Rapid prototyping compliant arterial phantoms for in-vitro studies and device testing

Authors: Giovanni Biglino, Peter Verschueren, Raf Zegels, Andrew M Taylor, Silvia Schievano

Published in: Journal of Cardiovascular Magnetic Resonance | Issue 1/2013

Login to get access

Abstract

Background

Compliant vascular phantoms are desirable for in-vitro patient-specific experiments and device testing. TangoPlus FullCure 930® is a commercially available rubber-like material that can be used for PolyJet rapid prototyping. This work aims to gather preliminary data on the distensibility of this material, in order to assess the feasibility of its use in the context of experimental cardiovascular modelling.

Methods

The descending aorta anatomy of a volunteer was modelled in 3D from cardiovascular magnetic resonance (CMR) images and rapid prototyped using TangoPlus. The model was printed with a range of increasing wall thicknesses (0.6, 0.7, 0.8, 1.0 and 1.5 mm), keeping the lumen of the vessel constant. Models were also printed in both vertical and horizontal orientations, thus resulting in a total of ten specimens. Compliance tests were performed by monitoring pressure variations while gradually increasing and decreasing internal volume. Knowledge of distensibility was thus derived and then implemented with CMR data to test two applications. Firstly, a patient-specific compliant model of hypoplastic aorta suitable for connection in a mock circulatory loop for in-vitro tests was manufactured. Secondly, the right ventricular outflow tract (RVOT) of a patient necessitating pulmonary valve replacement was printed in order to physically test device insertion and assess patient’s suitability for percutaneous pulmonary valve intervention.

Results

The distensibility of the material was identified in a range from 6.5 × 10-3 mmHg-1 for the 0.6 mm case, to 3.0 × 10-3 mmHg-1 for the 1.5 mm case. The models printed in the vertical orientation were always more compliant than their horizontal counterpart. Rapid prototyping of a compliant hypoplastic aorta and of a RVOT anatomical model were both feasible. Device insertion in the RVOT model was successful.

Conclusion

Values of distensibility, compared with literature data, show that TangoPlus is suitable for manufacturing arterial phantoms, with the added benefit of being compatible with PolyJet printing, thus guaranteeing representative anatomical finishing, and quick and inexpensive fabrication. The appealing possibility of printing models of non-uniform wall thickness, resembling more closely certain physiological scenarios, can also be explored. However, this material appears to be too stiff for modelling the more compliant systemic venous system.
Appendix
Available only for authorised users
Literature
1.
go back to reference Skalak R: Synthesis of a complete circulation. Cardiovascular fluid dynamics, volume 2. Edited by: Bergel DH. 1972, London: Academic press, 341-376. Skalak R: Synthesis of a complete circulation. Cardiovascular fluid dynamics, volume 2. Edited by: Bergel DH. 1972, London: Academic press, 341-376.
2.
go back to reference Matthys KS, Alastruey J, Peiró J, Khir AW, Segers P, Verdonck PR, Parker KH, Sherwin SJ: Pulse wave propagation in a model human arterial network: assessment of 1-D numerical simulations against in vitro measurements. J Biomech. 2007, 40 (15): 3476-86. 10.1016/j.jbiomech.2007.05.027.CrossRefPubMed Matthys KS, Alastruey J, Peiró J, Khir AW, Segers P, Verdonck PR, Parker KH, Sherwin SJ: Pulse wave propagation in a model human arterial network: assessment of 1-D numerical simulations against in vitro measurements. J Biomech. 2007, 40 (15): 3476-86. 10.1016/j.jbiomech.2007.05.027.CrossRefPubMed
3.
go back to reference Vismara R, Laganà K, Migliavacca F, Schievano S, Coats L, Taylor AM, Bonhoeffer P: Experimental setup to evaluate the performance of percutaneous pulmonary valved stent in different outflow tract morphologies. Artif Organs. 2009, 33 (1): 46-53. 10.1111/j.1525-1594.2008.00673.x.CrossRefPubMed Vismara R, Laganà K, Migliavacca F, Schievano S, Coats L, Taylor AM, Bonhoeffer P: Experimental setup to evaluate the performance of percutaneous pulmonary valved stent in different outflow tract morphologies. Artif Organs. 2009, 33 (1): 46-53. 10.1111/j.1525-1594.2008.00673.x.CrossRefPubMed
4.
go back to reference Poepping TL, Nikolov HN, Thorne ML, Holdsworth DW: A thin-walled carotid vessel phantom for Doppler ultrasound flow studies. Ultrasound Med Biol. 2004, 30 (8): 1067-78. 10.1016/j.ultrasmedbio.2004.06.003.CrossRefPubMed Poepping TL, Nikolov HN, Thorne ML, Holdsworth DW: A thin-walled carotid vessel phantom for Doppler ultrasound flow studies. Ultrasound Med Biol. 2004, 30 (8): 1067-78. 10.1016/j.ultrasmedbio.2004.06.003.CrossRefPubMed
5.
go back to reference Kitajima HD, Sundareswaran KS, Teisseyre TZ, Astary GW, Parks WJ, Skrinjar O, Oshinski JN, Yoganathan AP: Comparison of particle image velocimetry and phase contrast MRI in a patient-specific extracardiac total cavopulmonary connection. J Biomech Eng. 2008, 130 (4): 041004-10.1115/1.2900725.CrossRefPubMed Kitajima HD, Sundareswaran KS, Teisseyre TZ, Astary GW, Parks WJ, Skrinjar O, Oshinski JN, Yoganathan AP: Comparison of particle image velocimetry and phase contrast MRI in a patient-specific extracardiac total cavopulmonary connection. J Biomech Eng. 2008, 130 (4): 041004-10.1115/1.2900725.CrossRefPubMed
6.
go back to reference Figliola RS, Giardini A, Conover T, Camp TA, Biglino G, Chiulli J, Hsia T-Y: In vitro simulation and validation of the circulation with congenital heart defects. Prog Pediatr Cardiol. 2010, 30 (1–2): 71-80.PubMedCentralCrossRefPubMed Figliola RS, Giardini A, Conover T, Camp TA, Biglino G, Chiulli J, Hsia T-Y: In vitro simulation and validation of the circulation with congenital heart defects. Prog Pediatr Cardiol. 2010, 30 (1–2): 71-80.PubMedCentralCrossRefPubMed
7.
go back to reference Biglino G, Giardini A, Baker C, Figliola RS, Hsia T-Y, Taylor AM, Schievano S: In-vitro study of the Norwood palliation: a patient-specific mock circulatory system. ASAIO J. 2012, 58 (1): 25-31. 10.1097/MAT.0b013e3182396847.CrossRefPubMed Biglino G, Giardini A, Baker C, Figliola RS, Hsia T-Y, Taylor AM, Schievano S: In-vitro study of the Norwood palliation: a patient-specific mock circulatory system. ASAIO J. 2012, 58 (1): 25-31. 10.1097/MAT.0b013e3182396847.CrossRefPubMed
8.
go back to reference de Zélicourt DA, Pekkan K, Wills L, Kanter K, Forbess J, Sharma S, Fogel M, Yoganathan AP: In vitro flow analysis of a patient-specific intraatrial total cavopulmonary connection. Ann Thorac Surg. 2005, 79 (6): 2094-102. 10.1016/j.athoracsur.2004.12.052.CrossRefPubMed de Zélicourt DA, Pekkan K, Wills L, Kanter K, Forbess J, Sharma S, Fogel M, Yoganathan AP: In vitro flow analysis of a patient-specific intraatrial total cavopulmonary connection. Ann Thorac Surg. 2005, 79 (6): 2094-102. 10.1016/j.athoracsur.2004.12.052.CrossRefPubMed
9.
go back to reference Ensley AE, Lynch P, Chatzimavroudis GP, Lucas C, Sharma S, Yoganathan AP: Toward designing the optimal cavopulmonary connection: an in vitro study. Ann Thorac Surg. 1999, 68 (4): 1384-90. 10.1016/S0003-4975(99)00560-3.CrossRefPubMed Ensley AE, Lynch P, Chatzimavroudis GP, Lucas C, Sharma S, Yoganathan AP: Toward designing the optimal cavopulmonary connection: an in vitro study. Ann Thorac Surg. 1999, 68 (4): 1384-90. 10.1016/S0003-4975(99)00560-3.CrossRefPubMed
10.
go back to reference Knierbein B, Rosarius N, Unger A, Reul H, Rau G: CAD-design, stress analysis and in vitro evaluation of three leaflet blood-pump valves. J Biomed Eng. 1992, 14 (4): 275-86. 10.1016/0141-5425(92)90001-2.CrossRefPubMed Knierbein B, Rosarius N, Unger A, Reul H, Rau G: CAD-design, stress analysis and in vitro evaluation of three leaflet blood-pump valves. J Biomed Eng. 1992, 14 (4): 275-86. 10.1016/0141-5425(92)90001-2.CrossRefPubMed
11.
go back to reference Tanné D, Bertrand E, Kadem L, Pibarot P, Rieu R: Assessment of left heart and pulmonary circulation flow dynamics by a new pulsed mock circulatory system. Exp Fluids. 2010, 48 (5): 837-850. 10.1007/s00348-009-0771-x.CrossRef Tanné D, Bertrand E, Kadem L, Pibarot P, Rieu R: Assessment of left heart and pulmonary circulation flow dynamics by a new pulsed mock circulatory system. Exp Fluids. 2010, 48 (5): 837-850. 10.1007/s00348-009-0771-x.CrossRef
12.
go back to reference Gregory S, Timms D, Pearcy MJ, Tansley G: A naturally shaped silicone ventricle evaluated in a mock circulation loop: a preliminary study. J Med Eng Technol. 2009, 33 (3): 185-91. 10.1080/03091900802184072.CrossRefPubMed Gregory S, Timms D, Pearcy MJ, Tansley G: A naturally shaped silicone ventricle evaluated in a mock circulation loop: a preliminary study. J Med Eng Technol. 2009, 33 (3): 185-91. 10.1080/03091900802184072.CrossRefPubMed
13.
go back to reference Kolyva C, Biglino G, Pepper JR, Khir AW: A mock circulatory system with physiological distribution of terminal resistance and compliance: application for testing the intra-aortic balloon pump. Artif Organs. 2012, 36 (3): E62-70. 10.1111/j.1525-1594.2010.01071.x.CrossRefPubMed Kolyva C, Biglino G, Pepper JR, Khir AW: A mock circulatory system with physiological distribution of terminal resistance and compliance: application for testing the intra-aortic balloon pump. Artif Organs. 2012, 36 (3): E62-70. 10.1111/j.1525-1594.2010.01071.x.CrossRefPubMed
14.
go back to reference Segers P, Dubois F, De Wachter D, Verdonck P: Role and relevancy of a cardiovascular simulator. Cardiovasc Eng. 1998, 3 (1): 48-56. Segers P, Dubois F, De Wachter D, Verdonck P: Role and relevancy of a cardiovascular simulator. Cardiovasc Eng. 1998, 3 (1): 48-56.
15.
go back to reference Arcaute K, Wicker RB: Patient-specific compliant vessel manufacturing using dip-spin coating of rapid prototyped molds. J Manuf Sci Eng. 2008, 130: 051008-10.1115/1.2898839.CrossRef Arcaute K, Wicker RB: Patient-specific compliant vessel manufacturing using dip-spin coating of rapid prototyped molds. J Manuf Sci Eng. 2008, 130: 051008-10.1115/1.2898839.CrossRef
16.
go back to reference Sulaiman A, Boussel L, Taconnet F, Serfaty JM, Alsaid H, Attia C, Huet L, Douek P: In vitro non-rigid life-size model of aortic arch aneurysm for endovascular prosthesis assessment. Eur J Cardiothorac Surg. 2008, 33 (1): 53-7. 10.1016/j.ejcts.2007.10.016.CrossRefPubMed Sulaiman A, Boussel L, Taconnet F, Serfaty JM, Alsaid H, Attia C, Huet L, Douek P: In vitro non-rigid life-size model of aortic arch aneurysm for endovascular prosthesis assessment. Eur J Cardiothorac Surg. 2008, 33 (1): 53-7. 10.1016/j.ejcts.2007.10.016.CrossRefPubMed
17.
go back to reference Ibrahim D, Broilo TL, Heitz C, de Oliveira MG, de Oliveira HW, Nobre SM, Dos Santos Filho JH, Silva DN: Dimensional error of selective laser sintering, three-dimensional printing and PolyJet models in the reproduction of mandibular anatomy. J Craniomaxillofac Surg. 2009, 37 (3): 167-73. 10.1016/j.jcms.2008.10.008.CrossRefPubMed Ibrahim D, Broilo TL, Heitz C, de Oliveira MG, de Oliveira HW, Nobre SM, Dos Santos Filho JH, Silva DN: Dimensional error of selective laser sintering, three-dimensional printing and PolyJet models in the reproduction of mandibular anatomy. J Craniomaxillofac Surg. 2009, 37 (3): 167-73. 10.1016/j.jcms.2008.10.008.CrossRefPubMed
18.
go back to reference Rajagopal V, Janardan R, Miyaoka R, Monga M, Sweet RM: Modeling and simulation for flexible ureteroscopy. In Proceedings of the 24th Engineering and urology society annual meeting: 25 April 2009, Chicago, J Endourol. 2009, 23 (6): 1035-1036. Rajagopal V, Janardan R, Miyaoka R, Monga M, Sweet RM: Modeling and simulation for flexible ureteroscopy. In Proceedings of the 24th Engineering and urology society annual meeting: 25 April 2009, Chicago, J Endourol. 2009, 23 (6): 1035-1036.
19.
go back to reference Kimura T, Morita A, Nishimura K, Aiyama H, Itoh H, Fukaya S, Sora S, Ochiai C: Simulation of and training for cerebral aneurysm clipping with 3-dimensional models. Neurosurgery. 2009, 65 (4): 719-25. 10.1227/01.NEU.0000354350.88899.07.CrossRefPubMed Kimura T, Morita A, Nishimura K, Aiyama H, Itoh H, Fukaya S, Sora S, Ochiai C: Simulation of and training for cerebral aneurysm clipping with 3-dimensional models. Neurosurgery. 2009, 65 (4): 719-25. 10.1227/01.NEU.0000354350.88899.07.CrossRefPubMed
20.
go back to reference Schievano S, Migliavacca F, Coats L, Khambadkone S, Carminati M, Wilson N, Deanfield JE, Bonhoeffer P, Taylor AM: Percutaneous pulmonary valve implantation based on rapid prototyping of right ventricular outflow tract and pulmonary trunk from MR data. Radiology. 2007, 242 (2): 490-7. 10.1148/radiol.2422051994.CrossRefPubMed Schievano S, Migliavacca F, Coats L, Khambadkone S, Carminati M, Wilson N, Deanfield JE, Bonhoeffer P, Taylor AM: Percutaneous pulmonary valve implantation based on rapid prototyping of right ventricular outflow tract and pulmonary trunk from MR data. Radiology. 2007, 242 (2): 490-7. 10.1148/radiol.2422051994.CrossRefPubMed
21.
go back to reference Figliola RS, Beasley DE: Theory and design for mechanical measurements. 2006, Hoboken: Wiley, 4 Figliola RS, Beasley DE: Theory and design for mechanical measurements. 2006, Hoboken: Wiley, 4
22.
go back to reference Biglino G, Capelli C, Binazzi A, Reggiani R, Cosentino D, Migliavacca F, Bonhoeffer P, Taylor AM, Schievano S: Virtual and real bench testing of a new percutaneous valve device: a case study. EuroIntervention. 2012, 8 (1): 120-8. 10.4244/EIJV8I1A19.CrossRefPubMed Biglino G, Capelli C, Binazzi A, Reggiani R, Cosentino D, Migliavacca F, Bonhoeffer P, Taylor AM, Schievano S: Virtual and real bench testing of a new percutaneous valve device: a case study. EuroIntervention. 2012, 8 (1): 120-8. 10.4244/EIJV8I1A19.CrossRefPubMed
23.
go back to reference Capelli C, Biglino G, Petrini L, Migliavacca F, Cosentino D, Bonhoeffer P, Taylor AM, Schievano S: Finite elements strategies to satisfy clinical and engineering requirements in field of percutaneous valves. Ann Biomed Eng. 2012, 10.1007/s10439-012-0617-1. epub ahead of print Capelli C, Biglino G, Petrini L, Migliavacca F, Cosentino D, Bonhoeffer P, Taylor AM, Schievano S: Finite elements strategies to satisfy clinical and engineering requirements in field of percutaneous valves. Ann Biomed Eng. 2012, 10.1007/s10439-012-0617-1. epub ahead of print
24.
go back to reference Markl M, Schumacher R, Kuffer J, Bley TA, Hennig J: Rapid vessel prototyping: vascular modeling using 3t magnetic resonance angiography and rapid prototyping technology. MAGMA. 2005, 18 (6): 288-92. 10.1007/s10334-005-0019-6.CrossRefPubMed Markl M, Schumacher R, Kuffer J, Bley TA, Hennig J: Rapid vessel prototyping: vascular modeling using 3t magnetic resonance angiography and rapid prototyping technology. MAGMA. 2005, 18 (6): 288-92. 10.1007/s10334-005-0019-6.CrossRefPubMed
25.
go back to reference Canstein C, Cachot P, Faust A, Stalder AF, Bock J, Frydrychowicz A, Kuffer J, Hennig J, Markl M: 3D MR flow analysis in realistic rapid-prototyping model systems of the thoracic aorta: comparison with in vivo data and computational fluid dynamics in identical vessel geometries. Magn Reson Med. 2008, 59 (3): 535-46. 10.1002/mrm.21331.CrossRefPubMed Canstein C, Cachot P, Faust A, Stalder AF, Bock J, Frydrychowicz A, Kuffer J, Hennig J, Markl M: 3D MR flow analysis in realistic rapid-prototyping model systems of the thoracic aorta: comparison with in vivo data and computational fluid dynamics in identical vessel geometries. Magn Reson Med. 2008, 59 (3): 535-46. 10.1002/mrm.21331.CrossRefPubMed
26.
go back to reference Rose JL, Lalande A, Bouchot O, Walker PM, Ugolini P, Revol-Muller C, Cartier R, Brunotte F, Bourennane el-B: Influence of age and sex on aortic distensibility assessed by MRI in healthy subjects. Magn Reson Imaging. 2010, 28 (2): 255-263. 10.1016/j.mri.2009.07.001.CrossRefPubMed Rose JL, Lalande A, Bouchot O, Walker PM, Ugolini P, Revol-Muller C, Cartier R, Brunotte F, Bourennane el-B: Influence of age and sex on aortic distensibility assessed by MRI in healthy subjects. Magn Reson Imaging. 2010, 28 (2): 255-263. 10.1016/j.mri.2009.07.001.CrossRefPubMed
27.
go back to reference Chue CD, Edwards NC, Ferro CJ, Jones CM, Townend JN, Steeds RP: Insights into differential regional aortic distensibility in early chronic kidney disease: a cardiac magnetic resonance study. Eur Heart J. 2010, 31: 143-10.1093/eurheartj/ehp390.CrossRef Chue CD, Edwards NC, Ferro CJ, Jones CM, Townend JN, Steeds RP: Insights into differential regional aortic distensibility in early chronic kidney disease: a cardiac magnetic resonance study. Eur Heart J. 2010, 31: 143-10.1093/eurheartj/ehp390.CrossRef
28.
go back to reference Gan CT, Lankhaar JW, Westerhof N, Marcus JT, Becker A, Twisk JW, Boonstra A, Postmus PE, Vonk-Noordegraaf A: Noninvasively assessed pulmonary stiffness predicts mortality in pulmonary arterial hypertension. Chest. 2007, 132 (6): 1906-12. 10.1378/chest.07-1246.CrossRefPubMed Gan CT, Lankhaar JW, Westerhof N, Marcus JT, Becker A, Twisk JW, Boonstra A, Postmus PE, Vonk-Noordegraaf A: Noninvasively assessed pulmonary stiffness predicts mortality in pulmonary arterial hypertension. Chest. 2007, 132 (6): 1906-12. 10.1378/chest.07-1246.CrossRefPubMed
29.
go back to reference Lénárd Z, Fülöp D, Visontai Z, Jokkel G, Reneman R, Kollai M: Static versus dynamic distensibility of the carotid artery in humans. J Vasc Res. 2000, 37 (2): 103-11. 10.1159/000025721.CrossRefPubMed Lénárd Z, Fülöp D, Visontai Z, Jokkel G, Reneman R, Kollai M: Static versus dynamic distensibility of the carotid artery in humans. J Vasc Res. 2000, 37 (2): 103-11. 10.1159/000025721.CrossRefPubMed
30.
go back to reference Guyton AC, Hall JE: Textbook of medical physiology. 2006, Philadelphia: Elsevier Saunders, 11 Guyton AC, Hall JE: Textbook of medical physiology. 2006, Philadelphia: Elsevier Saunders, 11
31.
go back to reference Diamantopoulos I, Lever MJ: Can an intermittent pneumatic compression system monitor venous filling in the leg?. J Med Eng Technol. 2008, 32 (3): 221-7. 10.1080/03091900701559424.CrossRefPubMed Diamantopoulos I, Lever MJ: Can an intermittent pneumatic compression system monitor venous filling in the leg?. J Med Eng Technol. 2008, 32 (3): 221-7. 10.1080/03091900701559424.CrossRefPubMed
32.
go back to reference Voges I, Jerosch-Herold M, Hedderich J, Westphal C, Hart C, Helle M, Scheewe J, Pardun E, Kramer HH, Rickers C: Maladaptive aortic properties in children after palliation of hypoplastic left heart syndrome assessed by cardiovascular magnetic resonance imaging. Circulation. 2010, 122 (11): 1068-76. 10.1161/CIRCULATIONAHA.109.889733.CrossRefPubMed Voges I, Jerosch-Herold M, Hedderich J, Westphal C, Hart C, Helle M, Scheewe J, Pardun E, Kramer HH, Rickers C: Maladaptive aortic properties in children after palliation of hypoplastic left heart syndrome assessed by cardiovascular magnetic resonance imaging. Circulation. 2010, 122 (11): 1068-76. 10.1161/CIRCULATIONAHA.109.889733.CrossRefPubMed
33.
go back to reference Schievano S, Taylor AM, Capelli C, Coats L, Walker F, Lurz P, Nordmeyer J, Wright S, Khambadkone S, Tsang V, Carminati M, Bonhoeffer P: First-in-man implantation of a novel percutaneous valve: a new approach to medical device development. EuroIntervention. 2010, 5 (6): 745-50. 10.4244/EIJV5I6A122.CrossRefPubMed Schievano S, Taylor AM, Capelli C, Coats L, Walker F, Lurz P, Nordmeyer J, Wright S, Khambadkone S, Tsang V, Carminati M, Bonhoeffer P: First-in-man implantation of a novel percutaneous valve: a new approach to medical device development. EuroIntervention. 2010, 5 (6): 745-50. 10.4244/EIJV5I6A122.CrossRefPubMed
Metadata
Title
Rapid prototyping compliant arterial phantoms for in-vitro studies and device testing
Authors
Giovanni Biglino
Peter Verschueren
Raf Zegels
Andrew M Taylor
Silvia Schievano
Publication date
01-12-2013
Publisher
BioMed Central
Published in
Journal of Cardiovascular Magnetic Resonance / Issue 1/2013
Electronic ISSN: 1532-429X
DOI
https://doi.org/10.1186/1532-429X-15-2

Other articles of this Issue 1/2013

Journal of Cardiovascular Magnetic Resonance 1/2013 Go to the issue