01-12-2014 | original article

Stress distribution and bone–implant interface behavior in pick-up implant impression

A finite element model analysis

Authors: Piero Simeone, DDS, MSc, CDT, Pier Paolo Valentini, PhD, Roberto Pizzoferrato, PhD, Massimo Marinelli, PhD

Published in: international journal of stomatology & occlusion medicine | Issue 4/2014

Login to get access

Abstract

Purpose

The purpose of this research was to study the stress distribution at the bone–implant contact in the pick-up implant impression technique. The mechanical properties and the curing dynamics of Pattern Resin LS were previously investigated.

Materials and methods

Measurements of the elastic modules of Pattern Resin LS were carried out after complete curing by standard stress–strain tests and a three-point bending technique. A specific laboratory set-up was implemented to monitor the curing dynamics and the linear contraction by using optical imaging methods. A finite element model (FEM) was created and a complete assembly was made with 33 solid bodies representing the main components of the system including the external portion of cortical bone, the internal portion of trabecular bone, six implants, six copings, one bridging structure (modular individual tray), six modules, and six pattern resin layers to fill up the interstitial spaces. Isotropic characteristics for cortical and trabecular bone were incorporated. The stress distribution was measured at the total body of the bone–implant interface, at the external cortical bone surface, at the coping–implant connection area, and at the tray–coping junction. Parametric analysis was also done to study the behavior of two types of internal connection in the impression phase: type A consisting of a 0.4 mm tube-in-tube connection with a triangular head by three cams (Cam-Log Biotechnologies, Germany); type B consisting of a 0.5-mm internal hexagonal connection (Biomet 3i, USA).

Results

The measured values of Young’s modulus and the flexural modulus of the resin were 1.58 ± 0.12 GPa and 2.07 ± 0.24 GPa, respectively. The final volume contraction was found to be strongly dependent on the powder/liquid ratio; specifically, the tests suggested that the highest value of the powder/liquid ratio (2:1) was the most favorable one. In all cases, however, the Pattern Resin behavior was within the linear elastic range, without any permanent deformation or detachment. At individual tray–Pattern Resin interfaces, a stress of about 20.0 MPa was produced for each component; FEM analysis showed that the stress level at the implant–transfer copings connection was higher than that at the bone–implant interface. The trabecular and cortical bone showed the lowest level of stress concentration. B-type transfer showed a higher stress at the transfer coping–implant connection; A-type transfer coping allowed for a higher deformation area, increasing the risk of detachment.

Conclusions

The mechanical properties of Pattern Resin LS seem to be appropriate for use in the pick-up impression technique thanks to the low-volume contraction, the capabilities to dampen misfits, and permanent deformation effects. The shape of the copings affects the values of the stress both in the transfer coping–implant connection zones and at resin–transfer coping interfaces. A thicker clearance of the module–tray connection helps in reducing the mechanical stress. The FEM analysis showed that the removal procedure of the pick-up implant impression tray does not produce any irreversible damage on the cortical and trabecular bone structure. Since the yield threshold was high, the stress distribution at the bone–implant–coping configurations was largely within the safety range for the multiple internal-connection implants, in combination with a modular individual impression tray.
Literature
1.
go back to reference Vigolo P, Millstein PL. Evaluation of master cast technique for multiple abutment implant prostheses. Int J Oral Maxillofac Implant. 1993;8:439–45. Vigolo P, Millstein PL. Evaluation of master cast technique for multiple abutment implant prostheses. Int J Oral Maxillofac Implant. 1993;8:439–45.
2.
go back to reference Barrett MG, de Rijk WG, Burges JO. The accuracy of six impression techniques for osseointegrate implants. J Prosthodont. 1993;2:75–82.CrossRefPubMed Barrett MG, de Rijk WG, Burges JO. The accuracy of six impression techniques for osseointegrate implants. J Prosthodont. 1993;2:75–82.CrossRefPubMed
3.
go back to reference Hussaini S, Wung T. One clinical visit for a multiple implant restoration master cast fabrication. J Prosthet Dent. 1997;78:550–3.CrossRefPubMed Hussaini S, Wung T. One clinical visit for a multiple implant restoration master cast fabrication. J Prosthet Dent. 1997;78:550–3.CrossRefPubMed
4.
5.
go back to reference Carr AB. Comparison of impression techniques for a five-implants mandibular model. Int J Oral Maxillofac Implants. 1991;6:448–55.PubMed Carr AB. Comparison of impression techniques for a five-implants mandibular model. Int J Oral Maxillofac Implants. 1991;6:448–55.PubMed
6.
go back to reference Kim S, Nicoholls JI, Han CH, Lee KW. Displacement of implant component from impression to definitive cast. Int J Maxillofac Implants. 2006;21:747–55. Kim S, Nicoholls JI, Han CH, Lee KW. Displacement of implant component from impression to definitive cast. Int J Maxillofac Implants. 2006;21:747–55.
7.
go back to reference Simeone P, Valentini PP, Pizzoferrato R, Scudieri F. Dimensional accuracy of pick-up implant impression: an in vitro comparison of modular versus standard custom trays. Int J Oral Maxillofac Implants. 2011;26:538–46.PubMed Simeone P, Valentini PP, Pizzoferrato R, Scudieri F. Dimensional accuracy of pick-up implant impression: an in vitro comparison of modular versus standard custom trays. Int J Oral Maxillofac Implants. 2011;26:538–46.PubMed
8.
go back to reference Takahashi J, Kitahara K, Teraoka F, Kubo F. Resin pattern material with low polymerization shrinkage. Int J Prosthodont. 1999;12(4):325–9.PubMed Takahashi J, Kitahara K, Teraoka F, Kubo F. Resin pattern material with low polymerization shrinkage. Int J Prosthodont. 1999;12(4):325–9.PubMed
9.
go back to reference McDonnell T, Houston F, Byrne D, Gorman C, Claffey N. The effect of time lapse on the accuracy of two acrylic resins used to assemble an implant framework for soldering. J Prosthet Dent. 2004;91(6):538–40.CrossRefPubMed McDonnell T, Houston F, Byrne D, Gorman C, Claffey N. The effect of time lapse on the accuracy of two acrylic resins used to assemble an implant framework for soldering. J Prosthet Dent. 2004;91(6):538–40.CrossRefPubMed
10.
go back to reference Sato Y, Teixeira ER, Tsuga K, Shindoi N. The effectiveness of a new algorithm on a three-dimensional finite element model construction of bone trabeculae in implant biomechanics. J Oral Rehabil. 1999;26:640–3.CrossRefPubMed Sato Y, Teixeira ER, Tsuga K, Shindoi N. The effectiveness of a new algorithm on a three-dimensional finite element model construction of bone trabeculae in implant biomechanics. J Oral Rehabil. 1999;26:640–3.CrossRefPubMed
11.
go back to reference O’Mahony AM, Williams JL, Katz JO, Spencer P. Anisotropic elastic properties of cancellous bone from a human edentulous mandible. Clin Oral Implants Res. 2000;11:415–21.CrossRefPubMed O’Mahony AM, Williams JL, Katz JO, Spencer P. Anisotropic elastic properties of cancellous bone from a human edentulous mandible. Clin Oral Implants Res. 2000;11:415–21.CrossRefPubMed
12.
go back to reference Al-Sukhum J, Kelleway J, Helenius M. Devolopment of a three-dimensional finite element model of a human mandible containing endosseous dental implants. I. Mathematical validation and experimental verification. J Biomed Mater Res A. 2007;80:234–46.CrossRef Al-Sukhum J, Kelleway J, Helenius M. Devolopment of a three-dimensional finite element model of a human mandible containing endosseous dental implants. I. Mathematical validation and experimental verification. J Biomed Mater Res A. 2007;80:234–46.CrossRef
13.
go back to reference O’Mahony AM, Williams JL, Spencer P. Anisotropic elasticity of cortical and cancellous bone in the posterior mandible increases per-implant stress and strain under oblique loading. Clin Oral Implants Res. 2001;12:648–57.CrossRefPubMed O’Mahony AM, Williams JL, Spencer P. Anisotropic elasticity of cortical and cancellous bone in the posterior mandible increases per-implant stress and strain under oblique loading. Clin Oral Implants Res. 2001;12:648–57.CrossRefPubMed
14.
go back to reference Schwartz-Dabney CL, Dechow PC. Edentulation alters material properties of cortical bone in the human mandible. J Dent Res. 2002;81:613–7.CrossRefPubMed Schwartz-Dabney CL, Dechow PC. Edentulation alters material properties of cortical bone in the human mandible. J Dent Res. 2002;81:613–7.CrossRefPubMed
15.
go back to reference Himmlová L, Dostálová T, Kácovský A, Konvicˇková S. Influence of implant length and diameter on stress distribution: a finite element analysis. J Prosthet Dent. 2004;91(1):20–5.CrossRefPubMed Himmlová L, Dostálová T, Kácovský A, Konvicˇková S. Influence of implant length and diameter on stress distribution: a finite element analysis. J Prosthet Dent. 2004;91(1):20–5.CrossRefPubMed
16.
go back to reference Ekici B. Numerical analysis of a dental implant system in three-dimensions. Adv Eng Softw. 2002;33:109–13.CrossRef Ekici B. Numerical analysis of a dental implant system in three-dimensions. Adv Eng Softw. 2002;33:109–13.CrossRef
17.
go back to reference Meng RT, Jr., Latta MA. Physical properties of four acrylic denture based resins. J Contemp Dent Pract. 2005;6:1–15. Meng RT, Jr., Latta MA. Physical properties of four acrylic denture based resins. J Contemp Dent Pract. 2005;6:1–15.
18.
go back to reference Barbosa DB, Freitas de Souza R, Pero AC, Marra J, Compagnoni MA. Flexural strength of acrylic resins polymerized by different cycles. J Appl Oral Sci. 2007;15:424–8.CrossRefPubMedCentralPubMed Barbosa DB, Freitas de Souza R, Pero AC, Marra J, Compagnoni MA. Flexural strength of acrylic resins polymerized by different cycles. J Appl Oral Sci. 2007;15:424–8.CrossRefPubMedCentralPubMed
19.
go back to reference Vallittu PK. Flexural properties of acrylic resin polymers reinforced with unidirectional and woven glass fibers. J Prosthet Dent. 1999;81:318–26.CrossRefPubMed Vallittu PK. Flexural properties of acrylic resin polymers reinforced with unidirectional and woven glass fibers. J Prosthet Dent. 1999;81:318–26.CrossRefPubMed
21.
go back to reference Kwon T-Y, Han J-Y, Chun J-N, Son JS, Kyo-Han Kim K-H. Effects of prepolymerized particle size and polymerization kinetics on volumetric shrinkage of dental modeling resins. BioMed Res Int. 2014;2014:914739.PubMedCentralPubMed Kwon T-Y, Han J-Y, Chun J-N, Son JS, Kyo-Han Kim K-H. Effects of prepolymerized particle size and polymerization kinetics on volumetric shrinkage of dental modeling resins. BioMed Res Int. 2014;2014:914739.PubMedCentralPubMed
22.
go back to reference Gibbs SB, Versluis V, Tantbirojn D, Ahuja S. Comparison of polymerization shrinkage of Pattern Resins. J Prosthet Dent. 2014;112(2):293–8.CrossRefPubMed Gibbs SB, Versluis V, Tantbirojn D, Ahuja S. Comparison of polymerization shrinkage of Pattern Resins. J Prosthet Dent. 2014;112(2):293–8.CrossRefPubMed
23.
go back to reference Rajagopal P, Chitre V, Aras MA. A comparison of the accuracy of patterns processed from an inlay casting wax, an auto-polymerized resin and a light-cured resin pattern material. Indian J Dent Res. 2012;23:152–6.CrossRefPubMed Rajagopal P, Chitre V, Aras MA. A comparison of the accuracy of patterns processed from an inlay casting wax, an auto-polymerized resin and a light-cured resin pattern material. Indian J Dent Res. 2012;23:152–6.CrossRefPubMed
24.
go back to reference Cerqueira NM, Ozcan M, Gonçalves M, da Rocha DM, Vasconcellos DK, Bottino MA, Yener-Salihog˘lu E. A strain gauge analysis of microstrain induced by various splinting methods and acrylic resin types for implant impressions. Int J Oral Maxillofac Implants. 2012;27(2):341–5.PubMed Cerqueira NM, Ozcan M, Gonçalves M, da Rocha DM, Vasconcellos DK, Bottino MA, Yener-Salihog˘lu E. A strain gauge analysis of microstrain induced by various splinting methods and acrylic resin types for implant impressions. Int J Oral Maxillofac Implants. 2012;27(2):341–5.PubMed
Metadata
Title
Stress distribution and bone–implant interface behavior in pick-up implant impression
A finite element model analysis
Authors
Piero Simeone, DDS, MSc, CDT
Pier Paolo Valentini, PhD
Roberto Pizzoferrato, PhD
Massimo Marinelli, PhD
Publication date
01-12-2014
Publisher
Springer Vienna
Published in
international journal of stomatology & occlusion medicine / Issue 4/2014
Print ISSN: 1867-2221
Electronic ISSN: 1867-223X
DOI
https://doi.org/10.1007/s12548-014-0114-2

Other articles of this Issue 4/2014

international journal of stomatology & occlusion medicine 4/2014 Go to the issue