Published in: BMC Medical Imaging 1/2022

Open Access 01-12-2022 | Ultrasound | Research

The values of elastic quantitative and semi-quantitative indexes measured from different frequencies in the establishment of prediction models for breast tumor diagnosis

Authors: Xiao Xie, Yibo Ma, Xiaoxiao Xing, Haixia Zhou, Shuiqing Liu, Yanyan Zhang, Min Xu

Published in: BMC Medical Imaging | Issue 1/2022

Login to get access

Abstract

Objective

To obtain the elastic quantitative and semi-quantitative indexes of solid breast masses using ultrasound linear array probes with two different frequencies, and to construct prediction models and evaluate their diagnostic values.

Methods

A total of 92 patients who were scheduled for surgical treatment on solid breast masses were enrolled in this study. Linear array probes with two frequencies, 9-3 MHz (L9 group) and 14-5 MHz (L14 group), were used for sound touch elastography and strain elastography before surgery, and the maximum elasticity value (Emax), average elasticity value (Emean), minimum elasticity value (Emin), standard deviation (SD)(in kPa), elasticity ratio (E), and strain ratio to fat (SRf) were recorded and calculated for the breast mass (A) and surrounding tissues (Shell). The elastic characteristic indexes of the L9 group and L14 group were compared, and the prediction models of these two groups were constructed using Logistic regression method.

Results

The diagnostic performance of the prediction model based on L9 group was better than the model based on L14 group (AUC: 0.904 vs. 0.810, P = 0.0343, z = 2.116) and the best single index EMax-shell-L9 (P = 0.0398, z = 2.056). The sensitivity of L9 based model was 85.19% and the specificity was 84.21%.

Conclusion

The prediction model based on quantitative and semi-quantitative elastic ultrasound indexes from L9-3 probe exhibited better performance, which could improve the diagnostic accuracy for malignant breast tumors.
Appendix
Available only for authorised users
Literature
1.
go back to reference Y Z, Pa X. A Z, et al. Preliminary Study on the Application Value of Sound Touch Elastography in Benign Breast Lesions[J]. Chin J Ultrasound Med. 2020;36(01):18–21. doi. Y Z, Pa X. A Z, et al. Preliminary Study on the Application Value of Sound Touch Elastography in Benign Breast Lesions[J]. Chin J Ultrasound Med. 2020;36(01):18–21. doi.
2.
go back to reference Q LW. Y, J B, et al. Experimental Study on the Measurement of Elastography Index of Modules by Using Shear Wave Technique[J]. Chin J Ultrasound Med. 2018;34(07):655–9. doi. Q LW. Y, J B, et al. Experimental Study on the Measurement of Elastography Index of Modules by Using Shear Wave Technique[J]. Chin J Ultrasound Med. 2018;34(07):655–9. doi.
4.
go back to reference Kehe Z, Fanghong C, Haiping Z. Analysis of the Influence of Mass Depth on Mammography Results of Ultrasonography[J]. J Med Imaging. 2021;31(4):596–9. doi. Kehe Z, Fanghong C, Haiping Z. Analysis of the Influence of Mass Depth on Mammography Results of Ultrasonography[J]. J Med Imaging. 2021;31(4):596–9. doi.
9.
go back to reference H X, Y H, K L. Variable selection based on regularized regression of omics data[J]. Chinese Journal of Health Statistics, 2016, 33(04): 733–736. doi. H X, Y H, K L. Variable selection based on regularized regression of omics data[J]. Chinese Journal of Health Statistics, 2016, 33(04): 733–736. doi.
11.
go back to reference Chee C, Lombardo P, Schneider M, et al. Comparison of the Fat-to-Lesion Strain Ratio and the Gland-to-Lesion Strain Ratio With Controlled Precompression in Characterizing Indeterminate and Suspicious Breast Lesions on Ultrasound Imaging[J]. J Ultrasound Med. 2019;38(12):3257–66. doi:https://doi.org/10.1002/jum.15037.CrossRefPubMed Chee C, Lombardo P, Schneider M, et al. Comparison of the Fat-to-Lesion Strain Ratio and the Gland-to-Lesion Strain Ratio With Controlled Precompression in Characterizing Indeterminate and Suspicious Breast Lesions on Ultrasound Imaging[J]. J Ultrasound Med. 2019;38(12):3257–66. doi:https://​doi.​org/​10.​1002/​jum.​15037.CrossRefPubMed
14.
go back to reference T S, K M, AE B, et al. Tissue elasticity estimated by acoustic radiation force impulse quantification depends on the applied transducer force: an experimental study in kidney transplant patients[J]. European radiology, 2012, 22(10): 2130–2137. doi: https://doi.org/10.1007/s00330-012-2476-4. T S, K M, AE B, et al. Tissue elasticity estimated by acoustic radiation force impulse quantification depends on the applied transducer force: an experimental study in kidney transplant patients[J]. European radiology, 2012, 22(10): 2130–2137. doi: https://​doi.​org/​10.​1007/​s00330-012-2476-4.
Metadata
Title
The values of elastic quantitative and semi-quantitative indexes measured from different frequencies in the establishment of prediction models for breast tumor diagnosis
Authors
Xiao Xie
Yibo Ma
Xiaoxiao Xing
Haixia Zhou
Shuiqing Liu
Yanyan Zhang
Min Xu
Publication date
01-12-2022
Publisher
BioMed Central
Published in
BMC Medical Imaging / Issue 1/2022
Electronic ISSN: 1471-2342
DOI
https://doi.org/10.1186/s12880-022-00915-1

Other articles of this Issue 1/2022

BMC Medical Imaging 1/2022 Go to the issue