Published in: BMC Medical Imaging 1/2014

Open Access 01-12-2014 | Research article

Influence of trigger type, tube voltage and heart rate on calcified plaque imaging in dual source cardiac computed tomography: phantom study

Authors: Tobias Penzkofer, Eva Donandt, Peter Isfort, Thomas Allmendinger, Christiane K Kuhl, Andreas H Mahnken, Philipp Bruners

Published in: BMC Medical Imaging | Issue 1/2014

Login to get access

Abstract

Background

To investigate the impact of high pitch cardiac CT vs. retrospective ECG gated CT on the quantification of calcified vessel stenoses, with assessment of the influence of tube voltage, reconstruction kernel and heart rate.

Methods

A 4D cardiac movement phantom equipped with three different plaque phantoms (12.5%, 25% and 50% stenosis at different calcification levels), was scanned with a 128-row dual source CT scanner, applying different trigger types (gated vs. prospectively triggered high pitch), tube voltages (100-120 kV) and heart rates (50–90 beats per minute, bpm). Images were reconstructed using different standard (B26f, B46f, B70f) and iterative (I26f, I70f) convolution kernels. Absolute and relative plaque sizes were measured and statistically compared. Radiation dose associated with the different methods (gated vs. high pitch, 100 kV vs. 120 kV) were compared.

Results

Compared to the known diameters of the phantom plaques and vessels both CT-examination techniques overestimated the degrees of stenoses. Using the high pitch CT-protocol plaques appeared larger (0.09 ± 0.31 mm, 2 ± 8 percent points, PP) in comparison to the ECG-gated CT-scans. Reducing tube voltage had a similar effect, resulting in higher grading of the same stenoses by 3 ± 8 PP. In turn, sharper convolution kernels lead to a lower grading of stenoses (differences of up to 5%). Pairwise comparison of B26f and I26f, B46f and B70f, and B70f and I70f showed differences of 0–1 ± 6–8 PP of the plaque depiction. Motion artifacts were present only at 90 bpm high pitch experiments. High-pitch protocols were associated with significantly lower radiation doses compared with the ECG-gated protocols (258.0 mGy vs. 2829.8 mGy CTDIvol, p ≤ 0.0001).

Conclusion

Prospectively triggered high-pitch cardiac CT led to an overestimation of plaque diameter and degree of stenoses in a coronary phantom. This overestimation is only slight and probably negligible in a clinical situation. Even at higher heart rates high pitch CT-scanning allowed reliable measurements of plaque and vessel diameters with only slight differences compared ECG-gated protocols, although motion artifacts were present at 90 bpm using the high pitch protocols.
Appendix
Available only for authorised users
Literature
1.
go back to reference Vorobiof G, Achenbach S, Narula J: Minimizing radiation dose for coronary CT angiography. Cardiol Clin. 2012, 30 (1): 9-17. 10.1016/j.ccl.2011.11.003.CrossRefPubMed Vorobiof G, Achenbach S, Narula J: Minimizing radiation dose for coronary CT angiography. Cardiol Clin. 2012, 30 (1): 9-17. 10.1016/j.ccl.2011.11.003.CrossRefPubMed
2.
go back to reference Alkadhi H: Radiation dose of cardiac CT–what is the evidence?. Eur Radiol. 2009, 19 (6): 1311-1315. 10.1007/s00330-009-1312-y.CrossRefPubMed Alkadhi H: Radiation dose of cardiac CT–what is the evidence?. Eur Radiol. 2009, 19 (6): 1311-1315. 10.1007/s00330-009-1312-y.CrossRefPubMed
3.
go back to reference Einstein AJ, Moser KW, Thompson RC, Cerqueira MD, Henzlova MJ: Radiation dose to patients from cardiac diagnostic imaging. Circulation. 2007, 116 (11): 1290-1305. 10.1161/CIRCULATIONAHA.107.688101.CrossRefPubMed Einstein AJ, Moser KW, Thompson RC, Cerqueira MD, Henzlova MJ: Radiation dose to patients from cardiac diagnostic imaging. Circulation. 2007, 116 (11): 1290-1305. 10.1161/CIRCULATIONAHA.107.688101.CrossRefPubMed
4.
go back to reference Achenbach S, Marwan M, Ropers D, Schepis T, Pflederer T, Anders K, Kuettner A, Daniel WG, Uder M, Lell MM: Coronary computed tomography angiography with a consistent dose below 1 mSv using prospectively electrocardiogram-triggered high-pitch spiral acquisition. Eur Heart J. 2010, 31 (3): 340-346. 10.1093/eurheartj/ehp470.CrossRefPubMed Achenbach S, Marwan M, Ropers D, Schepis T, Pflederer T, Anders K, Kuettner A, Daniel WG, Uder M, Lell MM: Coronary computed tomography angiography with a consistent dose below 1 mSv using prospectively electrocardiogram-triggered high-pitch spiral acquisition. Eur Heart J. 2010, 31 (3): 340-346. 10.1093/eurheartj/ehp470.CrossRefPubMed
5.
go back to reference Flohr TG, Leng S, Yu L, Allmendinger T, Bruder H, Petersilka M, Eusemann CD, Stierstorfer K, Schmidt B, McCollough CH: Dual-source spiral CT with pitch up to 3.2 and 75 ms temporal resolution: image reconstruction and assessment of image quality. Med Phys. 2009, 36 (12): 5641-5653. 10.1118/1.3259739.CrossRefPubMed Flohr TG, Leng S, Yu L, Allmendinger T, Bruder H, Petersilka M, Eusemann CD, Stierstorfer K, Schmidt B, McCollough CH: Dual-source spiral CT with pitch up to 3.2 and 75 ms temporal resolution: image reconstruction and assessment of image quality. Med Phys. 2009, 36 (12): 5641-5653. 10.1118/1.3259739.CrossRefPubMed
6.
go back to reference Flohr TG, McCollough CH, Bruder H, Petersilka M, Gruber K, Suss C, Grasruck M, Stierstorfer K, Krauss B, Raupach R, Primak AN, Kuttner A, Achenbach S, Becker C, Kopp A, Ohnesorge BM: First performance evaluation of a dual-source CT (DSCT) system. Eur Radiol. 2006, 16 (2): 256-268. 10.1007/s00330-005-2919-2.CrossRefPubMed Flohr TG, McCollough CH, Bruder H, Petersilka M, Gruber K, Suss C, Grasruck M, Stierstorfer K, Krauss B, Raupach R, Primak AN, Kuttner A, Achenbach S, Becker C, Kopp A, Ohnesorge BM: First performance evaluation of a dual-source CT (DSCT) system. Eur Radiol. 2006, 16 (2): 256-268. 10.1007/s00330-005-2919-2.CrossRefPubMed
7.
go back to reference Morsbach F, Gordic S, Desbiolles L, Husarik D, Frauenfelder T, Schmidt B, Allmendinger T, Wildermuth S, Alkadhi H, Leschka S: Performance of turbo high-pitch dual-source CT for coronary CT angiography: first ex vivo and patient experience. Eur Radiol. 2014, 24 (8): 1889-1895. 10.1007/s00330-014-3209-7.CrossRefPubMed Morsbach F, Gordic S, Desbiolles L, Husarik D, Frauenfelder T, Schmidt B, Allmendinger T, Wildermuth S, Alkadhi H, Leschka S: Performance of turbo high-pitch dual-source CT for coronary CT angiography: first ex vivo and patient experience. Eur Radiol. 2014, 24 (8): 1889-1895. 10.1007/s00330-014-3209-7.CrossRefPubMed
8.
go back to reference Gordic S, Husarik DB, Desbiolles L, Leschka S, Frauenfelder T, Alkadhi H: High-pitch coronary CT angiography with third generation dual-source CT: limits of heart rate. Int J Cardiovasc Imaging. 2014, 30 (6): 1173-1179. 10.1007/s10554-014-0445-5.CrossRefPubMed Gordic S, Husarik DB, Desbiolles L, Leschka S, Frauenfelder T, Alkadhi H: High-pitch coronary CT angiography with third generation dual-source CT: limits of heart rate. Int J Cardiovasc Imaging. 2014, 30 (6): 1173-1179. 10.1007/s10554-014-0445-5.CrossRefPubMed
9.
go back to reference Ulzheimer S, Kalender WA: Assessment of calcium scoring performance in cardiac computed tomography. Eur Radiol. 2003, 13 (3): 484-497.PubMed Ulzheimer S, Kalender WA: Assessment of calcium scoring performance in cardiac computed tomography. Eur Radiol. 2003, 13 (3): 484-497.PubMed
10.
go back to reference Hausleiter J, Meyer T, Hermann F, Hadamitzky M, Krebs M, Gerber TC, McCollough C, Martinoff S, Kastrati A, Schomig A, Achenbach S: Estimated radiation dose associated with cardiac CT angiography. JAMA. 2009, 301 (5): 500-507. 10.1001/jama.2009.54.CrossRefPubMed Hausleiter J, Meyer T, Hermann F, Hadamitzky M, Krebs M, Gerber TC, McCollough C, Martinoff S, Kastrati A, Schomig A, Achenbach S: Estimated radiation dose associated with cardiac CT angiography. JAMA. 2009, 301 (5): 500-507. 10.1001/jama.2009.54.CrossRefPubMed
11.
go back to reference Coles DR, Smail MA, Negus IS, Wilde P, Oberhoff M, Karsch KR, Baumbach A: Comparison of radiation doses from multislice computed tomography coronary angiography and conventional diagnostic angiography. J Am Coll Cardiol. 2006, 47 (9): 1840-1845. 10.1016/j.jacc.2005.11.078.CrossRefPubMed Coles DR, Smail MA, Negus IS, Wilde P, Oberhoff M, Karsch KR, Baumbach A: Comparison of radiation doses from multislice computed tomography coronary angiography and conventional diagnostic angiography. J Am Coll Cardiol. 2006, 47 (9): 1840-1845. 10.1016/j.jacc.2005.11.078.CrossRefPubMed
12.
go back to reference Hoffman U, Venkatesh V, White RD, Woodard PK, Carr JJ, Dorbala S, Earls JP, Jacobs JE, Mammen L, Martin ET, Ryan T, White CS: ACR Appropriateness Criteria: Acute Nonspecific Chest Pain - low Probability of Coronary Artery Disease. 2011, Reston, VA, USA: American College of Radiology, ACR Appropriateness Criteria Hoffman U, Venkatesh V, White RD, Woodard PK, Carr JJ, Dorbala S, Earls JP, Jacobs JE, Mammen L, Martin ET, Ryan T, White CS: ACR Appropriateness Criteria: Acute Nonspecific Chest Pain - low Probability of Coronary Artery Disease. 2011, Reston, VA, USA: American College of Radiology, ACR Appropriateness Criteria
13.
go back to reference Pflederer T, Rudofsky L, Ropers D, Bachmann S, Marwan M, Daniel WG, Achenbach S: Image quality in a low radiation exposure protocol for retrospectively ECG-gated coronary CT angiography. AJR Am J Roentgenol. 2009, 192 (4): 1045-1050. 10.2214/AJR.08.1025.CrossRefPubMed Pflederer T, Rudofsky L, Ropers D, Bachmann S, Marwan M, Daniel WG, Achenbach S: Image quality in a low radiation exposure protocol for retrospectively ECG-gated coronary CT angiography. AJR Am J Roentgenol. 2009, 192 (4): 1045-1050. 10.2214/AJR.08.1025.CrossRefPubMed
14.
go back to reference Paul JF, Abada HT: Strategies for reduction of radiation dose in cardiac multislice CT. Eur Radiol. 2007, 17 (8): 2028-2037. 10.1007/s00330-007-0584-3.CrossRefPubMed Paul JF, Abada HT: Strategies for reduction of radiation dose in cardiac multislice CT. Eur Radiol. 2007, 17 (8): 2028-2037. 10.1007/s00330-007-0584-3.CrossRefPubMed
15.
go back to reference Maruyama T, Takada M, Hasuike T, Yoshikawa A, Namimatsu E, Yoshizumi T: Radiation dose reduction and coronary assessability of prospective electrocardiogram-gated computed tomography coronary angiography: comparison with retrospective electrocardiogram-gated helical scan. J Am Coll Cardiol. 2008, 52 (18): 1450-1455. 10.1016/j.jacc.2008.07.048.CrossRefPubMed Maruyama T, Takada M, Hasuike T, Yoshikawa A, Namimatsu E, Yoshizumi T: Radiation dose reduction and coronary assessability of prospective electrocardiogram-gated computed tomography coronary angiography: comparison with retrospective electrocardiogram-gated helical scan. J Am Coll Cardiol. 2008, 52 (18): 1450-1455. 10.1016/j.jacc.2008.07.048.CrossRefPubMed
16.
go back to reference Flohr TG, Bruder H, Stierstorfer K, Petersilka M, Schmidt B, McCollough CH: Image reconstruction and image quality evaluation for a dual source CT scanner. Med Phys. 2008, 35 (12): 5882-5897. 10.1118/1.3020756.CrossRefPubMedPubMedCentral Flohr TG, Bruder H, Stierstorfer K, Petersilka M, Schmidt B, McCollough CH: Image reconstruction and image quality evaluation for a dual source CT scanner. Med Phys. 2008, 35 (12): 5882-5897. 10.1118/1.3020756.CrossRefPubMedPubMedCentral
17.
go back to reference Achenbach S, Marwan M, Schepis T, Pflederer T, Bruder H, Allmendinger T, Petersilka M, Anders K, Lell M, Kuettner A, Ropers D, Daniel WG, Flohr T: High-pitch spiral acquisition: a new scan mode for coronary CT angiography. J Cardiovasc Comput Tomogr. 2009, 3 (2): 117-121. 10.1016/j.jcct.2009.02.008.CrossRefPubMed Achenbach S, Marwan M, Schepis T, Pflederer T, Bruder H, Allmendinger T, Petersilka M, Anders K, Lell M, Kuettner A, Ropers D, Daniel WG, Flohr T: High-pitch spiral acquisition: a new scan mode for coronary CT angiography. J Cardiovasc Comput Tomogr. 2009, 3 (2): 117-121. 10.1016/j.jcct.2009.02.008.CrossRefPubMed
18.
go back to reference Achenbach S, Goroll T, Seltmann M, Pflederer T, Anders K, Ropers D, Daniel WG, Uder M, Lell M, Marwan M: Detection of coronary artery stenoses by low-dose, prospectively ECG-triggered, high-pitch spiral coronary CT angiography. JACC Cardiovasc Imaging. 2011, 4 (4): 328-337. 10.1016/j.jcmg.2011.01.012.CrossRefPubMed Achenbach S, Goroll T, Seltmann M, Pflederer T, Anders K, Ropers D, Daniel WG, Uder M, Lell M, Marwan M: Detection of coronary artery stenoses by low-dose, prospectively ECG-triggered, high-pitch spiral coronary CT angiography. JACC Cardiovasc Imaging. 2011, 4 (4): 328-337. 10.1016/j.jcmg.2011.01.012.CrossRefPubMed
19.
go back to reference Kropil P, Rojas CA, Ghoshhajra B, Lanzman RS, Miese FR, Scherer A, Kalra M, Abbara S: Prospectively ECG-triggered high-pitch spiral acquisition for cardiac CT angiography in routine clinical practice: initial results. J Thorac Imaging. 2012, 27 (3): 194-201. 10.1097/RTI.0b013e318228223f.CrossRefPubMedPubMedCentral Kropil P, Rojas CA, Ghoshhajra B, Lanzman RS, Miese FR, Scherer A, Kalra M, Abbara S: Prospectively ECG-triggered high-pitch spiral acquisition for cardiac CT angiography in routine clinical practice: initial results. J Thorac Imaging. 2012, 27 (3): 194-201. 10.1097/RTI.0b013e318228223f.CrossRefPubMedPubMedCentral
20.
go back to reference Dey D, Slomka P, Chien D, Fieno D, Abidov A, Saouaf R, Thomson L, Friedman JD, Berman DS: Direct quantitative in vivo comparison of calcified atherosclerotic plaque on vascular MRI and CT by multimodality image registration. J Magn Reson Imaging. 2006, 23 (3): 345-354. 10.1002/jmri.20520.CrossRefPubMed Dey D, Slomka P, Chien D, Fieno D, Abidov A, Saouaf R, Thomson L, Friedman JD, Berman DS: Direct quantitative in vivo comparison of calcified atherosclerotic plaque on vascular MRI and CT by multimodality image registration. J Magn Reson Imaging. 2006, 23 (3): 345-354. 10.1002/jmri.20520.CrossRefPubMed
21.
go back to reference Joseph PM, Spital RD: The exponential edge-gradient effect in x-ray computed tomography. Phys Med Biol. 1981, 26 (3): 473-487. 10.1088/0031-9155/26/3/010.CrossRefPubMed Joseph PM, Spital RD: The exponential edge-gradient effect in x-ray computed tomography. Phys Med Biol. 1981, 26 (3): 473-487. 10.1088/0031-9155/26/3/010.CrossRefPubMed
22.
go back to reference Scharf M, Bink R, May MS, Hentschke C, Achenbach S, Uder M, Lell MM: High-pitch thoracic CT with simultaneous assessment of coronary arteries: effect of heart rate and heart rate variability on image quality and diagnostic accuracy. JACC Cardiovasc Imaging. 2011, 4 (6): 602-609. 10.1016/j.jcmg.2011.02.014.CrossRefPubMed Scharf M, Bink R, May MS, Hentschke C, Achenbach S, Uder M, Lell MM: High-pitch thoracic CT with simultaneous assessment of coronary arteries: effect of heart rate and heart rate variability on image quality and diagnostic accuracy. JACC Cardiovasc Imaging. 2011, 4 (6): 602-609. 10.1016/j.jcmg.2011.02.014.CrossRefPubMed
23.
go back to reference Feuchtner G, Goetti R, Plass A, Baumueller S, Stolzmann P, Scheffel H, Wieser M, Marincek B, Alkadhi H, Leschka S: Dual-step prospective ECG-triggered 128-slice dual-source CT for evaluation of coronary arteries and cardiac function without heart rate control: a technical note. Eur Radiol. 2010, 20 (9): 2092-2099. 10.1007/s00330-010-1794-7.CrossRefPubMed Feuchtner G, Goetti R, Plass A, Baumueller S, Stolzmann P, Scheffel H, Wieser M, Marincek B, Alkadhi H, Leschka S: Dual-step prospective ECG-triggered 128-slice dual-source CT for evaluation of coronary arteries and cardiac function without heart rate control: a technical note. Eur Radiol. 2010, 20 (9): 2092-2099. 10.1007/s00330-010-1794-7.CrossRefPubMed
24.
go back to reference Lell M, Marwan M, Schepis T, Pflederer T, Anders K, Flohr T, Allmendinger T, Kalender W, Ertel D, Thierfelder C, Kuettner A, Ropers D, Daniel WG, Achenbach S: Prospectively ECG-triggered high-pitch spiral acquisition for coronary CT angiography using dual source CT: technique and initial experience. Eur Radiol. 2009, 19 (11): 2576-2583. 10.1007/s00330-009-1558-4.CrossRefPubMed Lell M, Marwan M, Schepis T, Pflederer T, Anders K, Flohr T, Allmendinger T, Kalender W, Ertel D, Thierfelder C, Kuettner A, Ropers D, Daniel WG, Achenbach S: Prospectively ECG-triggered high-pitch spiral acquisition for coronary CT angiography using dual source CT: technique and initial experience. Eur Radiol. 2009, 19 (11): 2576-2583. 10.1007/s00330-009-1558-4.CrossRefPubMed
Metadata
Title
Influence of trigger type, tube voltage and heart rate on calcified plaque imaging in dual source cardiac computed tomography: phantom study
Authors
Tobias Penzkofer
Eva Donandt
Peter Isfort
Thomas Allmendinger
Christiane K Kuhl
Andreas H Mahnken
Philipp Bruners
Publication date
01-12-2014
Publisher
BioMed Central
Published in
BMC Medical Imaging / Issue 1/2014
Electronic ISSN: 1471-2342
DOI
https://doi.org/10.1186/1471-2342-14-30

Other articles of this Issue 1/2014

BMC Medical Imaging 1/2014 Go to the issue