A study on the mechanical characteristics of the EBM-printed Ti-6Al-4V LCP plates in vitro

Authors: Peng-cheng Liu, Yun-ji Yang, Run Liu, He-xi Shu, Jin-peng Gong, Yong Yang, Qi Sun, Xing Wu, Ming Cai

Published in: Journal of Orthopaedic Surgery and Research | Issue 1/2014

Login to get access

Abstract

Purpose

The electron beam melting (EBM) Ti-6Al-4V material technology has been developed over a short time period. It was introduced through a research to develop Ti-6Al-4V implants for patients, but EBM printed locking compression plates have not been used for clinical implants. The main purpose of this study is to find whether the EBM Ti-6Al-4V plate suit for clinical implants.

Methods

First, we scanned an AO-locking compression plate (LCP) and printed LCP samples using EBM. Next, we evaluated the EBM plate surface roughness through optical microscopy as well as the LCP and EBM plates’ mechanical characteristics using the ASTM standard, which is commonly used to test the mechanical properties of bone plates subject to bending. Each sample was examined using a single-cycle four-point bending test and hardness testing to acquire data on bending stiffness, bending strength, bending structural stiffness, and hardness.

Results

The results show significant differences in bending stiffness, bending strength, bending structural stiffness, and hardness between the samples using EBM and the original LCP plates. The EBM-printed samples’ surface roughness was 0.49 ± 0.02 μm. The mean hardness of the LCP sample was 266.67 HV10 ± 5.8, and the EBM-printed sample mean hardness was 341.1 HV10 ± 1.93. The EBM samples' bending stiffness was 87.67%, which is greater than using the LCP plates'; and the bending strength was 190.7% greater, the bending structural stiffness was 73.2% greater, and the hardness was 27.9% greater.

Conclusions

The results show that the EBM plates' general mechanical strength was significantly greater than the LCP plates. An EBM plate is advantageous for clinical implants because it can be customized with great potential for improvement.
Appendix
Available only for authorised users
Literature
1.
go back to reference Sommer C, Gautier E, Muller M, Helfet DL, Wagner M: First clinical results of the locking compression plate (LCP). Injury. 2003, 34 (Suppl 2): B43-B54. 10.1016/j.injury.2003.09.024.CrossRefPubMed Sommer C, Gautier E, Muller M, Helfet DL, Wagner M: First clinical results of the locking compression plate (LCP). Injury. 2003, 34 (Suppl 2): B43-B54. 10.1016/j.injury.2003.09.024.CrossRefPubMed
2.
go back to reference Zeman J, Matejka J, Matejka T, Salasek M, Zeman P, Nepras P: Open reduction and plate fixation (ORIF LCP) for treatment of bilateral calcaneal fractures. Acta Chir Orthop Traumatol Cech. 2013, 80 (2): 142-147.PubMed Zeman J, Matejka J, Matejka T, Salasek M, Zeman P, Nepras P: Open reduction and plate fixation (ORIF LCP) for treatment of bilateral calcaneal fractures. Acta Chir Orthop Traumatol Cech. 2013, 80 (2): 142-147.PubMed
3.
go back to reference Umer M, Abbas K, Khan S, Rashid HU: Locking compression plate in musculoskeletal oncology `a friend in need’. Clin Orthop Surg. 2013, 5 (4): 321-326. 10.4055/cios.2013.5.4.321.PubMedCentralCrossRefPubMed Umer M, Abbas K, Khan S, Rashid HU: Locking compression plate in musculoskeletal oncology `a friend in need’. Clin Orthop Surg. 2013, 5 (4): 321-326. 10.4055/cios.2013.5.4.321.PubMedCentralCrossRefPubMed
4.
go back to reference Kumar MN, Ravindranath VP, Ravishankar M: Outcome of locking compression plates in humeral shaft nonunions. Indian J Orthop. 2013, 47 (2): 150-155. 10.4103/0019-5413.108899.PubMedCentralCrossRefPubMed Kumar MN, Ravindranath VP, Ravishankar M: Outcome of locking compression plates in humeral shaft nonunions. Indian J Orthop. 2013, 47 (2): 150-155. 10.4103/0019-5413.108899.PubMedCentralCrossRefPubMed
5.
go back to reference Khan SA, Shamshery P, Gupta V, Trikha V, Varshney MK, Kumar A: Locking compression plate in long standing clavicular nonunions with poor bone stock. J Trauma. 2008, 64 (2): 439-441. 10.1097/01.ta.0000238716.97303.b3.CrossRefPubMed Khan SA, Shamshery P, Gupta V, Trikha V, Varshney MK, Kumar A: Locking compression plate in long standing clavicular nonunions with poor bone stock. J Trauma. 2008, 64 (2): 439-441. 10.1097/01.ta.0000238716.97303.b3.CrossRefPubMed
6.
go back to reference Harrysson OLA, Cansizoglu O, Marcellin-Little DJ, Cormier DR, West HA: Direct metal fabrication of titanium implants with tailored materials and mechanical properties using electron beam melting technology. Mat Sci Eng C. 2008, 28 (3): 366-373. 10.1016/j.msec.2007.04.022.CrossRef Harrysson OLA, Cansizoglu O, Marcellin-Little DJ, Cormier DR, West HA: Direct metal fabrication of titanium implants with tailored materials and mechanical properties using electron beam melting technology. Mat Sci Eng C. 2008, 28 (3): 366-373. 10.1016/j.msec.2007.04.022.CrossRef
7.
go back to reference Murr LE, Amato KN, Li SJ, Tian YX, Cheng XY, Gaytan SM, Martinez E, Shindo PW, Medina F, Wicker RB: Microstructure and mechanical properties of open-cellular biomaterials prototypes for total knee replacement implants fabricated by electron beam melting. J Mech Behav Biomed Mater. 2011, 4 (7): 1396-1411. 10.1016/j.jmbbm.2011.05.010.CrossRefPubMed Murr LE, Amato KN, Li SJ, Tian YX, Cheng XY, Gaytan SM, Martinez E, Shindo PW, Medina F, Wicker RB: Microstructure and mechanical properties of open-cellular biomaterials prototypes for total knee replacement implants fabricated by electron beam melting. J Mech Behav Biomed Mater. 2011, 4 (7): 1396-1411. 10.1016/j.jmbbm.2011.05.010.CrossRefPubMed
8.
go back to reference Rapuano BE, Singh H, Boskey AL, Doty SB, MacDonald DE: Heat and radiofrequency plasma glow discharge pretreatment of a titanium alloy: evidence [corrected] for enhanced osteoinductive properties. J Cell Biochem. 2013, 114 (8): 1917-1927. 10.1002/jcb.24536.CrossRefPubMed Rapuano BE, Singh H, Boskey AL, Doty SB, MacDonald DE: Heat and radiofrequency plasma glow discharge pretreatment of a titanium alloy: evidence [corrected] for enhanced osteoinductive properties. J Cell Biochem. 2013, 114 (8): 1917-1927. 10.1002/jcb.24536.CrossRefPubMed
9.
go back to reference Cheng XY, Li SJ, Murr LE, Zhang ZB, Hao YL, Yang R, Medina F, Wicker RB: Compression deformation behavior of Ti-6Al-4 V alloy with cellular structures fabricated by electron beam melting. J Mech Behav Biomed Mater. 2012, 16: 153-162. 10.1016/j.jmbbm.2012.10.005.CrossRefPubMed Cheng XY, Li SJ, Murr LE, Zhang ZB, Hao YL, Yang R, Medina F, Wicker RB: Compression deformation behavior of Ti-6Al-4 V alloy with cellular structures fabricated by electron beam melting. J Mech Behav Biomed Mater. 2012, 16: 153-162. 10.1016/j.jmbbm.2012.10.005.CrossRefPubMed
10.
go back to reference Derand P, Rannar LE, Hirsch JM: Imaging, virtual planning, design, and production of patient-specific implants and clinical validation in craniomaxillofacial surgery. Craniomaxillofac Trauma Reconstr. 2012, 5 (3): 137-144. 10.1055/s-0032-1313357.PubMedCentralCrossRefPubMed Derand P, Rannar LE, Hirsch JM: Imaging, virtual planning, design, and production of patient-specific implants and clinical validation in craniomaxillofacial surgery. Craniomaxillofac Trauma Reconstr. 2012, 5 (3): 137-144. 10.1055/s-0032-1313357.PubMedCentralCrossRefPubMed
11.
go back to reference Specification and Test Method for Metallic Bone Plates. ASTM International; 2008. Specification and Test Method for Metallic Bone Plates. ASTM International; 2008.
12.
go back to reference DeTora M, Kraus K: Mechanical testing of 3.5 mm locking and non-locking bone plates. VCOT. 2008, 21 (4): 318-322.PubMed DeTora M, Kraus K: Mechanical testing of 3.5 mm locking and non-locking bone plates. VCOT. 2008, 21 (4): 318-322.PubMed
13.
go back to reference Eden L, Doht S, Frey SP, Ziegler D, Stoyhe J, Fehske K, Blunk T, Meffert RH: Biomechanical comparison of the locking compression superior anterior clavicle plate with seven and ten hole reconstruction plates in midshaft clavicle fracture stabilisation. Int Orthop. 2012, 36 (12): 2537-2543. 10.1007/s00264-012-1671-x.PubMedCentralCrossRefPubMed Eden L, Doht S, Frey SP, Ziegler D, Stoyhe J, Fehske K, Blunk T, Meffert RH: Biomechanical comparison of the locking compression superior anterior clavicle plate with seven and ten hole reconstruction plates in midshaft clavicle fracture stabilisation. Int Orthop. 2012, 36 (12): 2537-2543. 10.1007/s00264-012-1671-x.PubMedCentralCrossRefPubMed
14.
go back to reference Peivandi MT, Yusof-Sani MR, Amel-Farzad H: Exploring the reasons for orthopedic implant failure in traumatic fractures of the lower limb. Arch Iran Med. 2013, 16 (8): 478-482.PubMed Peivandi MT, Yusof-Sani MR, Amel-Farzad H: Exploring the reasons for orthopedic implant failure in traumatic fractures of the lower limb. Arch Iran Med. 2013, 16 (8): 478-482.PubMed
15.
go back to reference Harrysson OLA, Cormier DR: Direct Fabrication of Custom Orthopedic Implants Using Electron Beam Melting Technology. 2005, John Wiley & Sons, LtdCrossRef Harrysson OLA, Cormier DR: Direct Fabrication of Custom Orthopedic Implants Using Electron Beam Melting Technology. 2005, John Wiley & Sons, LtdCrossRef
16.
go back to reference Toksvig-Larsen S, Ryd L: Surface characteristics following tibial preparation during total knee arthroplasty. J Arthroplasty. 1994, 9 (1): 63-66. 10.1016/0883-5403(94)90138-4.CrossRefPubMed Toksvig-Larsen S, Ryd L: Surface characteristics following tibial preparation during total knee arthroplasty. J Arthroplasty. 1994, 9 (1): 63-66. 10.1016/0883-5403(94)90138-4.CrossRefPubMed
17.
go back to reference Parthasarathy J, Starly B, Raman S, Christensen A: Mechanical evaluation of porous titanium (Ti6Al4V) structures with electron beam melting (EBM). J Mech Behav Biomed Mater. 2010, 3 (3): 249-259. 10.1016/j.jmbbm.2009.10.006.CrossRefPubMed Parthasarathy J, Starly B, Raman S, Christensen A: Mechanical evaluation of porous titanium (Ti6Al4V) structures with electron beam melting (EBM). J Mech Behav Biomed Mater. 2010, 3 (3): 249-259. 10.1016/j.jmbbm.2009.10.006.CrossRefPubMed
18.
go back to reference Bandyopadhyay A, Espana F, Balla VK, Bose S, Ohgami Y, Davies NM: Influence of porosity on mechanical properties and in vivo response of Ti6Al4V implants. Acta Biomater. 2010, 6 (4): 1640-1648. 10.1016/j.actbio.2009.11.011.PubMedCentralCrossRefPubMed Bandyopadhyay A, Espana F, Balla VK, Bose S, Ohgami Y, Davies NM: Influence of porosity on mechanical properties and in vivo response of Ti6Al4V implants. Acta Biomater. 2010, 6 (4): 1640-1648. 10.1016/j.actbio.2009.11.011.PubMedCentralCrossRefPubMed
19.
go back to reference Wu SH, Li Y, Zhang YQ, Li XK, Yuan CF, Hao YL, Zhang ZY, Guo Z: Porous titanium-6 aluminum-4 vanadium cage has better osseointegration and less micromotion than a poly-ether-ether-ketone cage in sheep vertebral fusion. Artif Organs. 2013, 37 (12): E191-E201. 10.1111/aor.12153.CrossRefPubMed Wu SH, Li Y, Zhang YQ, Li XK, Yuan CF, Hao YL, Zhang ZY, Guo Z: Porous titanium-6 aluminum-4 vanadium cage has better osseointegration and less micromotion than a poly-ether-ether-ketone cage in sheep vertebral fusion. Artif Organs. 2013, 37 (12): E191-E201. 10.1111/aor.12153.CrossRefPubMed
Metadata
Title
A study on the mechanical characteristics of the EBM-printed Ti-6Al-4V LCP plates in vitro
Authors
Peng-cheng Liu
Yun-ji Yang
Run Liu
He-xi Shu
Jin-peng Gong
Yong Yang
Qi Sun
Xing Wu
Ming Cai
Publication date
01-12-2014
Publisher
BioMed Central
Published in
Journal of Orthopaedic Surgery and Research / Issue 1/2014
Electronic ISSN: 1749-799X
DOI
https://doi.org/10.1186/s13018-014-0106-3

Other articles of this Issue 1/2014

Journal of Orthopaedic Surgery and Research 1/2014 Go to the issue