff WJ: Pathology of intraocular lenses in 33 eyes examined postmortem. Ophthalmology. 1983, 90 (4): 386-403. 10.1016/S0161-6420(83)34558-9.CrossRefPubMed McDonnell PJ, Green WR, Maumenee AE, Iliff WJ: Pathology of intraocular lenses in 33 eyes examined postmortem. Ophthalmology. 1983, 90 (4): 386-403. 10.1016/S0161-6420(83)34558-9.CrossRefPubMed
6.
go back to reference Roy FH: After-cataract: clinical and pathologic evaluation. Ann Ophthalmol. 1971, 3 (12): 1364-1366. passimPubMed Roy FH: After-cataract: clinical and pathologic evaluation. Ann Ophthalmol. 1971, 3 (12): 1364-1366. passimPubMed
7.
go back to reference Awasthi N, Guo S, Wagner BJ: Posterior capsular opacification: a problem reduced but not yet eradicated. Arch Ophthalmol. 2009, 127 (4): 555-562. 10.1001/archophthalmol.2009.3.CrossRefPubMed Awasthi N, Guo S, Wagner BJ: Posterior capsular opacification: a problem reduced but not yet eradicated. Arch Ophthalmol. 2009, 127 (4): 555-562. 10.1001/archophthalmol.2009.3.CrossRefPubMed
8.
go back to reference Mcdonnell PJ, Zarbin MA, Green WR: Posterior capsule opacification in pseudophakic eyes. Ophthalmology. 1983, 90 (12): 1548-1553. 10.1016/S0161-6420(83)34350-5.CrossRefPubMed Mcdonnell PJ, Zarbin MA, Green WR: Posterior capsule opacification in pseudophakic eyes. Ophthalmology. 1983, 90 (12): 1548-1553. 10.1016/S0161-6420(83)34350-5.CrossRefPubMed
9.
go back to reference Saika S: Relationship between posterior capsule opacification and intraocular lens biocompatibility. Prog Retin Eye Res. 2004, 23 (3): 283-305. 10.1016/j.preteyeres.2004.02.004.CrossRefPubMed Saika S: Relationship between posterior capsule opacification and intraocular lens biocompatibility. Prog Retin Eye Res. 2004, 23 (3): 283-305. 10.1016/j.preteyeres.2004.02.004.CrossRefPubMed
10.
go back to reference Miyake K, Ota I, Ichihashi S, Miyake S, Tanaka Y, Terasaki H: New classification of capsular block syndrome. J Cataract Refract Surg. 1998, 24 (9): 1230-1234. 10.1016/S0886-3350(98)80017-5.CrossRefPubMed Miyake K, Ota I, Ichihashi S, Miyake S, Tanaka Y, Terasaki H: New classification of capsular block syndrome. J Cataract Refract Surg. 1998, 24 (9): 1230-1234. 10.1016/S0886-3350(98)80017-5.CrossRefPubMed
11.
go back to reference Apple DJ, Peng Q, Visessook N, Werner L, Pandey SK, Escobar-Gomez M, Ram J, Auffarth GU: Eradication of posterior capsule opacification: documentation of a marked decrease in Nd:YAG laser posterior capsulotomy rates noted in an analysis of 5416 pseudophakic human eyes obtained postmortem. Ophthalmology. 2001, 108 (3): 505-518. 10.1016/S0161-6420(00)00589-3.CrossRefPubMed Apple DJ, Peng Q, Visessook N, Werner L, Pandey SK, Escobar-Gomez M, Ram J, Auffarth GU: Eradication of posterior capsule opacification: documentation of a marked decrease in Nd:YAG laser posterior capsulotomy rates noted in an analysis of 5416 pseudophakic human eyes obtained postmortem. Ophthalmology. 2001, 108 (3): 505-518. 10.1016/S0161-6420(00)00589-3.CrossRefPubMed
12.
go back to reference Aron-Rosa DS, Aron JJ, Cohn HC: Use of a pulsed picosecond Nd: YAG laser in 6,664 cases. J Am Intraocul Implant Soc. 1984, 10 (1): 35-39. 10.1016/S0146-2776(84)80074-9.CrossRefPubMed Aron-Rosa DS, Aron JJ, Cohn HC: Use of a pulsed picosecond Nd: YAG laser in 6,664 cases. J Am Intraocul Implant Soc. 1984, 10 (1): 35-39. 10.1016/S0146-2776(84)80074-9.CrossRefPubMed
13.
go back to reference Channell MM, Beckman H: Intraocular pressure changes after neodymium-YAG laser posterior capsulotomy. Arch Ophthalmol. 1984, 102 (7): 1024-1026. 10.1001/archopht.1984.01040030826025.CrossRefPubMed Channell MM, Beckman H: Intraocular pressure changes after neodymium-YAG laser posterior capsulotomy. Arch Ophthalmol. 1984, 102 (7): 1024-1026. 10.1001/archopht.1984.01040030826025.CrossRefPubMed
14.
go back to reference Ficker LA, Steele AD: Complications of Nd:YAG laser posterior capsulotomy. Trans Ophthalmol Soc U K. 1985, 104 (Pt 5): 529-532.PubMed Ficker LA, Steele AD: Complications of Nd:YAG laser posterior capsulotomy. Trans Ophthalmol Soc U K. 1985, 104 (Pt 5): 529-532.PubMed
15.
go back to reference Geerling G, Roider J, Schmidt-Erfurt U, Nahen K, el-Hifnawi el S, Laqua H, Vogel A: Initial clinical experience with the picosecond Nd:YLF laser for intraocular therapeutic applications. Br J Ophthalmol. 1998, 82 (5): 504-509. 10.1136/bjo.82.5.504.CrossRefPubMedPubMedCentral Geerling G, Roider J, Schmidt-Erfurt U, Nahen K, el-Hifnawi el S, Laqua H, Vogel A: Initial clinical experience with the picosecond Nd:YLF laser for intraocular therapeutic applications. Br J Ophthalmol. 1998, 82 (5): 504-509. 10.1136/bjo.82.5.504.CrossRefPubMedPubMedCentral
16.
go back to reference Kraff MC, Sanders DR, Lieberman HL: Intraocular pressure and the corneal endothelium after neodymium-YAG laser posterior capsulotomy. Relative effects of aphakia and pseudophakia. Arch Ophthalmol. 1985, 103 (4): 511-514. 10.1001/archopht.1985.01050040053016.CrossRefPubMed Kraff MC, Sanders DR, Lieberman HL: Intraocular pressure and the corneal endothelium after neodymium-YAG laser posterior capsulotomy. Relative effects of aphakia and pseudophakia. Arch Ophthalmol. 1985, 103 (4): 511-514. 10.1001/archopht.1985.01050040053016.CrossRefPubMed
17.
go back to reference Slomovic AR, Parrish RK, Forster RK, Cubillas A: Neodymium-YAG laser posterior capsulotomy. Central corneal endothelial cell density. Arch Ophthalmol. 1986, 104 (4): 536-538. 10.1001/archopht.1986.01050160092019.CrossRefPubMed Slomovic AR, Parrish RK, Forster RK, Cubillas A: Neodymium-YAG laser posterior capsulotomy. Central corneal endothelial cell density. Arch Ophthalmol. 1986, 104 (4): 536-538. 10.1001/archopht.1986.01050160092019.CrossRefPubMed
18.
go back to reference Gomaa A, Liu C: Nd:YAG laser capsulotomy: a survey of UK practice and recommendations. Eur J Ophthalmol. 2011, 21 (4): 385-390. 10.5301/EJO.2010.6085.CrossRefPubMed Gomaa A, Liu C: Nd:YAG laser capsulotomy: a survey of UK practice and recommendations. Eur J Ophthalmol. 2011, 21 (4): 385-390. 10.5301/EJO.2010.6085.CrossRefPubMed
19.
go back to reference Masood A, Sarfraz M: Capturing outlines of 2D objects with Bézier cubic approximation. Image Vision Comput. 2009, 27 (6): 704-712. 10.1016/j.imavis.2008.07.012.CrossRef Masood A, Sarfraz M: Capturing outlines of 2D objects with Bézier cubic approximation. Image Vision Comput. 2009, 27 (6): 704-712. 10.1016/j.imavis.2008.07.012.CrossRef
20.
go back to reference Izatt JA, Hee MR, Swanson EA, Lin CP, Huang D, Schuman JS, Puliafito CA, Fujimoto JG: Micrometer-scale resolution imaging of the anterior eye in vivo with optical coherence tomography. Arch Ophthalmol. 1994, 112 (12): 1584-1589. 10.1001/archopht.1994.01090240090031.CrossRefPubMed Izatt JA, Hee MR, Swanson EA, Lin CP, Huang D, Schuman JS, Puliafito CA, Fujimoto JG: Micrometer-scale resolution imaging of the anterior eye in vivo with optical coherence tomography. Arch Ophthalmol. 1994, 112 (12): 1584-1589. 10.1001/archopht.1994.01090240090031.CrossRefPubMed
21.
go back to reference Wang Y, Jiang H, Shen M, Lam BL, DeBuc DC, Ye Y, Li M, Tao A, Shao Y, Wang J: Quantitative analysis of the intraretinal layers and optic nerve head using ultra-high resolution optical coherence tomography. J Biomed Opt. 2012, 17 (6): 066013-10.1117/1.JBO.17.6.066013.CrossRefPubMedPubMedCentral Wang Y, Jiang H, Shen M, Lam BL, DeBuc DC, Ye Y, Li M, Tao A, Shao Y, Wang J: Quantitative analysis of the intraretinal layers and optic nerve head using ultra-high resolution optical coherence tomography. J Biomed Opt. 2012, 17 (6): 066013-10.1117/1.JBO.17.6.066013.CrossRefPubMedPubMedCentral
22.
go back to reference Loya N, Lichter H, Barash D, Goldenberg-Cohen N, Strassmann E, Weinberger D: Posterior chamber intraocular lens implantation after capsular tear: ultrasound biomicroscopy evaluation. J Cataract Refract Surg. 2001, 27 (9): 1423-1427. 10.1016/S0886-3350(01)00786-6.CrossRefPubMed Loya N, Lichter H, Barash D, Goldenberg-Cohen N, Strassmann E, Weinberger D: Posterior chamber intraocular lens implantation after capsular tear: ultrasound biomicroscopy evaluation. J Cataract Refract Surg. 2001, 27 (9): 1423-1427. 10.1016/S0886-3350(01)00786-6.CrossRefPubMed
23.
go back to reference Muftuoglu O, Hosal BM, Karel F, Zilelioglu G: Drug-induced intraocular lens movement and near visual acuity after AcrySof intraocular lens implantation. J Cataract Refract Surg. 2005, 31 (7): 1298-1305. 10.1016/j.jcrs.2004.11.042.CrossRefPubMed Muftuoglu O, Hosal BM, Karel F, Zilelioglu G: Drug-induced intraocular lens movement and near visual acuity after AcrySof intraocular lens implantation. J Cataract Refract Surg. 2005, 31 (7): 1298-1305. 10.1016/j.jcrs.2004.11.042.CrossRefPubMed
24.
go back to reference Sorenson AL, Holladay JT, Kim T, Kendall CJ, Carlson AN: Ultrasonographic measurement of induced myopia associated with capsular bag distention syndrome. Ophthalmology. 2000, 107 (5): 902-908. 10.1016/S0161-6420(00)00020-8.CrossRefPubMed Sorenson AL, Holladay JT, Kim T, Kendall CJ, Carlson AN: Ultrasonographic measurement of induced myopia associated with capsular bag distention syndrome. Ophthalmology. 2000, 107 (5): 902-908. 10.1016/S0161-6420(00)00020-8.CrossRefPubMed
25.
go back to reference Hayashi K, Hayashi H, Nakao F, Hayashi F: Reproducibility of posterior capsule opacification measurement using Scheimpflug videophotography. J Cataract Refract Surg. 1998, 24 (12): 1632-1635. 10.1016/S0886-3350(98)80355-6.CrossRefPubMed Hayashi K, Hayashi H, Nakao F, Hayashi F: Reproducibility of posterior capsule opacification measurement using Scheimpflug videophotography. J Cataract Refract Surg. 1998, 24 (12): 1632-1635. 10.1016/S0886-3350(98)80355-6.CrossRefPubMed
26.
go back to reference Jain R, Grewal D, Gupta R, Grewal SP: Scheimpflug imaging in late Capsular Bag Distention syndrome after phacoemulsification. Am J Ophthalmol. 2006, 142 (6): 1083-1085. 10.1016/j.ajo.2006.06.055.CrossRefPubMed Jain R, Grewal D, Gupta R, Grewal SP: Scheimpflug imaging in late Capsular Bag Distention syndrome after phacoemulsification. Am J Ophthalmol. 2006, 142 (6): 1083-1085. 10.1016/j.ajo.2006.06.055.CrossRefPubMed
27.
go back to reference Elgohary MA, Chauhan DS, Dowler JG: Optical coherence tomography of intraocular lens implants and their relationship to the posterior capsule: a pilot study comparing a hydrophobic acrylic to a plate-haptic silicone type. Ophthalmic Res. 2006, 38 (3): 116-124. 10.1159/000090532.CrossRefPubMed Elgohary MA, Chauhan DS, Dowler JG: Optical coherence tomography of intraocular lens implants and their relationship to the posterior capsule: a pilot study comparing a hydrophobic acrylic to a plate-haptic silicone type. Ophthalmic Res. 2006, 38 (3): 116-124. 10.1159/000090532.CrossRefPubMed
28.
go back to reference Moreno-Montañes J, Alvarez A, Bes-Rastrollo M, Garcia-Layana A: Optical coherence tomography evaluation of posterior capsule opacification related to intraocular lens design. J Cataract Refract Surg. 2008, 34 (4): 643-650. 10.1016/j.jcrs.2007.11.035.CrossRefPubMed Moreno-Montañes J, Alvarez A, Bes-Rastrollo M, Garcia-Layana A: Optical coherence tomography evaluation of posterior capsule opacification related to intraocular lens design. J Cataract Refract Surg. 2008, 34 (4): 643-650. 10.1016/j.jcrs.2007.11.035.CrossRefPubMed
29.
go back to reference Moreno-Montañes J, Alvarez A, Maldonado MJ: Objective quantification of posterior capsule opacification after cataract surgery, with optical coherence tomography. Invest Ophthalmol Vis Sci. 2005, 46 (11): 3999-4006. 10.1167/iovs.04-1531.CrossRefPubMed Moreno-Montañes J, Alvarez A, Maldonado MJ: Objective quantification of posterior capsule opacification after cataract surgery, with optical coherence tomography. Invest Ophthalmol Vis Sci. 2005, 46 (11): 3999-4006. 10.1167/iovs.04-1531.CrossRefPubMed
30.
go back to reference Puliafito CA, Hee MR, Lin CP, Reichel E, Schuman JS, Duker JS, Izatt JA, Swanson EA, Fujimoto JG: Imaging of macular diseases with optical coherence tomography. Ophthalmology. 1995, 102 (2): 217-229. 10.1016/S0161-6420(95)31032-9.CrossRefPubMed Puliafito CA, Hee MR, Lin CP, Reichel E, Schuman JS, Duker JS, Izatt JA, Swanson EA, Fujimoto JG: Imaging of macular diseases with optical coherence tomography. Ophthalmology. 1995, 102 (2): 217-229. 10.1016/S0161-6420(95)31032-9.CrossRefPubMed
31.
go back to reference Schuman JS, Pedut-Kloizman T, Hertzmark E, Hee MR, Wilkins JR, Coker JG, Puliafito CA, Fujimoto JG, Swanson EA: Reproducibility of nerve fiber layer thickness measurements using optical coherence tomography. Ophthalmology. 1996, 103 (11): 1889-1898. 10.1016/S0161-6420(96)30410-7.CrossRefPubMedPubMedCentral Schuman JS, Pedut-Kloizman T, Hertzmark E, Hee MR, Wilkins JR, Coker JG, Puliafito CA, Fujimoto JG, Swanson EA: Reproducibility of nerve fiber layer thickness measurements using optical coherence tomography. Ophthalmology. 1996, 103 (11): 1889-1898. 10.1016/S0161-6420(96)30410-7.CrossRefPubMedPubMedCentral
32.
go back to reference Nozaki M, Kimura H, Kojima M, Ogura Y: Optical coherence tomographic findings of the anterior segment after nonpenetrating deep sclerectomy. Am J Ophthalmol. 2002, 133 (6): 837-839. 10.1016/S0002-9394(02)01411-3.CrossRefPubMed Nozaki M, Kimura H, Kojima M, Ogura Y: Optical coherence tomographic findings of the anterior segment after nonpenetrating deep sclerectomy. Am J Ophthalmol. 2002, 133 (6): 837-839. 10.1016/S0002-9394(02)01411-3.CrossRefPubMed
33.
go back to reference Wang J, Fonn D, Simpson TL, Jones L: Precorneal and pre- and postlens tear film thickness measured indirectly with optical coherence tomography. Invest Ophthalmol Vis Sci. 2003, 44 (6): 2524-2528. 10.1167/iovs.02-0731.CrossRefPubMed Wang J, Fonn D, Simpson TL, Jones L: Precorneal and pre- and postlens tear film thickness measured indirectly with optical coherence tomography. Invest Ophthalmol Vis Sci. 2003, 44 (6): 2524-2528. 10.1167/iovs.02-0731.CrossRefPubMed
34.
go back to reference Wirbelauer C, Winkler J, Scholz C, Haberle H, Pham DT: Experimental imaging of intracorneal ring segments with optical coherence tomography. J Refract Surg. 2003, 19 (3): 367-371.PubMed Wirbelauer C, Winkler J, Scholz C, Haberle H, Pham DT: Experimental imaging of intracorneal ring segments with optical coherence tomography. J Refract Surg. 2003, 19 (3): 367-371.PubMed
35.
go back to reference Linnola RJ, Findl O, Hermann B, Sattmann H, Unterhuber A, Happonen RP, Drexler W: Intraocular lens-capsular bag imaging with ultrahigh-resolution optical coherence tomography Pseudophakic human autopsy eyes. J Cataract Refract Surg. 2005, 31 (4): 818-823. 10.1016/j.jcrs.2004.06.050.CrossRefPubMed Linnola RJ, Findl O, Hermann B, Sattmann H, Unterhuber A, Happonen RP, Drexler W: Intraocular lens-capsular bag imaging with ultrahigh-resolution optical coherence tomography Pseudophakic human autopsy eyes. J Cataract Refract Surg. 2005, 31 (4): 818-823. 10.1016/j.jcrs.2004.06.050.CrossRefPubMed
36.
go back to reference Zhao Y, Li J, Lu W, Chang P, Lu P, Yu F, Xing X, Ding X, Lu F: Capsular adhesion to intraocular lens in highly myopic eyes evaluated in vivo using ultralong-scan-depth optical coherence tomography. Am J Ophthalmol. 2013, 155 (3): 484-491. 10.1016/j.ajo.2012.08.019. e481CrossRefPubMed Zhao Y, Li J, Lu W, Chang P, Lu P, Yu F, Xing X, Ding X, Lu F: Capsular adhesion to intraocular lens in highly myopic eyes evaluated in vivo using ultralong-scan-depth optical coherence tomography. Am J Ophthalmol. 2013, 155 (3): 484-491. 10.1016/j.ajo.2012.08.019. e481CrossRefPubMed
37.
go back to reference Cheng CY, Yen MY, Chen SJ, Kao SC, Hsu WM, Liu JH: Visual acuity and contrast sensitivity in different types of posterior capsule opacification. J Cataract Refract Surg. 2001, 27 (7): 1055-1060. 10.1016/S0886-3350(00)00867-1.CrossRefPubMed Cheng CY, Yen MY, Chen SJ, Kao SC, Hsu WM, Liu JH: Visual acuity and contrast sensitivity in different types of posterior capsule opacification. J Cataract Refract Surg. 2001, 27 (7): 1055-1060. 10.1016/S0886-3350(00)00867-1.CrossRefPubMed
38.
go back to reference Miyake K, Ota I, Miyake S, Horiguchi M: Liquefied aftercataract: a complication of continuous curvilinear capsulorhexis and intraocular lens implantation in the lens capsule. Am J Ophthalmol. 1998, 125 (4): 429-435. 10.1016/S0002-9394(99)80182-2.CrossRefPubMed Miyake K, Ota I, Miyake S, Horiguchi M: Liquefied aftercataract: a complication of continuous curvilinear capsulorhexis and intraocular lens implantation in the lens capsule. Am J Ophthalmol. 1998, 125 (4): 429-435. 10.1016/S0002-9394(99)80182-2.CrossRefPubMed
39.
go back to reference Tan YL, Mohanram LS, Ti SE, Aung T, Perera S: Imaging late capsular bag distension syndrome: an anterior segment optical coherence tomography study. Clin Ophthalmol. 2012, 6: 1455-1458.CrossRefPubMedPubMedCentral Tan YL, Mohanram LS, Ti SE, Aung T, Perera S: Imaging late capsular bag distension syndrome: an anterior segment optical coherence tomography study. Clin Ophthalmol. 2012, 6: 1455-1458.CrossRefPubMedPubMedCentral
40.
go back to reference Gregorčič P, Petkovšek R, Možina J, Močnik G: Measurements of cavitation bubble dynamics based on beam-deflection probe. Appl Phys A. 2008, 93 (4): 901-905. 10.1007/s00339-008-4751-4. Gregorčič P, Petkovšek R, Možina J, Močnik G: Measurements of cavitation bubble dynamics based on beam-deflection probe. Appl Phys A. 2008, 93 (4): 901-905. 10.1007/s00339-008-4751-4.
41.
go back to reference Orthaber U, Petkovšek R, Schille J, Hartwig L, Hawlina G, Drnovšek-Olup B, Vrečko A, Poberaj I: Effect of laser-induced cavitation bubble on a thin elastic membrane. Opt Laser Technol. 2014, 64: 94-100.CrossRef Orthaber U, Petkovšek R, Schille J, Hartwig L, Hawlina G, Drnovšek-Olup B, Vrečko A, Poberaj I: Effect of laser-induced cavitation bubble on a thin elastic membrane. Opt Laser Technol. 2014, 64: 94-100.CrossRef
42.
go back to reference Vogel A, Hentschel W, Holzfuss J, Lauterborn W: Cavitation bubble dynamics and acoustic transient generation in ocular surgery with pulsed neodymium: YAG lasers. Ophthalmology. 1986, 93 (10): 1259-1269. 10.1016/S0161-6420(86)33576-0.CrossRefPubMed Vogel A, Hentschel W, Holzfuss J, Lauterborn W: Cavitation bubble dynamics and acoustic transient generation in ocular surgery with pulsed neodymium: YAG lasers. Ophthalmology. 1986, 93 (10): 1259-1269. 10.1016/S0161-6420(86)33576-0.CrossRefPubMed
43.
go back to reference Hayashi H, Hayashi K, Nakao F, Hayashi F: Quantitative comparison of posterior capsule opacification after polymethylmethacrylate, silicone, and soft acrylic intraocular lens implantation. Arch Ophthalmol. 1998, 116 (12): 1579-1582. 10.1001/archopht.116.12.1579.CrossRefPubMed Hayashi H, Hayashi K, Nakao F, Hayashi F: Quantitative comparison of posterior capsule opacification after polymethylmethacrylate, silicone, and soft acrylic intraocular lens implantation. Arch Ophthalmol. 1998, 116 (12): 1579-1582. 10.1001/archopht.116.12.1579.CrossRefPubMed
44.
go back to reference Hayashi K, Hayashi H, Nakao F, Hayashi F: Changes in posterior capsule opacification after poly(methyl methacrylate), silicone, and acrylic intraocular lens implantation. J Cataract Refract Surg. 2001, 27 (6): 817-824. 10.1016/S0886-3350(01)00787-8.CrossRefPubMed Hayashi K, Hayashi H, Nakao F, Hayashi F: Changes in posterior capsule opacification after poly(methyl methacrylate), silicone, and acrylic intraocular lens implantation. J Cataract Refract Surg. 2001, 27 (6): 817-824. 10.1016/S0886-3350(01)00787-8.CrossRefPubMed
45.
go back to reference Johnston RL, Spalton DJ, Hussain A, Marshall J: In vitro protein adsorption to 2 intraocular lens materials. J Cataract Refract Surg. 1999, 25 (8): 1109-1115. 10.1016/S0886-3350(99)00137-6.CrossRefPubMed Johnston RL, Spalton DJ, Hussain A, Marshall J: In vitro protein adsorption to 2 intraocular lens materials. J Cataract Refract Surg. 1999, 25 (8): 1109-1115. 10.1016/S0886-3350(99)00137-6.CrossRefPubMed
46.
go back to reference Ursell PG, Spalton DJ, Pande MV, Hollick EJ, Barman S, Boyce J, Tilling K: Relationship between intraocular lens biomaterials and posterior capsule opacification. J Cataract Refract Surg. 1998, 24 (3): 352-360. 10.1016/S0886-3350(98)80323-4.CrossRefPubMed Ursell PG, Spalton DJ, Pande MV, Hollick EJ, Barman S, Boyce J, Tilling K: Relationship between intraocular lens biomaterials and posterior capsule opacification. J Cataract Refract Surg. 1998, 24 (3): 352-360. 10.1016/S0886-3350(98)80323-4.CrossRefPubMed
47.
go back to reference Kanellopoulos AJ, Asimellis G: Digital analysis of flap parameter accuracy and objective assessment of opaque bubble layer in femtosecond laser-assisted LASIK: a novel technique. Clin Ophthalmol. 2013, 7: 343-351.CrossRefPubMedPubMedCentral Kanellopoulos AJ, Asimellis G: Digital analysis of flap parameter accuracy and objective assessment of opaque bubble layer in femtosecond laser-assisted LASIK: a novel technique. Clin Ophthalmol. 2013, 7: 343-351.CrossRefPubMedPubMedCentral
Metadata
Title
Optical coherence tomography for an in-vivo study of posterior-capsule-opacification types and their influence on the total-pulse energy required for Nd:YAG capsulotomy: a case series
Authors
Gregor Hawlina
Darko Perovšek
Brigita Drnovšek-Olup
Janez Možina
Peter Gregorčič
Publication date
01-12-2014
Publisher
BioMed Central
Published in
BMC Ophthalmology / Issue 1/2014
Electronic ISSN: 1471-2415
DOI
https://doi.org/10.1186/1471-2415-14-131

Other articles of this Issue 1/2014

BMC Ophthalmology 1/2014 Go to the issue