Published in: Head & Face Medicine 1/2020

01-12-2020 | Research

Accuracy of intraoral scans in the mixed dentition: a prospective non-randomized comparative clinical trial

Authors: Konrad Liczmanski, Thomas Stamm, Cristina Sauerland, Moritz Blanck-Lubarsch

Published in: Head & Face Medicine | Issue 1/2020

Login to get access

Abstract

Objective

In-vivo accuracy of intraoral scans of complete mixed dentitions of patients in active treatment have not yet been investigated. The aim was to test the hypothesis that dimensional differences between intraoral scans and conventional alginate impressions in the mixed dentition are clinically irrelevant.

Methods

Trial design: Prospective non-randomized comparative clinical trial. Based on sample size calculation 44 evaluable mixed dentition jaws of patients in active orthodontic treatment were included. Each patient received an alginate impression following an intraoral scan (TRIOS® Ortho). Plaster cast was fabricated and scanned with an external scanner (ATOS-SO®). Both STL datasets were analyzed with the 3D inspection and mesh processing software GOM Inspect®. Statistical analysis comprised sample size calculation, t-test as well as nonparametric tests.

Results

The absolute mean difference between digital plaster casts and intraoral scans is 0.022 mm ± 0.027 mm (median 0.015 mm). The obtained measurements are in the range of comparable studies on full arch permanent dentitions. Gender, the size of the jaw represented by the dentition stage and upper respectively lower jaw, as well the malocclusion have no effect on the total deviations between digital plaster casts and intraoral scans. Detectable impression errors were bubbles in fissures and marginal ridges as well as incomplete alginate flow and detachment from the tray. Detectable scanning errors were incomplete distal surface of the most distal molar.

Conclusion

Dimensional differences between intraoral scans and conventional alginate impressions in the mixed dentition are clinically irrelevant for orthodontic purposes. In all clinical situations of active treatment in the mixed dentition, the intraoral scans are more detailed and less error-prone.
Appendix
Available only for authorised users
Literature
1.
go back to reference Cervino G, et al. Alginate materials and dental impression technique: A current state of the art and application to dental practice. Mar Drugs. 2019;17(1):18.CrossRef Cervino G, et al. Alginate materials and dental impression technique: A current state of the art and application to dental practice. Mar Drugs. 2019;17(1):18.CrossRef
2.
go back to reference Erbe C, Ruf S, Wöstmann B, Balkenhol M. Dimensional stability of contemporary irreversible hydrocolloids: humidor versus wet tissue storage. J Prosthet Dent. 2012;108(2):114–22.CrossRef Erbe C, Ruf S, Wöstmann B, Balkenhol M. Dimensional stability of contemporary irreversible hydrocolloids: humidor versus wet tissue storage. J Prosthet Dent. 2012;108(2):114–22.CrossRef
3.
go back to reference Bocklet C, et al. Effect of scan substrates on accuracy of 7 intraoral digital impression systems using human maxilla model. Orthod Craniofac Res. 2019;22:168–74.CrossRef Bocklet C, et al. Effect of scan substrates on accuracy of 7 intraoral digital impression systems using human maxilla model. Orthod Craniofac Res. 2019;22:168–74.CrossRef
4.
go back to reference Albdour EA, Shaheen E, Vranckx M, Mangano FG, Politis C, Jacobs R. A novel in vivo method to evaluate trueness of digital impressions. BMC Oral Health. 2018;18(1):117.CrossRef Albdour EA, Shaheen E, Vranckx M, Mangano FG, Politis C, Jacobs R. A novel in vivo method to evaluate trueness of digital impressions. BMC Oral Health. 2018;18(1):117.CrossRef
5.
go back to reference Zhang F, Suh K-J, Lee K-M. Validity of intraoral scans compared with plaster models: an in-vivo comparison of dental measurements and 3D surface analysis. PLoS One. 2016;11(6):e0157713.CrossRef Zhang F, Suh K-J, Lee K-M. Validity of intraoral scans compared with plaster models: an in-vivo comparison of dental measurements and 3D surface analysis. PLoS One. 2016;11(6):e0157713.CrossRef
6.
go back to reference Atieh MA, Ritter AV, Ko C-C, Duqum I. Accuracy evaluation of intraoral optical impressions: a clinical study using a reference appliance. J Prosthet Dent. 2017;118(3):400–5.CrossRef Atieh MA, Ritter AV, Ko C-C, Duqum I. Accuracy evaluation of intraoral optical impressions: a clinical study using a reference appliance. J Prosthet Dent. 2017;118(3):400–5.CrossRef
7.
go back to reference Zilberman O, Huggare J, Parikakis KA. Evaluation of the validity of tooth size and arch width measurements using conventional and three-dimensional virtual orthodontic models. Angle Orthod. 2003;73(3):301–6.PubMed Zilberman O, Huggare J, Parikakis KA. Evaluation of the validity of tooth size and arch width measurements using conventional and three-dimensional virtual orthodontic models. Angle Orthod. 2003;73(3):301–6.PubMed
8.
go back to reference D. R. Stevens, “Flores-Mir C, Nebbe B, Raboud DW, Heo G, Major PW. Validity, realiability, and reproducibility of plaster vc digital study models: comparison of peer assessment rating and Bolton analysis and their constituent measurements,” Am J Orthod Dentofac Orthop, vol. 129, no. 6, pp. 794–803, 2006.CrossRef D. R. Stevens, “Flores-Mir C, Nebbe B, Raboud DW, Heo G, Major PW. Validity, realiability, and reproducibility of plaster vc digital study models: comparison of peer assessment rating and Bolton analysis and their constituent measurements,” Am J Orthod Dentofac Orthop, vol. 129, no. 6, pp. 794–803, 2006.CrossRef
9.
go back to reference Naidu D, Freer TJ. Validity, reliability, and reproducibility of the iOC intraoral scanner: a comparison of tooth widths and Bolton ratios. Am J Orthod Dentofac Orthop. 2013;144(2):304–10.CrossRef Naidu D, Freer TJ. Validity, reliability, and reproducibility of the iOC intraoral scanner: a comparison of tooth widths and Bolton ratios. Am J Orthod Dentofac Orthop. 2013;144(2):304–10.CrossRef
10.
go back to reference Wiranto MG, Engelbrecht WP, Nolthenius HET, van der Meer WJ, Ren Y. Validity, reliability, and reproducibility of linear measurements on digital models obtained from intraoral and cone-beam computed tomography scans of alginate impressions. Am J Orthod Dentofac Orthop. 2013;143(1):140–7.CrossRef Wiranto MG, Engelbrecht WP, Nolthenius HET, van der Meer WJ, Ren Y. Validity, reliability, and reproducibility of linear measurements on digital models obtained from intraoral and cone-beam computed tomography scans of alginate impressions. Am J Orthod Dentofac Orthop. 2013;143(1):140–7.CrossRef
11.
go back to reference Grünheid T, McCarthy SD, Larson BE. Clinical use of a direct chairside oral scanner: an assessment of accuracy, time, and patient acceptance. Am J Orthod Dentofac Orthop. 2014;146(5):673–82.CrossRef Grünheid T, McCarthy SD, Larson BE. Clinical use of a direct chairside oral scanner: an assessment of accuracy, time, and patient acceptance. Am J Orthod Dentofac Orthop. 2014;146(5):673–82.CrossRef
12.
go back to reference Park J-M, Choi S, Myung J-Y, Chun Y-S, Kim M. Impact of orthodontic brackets on the intraoral scan data accuracy. BioMed Res Int. 2016;2–3. Park J-M, Choi S, Myung J-Y, Chun Y-S, Kim M. Impact of orthodontic brackets on the intraoral scan data accuracy. BioMed Res Int. 2016;2–3.
13.
go back to reference Ender A, Attin T, Mehl A. In vivo precision of conventional and digital methods of obtaining complete-arch dental impressions. J Prosthet Dent. 2016;115(3):313–20.CrossRef Ender A, Attin T, Mehl A. In vivo precision of conventional and digital methods of obtaining complete-arch dental impressions. J Prosthet Dent. 2016;115(3):313–20.CrossRef
14.
go back to reference Zimmermann M, Koller C, Rumetsch M, Ender A, Mehl A. Precision of guided scanning procedures for full-arch digital impressions in vivo. J Orofac Orthop Kieferorthopädie. 2017;78(6):466–71.CrossRef Zimmermann M, Koller C, Rumetsch M, Ender A, Mehl A. Precision of guided scanning procedures for full-arch digital impressions in vivo. J Orofac Orthop Kieferorthopädie. 2017;78(6):466–71.CrossRef
15.
go back to reference San José V, Bellot-Arcís C, Tarazona B, Zamora N, Lagravère MO, Paredes-Gallardo V. Dental measurements and Bolton index reliability and accuracy obtained from 2D digital, 3D segmented CBCT, and 3d intraoral laser scanner. J Clin Exp Dent. 2017;9(12):e1466.PubMedPubMedCentral San José V, Bellot-Arcís C, Tarazona B, Zamora N, Lagravère MO, Paredes-Gallardo V. Dental measurements and Bolton index reliability and accuracy obtained from 2D digital, 3D segmented CBCT, and 3d intraoral laser scanner. J Clin Exp Dent. 2017;9(12):e1466.PubMedPubMedCentral
16.
go back to reference Lee K-M. Comparison of two intraoral scanners based on three-dimensional surface analysis. Prog. Orthod. 2018;19(1):6.CrossRef Lee K-M. Comparison of two intraoral scanners based on three-dimensional surface analysis. Prog. Orthod. 2018;19(1):6.CrossRef
17.
go back to reference Patzelt SB, Bishti S, Stampf S, Att W. Accuracy of computer-aided design/computer-aided manufacturing–generated dental casts based on intraoral scanner data. J Am Dent Assoc. 2014;145(11):1133–40.CrossRef Patzelt SB, Bishti S, Stampf S, Att W. Accuracy of computer-aided design/computer-aided manufacturing–generated dental casts based on intraoral scanner data. J Am Dent Assoc. 2014;145(11):1133–40.CrossRef
18.
go back to reference Patzelt SB, Emmanouilidi A, Stampf S, Strub JR, Att W. Accuracy of full-arch scans using intraoral scanners. Clin Oral Investig. 2014;18(6):1687–94.CrossRef Patzelt SB, Emmanouilidi A, Stampf S, Strub JR, Att W. Accuracy of full-arch scans using intraoral scanners. Clin Oral Investig. 2014;18(6):1687–94.CrossRef
19.
go back to reference Medina-Sotomayor P, Pascual A, Camps I. Accuracy of four digital scanners according to scanning strategy in complete-arch impressions. PLoS One. 2018;13(9):e0202916.CrossRef Medina-Sotomayor P, Pascual A, Camps I. Accuracy of four digital scanners according to scanning strategy in complete-arch impressions. PLoS One. 2018;13(9):e0202916.CrossRef
20.
go back to reference Yoon JH, Yu H-S, Choi Y, Choi T-H, Choi S-H, Cha J-Y. Model Analysis of Digital Models in Moderate to Severe Crowding: In Vivo Validation and Clinical Application. BioMed Res Int. 2018;2018:2–5. Yoon JH, Yu H-S, Choi Y, Choi T-H, Choi S-H, Cha J-Y. Model Analysis of Digital Models in Moderate to Severe Crowding: In Vivo Validation and Clinical Application. BioMed Res Int. 2018;2018:2–5.
Metadata
Title
Accuracy of intraoral scans in the mixed dentition: a prospective non-randomized comparative clinical trial
Authors
Konrad Liczmanski
Thomas Stamm
Cristina Sauerland
Moritz Blanck-Lubarsch
Publication date
01-12-2020
Publisher
BioMed Central
Published in
Head & Face Medicine / Issue 1/2020
Electronic ISSN: 1746-160X
DOI
https://doi.org/10.1186/s13005-020-00222-6

Other articles of this Issue 1/2020

Head & Face Medicine 1/2020 Go to the issue