Skip to main content
Top
Published in: Ophthalmology and Therapy 1/2019

Open Access 01-03-2019 | Cataract | Original Research

Preclinical Safety Evaluation of Ophthalmic Viscosurgical Devices in Rabbits and a Novel Mini-Pig Model

Authors: Ronika S. Leang, Lisa J. Kloft, Brad Gray, Arlene E. Gwon, Ling C. Huang

Published in: Ophthalmology and Therapy | Issue 1/2019

Login to get access

Abstract

Introduction

To establish the preclinical safety and equivalency of ophthalmic viscosurgical devices (OVDs) comprised of bacterially sourced sodium hyaluronate (HA) to animal sourced HA using pyrogenicity and aqueous exchange models in rabbits and a novel mini-pig model to evaluate corneal endothelial cell protection in vivo.

Methods

HEALON OVD and HEALON5 OVD containing animal-derived HA and HEALON PRO OVD and HEALON5 PRO OVD containing bacterial-derived HA were used. Two rabbit aqueous exchange studies were conducted where aqueous humor was exchanged with OVDs in six animals each to observe potential ocular inflammation, intraocular pressure (IOP) response, corneal health and pachymetry until 7 days post procedure, as well as overall assessment of the OVDs. Endothelial cell protection was evaluated in a Yucatan mini-pig cataract surgery model where HEALON PRO and HEALON5 PRO OVDs were compared to HEALON and HEALON5 OVDs, respectively. Following cataract surgery with use of OVDs in six animals per study, animals were evaluated for ocular and general health, IOP, corneal thickness, ocular inflammation, and endothelial cell protection on days 1, 3, 7 and 14 post-surgery.

Results

All rabbit studies demonstrated equivalence between bacterial-derived and animal-derived OVDs. Mild, post-surgical irritation, IOP increase, and corneal thickness measurements were not significantly different in HEALON PRO OVD and HEALON5 PRO OVD compared to HEALON and HEALON5 OVDs, respectively. The mini-pig model developed to investigate endothelial cell protection was successful in demonstrating equivalence between the OVDs studied. Changes in IOP mirrored actual surgical procedures, while corneal pachymetry and endothelial cell density remained constant for all OVDs used. Slit lamp observations showed expected inflammation following surgical procedures, likely due to challenges encountered during surgical procedures.

Conclusion

Rabbit pyrogenicity and aqueous exchanged paired with a novel simulated cataract surgery mini-pig model demonstrate equivalence of OVDs regardless of HA source. Albeit with challenges, the mini-pig model was shown to be a promising tool for endothelial cell evaluation during the development of new OVDs for ophthalmic use.

Funding

Johnson & Johnson Surgical Vision, Inc.
Literature
1.
go back to reference Arshinoff S. Ophthalmic viscosurgical devices. In: Kohnen T, Koch D, editors. Cataract and refractive surgery essentials in ophthalmology. Berlin, Heidelberg: Springer; 2005. p. 37–62.CrossRef Arshinoff S. Ophthalmic viscosurgical devices. In: Kohnen T, Koch D, editors. Cataract and refractive surgery essentials in ophthalmology. Berlin, Heidelberg: Springer; 2005. p. 37–62.CrossRef
2.
go back to reference Belda JI, Artola A, Garcia-Manzanares MD, et al. Hyaluronic acid combined with mannitol to improve protection against free-radical endothelial damage: experimental model. J Cataract Refract Surg. 2005;31(6):1213–8.CrossRef Belda JI, Artola A, Garcia-Manzanares MD, et al. Hyaluronic acid combined with mannitol to improve protection against free-radical endothelial damage: experimental model. J Cataract Refract Surg. 2005;31(6):1213–8.CrossRef
3.
go back to reference Buratto L, Giardini P, Bellucci R. Viscoelastic materials, chapter 2. In: Drummon A, Plummer L, editors. Viscoelastics in ophthalmic surgery. USA: SLACK Inc; 2000. pp. 5–9. Buratto L, Giardini P, Bellucci R. Viscoelastic materials, chapter 2. In: Drummon A, Plummer L, editors. Viscoelastics in ophthalmic surgery. USA: SLACK Inc; 2000. pp. 5–9.
4.
go back to reference Liesegang TJ. Viscoelastic substances in ophthalmology. Surv Ophthalmol. 1990;34(4):268–93.CrossRef Liesegang TJ. Viscoelastic substances in ophthalmology. Surv Ophthalmol. 1990;34(4):268–93.CrossRef
5.
go back to reference Schulze SD, Bertelmann T, Manojlovic I, Bodanowitz S, Irle S, Sekundo W. Changes in corneal endothelium cell characteristics after cataract surgery with and without use of viscoelastic substances during intraocular lens implantation. Clin Ophthalmol. 2015;9:2073–80.CrossRef Schulze SD, Bertelmann T, Manojlovic I, Bodanowitz S, Irle S, Sekundo W. Changes in corneal endothelium cell characteristics after cataract surgery with and without use of viscoelastic substances during intraocular lens implantation. Clin Ophthalmol. 2015;9:2073–80.CrossRef
6.
go back to reference Valdez-Garcia JE, Lozano-Ramirez JF, Zavala J. Adult white New Zealand rabbit as suitable model for corneal endothelial engineering. BMC Res Notes. 2015;8:28.CrossRef Valdez-Garcia JE, Lozano-Ramirez JF, Zavala J. Adult white New Zealand rabbit as suitable model for corneal endothelial engineering. BMC Res Notes. 2015;8:28.CrossRef
7.
go back to reference Higashide T, Sugiyama K. Use of viscoelastic substance in ophthalmic surgery—focus on sodium hyaluronate. Clin Ophthalmol. 2008;2(1):21–30.CrossRef Higashide T, Sugiyama K. Use of viscoelastic substance in ophthalmic surgery—focus on sodium hyaluronate. Clin Ophthalmol. 2008;2(1):21–30.CrossRef
8.
go back to reference Shimmura S, Tsubota K, Oguchi Y, Fukumura D, Suematsu M, Tsuchiya M. Oxiradical-dependent photoemission induced by a phacoemulsification probe. Invest Ophthalmol Vis Sci. 1992;33(10):2904–7.PubMed Shimmura S, Tsubota K, Oguchi Y, Fukumura D, Suematsu M, Tsuchiya M. Oxiradical-dependent photoemission induced by a phacoemulsification probe. Invest Ophthalmol Vis Sci. 1992;33(10):2904–7.PubMed
9.
go back to reference Suzuki H, Igarashi T, Shiwa T, Takahashi H. Efficacy of ophthalmic viscosurgical devices in preventing temperature rise at the corneal endothelium during phacoemulsification. Curr Eye Res. 2016;41(12):1548–52.CrossRef Suzuki H, Igarashi T, Shiwa T, Takahashi H. Efficacy of ophthalmic viscosurgical devices in preventing temperature rise at the corneal endothelium during phacoemulsification. Curr Eye Res. 2016;41(12):1548–52.CrossRef
10.
go back to reference Takahashi H. Free radical development in phacoemulsification cataract surgery. J Nippon Med Sch. 2005;72(1):4–12.CrossRef Takahashi H. Free radical development in phacoemulsification cataract surgery. J Nippon Med Sch. 2005;72(1):4–12.CrossRef
11.
go back to reference Arshinoff S, Carbonneau M. Using OVDs for glaucoma surgery. J Emmetropia. 2013;4:161–9. Arshinoff S, Carbonneau M. Using OVDs for glaucoma surgery. J Emmetropia. 2013;4:161–9.
12.
go back to reference Liu L, Liu Y, Li J, Du G, Chen J. Microbial production of hyaluronic acid: current state, challenges, and perspectives. Microb Cell Fact. 2011;10:99.CrossRef Liu L, Liu Y, Li J, Du G, Chen J. Microbial production of hyaluronic acid: current state, challenges, and perspectives. Microb Cell Fact. 2011;10:99.CrossRef
13.
go back to reference Abbott Medical Optics. Data on file—Evaluation of the rhelogical properties of Healon PRO OVD and Healon5 PRO OVD. 2016. Abbott Medical Optics. Data on file—Evaluation of the rhelogical properties of Healon PRO OVD and Healon5 PRO OVD. 2016.
14.
go back to reference Gwon A. The rabbit in cataract/IOL surgery. In: Tsonis P, editor. Animal models in eye research. Dayton, OH: Elsevier Ltd; 2008. p. 184–204.CrossRef Gwon A. The rabbit in cataract/IOL surgery. In: Tsonis P, editor. Animal models in eye research. Dayton, OH: Elsevier Ltd; 2008. p. 184–204.CrossRef
15.
go back to reference Van Horn DL, Sendele DD, Seideman S, Buco PJ. Regenerative capacity of the corneal endothelium in rabbit and cat. Invest Ophthalmol Vis Sci. 1977;16(7):597–613.PubMed Van Horn DL, Sendele DD, Seideman S, Buco PJ. Regenerative capacity of the corneal endothelium in rabbit and cat. Invest Ophthalmol Vis Sci. 1977;16(7):597–613.PubMed
16.
go back to reference Atzpodien EA, Jacobsen B, Funk J, et al. Advanced clinical imaging and tissue-based biomarkers of the eye for toxicology studies in Minipigs. Toxicol Pathol. 2016;44(3):398–413.CrossRef Atzpodien EA, Jacobsen B, Funk J, et al. Advanced clinical imaging and tissue-based biomarkers of the eye for toxicology studies in Minipigs. Toxicol Pathol. 2016;44(3):398–413.CrossRef
17.
go back to reference Nicholls SM, Mitchard LK, Laycock GM, et al. A model of corneal graft rejection in semi-inbred NIH miniature swine: significant T-cell infiltration of clinically accepted allografts. Invest Ophthalmol Vis Sci. 2012;53(6):3183–92.CrossRef Nicholls SM, Mitchard LK, Laycock GM, et al. A model of corneal graft rejection in semi-inbred NIH miniature swine: significant T-cell infiltration of clinically accepted allografts. Invest Ophthalmol Vis Sci. 2012;53(6):3183–92.CrossRef
18.
go back to reference Storr-Paulsen A, Norregaard JC, Farik G, Tarnhoj J. The influence of viscoelastic substances on the corneal endothelial cell population during cataract surgery: a prospective study of cohesive and dispersive viscoelastics. Acta Ophthalmol Scand. 2007;85(2):183–7.CrossRef Storr-Paulsen A, Norregaard JC, Farik G, Tarnhoj J. The influence of viscoelastic substances on the corneal endothelial cell population during cataract surgery: a prospective study of cohesive and dispersive viscoelastics. Acta Ophthalmol Scand. 2007;85(2):183–7.CrossRef
Metadata
Title
Preclinical Safety Evaluation of Ophthalmic Viscosurgical Devices in Rabbits and a Novel Mini-Pig Model
Authors
Ronika S. Leang
Lisa J. Kloft
Brad Gray
Arlene E. Gwon
Ling C. Huang
Publication date
01-03-2019
Publisher
Springer Healthcare
Keyword
Cataract
Published in
Ophthalmology and Therapy / Issue 1/2019
Print ISSN: 2193-8245
Electronic ISSN: 2193-6528
DOI
https://doi.org/10.1007/s40123-019-0167-9

Other articles of this Issue 1/2019

Ophthalmology and Therapy 1/2019 Go to the issue