Published in: The Journal of Obstetrics and Gynecology of India 5/2015

01-10-2015 | Invited Review Article

Surgical Efficacy Among Laparoscopic Ultrasonic Dissectors: Are We Advancing Safely? A Review of Literature

Authors: Rajesh Devassy, Sreelatha Gopalakrishnan, Rudy Leon De Wilde

Published in: The Journal of Obstetrics and Gynecology of India | Issue 5/2015

Login to get access

Abstract

The specialty of laparoscopy has evolved with the advent of new technologies over the last few years. Energy-based devices and Ultrasonic dissectors are used with a lot of factors in play-including ergonomics and economics during surgery. Here an attempt is based to review the surgical efficacy and safety of these dissectors with importance to plume production and lateral thermal damage. The factors contributing to adversities to the dissectors are also to be noted. The strategy adopted was aimed at finding relevant studies from PubMed from 1995 to 2014. The basic principle of plume production and thermal damage are studied in this review. Factors contributing to the same that can lead to adversities during laparoscopic surgeries are identified. Summarizing key points that increase lateral thermal damage and plume production amongst different ultrasonic shears and suggesting a technique to identify the right balance between the existing dissectors was possible. The RF Device and USS are both useful and widely used and are more safer than monopolar devices. RF Device is considerably slower than USS, as it cannot achieve coagulation and cutting at the same time. Although USS definitely improvises dissection and has less thermal injury than RF Device, the clinical implications in balancing dissection efficacy with hemostasis need to be investigated further. The ideal haemostatic energy-based shear device would be one with excellent hemostatic results and visual acuity while allowing none or minimal thermal energy escape at the point of application. In our current setting, a combined use of both RF and USS device usage as applied in the particular situations has potential.
Literature
1.
go back to reference Lallouf LB, Beri A, Klinger CH, et al. Practical hints for hemostasis in laparoscopic surgery. Minim Invasive Ther Allied Technol. 2007;16(1):45–51.CrossRef Lallouf LB, Beri A, Klinger CH, et al. Practical hints for hemostasis in laparoscopic surgery. Minim Invasive Ther Allied Technol. 2007;16(1):45–51.CrossRef
3.
go back to reference Den Boer KT, de Jong T, Dankelman J, et al. Problems with laparoscopic instruments: opinion of experts. J Laparoendosc Adv Surg Tech A. 2001;11(3):149–55.CrossRef Den Boer KT, de Jong T, Dankelman J, et al. Problems with laparoscopic instruments: opinion of experts. J Laparoendosc Adv Surg Tech A. 2001;11(3):149–55.CrossRef
5.
go back to reference Liberati A, Altman DG, Tetzlaff J, et al. The PRISMA statement for reporting systematic reviews and meta-analyses of studies that evaluate healthcare interventions: explanation and elaboration. BMJ. 2009;21:339. doi:10.1136/bmj.b2700. Liberati A, Altman DG, Tetzlaff J, et al. The PRISMA statement for reporting systematic reviews and meta-analyses of studies that evaluate healthcare interventions: explanation and elaboration. BMJ. 2009;21:339. doi:10.​1136/​bmj.​b2700.
7.
go back to reference Pogorelić Z, Perko Z, Družijanić N, et al. How to prevent lateral thermal damage to tissue using the harmonic scalpel: experimental study on pig small intestine and abdominal wall. Eur Surg Res. 2009;43:235–40.CrossRefPubMed Pogorelić Z, Perko Z, Družijanić N, et al. How to prevent lateral thermal damage to tissue using the harmonic scalpel: experimental study on pig small intestine and abdominal wall. Eur Surg Res. 2009;43:235–40.CrossRefPubMed
8.
go back to reference DesCoteaux J, Picard P, Poulin E, et al. Preliminary study of electrocautery smoke particles produced in vitro and during laparoscopic procedures. Surgical Endoscopy. 1996;10:152. doi:10.1007/BF00188362.CrossRefPubMed DesCoteaux J, Picard P, Poulin E, et al. Preliminary study of electrocautery smoke particles produced in vitro and during laparoscopic procedures. Surgical Endoscopy. 1996;10:152. doi:10.​1007/​BF00188362.CrossRefPubMed
9.
go back to reference Barrett WL, Garber SM. Surgical smoke: a review of the literature. Is this just a lot of hot air? Surg Endosc. 2003;17(6):979–87 (Epub 2003 Mar 19).CrossRefPubMed Barrett WL, Garber SM. Surgical smoke: a review of the literature. Is this just a lot of hot air? Surg Endosc. 2003;17(6):979–87 (Epub 2003 Mar 19).CrossRefPubMed
10.
go back to reference Matthew MD, Stevens E, Mark Taylor S, et al. The air that we breathe’: assessment of laser and electrosurgical dissection devices on operating theater air quality. J Otolaryngol Head Neck Surg. 2014;43(1):39. doi:10.1186/s40463-014-0039. Matthew MD, Stevens E, Mark Taylor S, et al. The air that we breathe’: assessment of laser and electrosurgical dissection devices on operating theater air quality. J Otolaryngol Head Neck Surg. 2014;43(1):39. doi:10.​1186/​s40463-014-0039.
11.
go back to reference Ball K. Survey of physicians attitudes toward surgical smoke. Can Oper Room Nurs J. 1995;13(4):18–21 (No abstract available). Ball K. Survey of physicians attitudes toward surgical smoke. Can Oper Room Nurs J. 1995;13(4):18–21 (No abstract available).
12.
go back to reference Hoglan M. Potential hazards from electrosurgery plume–recommendations for surgical smoke evacuation. Can Oper Room Nurs J. 1995;13(4):10–6.PubMed Hoglan M. Potential hazards from electrosurgery plume–recommendations for surgical smoke evacuation. Can Oper Room Nurs J. 1995;13(4):10–6.PubMed
13.
go back to reference Ott DE, Moss E, Martinez K. Aerosol exposure from an ultrasonically activated (Harmonic) device. J Am Assoc Gynecol Laparosc. 1998;5(1):29–32.CrossRefPubMed Ott DE, Moss E, Martinez K. Aerosol exposure from an ultrasonically activated (Harmonic) device. J Am Assoc Gynecol Laparosc. 1998;5(1):29–32.CrossRefPubMed
14.
go back to reference Weld KJ, Dryer S, Ames CD, et al. Analysis of surgical smoke produced by various energy-based instruments and effect on laparoscopic visibility. J Endourol. 2007;21(3):347–51.CrossRefPubMed Weld KJ, Dryer S, Ames CD, et al. Analysis of surgical smoke produced by various energy-based instruments and effect on laparoscopic visibility. J Endourol. 2007;21(3):347–51.CrossRefPubMed
15.
16.
go back to reference Kim FJ, Sehrt D, Pompeo A, et al. Comparison of surgical plume among laparoscopic ultrasonic dissectors using a real-time digital quantitative technology. Surg Endosc. 2012;26:3408–12 (PubMed).PubMedCentralCrossRefPubMed Kim FJ, Sehrt D, Pompeo A, et al. Comparison of surgical plume among laparoscopic ultrasonic dissectors using a real-time digital quantitative technology. Surg Endosc. 2012;26:3408–12 (PubMed).PubMedCentralCrossRefPubMed
17.
go back to reference Kim FJ, Sehrt D, Pompeo A, et al. Laminar and turbulent surgical plume characteristics generated from curved- and straight-blade laparoscopicultrasonic dissectors. Surg Endosc. 2014;28(5):1674–7. doi:10.1007/s00464-013-3369-6 (Epub 2014 Jan 8).CrossRefPubMed Kim FJ, Sehrt D, Pompeo A, et al. Laminar and turbulent surgical plume characteristics generated from curved- and straight-blade laparoscopicultrasonic dissectors. Surg Endosc. 2014;28(5):1674–7. doi:10.​1007/​s00464-013-3369-6 (Epub 2014 Jan 8).CrossRefPubMed
18.
go back to reference Sartori PV, De Fina S, Colombo G, et al. Ligasure versus Ultracision in thyroid surgery: a prospective randomized study. Langenbecks Arch Surg. 2008;393:655–8.CrossRefPubMed Sartori PV, De Fina S, Colombo G, et al. Ligasure versus Ultracision in thyroid surgery: a prospective randomized study. Langenbecks Arch Surg. 2008;393:655–8.CrossRefPubMed
19.
go back to reference Gözen AS, Teber D, Rassweiler JJ. Principles and initial experience of a new device for dissection and hemostasis. Minim Invasive Ther Allied Technol. 2007;16:58–65.CrossRefPubMed Gözen AS, Teber D, Rassweiler JJ. Principles and initial experience of a new device for dissection and hemostasis. Minim Invasive Ther Allied Technol. 2007;16:58–65.CrossRefPubMed
20.
go back to reference Perko Z, Pogorelić Z, Bilan K, et al. Lateral thermal damage to rat abdominal wall after harmonic scalpel application. Surg Endosc. 2006;20:322–4.CrossRefPubMed Perko Z, Pogorelić Z, Bilan K, et al. Lateral thermal damage to rat abdominal wall after harmonic scalpel application. Surg Endosc. 2006;20:322–4.CrossRefPubMed
21.
go back to reference Kadesky KM, Schopf B, Blair GK. Proximity injury by the ultrasonically activated scalpel during dissection. J Pediatr Surg. 1997;32:878–9.CrossRefPubMed Kadesky KM, Schopf B, Blair GK. Proximity injury by the ultrasonically activated scalpel during dissection. J Pediatr Surg. 1997;32:878–9.CrossRefPubMed
22.
go back to reference Amaral JF. Depth of thermal injury: ultracisionally activated scalpel vs. electrosurgery. Surg Endosc. 1995;9:226. Amaral JF. Depth of thermal injury: ultracisionally activated scalpel vs. electrosurgery. Surg Endosc. 1995;9:226.
23.
go back to reference Humes DJ, Ahmed I, Lobo DN. The pedicle effect and direct coupling: delayed thermal injuries to the bile duct after laparoscopic cholecystectomy. Arch Surg. 2010;145:96–8.PubMed Humes DJ, Ahmed I, Lobo DN. The pedicle effect and direct coupling: delayed thermal injuries to the bile duct after laparoscopic cholecystectomy. Arch Surg. 2010;145:96–8.PubMed
24.
go back to reference Diamantis T, Kontos M, Arvelakis A, et al. Comparison of monopolar electrocoagulation, bipolar electrocoagulation, Ultracision, and Ligasure. Surg Today. 2006;36:908–13.CrossRefPubMed Diamantis T, Kontos M, Arvelakis A, et al. Comparison of monopolar electrocoagulation, bipolar electrocoagulation, Ultracision, and Ligasure. Surg Today. 2006;36:908–13.CrossRefPubMed
25.
go back to reference Diamantis T, Gialikaris S, Kontos M, et al. Comparison of safety and efficacy of ultrasonic and bipolar thermal energy: an experimental study. Surg Laparosc Endosc Percutan Tech. 2008;18:384–90.CrossRefPubMed Diamantis T, Gialikaris S, Kontos M, et al. Comparison of safety and efficacy of ultrasonic and bipolar thermal energy: an experimental study. Surg Laparosc Endosc Percutan Tech. 2008;18:384–90.CrossRefPubMed
26.
go back to reference Newcomb WL, Hope WW, Schmelzer TM, et al. Comparison of blood vessel sealing among new electrosurgical and ultrasonic devices. Surg Endosc. 2009;23:90–6.CrossRefPubMed Newcomb WL, Hope WW, Schmelzer TM, et al. Comparison of blood vessel sealing among new electrosurgical and ultrasonic devices. Surg Endosc. 2009;23:90–6.CrossRefPubMed
27.
go back to reference Druzijanić N, Perko Z, Kraljević D, et al. Harmonic scalpel in transanal microsurgery. Hepatogastroenterology. 2008;55:356–8.PubMed Druzijanić N, Perko Z, Kraljević D, et al. Harmonic scalpel in transanal microsurgery. Hepatogastroenterology. 2008;55:356–8.PubMed
28.
go back to reference Smulders JF, de Hingh IH, Stavast J, et al. Exploring new technologies to facilitate laparoscopic surgery: creating intestinal anastomoses without sutures or staples, using a radio-frequency-energy-driven bipolar fusion device. Surg Endosc. 2007;21:2105–9.CrossRefPubMed Smulders JF, de Hingh IH, Stavast J, et al. Exploring new technologies to facilitate laparoscopic surgery: creating intestinal anastomoses without sutures or staples, using a radio-frequency-energy-driven bipolar fusion device. Surg Endosc. 2007;21:2105–9.CrossRefPubMed
29.
go back to reference Elemen L, Yazir Y, Tugay M, et al. LigaSure compared with ligatures and endoclips in experimental appendectomy: how safe is it? Pediatr Surg Int. 2010;26:539–45.CrossRefPubMed Elemen L, Yazir Y, Tugay M, et al. LigaSure compared with ligatures and endoclips in experimental appendectomy: how safe is it? Pediatr Surg Int. 2010;26:539–45.CrossRefPubMed
30.
go back to reference Družijanić N, Pogorelić Z, Perko Z, et al. Comparison of lateral thermal damage of the human peritoneum using monopolar diathermy, Harmonic scalpel and LigaSure. Can J Surg. 2012;55(5):317–21.PubMedCentralCrossRefPubMed Družijanić N, Pogorelić Z, Perko Z, et al. Comparison of lateral thermal damage of the human peritoneum using monopolar diathermy, Harmonic scalpel and LigaSure. Can J Surg. 2012;55(5):317–21.PubMedCentralCrossRefPubMed
31.
go back to reference Lamberton GR, Hsi RS, Jin DH, Lindler TU, et al. Prospective comparison of four laparoscopic vessel ligation devices. J Endourol. 2008;22:2307–12.CrossRefPubMed Lamberton GR, Hsi RS, Jin DH, Lindler TU, et al. Prospective comparison of four laparoscopic vessel ligation devices. J Endourol. 2008;22:2307–12.CrossRefPubMed
34.
go back to reference Sutton PA, Awad S, Perkins AC, et al. Comparison of lateral thermal spread using monopolar and bipolar diathermy, the Harmonic Scalpel and the Ligasure. Br J Surg. 2010;97:428–33.CrossRefPubMed Sutton PA, Awad S, Perkins AC, et al. Comparison of lateral thermal spread using monopolar and bipolar diathermy, the Harmonic Scalpel and the Ligasure. Br J Surg. 2010;97:428–33.CrossRefPubMed
35.
go back to reference Kwok A, Nevell D, Ferrier A, et al. Comparison of tissue injury between laparosonic coagulating shears and electrosurgical scissors in the sheep model. J Am Assoc Gynecol Laparosc. 2001;8:378–84.CrossRefPubMed Kwok A, Nevell D, Ferrier A, et al. Comparison of tissue injury between laparosonic coagulating shears and electrosurgical scissors in the sheep model. J Am Assoc Gynecol Laparosc. 2001;8:378–84.CrossRefPubMed
Metadata
Title
Surgical Efficacy Among Laparoscopic Ultrasonic Dissectors: Are We Advancing Safely? A Review of Literature
Authors
Rajesh Devassy
Sreelatha Gopalakrishnan
Rudy Leon De Wilde
Publication date
01-10-2015
Publisher
Springer India
Published in
The Journal of Obstetrics and Gynecology of India / Issue 5/2015
Print ISSN: 0971-9202
Electronic ISSN: 0975-6434
DOI
https://doi.org/10.1007/s13224-015-0774-x

Other articles of this Issue 5/2015

The Journal of Obstetrics and Gynecology of India 5/2015 Go to the issue