Preliminary experience with cardiovascular magnetic resonance in evaluation of fetal cardiovascular anomalies

Authors: Su-Zhen Dong, Ming Zhu, Fen Li

Published in: Journal of Cardiovascular Magnetic Resonance | Issue 1/2013

Login to get access

Abstract

Background

The cardiovascular system is the part of the fetal anatomy that most frequently suffers from congenital pathology. This study shows our preliminary experience with fetal cardiovascular magnetic resonance (CMR) to evaluate congenital cardiovascular abnormalities.

Methods

Between January 2006 and June 2011, Prenatal routine obstetric ultrasound (US), echocardiography and CMR data from 68 pregnant women carrying fetuses with congenital cardiovascular anomalies were compared with postnatal diagnoses (postnatal imagings, surgery and autopsy). All prenatal CMR was performed at 1.5 T. Imaging sequences included steady-state free-precession (SSFP) sequences, real-time SSFP and single-shot turbo spin echo (SSTSE) sequences. The images were analyzed with an anatomic segmental approach by two radiologists.

Results

Fetal CMR yielded the same diagnosis as postnatal findings in 79% (54/68) of patients. The diagnostic sensitivity of routine obstetric US for cardiac anomalies was 46% (31/68). The diagnostic sensitivity of fetal echocardiographic examination by a fetal cardiac specialist was 82% (56/68). In 2 (3%) of 68 cases, diagnoses with both echocardiography and CMR were incorrect when compared with postnatal diagnosis. In ten (15%) cases, diagnosis at echocardiography was incorrect and that at CMR was correct. In twelve (18%) cases, diagnosis at echocardiography was correct and that at CMR was incorrect. Ten cases missed or misdiagnosed by echocardiography but correctly diagnosed by fetal CMR included asplenia syndrome (n = 2), interrupted inferior vena cava of polysplenia syndrome (n = 1), tricuspid incompetence (n = 1), double outlet right ventricle (n = 2), double aortic arch (n = 1), right pulmonary artery hypoplasia (n = 1), right-sided aortic arch of tetralogy of Fallot (n = 1) and hypoplastic left heart syndrome of a twin fetus (n = 1).

Conclusion

Fetal CMR is a promising diagnostic tool for assessment of congenital cardiovascular abnormalities, especially in situations that limit echocardiography.
Appendix
Available only for authorised users
Literature
1.
go back to reference Chiappa E: The impact of prenatal diagnosis of congenital heart disease on pediatric cardiology and cardiac surgery. J Cardiovasc Med. 2007, 8: 12-16. 10.2459/01.JCM.0000247429.28957.80.CrossRef Chiappa E: The impact of prenatal diagnosis of congenital heart disease on pediatric cardiology and cardiac surgery. J Cardiovasc Med. 2007, 8: 12-16. 10.2459/01.JCM.0000247429.28957.80.CrossRef
2.
go back to reference Ruckhäberle E, Nekolla SG, Ganter C, Schneider KT, Peter A, Raidt A, Kovacs L, Brockmeier SJ, Schwaiger M, Oberhoffer R, Papadopulos NA: In vivo intrauterine sound pressure and temperature measurements during magnetic resonance imaging (1.5 T) in pregnant ewes. Fetal Diagn. Ther. 2008, 24: 203-210. 10.1159/000151339.CrossRefPubMed Ruckhäberle E, Nekolla SG, Ganter C, Schneider KT, Peter A, Raidt A, Kovacs L, Brockmeier SJ, Schwaiger M, Oberhoffer R, Papadopulos NA: In vivo intrauterine sound pressure and temperature measurements during magnetic resonance imaging (1.5 T) in pregnant ewes. Fetal Diagn. Ther. 2008, 24: 203-210. 10.1159/000151339.CrossRefPubMed
3.
go back to reference Gorincour G, Bourlière-Najean B, Bonello B, Fraisse A, Philip N, Potier A, Kreitmann B, Petit P: Feasibility of fetal cardiac magnetic resonance imaging: preliminary experience. Ultrasound Obstet. Gynecol. 2007, 29: 105-108. 10.1002/uog.3924.CrossRefPubMed Gorincour G, Bourlière-Najean B, Bonello B, Fraisse A, Philip N, Potier A, Kreitmann B, Petit P: Feasibility of fetal cardiac magnetic resonance imaging: preliminary experience. Ultrasound Obstet. Gynecol. 2007, 29: 105-108. 10.1002/uog.3924.CrossRefPubMed
4.
go back to reference Saleem SN: Feasibility of MRI of the fetal heart with balanced steady-state free precession sequence along fetal body and cardiac planes. AJR. 2008, 191: 1208-1215. 10.2214/AJR.07.3839.CrossRefPubMed Saleem SN: Feasibility of MRI of the fetal heart with balanced steady-state free precession sequence along fetal body and cardiac planes. AJR. 2008, 191: 1208-1215. 10.2214/AJR.07.3839.CrossRefPubMed
5.
go back to reference Manganaro L, Savelli S, Di Maurizio M, Perrone A, Tesei J, Francioso A, Angeletti M, Coratella F, Irimia D, Fierro F, Ventriglia F, Ballesio L: Potential role of fetal cardiac evaluation with magnetic resonance imaging: preliminary experience. Prenat. Diagn. 2008, 28: 148-156. 10.1002/pd.1945.CrossRefPubMed Manganaro L, Savelli S, Di Maurizio M, Perrone A, Tesei J, Francioso A, Angeletti M, Coratella F, Irimia D, Fierro F, Ventriglia F, Ballesio L: Potential role of fetal cardiac evaluation with magnetic resonance imaging: preliminary experience. Prenat. Diagn. 2008, 28: 148-156. 10.1002/pd.1945.CrossRefPubMed
6.
go back to reference Muhler MR, Rake A, Schwabe M, Chaoui R, Heling KS, Planke C, Lembcke A, Fischer T, Kivelitz D: Truncus arteriosus communis inamidtrimester fetus: comparison of prenatal ultrasound and MRI with postmortem MRI and autopsy. Eur. Radiol. 2004, 14: 2120-2124. 10.1007/s00330-004-2419-9.CrossRefPubMed Muhler MR, Rake A, Schwabe M, Chaoui R, Heling KS, Planke C, Lembcke A, Fischer T, Kivelitz D: Truncus arteriosus communis inamidtrimester fetus: comparison of prenatal ultrasound and MRI with postmortem MRI and autopsy. Eur. Radiol. 2004, 14: 2120-2124. 10.1007/s00330-004-2419-9.CrossRefPubMed
7.
go back to reference McMahon CJ, Taylor MD, Cassady CI, Olutoye OO, Bezold LI: Diagnosis of pentalogyof cantrell in the fetus using magnetic resonance imaging and ultrasound. Pediatr. Cardiol. 2007, 28: 172-175. 10.1007/s00246-006-0032-1.CrossRefPubMed McMahon CJ, Taylor MD, Cassady CI, Olutoye OO, Bezold LI: Diagnosis of pentalogyof cantrell in the fetus using magnetic resonance imaging and ultrasound. Pediatr. Cardiol. 2007, 28: 172-175. 10.1007/s00246-006-0032-1.CrossRefPubMed
8.
go back to reference Fogel MA, Wilson RD, Flake A, Johnson M, Cohen D, McNeal G, Tian ZY, Rychik J: Preliminary investigations into a new method of functional assessment of the fetal heart usinga novel application of “real- time” cardiac magnetic resonance imaging. Fetal Diagn. Ther. 2005, 20: 475-480. 10.1159/000086837.CrossRefPubMed Fogel MA, Wilson RD, Flake A, Johnson M, Cohen D, McNeal G, Tian ZY, Rychik J: Preliminary investigations into a new method of functional assessment of the fetal heart usinga novel application of “real- time” cardiac magnetic resonance imaging. Fetal Diagn. Ther. 2005, 20: 475-480. 10.1159/000086837.CrossRefPubMed
9.
go back to reference Manganaro L, Savelli S, Di Maurizio M, Perrone A, Francioso A, La Barbera L, Totaro P, Fierro F, Tomei A, Coratella F, Giancotti A, Ballesio L, Ventriglia F: Assessment of congenital heart disease (CHD): is there a role for fetal magnetic resonance imaging (MRI)?. Eur. J. Radiol. 2009, 72: 172-180. 10.1016/j.ejrad.2008.06.016.CrossRefPubMed Manganaro L, Savelli S, Di Maurizio M, Perrone A, Francioso A, La Barbera L, Totaro P, Fierro F, Tomei A, Coratella F, Giancotti A, Ballesio L, Ventriglia F: Assessment of congenital heart disease (CHD): is there a role for fetal magnetic resonance imaging (MRI)?. Eur. J. Radiol. 2009, 72: 172-180. 10.1016/j.ejrad.2008.06.016.CrossRefPubMed
10.
go back to reference Osada H, Kaku K, Masuda K, Iitsuka Y, Seki K, Sekiya S: Quantitative and qualitative evaluations of fetal lung with MR imaging. Radiology. 2004, 231: 887-892. 10.1148/radiol.2313021689.CrossRefPubMed Osada H, Kaku K, Masuda K, Iitsuka Y, Seki K, Sekiya S: Quantitative and qualitative evaluations of fetal lung with MR imaging. Radiology. 2004, 231: 887-892. 10.1148/radiol.2313021689.CrossRefPubMed
11.
go back to reference Carvalho JS, Ho SY, Shinebourne EA: Sequential segmental analysis in complex fetal cardiac abnormalities: a logical approach to diagnosis. Ultrasound Obstet. Gynecol. 2005, 26: 105-111. 10.1002/uog.1970.CrossRefPubMed Carvalho JS, Ho SY, Shinebourne EA: Sequential segmental analysis in complex fetal cardiac abnormalities: a logical approach to diagnosis. Ultrasound Obstet. Gynecol. 2005, 26: 105-111. 10.1002/uog.1970.CrossRefPubMed
12.
go back to reference Pugash D, Brugger PC, Bettelheim D, Prayer D: Prenatal ultrasound and fetal MRI: the comparative value of each modality in prenatal diagnosis. Eur. J. Radiol. 2008, 68: 214-226. 10.1016/j.ejrad.2008.06.031.CrossRefPubMed Pugash D, Brugger PC, Bettelheim D, Prayer D: Prenatal ultrasound and fetal MRI: the comparative value of each modality in prenatal diagnosis. Eur. J. Radiol. 2008, 68: 214-226. 10.1016/j.ejrad.2008.06.031.CrossRefPubMed
13.
go back to reference Bulas D: Fetal magnetic resonance imaging as a complement to fetal ultrasonography. Ultrasound Q. 2007, 23: 3-22. 10.1097/01.ruq.0000263841.69689.a7.CrossRefPubMed Bulas D: Fetal magnetic resonance imaging as a complement to fetal ultrasonography. Ultrasound Q. 2007, 23: 3-22. 10.1097/01.ruq.0000263841.69689.a7.CrossRefPubMed
14.
go back to reference Santos XM, Papanna R, Johnson A, Cass DL, Olutoye OO, Moise KJ, Belleza-Bascon B, Cassady CI: The use of combined ultrasound and magnetic resonance imaging in the detection of fetal anomalies. Prenat. Diagn. 2010, 30: 402-407.PubMed Santos XM, Papanna R, Johnson A, Cass DL, Olutoye OO, Moise KJ, Belleza-Bascon B, Cassady CI: The use of combined ultrasound and magnetic resonance imaging in the detection of fetal anomalies. Prenat. Diagn. 2010, 30: 402-407.PubMed
15.
go back to reference Habas PA, Kim K, Rousseau F, Glenn OA, Barkovich AJ, Studholme C: Atlas-based segmentation of developing tissues in the human brain with quantitative validation in young fetuses. Hum. Brain Mapp. 2010, 31: 1348-1358. 10.1002/hbm.20935.PubMedCentralCrossRefPubMed Habas PA, Kim K, Rousseau F, Glenn OA, Barkovich AJ, Studholme C: Atlas-based segmentation of developing tissues in the human brain with quantitative validation in young fetuses. Hum. Brain Mapp. 2010, 31: 1348-1358. 10.1002/hbm.20935.PubMedCentralCrossRefPubMed
16.
go back to reference De Wilde JP, Rivers AW, Price DL: A review of the current use of magnetic resonance imaging in pregnancyand safety implications for the fetus. Prog BiophysMol Biol. 2005, 87: 335-353. 10.1016/j.pbiomolbio.2004.08.010.CrossRef De Wilde JP, Rivers AW, Price DL: A review of the current use of magnetic resonance imaging in pregnancyand safety implications for the fetus. Prog BiophysMol Biol. 2005, 87: 335-353. 10.1016/j.pbiomolbio.2004.08.010.CrossRef
17.
go back to reference Levine D, Zuo C, Faro CB, Chen Q: Potential heating effect in the gravid uterus during MR HASTE imaging. J. Magn. Reson. Imaging. 2001, 13: 856-861. 10.1002/jmri.1122.CrossRefPubMed Levine D, Zuo C, Faro CB, Chen Q: Potential heating effect in the gravid uterus during MR HASTE imaging. J. Magn. Reson. Imaging. 2001, 13: 856-861. 10.1002/jmri.1122.CrossRefPubMed
18.
go back to reference Loomba RS, Chandrasekar S, Shah PH, Sanan P: The developing role of fetal magnetic resonance imaging in the diagnosis of congenital cardiac anomalies: A systematic review. Ann Pediatr Cardiol. 2011, 4: 172-176. 10.4103/0974-2069.84665.PubMedCentralCrossRefPubMed Loomba RS, Chandrasekar S, Shah PH, Sanan P: The developing role of fetal magnetic resonance imaging in the diagnosis of congenital cardiac anomalies: A systematic review. Ann Pediatr Cardiol. 2011, 4: 172-176. 10.4103/0974-2069.84665.PubMedCentralCrossRefPubMed
19.
go back to reference Sklansky M: Advances in fetal cardiac imaging. Pediatr. Cardiol. 2004, 25: 307-321.PubMed Sklansky M: Advances in fetal cardiac imaging. Pediatr. Cardiol. 2004, 25: 307-321.PubMed
20.
go back to reference Kasprian G, Balassy C, Brugger PC, Prayer D: MRI of normal and pathological fetal lung development. Eur. J. Radiol. 2006, 57: 261-270. 10.1016/j.ejrad.2005.11.031.CrossRefPubMed Kasprian G, Balassy C, Brugger PC, Prayer D: MRI of normal and pathological fetal lung development. Eur. J. Radiol. 2006, 57: 261-270. 10.1016/j.ejrad.2005.11.031.CrossRefPubMed
21.
go back to reference Hörmann M, Brugger PC, Balassy C, Witzani L, Prayer D: Fetal MRI of the urinary system. Eur. J. Radiol. 2006, 57: 303-311. 10.1016/j.ejrad.2005.11.028.CrossRefPubMed Hörmann M, Brugger PC, Balassy C, Witzani L, Prayer D: Fetal MRI of the urinary system. Eur. J. Radiol. 2006, 57: 303-311. 10.1016/j.ejrad.2005.11.028.CrossRefPubMed
22.
go back to reference Yamamura J, Schnackenburg B, Kooijmann H, Frisch M, Hecher K, Adam G, Wedegärtner U: High resolution MR imaging of the fetal heart with cardiac triggering: a feasibility study in the sheep fetus. Eur. Radiol. 2009, 19: 2383-2390. 10.1007/s00330-009-1420-8.CrossRefPubMed Yamamura J, Schnackenburg B, Kooijmann H, Frisch M, Hecher K, Adam G, Wedegärtner U: High resolution MR imaging of the fetal heart with cardiac triggering: a feasibility study in the sheep fetus. Eur. Radiol. 2009, 19: 2383-2390. 10.1007/s00330-009-1420-8.CrossRefPubMed
23.
go back to reference Yamamura J, Frisch M, Ecker H, Graessner J, Hecher K, Adam G, Wedegärtner U: Self-gating MR imaging of the fetal heart: comparison with real cardiac triggering. Eur. Radiol. 2011, 21: 142-149. 10.1007/s00330-010-1911-7.CrossRefPubMed Yamamura J, Frisch M, Ecker H, Graessner J, Hecher K, Adam G, Wedegärtner U: Self-gating MR imaging of the fetal heart: comparison with real cardiac triggering. Eur. Radiol. 2011, 21: 142-149. 10.1007/s00330-010-1911-7.CrossRefPubMed
24.
go back to reference Jansz MS, Seed M, van Amerom JF, Wong D, Grosse-Wortmann L, Yoo SJ, Macgowan CK: Metric optimized gating for fetal cardiac MRI. Magn. Reson. Med. 2010, 64: 1304-1314. 10.1002/mrm.22542.CrossRefPubMed Jansz MS, Seed M, van Amerom JF, Wong D, Grosse-Wortmann L, Yoo SJ, Macgowan CK: Metric optimized gating for fetal cardiac MRI. Magn. Reson. Med. 2010, 64: 1304-1314. 10.1002/mrm.22542.CrossRefPubMed
Metadata
Title
Preliminary experience with cardiovascular magnetic resonance in evaluation of fetal cardiovascular anomalies
Authors
Su-Zhen Dong
Ming Zhu
Fen Li
Publication date
01-12-2013
Publisher
BioMed Central
Published in
Journal of Cardiovascular Magnetic Resonance / Issue 1/2013
Electronic ISSN: 1532-429X
DOI
https://doi.org/10.1186/1532-429X-15-40

Other articles of this Issue 1/2013

Journal of Cardiovascular Magnetic Resonance 1/2013 Go to the issue