Evaluation of emissivity and temperature profile of laparoscopic ultrasonic devices (blades and passive jaws)

Authors: Fernando J. Kim, David Sehrt, Rodrigo Donalisio da Silva, Diedra Gustafson, Leticia Nogueira, Wilson R. Molina

Published in: Surgical Endoscopy | Issue 5/2015

Login to get access

Abstract

Objectives

We examined the emissivity and temperature profile of passive and active jaws of various laparoscopic ultrasonic devices during cutting, coagulation, and cooling time.

Methods

The Harmonic ACE™ (ACE), Covidien Sonicision™ (SNC), and Olympus SonoSurg™ (SS) were applied using pre-set factory cutting and coagulation settings to Bovine mesentery and Lamb renal veins, respectively. The maximum temperature and cooling time to reach 60 °C were recorded using an infrared Fluke Ti55 thermal imager. Histological examination was evaluated after application of energy.

Results

The ACE, Sonicision, and SonoSurg had emissivity measurements of 0.49 ± 0.01, 0.40 ± 0.00, and 0.39 ± 0.01, respectively. Maximum cutting temperatures were: ACE = 191.1°, SNC = 227.1°, and SNS 184.8° * (*p < 0.001). Maximum coagulation temperatures did not differ significantly among devices (p = 0.490). The cooling time to reach 60 °C after activation were 35.7 s (ACE), 38.7 s (SNC), and 27.4 s* (SS) (*p < 0.001). The cooling time of passive jaws to reach 60 °C after activation were 25.4 s* (ACE), 5.7 s (SNC), and 15.4 s (SS) (*p < 0.001).

Conclusion

Laparoscopic ultrasonic instruments obtain the same cutting and coagulation objectives but in different manners. The Sonicision improves cutting by getting the blade hotter while the SonoSurg has more precise coagulation effects by heating slower. Emissivity values varied among instruments, providing equally varied results. Depending on the purpose of the devices, a certain device may be more appropriate. Based on emissivity, more studies are needed to identify the ideal material that can predictably and effectively perform in clinical settings. Although different blade geometry is apparent between instruments, the jaws are also designed differently between the generations of instruments.
Literature
1.
go back to reference Kim FJ, Chammas MF, Gewehr E et al (2008) Temperature safety profile of laparoscopic devices: Harmonic ACE (ACE), Ligasure V(LV), and plasma trisector (PT). Surg Endosc 22:1464CrossRefPubMed Kim FJ, Chammas MF, Gewehr E et al (2008) Temperature safety profile of laparoscopic devices: Harmonic ACE (ACE), Ligasure V(LV), and plasma trisector (PT). Surg Endosc 22:1464CrossRefPubMed
2.
go back to reference Kadesky KM, Schopf B, Magee JF et al (1997) Proximity injury by the ultrasonically activated scalpel during dissection. J Pediatr Surg 32:878CrossRefPubMed Kadesky KM, Schopf B, Magee JF et al (1997) Proximity injury by the ultrasonically activated scalpel during dissection. J Pediatr Surg 32:878CrossRefPubMed
3.
go back to reference Homayounfar K, Meis J, Jung K et al (2012) Ultrasonic scalpel causes greater depth of soft tissue necrosis compared to monopolar electrocautery at standard power level settings in a pig model. BMC Surg 12:3CrossRefPubMedCentralPubMed Homayounfar K, Meis J, Jung K et al (2012) Ultrasonic scalpel causes greater depth of soft tissue necrosis compared to monopolar electrocautery at standard power level settings in a pig model. BMC Surg 12:3CrossRefPubMedCentralPubMed
4.
go back to reference Awwad JT, Isaacson K (1996) The harmonic scalpel: an intraoperative complication. Obstet Gynecol 88:718CrossRefPubMed Awwad JT, Isaacson K (1996) The harmonic scalpel: an intraoperative complication. Obstet Gynecol 88:718CrossRefPubMed
5.
go back to reference Kim FJ, Sehrt D, Pompeo A et al (2012) Comparison of surgical plume among laparoscopic ultrasonic dissectors using a real-time digital quantitative technology. Surg Endosc 26:3408CrossRefPubMedCentralPubMed Kim FJ, Sehrt D, Pompeo A et al (2012) Comparison of surgical plume among laparoscopic ultrasonic dissectors using a real-time digital quantitative technology. Surg Endosc 26:3408CrossRefPubMedCentralPubMed
6.
go back to reference Ma B (1996) Infrared Radiation, a handbook for applications. Plenum press, New York Ma B (1996) Infrared Radiation, a handbook for applications. Plenum press, New York
7.
go back to reference Cimino WW, Bond LJ (1996) Physics of ultrasonic surgery using tissue fragmentation: Part I. Ultrasound Med Biol 22:89CrossRefPubMed Cimino WW, Bond LJ (1996) Physics of ultrasonic surgery using tissue fragmentation: Part I. Ultrasound Med Biol 22:89CrossRefPubMed
8.
go back to reference Ka Z (2001) Surgical Laparoscopy. Lippincott Williams & Wilkins, Philadelphia Ka Z (2001) Surgical Laparoscopy. Lippincott Williams & Wilkins, Philadelphia
9.
go back to reference Tsirline VB, Lau KN, Swan RZ et al (2013) Evaluation of an innovative, cordless ultrasonic dissector. Surg Innov 20:524CrossRefPubMed Tsirline VB, Lau KN, Swan RZ et al (2013) Evaluation of an innovative, cordless ultrasonic dissector. Surg Innov 20:524CrossRefPubMed
11.
go back to reference Seehofer D, Mogl M, Boas-Knoop S et al (2012) Safety and efficacy of new integrated bipolar and ultrasonic scissors compared to conventional laparoscopic 5-mm sealing and cutting instruments. Surg Endosc 26:2541CrossRefPubMedCentralPubMed Seehofer D, Mogl M, Boas-Knoop S et al (2012) Safety and efficacy of new integrated bipolar and ultrasonic scissors compared to conventional laparoscopic 5-mm sealing and cutting instruments. Surg Endosc 26:2541CrossRefPubMedCentralPubMed
12.
go back to reference Song C, Tang B, Campbell PA et al (2009) Thermal spread and heat absorbance differences between open and laparoscopic surgeries during energized dissections by electrosurgical instruments. Surg Endosc 23:2480CrossRefPubMed Song C, Tang B, Campbell PA et al (2009) Thermal spread and heat absorbance differences between open and laparoscopic surgeries during energized dissections by electrosurgical instruments. Surg Endosc 23:2480CrossRefPubMed
Metadata
Title
Evaluation of emissivity and temperature profile of laparoscopic ultrasonic devices (blades and passive jaws)
Authors
Fernando J. Kim
David Sehrt
Rodrigo Donalisio da Silva
Diedra Gustafson
Leticia Nogueira
Wilson R. Molina
Publication date
01-05-2015
Publisher
Springer US
Published in
Surgical Endoscopy / Issue 5/2015
Print ISSN: 0930-2794
Electronic ISSN: 1432-2218
DOI
https://doi.org/10.1007/s00464-014-3787-0

Other articles of this Issue 5/2015

Surgical Endoscopy 5/2015 Go to the issue