Published in: Radiation Oncology 1/2016

Open Access 01-12-2016 | Research

Re-irradiating spinal column metastases using IMRT and VMAT with and without flattening filter - a treatment planning study

Authors: Barbara Dobler, Amine Khemissi, Tina Obermeier, Matthias G. Hautmann, Zaira Katsilieri, Oliver Kölbl

Published in: Radiation Oncology | Issue 1/2016

Login to get access

Abstract

Background

The aim of this study was to investigate the potential of the flattening filter free (FFF) mode of a linear accelerator for intensity modulated radiation therapy (IMRT) and volumetric modulated arc therapy (VMAT) for patients with in-field recurrence of vertebral metastases.

Methods

An Elekta Synergy Linac with Agility™ head is used to simulate the treatment of ten patients with locally recurrent spinal column metastases. Four plans were generated for each patient treating the vertebrae sparing the spinal cord: Dual arc VMAT and nine field step and shoot IMRT each with and without flattening filter. Plan quality was assessed considering target coverage and sparing of the spinal cord and normal tissue. All plans were verified by a 2D-ionisation-chamber-array, peripheral doses were measured and compared to calculations. Delivery times were measured and compared. The Wilcoxon test was used for statistical analysis with a significance level of 0.05.

Results

Target coverage, homogeneity index and conformity index were comparable for both flat and flattening filter free beams. The volume of the spinal cord receiving the allowed maximum dose to keep the risk of radiation myelopathy at 0 % was at the same time significantly reduced to below the clinically relevant 1 ccm using FFF mode. In addition the mean dose deposited in the surrounding healthy tissue was significantly reduced in the FFF mode. All four techniques showed equally good gamma scores for plan verification. FFF plans required considerably more MU per fraction dose. Regardless of the large number of MU, out-of-field point dose was significantly lower for FFF plans, with an average reduction of 33 % and mean delivery time was significantly reduced by 22 % using FFF beams. When compared to IMRT FF, VMAT FFF offered even a reduction of 71 % in delivery time and 45 % in peripheral dose.

Conclusions

FFF plans showed a significant improvement in sparing of normal tissue and the spinal cord, keeping target coverage and homogeneity comparable. In addition, delivery times were significantly reduced for FFF treatments, minimizing intrafractional motion as well as strain for the patient. Shortest delivery times were achieved using VMAT FFF. For radiotherapy of spinal column metastases VMAT FFF may therefore be considered the preferable treatment option for the combination of Elekta Synergy Linacs and Oncentra® External Beam v4.5 treatment planning system.
Literature
1.
go back to reference Georg D, Knoos T, McClean B. Current status and future perspective of flattening filter free photon beams. Med Phys. 2011;38(3):1280–93.CrossRefPubMed Georg D, Knoos T, McClean B. Current status and future perspective of flattening filter free photon beams. Med Phys. 2011;38(3):1280–93.CrossRefPubMed
4.
go back to reference Ponisch F, Titt U, Vassiliev ON, Kry SF, Mohan R. Properties of unflattened photon beams shaped by a multileaf collimator. Med Phys. 2006;33(6):1738–46.CrossRefPubMed Ponisch F, Titt U, Vassiliev ON, Kry SF, Mohan R. Properties of unflattened photon beams shaped by a multileaf collimator. Med Phys. 2006;33(6):1738–46.CrossRefPubMed
5.
go back to reference Titt U, Vassiliev ON, Ponisch F, Dong L, Liu H, Mohan R. A flattening filter free photon treatment concept evaluation with Monte Carlo. Med Phys. 2006;33(6):1595–602.CrossRefPubMed Titt U, Vassiliev ON, Ponisch F, Dong L, Liu H, Mohan R. A flattening filter free photon treatment concept evaluation with Monte Carlo. Med Phys. 2006;33(6):1595–602.CrossRefPubMed
9.
go back to reference Parsai EI, Pearson D, Kvale T. Consequences of removing the flattening filter from linear accelerators in generating high dose rate photon beams for clinical applications: A Monte Carlo study verified by measurement. Nucl Inst Methods Phys Res B-Beam Interactions with Materials and Atoms. 2007;26(1–2):755–9. doi:10.1016/j.nimb.2007.03.020.CrossRef Parsai EI, Pearson D, Kvale T. Consequences of removing the flattening filter from linear accelerators in generating high dose rate photon beams for clinical applications: A Monte Carlo study verified by measurement. Nucl Inst Methods Phys Res B-Beam Interactions with Materials and Atoms. 2007;26(1–2):755–9. doi:10.​1016/​j.​nimb.​2007.​03.​020.CrossRef
11.
go back to reference Kry SF, Howell RM, Titt U, Salehpour M, Mohan R, Vassiliev ON. Energy spectra, sources, and shielding considerations for neutrons generated by a flattening filter-free Clinac. Med Phys. 2008;35(5):1906–11.CrossRefPubMed Kry SF, Howell RM, Titt U, Salehpour M, Mohan R, Vassiliev ON. Energy spectra, sources, and shielding considerations for neutrons generated by a flattening filter-free Clinac. Med Phys. 2008;35(5):1906–11.CrossRefPubMed
13.
go back to reference Almberg SS, Frengen J, Lindmo T. Monte Carlo study of in-field and out-of-field dose distributions from a linear accelerator operating with and without a flattening-filter. Med Phys. 2012;39(8):5194–203. doi:10.1118/1.4738963.CrossRefPubMed Almberg SS, Frengen J, Lindmo T. Monte Carlo study of in-field and out-of-field dose distributions from a linear accelerator operating with and without a flattening-filter. Med Phys. 2012;39(8):5194–203. doi:10.​1118/​1.​4738963.CrossRefPubMed
14.
15.
go back to reference Nicolini G, Ghosh-Laskar S, Shrivastava SK, Banerjee S, Chaudhary S, Agarwal JP, et al. Volumetric modulation arc radiotherapy with flattening filter-free beams compared with static gantry IMRT and 3D conformal radiotherapy for advanced esophageal cancer: a feasibility study. Int J Radiat Oncol Biol Phys. 2012;84(2):553–60. doi:10.1016/j.ijrobp.2011.12.041.CrossRefPubMed Nicolini G, Ghosh-Laskar S, Shrivastava SK, Banerjee S, Chaudhary S, Agarwal JP, et al. Volumetric modulation arc radiotherapy with flattening filter-free beams compared with static gantry IMRT and 3D conformal radiotherapy for advanced esophageal cancer: a feasibility study. Int J Radiat Oncol Biol Phys. 2012;84(2):553–60. doi:10.​1016/​j.​ijrobp.​2011.​12.​041.CrossRefPubMed
17.
go back to reference Shi LW, Lai YQ, Lin Q, Ha HM, Fu LR. The Effect of Flattening Filter Free on Three-dimensional Conformal Radiation Therapy (3D-CRT), Intensity-Modulated Radiation Therapy (IMRT), and Volumetric Modulated Arc Therapy (VMAT) Plans for Metastatic Brain Tumors from Non-small Cell Lung Cancer. Health Phys. 2015;109(1):1–9. doi:10.1097/HP.0000000000000278.CrossRefPubMed Shi LW, Lai YQ, Lin Q, Ha HM, Fu LR. The Effect of Flattening Filter Free on Three-dimensional Conformal Radiation Therapy (3D-CRT), Intensity-Modulated Radiation Therapy (IMRT), and Volumetric Modulated Arc Therapy (VMAT) Plans for Metastatic Brain Tumors from Non-small Cell Lung Cancer. Health Phys. 2015;109(1):1–9. doi:10.​1097/​HP.​0000000000000278​.CrossRefPubMed
19.
go back to reference Hrbacek J, Lang S, Graydon SN, Klock S, Riesterer O. Dosimetric comparison of flattened and unflattened beams for stereotactic ablative radiotherapy of stage I non-small cell lung cancer. Med Phys. 2014;41(3):031709. doi:10.1118/1.4866231.CrossRefPubMed Hrbacek J, Lang S, Graydon SN, Klock S, Riesterer O. Dosimetric comparison of flattened and unflattened beams for stereotactic ablative radiotherapy of stage I non-small cell lung cancer. Med Phys. 2014;41(3):031709. doi:10.​1118/​1.​4866231.CrossRefPubMed
21.
go back to reference Subramaniam S, Thirumalaiswamy S, Srinivas C, Gandhi GA, Kathirvel M, Kumar KK, et al. Chest wall radiotherapy with volumetric modulated arcs and the potential role of flattening filter free photon beams. Strahlenther Onkol. 2012;188(6):484–90. doi:10.1007/s00066-012-0075-6.CrossRefPubMed Subramaniam S, Thirumalaiswamy S, Srinivas C, Gandhi GA, Kathirvel M, Kumar KK, et al. Chest wall radiotherapy with volumetric modulated arcs and the potential role of flattening filter free photon beams. Strahlenther Onkol. 2012;188(6):484–90. doi:10.​1007/​s00066-012-0075-6.CrossRefPubMed
22.
24.
go back to reference Boda-Heggemann J, Mai S, Fleckenstein J, Siebenlist K, Simeonova A, Ehmann M, et al. Flattening-filter-free intensity modulated breath-hold image-guided SABR (Stereotactic ABlative Radiotherapy) can be applied in a 15-min treatment slot. Radiother Oncol. 2013;109(3):505–9. doi:10.1016/j.radonc.2013.09.014.CrossRefPubMed Boda-Heggemann J, Mai S, Fleckenstein J, Siebenlist K, Simeonova A, Ehmann M, et al. Flattening-filter-free intensity modulated breath-hold image-guided SABR (Stereotactic ABlative Radiotherapy) can be applied in a 15-min treatment slot. Radiother Oncol. 2013;109(3):505–9. doi:10.​1016/​j.​radonc.​2013.​09.​014.CrossRefPubMed
25.
go back to reference Murray LJ, Thompson CM, Lilley J, Cosgrove V, Franks K, Sebag-Montefiore D, et al. Radiation-induced second primary cancer risks from modern external beam radiotherapy for early prostate cancer: impact of stereotactic ablative radiotherapy (SABR), volumetric modulated arc therapy (VMAT) and flattening filter free (FFF) radiotherapy. Phys Med Biol. 2015;60(3):1237–57. doi:10.1088/0031-9155/60/3/1237.CrossRefPubMed Murray LJ, Thompson CM, Lilley J, Cosgrove V, Franks K, Sebag-Montefiore D, et al. Radiation-induced second primary cancer risks from modern external beam radiotherapy for early prostate cancer: impact of stereotactic ablative radiotherapy (SABR), volumetric modulated arc therapy (VMAT) and flattening filter free (FFF) radiotherapy. Phys Med Biol. 2015;60(3):1237–57. doi:10.​1088/​0031-9155/​60/​3/​1237.CrossRefPubMed
28.
36.
go back to reference Dobler B, Streck N, Klein E, Loeschel R, Haertl P, Koelbl O. Hybrid plan verification for intensity-modulated radiation therapy (IMRT) using the 2D ionization chamber array I'mRT MatriXX--a feasibility study. Phys Med Biol. 2010;55(2):N39–55. doi:10.1088/0031-9155/55/2/N02.CrossRefPubMed Dobler B, Streck N, Klein E, Loeschel R, Haertl P, Koelbl O. Hybrid plan verification for intensity-modulated radiation therapy (IMRT) using the 2D ionization chamber array I'mRT MatriXX--a feasibility study. Phys Med Biol. 2010;55(2):N39–55. doi:10.​1088/​0031-9155/​55/​2/​N02.CrossRefPubMed
38.
go back to reference International Commission On Radiation Units And Measurements. Prescribing, Recording and Reporting Photon Beam Therapy. Bethesda, Maryland, USA: ICRU; 1993. Report No. 50. ISBN 0-913394-48-3. International Commission On Radiation Units And Measurements. Prescribing, Recording and Reporting Photon Beam Therapy. Bethesda, Maryland, USA: ICRU; 1993. Report No. 50. ISBN 0-913394-48-3.
39.
go back to reference International Commission On Radiation Units And Measurements. Prescribing, Recording, and Reporting Photon-Beam Intensity-Modulated Radiation Therapy (IMRT); Report 83. Journal of the ICRU. 2010;10(1). doi:10.1093/jicru/ndq002. International Commission On Radiation Units And Measurements. Prescribing, Recording, and Reporting Photon-Beam Intensity-Modulated Radiation Therapy (IMRT); Report 83. Journal of the ICRU. 2010;10(1). doi:10.​1093/​jicru/​ndq002.
41.
go back to reference Low DA, Harms WB, Mutic S, Purdy JA. A technique for the quantitative evaluation of dose distributions. Med Phys. 1998;25(5):656–61.CrossRefPubMed Low DA, Harms WB, Mutic S, Purdy JA. A technique for the quantitative evaluation of dose distributions. Med Phys. 1998;25(5):656–61.CrossRefPubMed
42.
go back to reference Ezzell GA, Burmeister JW, Dogan N, LoSasso TJ, Mechalakos JG, Mihailidis D, et al. IMRT commissioning: multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119. Med Phys. 2009;36(11):5359–73.CrossRefPubMed Ezzell GA, Burmeister JW, Dogan N, LoSasso TJ, Mechalakos JG, Mihailidis D, et al. IMRT commissioning: multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119. Med Phys. 2009;36(11):5359–73.CrossRefPubMed
43.
go back to reference Ezzell GA, Galvin JM, Low D, Palta JR, Rosen I, Sharpe MB, et al. Guidance document on delivery, treatment planning, and clinical implementation of IMRT: report of the IMRT Subcommittee of the AAPM Radiation Therapy Committee. Med Phys. 2003;30(8):2089–115.CrossRefPubMed Ezzell GA, Galvin JM, Low D, Palta JR, Rosen I, Sharpe MB, et al. Guidance document on delivery, treatment planning, and clinical implementation of IMRT: report of the IMRT Subcommittee of the AAPM Radiation Therapy Committee. Med Phys. 2003;30(8):2089–115.CrossRefPubMed
44.
go back to reference Zhuang M, Huang L, Zhu D, Peng X, Lin Z. Reirradiation of nasopharyngeal carcinoma focusing on volumetric modulated arcs with flattening filter-free beams. Br J Radiol. 2015;88(1052):20140837. doi:10.1259/bjr.20140837.CrossRefPubMed Zhuang M, Huang L, Zhu D, Peng X, Lin Z. Reirradiation of nasopharyngeal carcinoma focusing on volumetric modulated arcs with flattening filter-free beams. Br J Radiol. 2015;88(1052):20140837. doi:10.​1259/​bjr.​20140837.CrossRefPubMed
Metadata
Title
Re-irradiating spinal column metastases using IMRT and VMAT with and without flattening filter - a treatment planning study
Authors
Barbara Dobler
Amine Khemissi
Tina Obermeier
Matthias G. Hautmann
Zaira Katsilieri
Oliver Kölbl
Publication date
01-12-2016
Publisher
BioMed Central
Published in
Radiation Oncology / Issue 1/2016
Electronic ISSN: 1748-717X
DOI
https://doi.org/10.1186/s13014-016-0603-0

Other articles of this Issue 1/2016

Radiation Oncology 1/2016 Go to the issue