Published in: International Orthopaedics 12/2012

01-12-2012 | Original Paper

Biomechanical comparison of the Locking Compression superior anterior clavicle plate with seven and ten hole reconstruction plates in midshaft clavicle fracture stabilisation

Authors: Lars Eden, Stefanie Doht, Sönke P. Frey, Dirk Ziegler, Jan Stoyhe, Kai Fehske, Torsten Blunk, Rainer H. Meffert

Published in: International Orthopaedics | Issue 12/2012

Login to get access

Abstract

Purpose

The purpose of this study was to characterise the biomechanical properties of the seven hole superior anterior clavicle LCP (locking compression plate) and to compare these with the properties of commonly applied implants used for the stabilisation of clavicular midshaft fractures such as the locking 7- and ten hole reconstruction plate.

Methods

Twenty-four synthetic clavicles were used. A transverse midshaft fracture was induced. The clavicles were fixed with angle stable clavicle LCPs, seven hole and ten hole reconstruction plates (n = 8 each). Twenty cycles of axial compression and torsion were performed for each sample, which was followed by 1,000 cycles of three point bending and ultimately bending to failure. Axial, torsional and cantilever bending stiffness were calculated from the data recorded.

Results

The clavicle LCP showed the highest overall stiffness compared to the seven and ten hole reconstruction plate. Significantly higher stiffness values were found for axial compression and external rotation. In the load-to-failure tests, the ten hole reconstruction plate especially showed early signs of plastic deformation, which might account for early plate insufficiency so frequently observed clinically.

Conclusion

The results indicate that the clavicle LCP, as compared to the reconstruction plates, leads to superior biomechanical stability in the treatment of midshaft clavicle fractures.
Literature
1.
go back to reference Robinson CM (1998) Fractures of the clavicle in the adult. Epidemiology and classification. J Bone Joint Surg Br 80(3):476–484PubMedCrossRef Robinson CM (1998) Fractures of the clavicle in the adult. Epidemiology and classification. J Bone Joint Surg Br 80(3):476–484PubMedCrossRef
2.
go back to reference Canadian Orthopaedic Trauma Society (2007) Nonoperative treatment compared with plate fixation of displaced midshaft clavicular fractures. A multicenter, randomized clinical trial. J Bone Joint Surg Am 89 (1):1–10 Canadian Orthopaedic Trauma Society (2007) Nonoperative treatment compared with plate fixation of displaced midshaft clavicular fractures. A multicenter, randomized clinical trial. J Bone Joint Surg Am 89 (1):1–10
3.
go back to reference Hill JM, McGuire MH, Crosby LA (1997) Closed treatment of displaced middle-third fractures of the clavicle gives poor results. J Bone Joint Surg Br 79(4):537–539PubMedCrossRef Hill JM, McGuire MH, Crosby LA (1997) Closed treatment of displaced middle-third fractures of the clavicle gives poor results. J Bone Joint Surg Br 79(4):537–539PubMedCrossRef
4.
5.
go back to reference Wijdicks FJ, Houwert M, Dijkgraaf M, de Lange D, Oosterhuis K, Clevers G, Verleisdonk EJ (2012) Complications after plate fixation and elastic stable intramedullary nailing of dislocated midshaft clavicle fractures: a retrospective comparison. Int Orthop. doi:10.1007/s00264-012-1615-5 Wijdicks FJ, Houwert M, Dijkgraaf M, de Lange D, Oosterhuis K, Clevers G, Verleisdonk EJ (2012) Complications after plate fixation and elastic stable intramedullary nailing of dislocated midshaft clavicle fractures: a retrospective comparison. Int Orthop. doi:10.​1007/​s00264-012-1615-5
6.
go back to reference Houwert RM, Wijdicks FJ, Steins Bisschop C, Verleisdonk EJ, Kruyt M (2012) Plate fixation versus intramedullary fixation for displaced mid-shaft clavicle fractures: a systematic review. Int Orthop 36(3):579–585. doi:10.1007/s00264-011-1422-4 PubMedCrossRef Houwert RM, Wijdicks FJ, Steins Bisschop C, Verleisdonk EJ, Kruyt M (2012) Plate fixation versus intramedullary fixation for displaced mid-shaft clavicle fractures: a systematic review. Int Orthop 36(3):579–585. doi:10.​1007/​s00264-011-1422-4 PubMedCrossRef
7.
go back to reference Ziegler D, Eden L, Meffert HR (2012) Preliminary, clinical results of a precontoured, angle stable plate in fixation of midshaft clavicle fractures (LCP). 19. Jahreskongress der Deutschen Vereinigung für Schulter-und Ellenbogenchirurgie (DVSE), Berlin, 10–12 May 2012 Ziegler D, Eden L, Meffert HR (2012) Preliminary, clinical results of a precontoured, angle stable plate in fixation of midshaft clavicle fractures (LCP). 19. Jahreskongress der Deutschen Vereinigung für Schulter-und Ellenbogenchirurgie (DVSE), Berlin, 10–12 May 2012
8.
go back to reference Robertson C, Celestre P, Mahar A, Schwartz A (2009) Reconstruction plates for stabilization of mid-shaft clavicle fractures: differences between nonlocked and locked plates in two different positions. J Should Elbow Surg 18(2):204–209. doi:10.1016/j.jse.2008.10.002 CrossRef Robertson C, Celestre P, Mahar A, Schwartz A (2009) Reconstruction plates for stabilization of mid-shaft clavicle fractures: differences between nonlocked and locked plates in two different positions. J Should Elbow Surg 18(2):204–209. doi:10.​1016/​j.​jse.​2008.​10.​002 CrossRef
10.
16.
go back to reference Iannotti MR, Crosby LA, Stafford P, Grayson G, Goulet R (2002) Effects of plate location and selection on the stability of midshaft clavicle osteotomies: a biomechanical study. J Should Elbow Surg 11(5):457–462CrossRef Iannotti MR, Crosby LA, Stafford P, Grayson G, Goulet R (2002) Effects of plate location and selection on the stability of midshaft clavicle osteotomies: a biomechanical study. J Should Elbow Surg 11(5):457–462CrossRef
Metadata
Title
Biomechanical comparison of the Locking Compression superior anterior clavicle plate with seven and ten hole reconstruction plates in midshaft clavicle fracture stabilisation
Authors
Lars Eden
Stefanie Doht
Sönke P. Frey
Dirk Ziegler
Jan Stoyhe
Kai Fehske
Torsten Blunk
Rainer H. Meffert
Publication date
01-12-2012
Publisher
Springer-Verlag
Published in
International Orthopaedics / Issue 12/2012
Print ISSN: 0341-2695
Electronic ISSN: 1432-5195
DOI
https://doi.org/10.1007/s00264-012-1671-x

Other articles of this Issue 12/2012

International Orthopaedics 12/2012 Go to the issue