Skip to main content
Top
Published in: Diabetes Therapy 3/2017

Open Access 01-06-2017 | Original Research

Indian Injection Technique Study: Population Characteristics and Injection Practices

Authors: Sanjay Kalra, Ambrish Mithal, Rakesh Sahay, Mathew John, A. G. Unnikrishnan, Banshi Saboo, Sujoy Ghosh, Debmalya Sanyal, Laurence J. Hirsch, Vandita Gupta, Kenneth W. Strauss

Published in: Diabetes Therapy | Issue 3/2017

Login to get access

Abstract

Introduction

It was estimated that 3.2 million Indians with diabetes injected insulin in 2010, but little is known about the techniques used.

Methods

In 2015 we conducted an injection technique questionnaire (ITQ) survey throughout India involving 1011 patients. Indian values were compared with those from 41 other countries participating in the ITQ, known here as rest of world (ROW).

Results

Mean HbA1c was 8.6. BMI values in India were 1.5–3 units lower than in ROW depending on patient group, meaning the risk of intramuscular (IM) injections is high in India. The mean total daily dose (TDD) of insulin was lower in every category of Indian patient than in ROW, perhaps reflecting the lower BMI. Needle reuse, whether with pens or syringes, is much higher in India than ROW and so is the number of times the needle is used. The majority (56.8%) of Indian insulin users performed only 2 injections/day as opposed to ROW where 45% of patients performed at least 4 injections/day. Indian patients inject insulin in the thighs more often than patients in ROW, a site where IM injections are more risky. Many patients do not have proper access to sharps containers or have other risk factors that could lead to blood-borne pathogen spread. More than 60% of used sharps in India go into the rubbish, with nearly 12% not even having the minimum protection of a cap.

Discussion

The shortest needles are very common in India; however, the level of needle reuse is high. Multiple daily injections therapy is not as common in India as ROW. More focus needs to be given to dwell times under the skin, reconstitution of cloudy insulins, skinfolds, and safe sharps disposal.
Literature
7.
go back to reference Tandon N, Kalra S, Balhara YS, et al. Forum for Injection Technique (FIT), India: The Indian recommendations 2.0, for best practice in insulin injection technique, 2015. Indian J Endocr Metab. 2015;19(3):317–331. Tandon N, Kalra S, Balhara YS, et al. Forum for Injection Technique (FIT), India: The Indian recommendations 2.0, for best practice in insulin injection technique, 2015. Indian J Endocr Metab. 2015;19(3):317–331.
8.
go back to reference Frid A, Hirsch L, Gaspar R, et al. New injection recommendations for patients with diabetes. Diabetes Metab. 2010;36(2):S3–18.CrossRefPubMed Frid A, Hirsch L, Gaspar R, et al. New injection recommendations for patients with diabetes. Diabetes Metab. 2010;36(2):S3–18.CrossRefPubMed
10.
go back to reference The worldwide data is labelled ROW (Rest Of World) and constitutes the mean values of the 41 other ITQ participating countries combined (excluding India). The worldwide data is labelled ROW (Rest Of World) and constitutes the mean values of the 41 other ITQ participating countries combined (excluding India).
11.
go back to reference Bantle JP, Neal L, Frankamp LM. Effects of the anatomical region used for insulin injections on glycaemia in type 1 diabetes subjects. Diabetes Care. 1993;16:1592–7.CrossRefPubMed Bantle JP, Neal L, Frankamp LM. Effects of the anatomical region used for insulin injections on glycaemia in type 1 diabetes subjects. Diabetes Care. 1993;16:1592–7.CrossRefPubMed
12.
go back to reference Frid A, Lindén B. Intraregional differences in the absorption of unmodified insulin from the abdominal wall. Diabet Med. 1992;9:236–9.CrossRefPubMed Frid A, Lindén B. Intraregional differences in the absorption of unmodified insulin from the abdominal wall. Diabet Med. 1992;9:236–9.CrossRefPubMed
13.
go back to reference Koivisto VA, Felig P. Alterations in insulin absorption and in blood glucose control associated with varying insulin injection sites in diabetic patients. Ann Intern Med. 1980;92:59–61.CrossRefPubMed Koivisto VA, Felig P. Alterations in insulin absorption and in blood glucose control associated with varying insulin injection sites in diabetic patients. Ann Intern Med. 1980;92:59–61.CrossRefPubMed
14.
go back to reference Annersten M, Willman A. Performing subcutaneous injections: a literature review. Worldviews Evidence Based Nurs. 2005;2:122–30.CrossRef Annersten M, Willman A. Performing subcutaneous injections: a literature review. Worldviews Evidence Based Nurs. 2005;2:122–30.CrossRef
15.
go back to reference Vidal M, Colungo C, Jansà M. Actualización sobre técnicas y sistemas de administración de la insulina (I) [Update on insulin administration techniques and devices (I)]. Av Diabetol. 2008;24:175–90. Vidal M, Colungo C, Jansà M. Actualización sobre técnicas y sistemas de administración de la insulina (I) [Update on insulin administration techniques and devices (I)]. Av Diabetol. 2008;24:175–90.
16.
go back to reference Gibney MA, Arce CH, Byron KJ, Hirsch LJ. Skin and subcutaneous adipose layer thickness in adults with diabetes at sites used for insulin injections: implications for needle length recommendations. Curr Med Res Opin. 2010;26:1519–30.CrossRefPubMed Gibney MA, Arce CH, Byron KJ, Hirsch LJ. Skin and subcutaneous adipose layer thickness in adults with diabetes at sites used for insulin injections: implications for needle length recommendations. Curr Med Res Opin. 2010;26:1519–30.CrossRefPubMed
17.
go back to reference Jain SM, Pandey K, Lahoti A, Rao PK. Evaluation of skin and subcutaneous tissue thickness at insulin injection sites in Indian, insulin naïve, type-2 diabetic adult population. Indian J Endocrinol Metab. 2013;17(5):864–70.CrossRefPubMedPubMedCentral Jain SM, Pandey K, Lahoti A, Rao PK. Evaluation of skin and subcutaneous tissue thickness at insulin injection sites in Indian, insulin naïve, type-2 diabetic adult population. Indian J Endocrinol Metab. 2013;17(5):864–70.CrossRefPubMedPubMedCentral
18.
go back to reference Hirsch L, Byron K, Gibney M. Intramuscular risk at insulin injection sites—measurement of the distance from skin to muscle and rationale for shorter-length needles for subcutaneous insulin therapy. Diabetes Technol Ther. 2014;16(12):867–73.CrossRefPubMed Hirsch L, Byron K, Gibney M. Intramuscular risk at insulin injection sites—measurement of the distance from skin to muscle and rationale for shorter-length needles for subcutaneous insulin therapy. Diabetes Technol Ther. 2014;16(12):867–73.CrossRefPubMed
19.
go back to reference Bergenstal RM, Strock ES, Peremislov D, Gibney MA, Parvu V, Hirsch LJ. Safety and efficacy of insulin therapy delivered via a 4 mm pen needle in obese patients with diabetes. Mayo Clin Proc. 2015;90(3):329–38. Bergenstal RM, Strock ES, Peremislov D, Gibney MA, Parvu V, Hirsch LJ. Safety and efficacy of insulin therapy delivered via a 4 mm pen needle in obese patients with diabetes. Mayo Clin Proc. 2015;90(3):329–38.
20.
go back to reference Birkebaek N, Solvig J, Hansen B, Jorgensen C, Smedegaard J, Christiansen J. A 4 mm needle reduces the risk of intramuscular injections without increasing backflow to skin surface in lean diabetic children and adults. Diabetes Care. 2008;22:e65.CrossRef Birkebaek N, Solvig J, Hansen B, Jorgensen C, Smedegaard J, Christiansen J. A 4 mm needle reduces the risk of intramuscular injections without increasing backflow to skin surface in lean diabetic children and adults. Diabetes Care. 2008;22:e65.CrossRef
21.
go back to reference Jamal R, Ross SA, Parkes JL, Pardo S, Ginsberg BH. Role of injection technique in use of insulin pens: prospective evaluation of a 31-gauge, 8 mm insulin pen needle. Endocr Pract. 1999;5:245–50.CrossRefPubMed Jamal R, Ross SA, Parkes JL, Pardo S, Ginsberg BH. Role of injection technique in use of insulin pens: prospective evaluation of a 31-gauge, 8 mm insulin pen needle. Endocr Pract. 1999;5:245–50.CrossRefPubMed
22.
go back to reference Kreugel G, Keers JC, Jongbloed A, Verweij-Gjaltema AH, Wolffenbuttel BHR. The influence of needle length on glycemic control and patient preference in obese diabetic patients. Diabetes. 2009;58:A117. Kreugel G, Keers JC, Jongbloed A, Verweij-Gjaltema AH, Wolffenbuttel BHR. The influence of needle length on glycemic control and patient preference in obese diabetic patients. Diabetes. 2009;58:A117.
23.
go back to reference Schwartz S, Hassman D, Shelmet J, et al. A multicenter, open-label, randomized, two-period crossover trial comparing glycemic control, satisfaction, and preference achieved with a 31 gauge × 6 mm needle versus a 29 gauge × 12.7 mm needle in obese patients with diabetes mellitus. Clin Ther. 2004;26:1663–78. Schwartz S, Hassman D, Shelmet J, et al. A multicenter, open-label, randomized, two-period crossover trial comparing glycemic control, satisfaction, and preference achieved with a 31 gauge × 6 mm needle versus a 29 gauge × 12.7 mm needle in obese patients with diabetes mellitus. Clin Ther. 2004;26:1663–78.
24.
go back to reference Kreugel G, Beijer HJM, Kerstens MN, ter Maaten JC, Sluiter WJ, Boot BS. Influence of needle size for SC insulin administration on metabolic control and patient acceptance. Eur Diabetes Nurs. 2007;4:1–5.CrossRef Kreugel G, Beijer HJM, Kerstens MN, ter Maaten JC, Sluiter WJ, Boot BS. Influence of needle size for SC insulin administration on metabolic control and patient acceptance. Eur Diabetes Nurs. 2007;4:1–5.CrossRef
25.
go back to reference Van Doorn LG, Alberda A, Lytzen L. Insulin leakage and pain perception with NovoFine 6 mm and NovoFine 12 mm needle lengths in patients with type 1 or type 2 diabetes. Diabet Med. 1998;1:S50. Van Doorn LG, Alberda A, Lytzen L. Insulin leakage and pain perception with NovoFine 6 mm and NovoFine 12 mm needle lengths in patients with type 1 or type 2 diabetes. Diabet Med. 1998;1:S50.
26.
go back to reference Miwa T, Itoh R, Kobayashi T, et al. Comparison of the effects of a new 32-gauge × 4-mm pen needle and a 32-gauge × 6-mm pen needle on glycemic control, safety, and patient ratings in Japanese adults with diabetes. Diabetes Technol Ther. 2012;14(12):1084–90.CrossRefPubMed Miwa T, Itoh R, Kobayashi T, et al. Comparison of the effects of a new 32-gauge × 4-mm pen needle and a 32-gauge × 6-mm pen needle on glycemic control, safety, and patient ratings in Japanese adults with diabetes. Diabetes Technol Ther. 2012;14(12):1084–90.CrossRefPubMed
27.
go back to reference Nagai Y, Ohshige T, Arai K, Kobayashi H, Sada Y, Ohmori S. Comparison between shorter straight and thinner microtapered insulin injection needles. Diabetes Technol Ther. 2013;15(7):550–5.CrossRefPubMed Nagai Y, Ohshige T, Arai K, Kobayashi H, Sada Y, Ohmori S. Comparison between shorter straight and thinner microtapered insulin injection needles. Diabetes Technol Ther. 2013;15(7):550–5.CrossRefPubMed
28.
go back to reference Hirose T, Ogihara T, Tozaka S, Kanderian S, Watada H. Identification and comparison of insulin pharmacokinetics injected with a new 4-mm needle vs 6- and 8-mm needles accounting for endogenous insulin and C-peptide secretion kinetics in non-diabetic adult males. J Diabetes Investig. 2013;4(3):287–96.CrossRefPubMedPubMedCentral Hirose T, Ogihara T, Tozaka S, Kanderian S, Watada H. Identification and comparison of insulin pharmacokinetics injected with a new 4-mm needle vs 6- and 8-mm needles accounting for endogenous insulin and C-peptide secretion kinetics in non-diabetic adult males. J Diabetes Investig. 2013;4(3):287–96.CrossRefPubMedPubMedCentral
29.
go back to reference de la Peña A, Yeo KP, Linnebjerg H, et al. Subcutaneous injection depth does not affect the pharmacokinetics or glucodynamics of insulin lispro in normal weight or healthy obese subjects. J Diabetes Sci Technol. 2015;9:1–7. de la Peña A, Yeo KP, Linnebjerg H, et al. Subcutaneous injection depth does not affect the pharmacokinetics or glucodynamics of insulin lispro in normal weight or healthy obese subjects. J Diabetes Sci Technol. 2015;9:1–7.
30.
go back to reference Birkebaek NH, Solvig J, Hansen B, Jorgensen C, Smedegaard J, Christiansen JS. A 4-mm needle reduces the risk of intramuscular injections without increasing backflow to skin surface in lean diabetic children and adults. Diabetes Care. 2008;31(9):e65.CrossRefPubMed Birkebaek NH, Solvig J, Hansen B, Jorgensen C, Smedegaard J, Christiansen JS. A 4-mm needle reduces the risk of intramuscular injections without increasing backflow to skin surface in lean diabetic children and adults. Diabetes Care. 2008;31(9):e65.CrossRefPubMed
31.
go back to reference Hirsch LJ, Gibney MA, Li L, Bérubé J. Glycemic control, reported pain and leakage with a 4 mm × 32 G pen needle in obese and non-obese adults with diabetes: a post hoc analysis. Curr Med Res Opin. 2012;28(8):1305–11.CrossRefPubMed Hirsch LJ, Gibney MA, Li L, Bérubé J. Glycemic control, reported pain and leakage with a 4 mm × 32 G pen needle in obese and non-obese adults with diabetes: a post hoc analysis. Curr Med Res Opin. 2012;28(8):1305–11.CrossRefPubMed
32.
go back to reference Lo Presti D, Ingegnosi C, Strauss K. Skin and subcutaneous thickness at injecting sites in children with diabetes: ultrasound findings and recommendations for giving injection. Pediatr Diabetes. 2012;13(7):525–33.CrossRefPubMed Lo Presti D, Ingegnosi C, Strauss K. Skin and subcutaneous thickness at injecting sites in children with diabetes: ultrasound findings and recommendations for giving injection. Pediatr Diabetes. 2012;13(7):525–33.CrossRefPubMed
33.
go back to reference Hofman PL, Derraik JG, Pinto TE, et al. Defining the ideal injection techniques when using 5-mm needles in children and adults. Diab Care. 2010;33(9):1940–4.CrossRef Hofman PL, Derraik JG, Pinto TE, et al. Defining the ideal injection techniques when using 5-mm needles in children and adults. Diab Care. 2010;33(9):1940–4.CrossRef
34.
go back to reference Hirsch LJ, Gibney MA, Albanese J, et al. Comparative glycemic control, safety and patient ratings for a new 4 mm × 32G insulin pen needle in adults with diabetes. Curr Med Res Opin. 2010;26(6):1531–41.CrossRefPubMed Hirsch LJ, Gibney MA, Albanese J, et al. Comparative glycemic control, safety and patient ratings for a new 4 mm × 32G insulin pen needle in adults with diabetes. Curr Med Res Opin. 2010;26(6):1531–41.CrossRefPubMed
35.
go back to reference Strock E, et al. Glycemic control, pain and leakage with 4 mm vs larger pen needles in obese patients treated with lantus or higher insulin doses: prespecified subgroup analyses. In: Poster 997-P. EASD, Barcelona, Spain. 2013. Strock E, et al. Glycemic control, pain and leakage with 4 mm vs larger pen needles in obese patients treated with lantus or higher insulin doses: prespecified subgroup analyses. In: Poster 997-P. EASD, Barcelona, Spain. 2013.
36.
go back to reference Ignaut D, Fu H. Comparison of insulin diluent leakage postinjection using two different needle lengths and injection volumes in obese patients with type 1 or type 2 diabetes mellitus. J Diabetes Sci Technol. 2012;6:389–93.CrossRefPubMedPubMedCentral Ignaut D, Fu H. Comparison of insulin diluent leakage postinjection using two different needle lengths and injection volumes in obese patients with type 1 or type 2 diabetes mellitus. J Diabetes Sci Technol. 2012;6:389–93.CrossRefPubMedPubMedCentral
37.
go back to reference Kreugel G, Keers JC, Kerstens MN, Wolffenbuttel BH. Randomized trial on the influence of the length of two insulin pen needles on glycemic control and patient preference in obese patients with diabetes. Diabetes Technol Ther. 2011;13(7):737–41.CrossRefPubMedPubMedCentral Kreugel G, Keers JC, Kerstens MN, Wolffenbuttel BH. Randomized trial on the influence of the length of two insulin pen needles on glycemic control and patient preference in obese patients with diabetes. Diabetes Technol Ther. 2011;13(7):737–41.CrossRefPubMedPubMedCentral
38.
go back to reference Vardar B, Kizilci S. Incidence of lipohypertrophy in diabetic patients and a study of influencing factors. Diabetes Res Clin Pract. 2007;77:231–6.CrossRefPubMed Vardar B, Kizilci S. Incidence of lipohypertrophy in diabetic patients and a study of influencing factors. Diabetes Res Clin Pract. 2007;77:231–6.CrossRefPubMed
39.
go back to reference Saez-de Ibarra L, Gallego F. Factors related to lipohypertrophy in insulin-treated diabetic patients: role of educational intervention. Pract Diab Int. 1998;15:9–11.CrossRef Saez-de Ibarra L, Gallego F. Factors related to lipohypertrophy in insulin-treated diabetic patients: role of educational intervention. Pract Diab Int. 1998;15:9–11.CrossRef
40.
go back to reference Blanco M, Hernández MT, Strauss KW, Amaya M. Prevalence and risk factors of lipohypertrophy in insulin-injecting patients with diabetes. Diabetes Metab. 2013;39(5):445–53.CrossRefPubMed Blanco M, Hernández MT, Strauss KW, Amaya M. Prevalence and risk factors of lipohypertrophy in insulin-injecting patients with diabetes. Diabetes Metab. 2013;39(5):445–53.CrossRefPubMed
41.
go back to reference Hirsch L, Ji L, Sun Z, et al. Lipohypertrophy—prevalence, risk factors and clinical characteristics of insulin-requiring patients in China. DTT. 2015;17(Suppl 1):A57–8. Hirsch L, Ji L, Sun Z, et al. Lipohypertrophy—prevalence, risk factors and clinical characteristics of insulin-requiring patients in China. DTT. 2015;17(Suppl 1):A57–8.
42.
go back to reference De Coninck C, Frid A, Gaspar R, et al. Results and analysis of the 2008–2009 insulin injection technique questionnaire survey. J Diabetes. 2010;2(3):168–79.CrossRefPubMed De Coninck C, Frid A, Gaspar R, et al. Results and analysis of the 2008–2009 insulin injection technique questionnaire survey. J Diabetes. 2010;2(3):168–79.CrossRefPubMed
43.
go back to reference Misnikova I, Dreval A, Gubkina V, Rusanova E. The risk of repeated use of insulin pen needles in patients with diabetes mellitus. J Diabetology. 2011;1:1–5. Misnikova I, Dreval A, Gubkina V, Rusanova E. The risk of repeated use of insulin pen needles in patients with diabetes mellitus. J Diabetology. 2011;1:1–5.
44.
go back to reference Schuler G, Pelz K, Kerp L. Is the reuse of needles for insulin injection systems associated with a higher risk of cutaneous complications? Diabetes Res Clin Pract. 1992;16:209–12.CrossRefPubMed Schuler G, Pelz K, Kerp L. Is the reuse of needles for insulin injection systems associated with a higher risk of cutaneous complications? Diabetes Res Clin Pract. 1992;16:209–12.CrossRefPubMed
45.
go back to reference Thomas DR, Fischer RG, Nicholas WC, Beghe C, Hatten KW, Thomas JN. Disposable insulin syringe reuse and aseptic practices in diabetic patients. J Gen Intern Med. 1989;4:97–100.CrossRefPubMed Thomas DR, Fischer RG, Nicholas WC, Beghe C, Hatten KW, Thomas JN. Disposable insulin syringe reuse and aseptic practices in diabetic patients. J Gen Intern Med. 1989;4:97–100.CrossRefPubMed
46.
go back to reference Puder J, Atar M, Muller B, Pavan M, Keller U. Using insulin pen needles up to five times does not affect needle tip shape nor increase pain intensity. Diabetes Res Clin Pract. 2005;67:119–23.CrossRefPubMed Puder J, Atar M, Muller B, Pavan M, Keller U. Using insulin pen needles up to five times does not affect needle tip shape nor increase pain intensity. Diabetes Res Clin Pract. 2005;67:119–23.CrossRefPubMed
47.
go back to reference Strauss K. Insulin injection techniques. Pract Diabetes Int. 1998;15:181–4.CrossRef Strauss K. Insulin injection techniques. Pract Diabetes Int. 1998;15:181–4.CrossRef
48.
go back to reference Galloway J, Spradlin C, Nelson R, Wentworth S, Davidson J, Swarner J. Factors influencing the absorption, serum insulin concentration, and blood glucose responses after injections of regular insulin and various insulin mixtures. Diabetes Care. 1981;4:366–76.CrossRefPubMed Galloway J, Spradlin C, Nelson R, Wentworth S, Davidson J, Swarner J. Factors influencing the absorption, serum insulin concentration, and blood glucose responses after injections of regular insulin and various insulin mixtures. Diabetes Care. 1981;4:366–76.CrossRefPubMed
49.
go back to reference Sindelka G, Heinemann L, Berger M, Frenck W, Chantelau E. Effect of insulin concentration, subcutaneous fat thickness and skin temperature on subcutaneous insulin absorption in healthy subjects. Diabetologia. 1994;37:377–80.CrossRefPubMed Sindelka G, Heinemann L, Berger M, Frenck W, Chantelau E. Effect of insulin concentration, subcutaneous fat thickness and skin temperature on subcutaneous insulin absorption in healthy subjects. Diabetologia. 1994;37:377–80.CrossRefPubMed
50.
go back to reference Henriksen JE, Djurhuus MS, Vaag A, et al. Impact of injection sites for soluble insulin on glycaemic control in Type 1 (insulin-dependent) diabetic patients treated with a multiple insulin injection regimen. Diabetologia. 1993;36:752–8.CrossRefPubMed Henriksen JE, Djurhuus MS, Vaag A, et al. Impact of injection sites for soluble insulin on glycaemic control in Type 1 (insulin-dependent) diabetic patients treated with a multiple insulin injection regimen. Diabetologia. 1993;36:752–8.CrossRefPubMed
51.
go back to reference Henriksen J, Vaag A, Ramsgaard Hansen I, Lauritzen M, Djurhuus M, Beck-Nielsen H. Absorption of NPH (Isophane) insulin in resting diabetic patients: evidence for subcutaneous injection in the thigh as the preferred site. Diabet Med. 1991;8:453–7.CrossRefPubMed Henriksen J, Vaag A, Ramsgaard Hansen I, Lauritzen M, Djurhuus M, Beck-Nielsen H. Absorption of NPH (Isophane) insulin in resting diabetic patients: evidence for subcutaneous injection in the thigh as the preferred site. Diabet Med. 1991;8:453–7.CrossRefPubMed
52.
go back to reference Koivisto V, Felig P. Alterations in insulin absorption and in blood glucose control associated with varying insulin injection sites in diabetic patients. Ann Intern Med. 1989;92:59–61.CrossRef Koivisto V, Felig P. Alterations in insulin absorption and in blood glucose control associated with varying insulin injection sites in diabetic patients. Ann Intern Med. 1989;92:59–61.CrossRef
53.
go back to reference De Meijer PH, Lutterman JA, van Lier HJ, van’t Laar A. The variability of the absorption of subcutaneously injected insulin: effect of injection technique and relation with brittleness. Diabet Med. 1990;7:499–505.CrossRefPubMed De Meijer PH, Lutterman JA, van Lier HJ, van’t Laar A. The variability of the absorption of subcutaneously injected insulin: effect of injection technique and relation with brittleness. Diabet Med. 1990;7:499–505.CrossRefPubMed
54.
go back to reference Berger M, Cüppers HJ, Hegner H, Jörgens V, Berchtold P. Absorption kinetics and biologic effects of subcutaneously injected insulin preparations. Diabetes Care. 1982;2:77–91.CrossRef Berger M, Cüppers HJ, Hegner H, Jörgens V, Berchtold P. Absorption kinetics and biologic effects of subcutaneously injected insulin preparations. Diabetes Care. 1982;2:77–91.CrossRef
55.
go back to reference Frid A, Gunnarson R, Guntner P, Linde P. Effects of accidental intramuscular injection on insulin absorption in IDDM. Diabetes Care. 1988;11:41–5.CrossRefPubMed Frid A, Gunnarson R, Guntner P, Linde P. Effects of accidental intramuscular injection on insulin absorption in IDDM. Diabetes Care. 1988;11:41–5.CrossRefPubMed
56.
go back to reference Vaag A, Handberg A, Lauritzen M, Henriksen JE, Pedersen KD, Beck-Nielsen H. Variation in absorption of NPH insulin due to intramuscular injection. Diabetes Care. 1990;13(1):74–6. Vaag A, Handberg A, Lauritzen M, Henriksen JE, Pedersen KD, Beck-Nielsen H. Variation in absorption of NPH insulin due to intramuscular injection. Diabetes Care. 1990;13(1):74–6.
57.
go back to reference Thow J, Johnson A, Fulcher G, Home P. Different absorption of Isophane (NPH) Insulin from subcutaneous and intramuscular sites suggests a need to reassess recommended insulin injection technique. Diabet Med. 1990;7(3):600–2.CrossRefPubMed Thow J, Johnson A, Fulcher G, Home P. Different absorption of Isophane (NPH) Insulin from subcutaneous and intramuscular sites suggests a need to reassess recommended insulin injection technique. Diabet Med. 1990;7(3):600–2.CrossRefPubMed
58.
go back to reference Rave K, Heise T, Weyer C, et al. Intramuscular versus subcutaneous injection of soluble and lispro insulins: comparison of metabolic effects in healthy subjects. Diabet Med. 1998;15:747–51.CrossRefPubMed Rave K, Heise T, Weyer C, et al. Intramuscular versus subcutaneous injection of soluble and lispro insulins: comparison of metabolic effects in healthy subjects. Diabet Med. 1998;15:747–51.CrossRefPubMed
59.
go back to reference Karges B, Boehm BO, Karges W. Early hypoglycemia after accidental intramuscular injection of insulin glargine. Diabet Med. 2005;22:1444–5.CrossRefPubMed Karges B, Boehm BO, Karges W. Early hypoglycemia after accidental intramuscular injection of insulin glargine. Diabet Med. 2005;22:1444–5.CrossRefPubMed
60.
go back to reference Karges B, Boehm BO, Karges W. Early hypoglycaemia after accidental intramuscular injection of insulin glargine. Diabetes Med. 2005;22(10):1444–5.CrossRef Karges B, Boehm BO, Karges W. Early hypoglycaemia after accidental intramuscular injection of insulin glargine. Diabetes Med. 2005;22(10):1444–5.CrossRef
61.
go back to reference Spraul M, Chantelau E, Koumoulidou J, Berger M. Subcutaneous or nonsubcutaneous injection of insulin. Diabetes Care. 1988;11:733–6.CrossRefPubMed Spraul M, Chantelau E, Koumoulidou J, Berger M. Subcutaneous or nonsubcutaneous injection of insulin. Diabetes Care. 1988;11:733–6.CrossRefPubMed
63.
go back to reference Hirsch L, Byron K, Gibney M. Intramuscular risk at insulin injection sites-measurement of the distance from skin to muscle and rationale for shorter-length needles for subcutaneous insulin therapy. Diabetes Technol Ther. 2014;16(12):867–73.CrossRefPubMed Hirsch L, Byron K, Gibney M. Intramuscular risk at insulin injection sites-measurement of the distance from skin to muscle and rationale for shorter-length needles for subcutaneous insulin therapy. Diabetes Technol Ther. 2014;16(12):867–73.CrossRefPubMed
64.
go back to reference de la Peña A, Ma X, Reddy S, Ovalle F, Bergenstal RM, Jackson JA. Application of PK/PD modeling and simulation to dosing regimen optimization of high-dose human regular U-500 insulin. J Diabetes Sci Technol. 2014;8:821–9.CrossRefPubMedPubMedCentral de la Peña A, Ma X, Reddy S, Ovalle F, Bergenstal RM, Jackson JA. Application of PK/PD modeling and simulation to dosing regimen optimization of high-dose human regular U-500 insulin. J Diabetes Sci Technol. 2014;8:821–9.CrossRefPubMedPubMedCentral
66.
go back to reference Ginsberg BH, Parkes JL, Sparacino C. The kinetics of insulin administration by insulin pens. Horm Metab Res. 1994;26:584–7.CrossRefPubMed Ginsberg BH, Parkes JL, Sparacino C. The kinetics of insulin administration by insulin pens. Horm Metab Res. 1994;26:584–7.CrossRefPubMed
67.
go back to reference Annersten M, Frid A. Insulin pens dribble from the tip of the needle after injection. Pract Diab Int. 2000;17:109–11.CrossRef Annersten M, Frid A. Insulin pens dribble from the tip of the needle after injection. Pract Diab Int. 2000;17:109–11.CrossRef
68.
go back to reference Rissler J, Jørgensen C, Rye Hansen M, Hansen NA. Evaluation of the injection force dynamics of a modified prefilled insulin pen. Expert Opin Pharmacother. 2008;9:2217–22.CrossRefPubMed Rissler J, Jørgensen C, Rye Hansen M, Hansen NA. Evaluation of the injection force dynamics of a modified prefilled insulin pen. Expert Opin Pharmacother. 2008;9:2217–22.CrossRefPubMed
69.
go back to reference Broadway CA. Prevention of insulin leakage after subcutaneous injection. Diabetes Educator. 1991;17:90.CrossRefPubMed Broadway CA. Prevention of insulin leakage after subcutaneous injection. Diabetes Educator. 1991;17:90.CrossRefPubMed
70.
go back to reference Le Floch JP, Herbreteau C, Lange F, Perlemuter L. Biologic material in needles and cartridges after insulin injection with a pen in diabetic patients. Diabetes Care. 1998;21:1502–4.CrossRef Le Floch JP, Herbreteau C, Lange F, Perlemuter L. Biologic material in needles and cartridges after insulin injection with a pen in diabetic patients. Diabetes Care. 1998;21:1502–4.CrossRef
71.
go back to reference Bärtsch U, Comtesse C, Wetekam B. Insulin pens for treatment of diabetes (article in German). Ther Umsch. 2006;63:398–404.CrossRefPubMed Bärtsch U, Comtesse C, Wetekam B. Insulin pens for treatment of diabetes (article in German). Ther Umsch. 2006;63:398–404.CrossRefPubMed
72.
go back to reference Byetta Pen User Manual. Eli Lilly and Company; 2007. Byetta Pen User Manual. Eli Lilly and Company; 2007.
73.
go back to reference Chantelau E, Heinemann L, Ross D. Air bubbles in insulin pens. Lancet. 1989;334:387–8.CrossRef Chantelau E, Heinemann L, Ross D. Air bubbles in insulin pens. Lancet. 1989;334:387–8.CrossRef
74.
go back to reference Strauss K, De Gols H, Letondeur C, Matyjaszczyk M, Frid A. The second injection technique event (SITE), May 2000, Barcelona, Spain. Pract Diab Int. 2002;19:17–21.CrossRef Strauss K, De Gols H, Letondeur C, Matyjaszczyk M, Frid A. The second injection technique event (SITE), May 2000, Barcelona, Spain. Pract Diab Int. 2002;19:17–21.CrossRef
75.
go back to reference Strauss K. Insulin injection techniques: report from the 1st International Insulin Injection Technique Workshop, Strasbourg, France—June 1997. Pract Diab Int. 1998;15:16–20.CrossRef Strauss K. Insulin injection techniques: report from the 1st International Insulin Injection Technique Workshop, Strasbourg, France—June 1997. Pract Diab Int. 1998;15:16–20.CrossRef
78.
go back to reference Chantelau E, Lee DM, Hemmann DM, Zipfel U, Echterhoff S. What makes insulin injections painful? Br Med J. 1991;303:26–7.CrossRef Chantelau E, Lee DM, Hemmann DM, Zipfel U, Echterhoff S. What makes insulin injections painful? Br Med J. 1991;303:26–7.CrossRef
79.
go back to reference Maljaars C. Scherpe studie naalden voor eenmalig gebruik [Sharp study needles for single use]. Diabetes Levery. 2002;4:36–37. Maljaars C. Scherpe studie naalden voor eenmalig gebruik [Sharp study needles for single use]. Diabetes Levery. 2002;4:36–37.
80.
go back to reference Torrance T. An unexpected hazard of insulin injection. Pract Diabetes Int. 2002;19:63.CrossRef Torrance T. An unexpected hazard of insulin injection. Pract Diabetes Int. 2002;19:63.CrossRef
81.
82.
go back to reference Jehle PM, Micheler C, Jehle DR, Breitig D, Boehm BO. Inadequate suspension of neutral protamine Hagendorn (NPH) insulin in pens. Lancet. 1999;354:1604–7.CrossRefPubMed Jehle PM, Micheler C, Jehle DR, Breitig D, Boehm BO. Inadequate suspension of neutral protamine Hagendorn (NPH) insulin in pens. Lancet. 1999;354:1604–7.CrossRefPubMed
83.
go back to reference Brown A, Steel JM, Duncan C, Duncun A, McBain AM. An assessment of the adequacy of suspension of insulin in pen injectors. Diab Med. 2004;21:604–8.CrossRef Brown A, Steel JM, Duncan C, Duncun A, McBain AM. An assessment of the adequacy of suspension of insulin in pen injectors. Diab Med. 2004;21:604–8.CrossRef
85.
go back to reference Springs MH. Shake, rattle, or roll? “Challenging traditional insulin injection practices”. Am J Nurs. 1999;99:14.PubMed Springs MH. Shake, rattle, or roll? “Challenging traditional insulin injection practices”. Am J Nurs. 1999;99:14.PubMed
86.
go back to reference Jehle PM, Micheler C, Jehle DR, Breitig D, Boehm BO. Inadequate suspension of neutral protamine Hagendorn (NPH) insulin in pens. Lancet. 1999;354(9190):1604–7.CrossRefPubMed Jehle PM, Micheler C, Jehle DR, Breitig D, Boehm BO. Inadequate suspension of neutral protamine Hagendorn (NPH) insulin in pens. Lancet. 1999;354(9190):1604–7.CrossRefPubMed
87.
go back to reference Brown A, Steel JM, Duncan C, Duncan A, McBain AM. An assessment of the adequacy of suspension of insulin in pen injectors. Diabet Med. 2004;21(6):604–8.CrossRefPubMed Brown A, Steel JM, Duncan C, Duncan A, McBain AM. An assessment of the adequacy of suspension of insulin in pen injectors. Diabet Med. 2004;21(6):604–8.CrossRefPubMed
88.
go back to reference Kaiser P, Maxeiner S, Weise A, et al. Assessment of the mixing efficiency of neutral protamine Hagedorn cartridges. J Diabetes Sci Technol. 2010;4(3):652–7.CrossRefPubMedPubMedCentral Kaiser P, Maxeiner S, Weise A, et al. Assessment of the mixing efficiency of neutral protamine Hagedorn cartridges. J Diabetes Sci Technol. 2010;4(3):652–7.CrossRefPubMedPubMedCentral
89.
go back to reference Polak M, Beregszaszi M, Belarbi N, et al. Subcutaneous or intra-muscular injections of insulin in children: are we injecting where we think we are? Diabetes Care. 1996;19(12):1434–6.CrossRefPubMed Polak M, Beregszaszi M, Belarbi N, et al. Subcutaneous or intra-muscular injections of insulin in children: are we injecting where we think we are? Diabetes Care. 1996;19(12):1434–6.CrossRefPubMed
90.
go back to reference Tubiana-Rufi N, Belarbi N, Du Pasquier-Fediaevsky L, et al. Reduction of the risk of intra-muscular insulin injection with the 8 mm length needles in thin diabetic children. Diabetologia 1998;41:A247. Tubiana-Rufi N, Belarbi N, Du Pasquier-Fediaevsky L, et al. Reduction of the risk of intra-muscular insulin injection with the 8 mm length needles in thin diabetic children. Diabetologia 1998;41:A247.
91.
go back to reference Kalra S, Mithal A, Sahay R, et al. Indian Injection Technique Study: Injecting Complications, Education, and the Health Care Professional. Diabetes Ther 2017. doi:10.1007/s13300-017-0244-9. Kalra S, Mithal A, Sahay R, et al. Indian Injection Technique Study: Injecting Complications, Education, and the Health Care Professional. Diabetes Ther 2017. doi:10.​1007/​s13300-017-0244-9.
Metadata
Title
Indian Injection Technique Study: Population Characteristics and Injection Practices
Authors
Sanjay Kalra
Ambrish Mithal
Rakesh Sahay
Mathew John
A. G. Unnikrishnan
Banshi Saboo
Sujoy Ghosh
Debmalya Sanyal
Laurence J. Hirsch
Vandita Gupta
Kenneth W. Strauss
Publication date
01-06-2017
Publisher
Springer Healthcare
Published in
Diabetes Therapy / Issue 3/2017
Print ISSN: 1869-6953
Electronic ISSN: 1869-6961
DOI
https://doi.org/10.1007/s13300-017-0243-x

Other articles of this Issue 3/2017

Diabetes Therapy 3/2017 Go to the issue
Live Webinar | 27-06-2024 | 18:00 (CEST)

Keynote webinar | Spotlight on medication adherence

Live: Thursday 27th June 2024, 18:00-19:30 (CEST)

WHO estimates that half of all patients worldwide are non-adherent to their prescribed medication. The consequences of poor adherence can be catastrophic, on both the individual and population level.

Join our expert panel to discover why you need to understand the drivers of non-adherence in your patients, and how you can optimize medication adherence in your clinics to drastically improve patient outcomes.

Prof. Kevin Dolgin
Prof. Florian Limbourg
Prof. Anoop Chauhan
Developed by: Springer Medicine
Obesity Clinical Trial Summary

At a glance: The STEP trials

A round-up of the STEP phase 3 clinical trials evaluating semaglutide for weight loss in people with overweight or obesity.

Developed by: Springer Medicine

Highlights from the ACC 2024 Congress

Year in Review: Pediatric cardiology

Watch Dr. Anne Marie Valente present the last year's highlights in pediatric and congenital heart disease in the official ACC.24 Year in Review session.

Year in Review: Pulmonary vascular disease

The last year's highlights in pulmonary vascular disease are presented by Dr. Jane Leopold in this official video from ACC.24.

Year in Review: Valvular heart disease

Watch Prof. William Zoghbi present the last year's highlights in valvular heart disease from the official ACC.24 Year in Review session.

Year in Review: Heart failure and cardiomyopathies

Watch this official video from ACC.24. Dr. Biykem Bozkurt discusses last year's major advances in heart failure and cardiomyopathies.