Open Access 01-04-2015 | Review Article

Managing patients with advisory defibrillator leads: what can we learn from published data?

Authors: F. A. Bracke, B. M. van Gelder

Published in: Netherlands Heart Journal | Issue 4/2015

Login to get access

Abstract

Defibrillator lead advisories stir a lot of emotions, both with patients and physicians, and this may influence lead management. We reviewed the literature for a more evidence-based approach to this issue.
From the complications of two of the current advisory leads, the Medtronic Sprint Fidelis and St. Jude Riata leads, and the consequences of possible interventions, we can conclude that a restrained approach to premature replacement is appropriate. It may be opportune to replace the leads during a scheduled generator replacement in case of a higher electrical failure rate, in order to prevent future premature interventions.
We found no support to extract non-functional advisory leads. In contrast, extraction is often more demanding than anticipated, and the risk substantially exceeds that of simply abandoning the leads.
Literature
1.
go back to reference Kay GN, Brinker JA, Kawanishi DT, et al. Risks of spontaneous injury and extraction of an active fixation pacemaker lead: report of the Accufix Multicenter Clinical Study and Worldwide Registry. Circulation. 1999;100:2344–52.CrossRefPubMed Kay GN, Brinker JA, Kawanishi DT, et al. Risks of spontaneous injury and extraction of an active fixation pacemaker lead: report of the Accufix Multicenter Clinical Study and Worldwide Registry. Circulation. 1999;100:2344–52.CrossRefPubMed
2.
go back to reference Maisel WH, Hauser RG, Hammill SC, et al. Recommendations from the Heart Rhythm Society task force on lead performance policies and guidelines: developed in collaboration with the American College of Cardiology and the American Heart Association. Heart Rhythm. 2009;6:869–85.CrossRefPubMed Maisel WH, Hauser RG, Hammill SC, et al. Recommendations from the Heart Rhythm Society task force on lead performance policies and guidelines: developed in collaboration with the American College of Cardiology and the American Heart Association. Heart Rhythm. 2009;6:869–85.CrossRefPubMed
3.
go back to reference Abdelhadi RH, Saba SF, Ellis CR, et al. Independent multicenter study of Riata and Riata ST implantable cardioverter-defibrillator leads. Heart Rhythm. 2013;10:361–5.CrossRefPubMed Abdelhadi RH, Saba SF, Ellis CR, et al. Independent multicenter study of Riata and Riata ST implantable cardioverter-defibrillator leads. Heart Rhythm. 2013;10:361–5.CrossRefPubMed
4.
go back to reference Hayes D, Freedman R, Curtis AB, et al. Prevalence of externalized conductors in Riata and Riata ST silicone leads: Results from the prospective, multicenter Riata Lead Evaluation Study. Heart Rhythm. 2013;10:1778–82.CrossRefPubMed Hayes D, Freedman R, Curtis AB, et al. Prevalence of externalized conductors in Riata and Riata ST silicone leads: Results from the prospective, multicenter Riata Lead Evaluation Study. Heart Rhythm. 2013;10:1778–82.CrossRefPubMed
5.
go back to reference Parkash R, Exner D, Champagne J, et al. Failure rate of the Riata lead under advisory: a report from the CHRS Device Committee. Heart Rhythm. 2013;10:692–5.CrossRefPubMed Parkash R, Exner D, Champagne J, et al. Failure rate of the Riata lead under advisory: a report from the CHRS Device Committee. Heart Rhythm. 2013;10:692–5.CrossRefPubMed
6.
go back to reference Barrett CD, Danik SB. The Riata lead recall—more data, but do we now know what we need to know? Heart Rhythm. 2013;10:1783–4.CrossRefPubMed Barrett CD, Danik SB. The Riata lead recall—more data, but do we now know what we need to know? Heart Rhythm. 2013;10:1783–4.CrossRefPubMed
7.
go back to reference Kleemann T, Becker T, Doenges K, et al. Annual rate of transvenous defibrillation lead defects in implantable cardioverter-defibrillators over a period of >10 years. Circulation. 2007;115:2474–80.CrossRefPubMed Kleemann T, Becker T, Doenges K, et al. Annual rate of transvenous defibrillation lead defects in implantable cardioverter-defibrillators over a period of >10 years. Circulation. 2007;115:2474–80.CrossRefPubMed
8.
go back to reference Demirel F, Adiyaman A, Delnoy PPHM, Smit JJJ, Ramdat Misier AR, Elvan A. Mechanical and electrical dysfunction of Riata implantable cardioverter-defibrillator leads. Europace. 2014;16:1787–94.CrossRefPubMed Demirel F, Adiyaman A, Delnoy PPHM, Smit JJJ, Ramdat Misier AR, Elvan A. Mechanical and electrical dysfunction of Riata implantable cardioverter-defibrillator leads. Europace. 2014;16:1787–94.CrossRefPubMed
9.
go back to reference Theuns DAMJ, Elvan A, Voogt W de, Cock CC de, Erven L van, Meine M. Prevalence and presentation of externalized conductors and electrical abnormalities in Riata defibrillator leads after fluoroscopic screening: report from The Netherlands Heart Rhythm Association Device Advisory Committee. Circ Arrhythm Electrophysiol. 2012;5:1059–63.CrossRefPubMed Theuns DAMJ, Elvan A, Voogt W de, Cock CC de, Erven L van, Meine M. Prevalence and presentation of externalized conductors and electrical abnormalities in Riata defibrillator leads after fluoroscopic screening: report from The Netherlands Heart Rhythm Association Device Advisory Committee. Circ Arrhythm Electrophysiol. 2012;5:1059–63.CrossRefPubMed
11.
go back to reference Birnie DH, Parkash R, Exner DV, et al. Clinical predictors of Fidelis lead failure: report from the Canadian Heart Rhythm Society Device Committee. Circulation. 2012;125:1217–25.CrossRefPubMed Birnie DH, Parkash R, Exner DV, et al. Clinical predictors of Fidelis lead failure: report from the Canadian Heart Rhythm Society Device Committee. Circulation. 2012;125:1217–25.CrossRefPubMed
12.
go back to reference Hauser RG, Abdelhadi R, McGriff D, Retel LK. Deaths caused by the failure of Riata and Riata ST implantable cardioverter-defibrillator leads. Heart Rhythm. 2012;9:1227–35.CrossRefPubMed Hauser RG, Abdelhadi R, McGriff D, Retel LK. Deaths caused by the failure of Riata and Riata ST implantable cardioverter-defibrillator leads. Heart Rhythm. 2012;9:1227–35.CrossRefPubMed
13.
go back to reference Morrison TB, Friedman PA, Kallinen LM, et al. Impact of implanted recalled sprint Fidelis lead on patient mortality. J Am Coll Cardiol. 2011;58:278–83.CrossRefPubMed Morrison TB, Friedman PA, Kallinen LM, et al. Impact of implanted recalled sprint Fidelis lead on patient mortality. J Am Coll Cardiol. 2011;58:278–83.CrossRefPubMed
14.
go back to reference van Welsenes GH, van Rees JB, Borleffs CJW, et al. Long-term follow-up of primary and secondary prevention implantable cardioverter defibrillator patients. Europace. 2011;13:389–94.CrossRefPubMed van Welsenes GH, van Rees JB, Borleffs CJW, et al. Long-term follow-up of primary and secondary prevention implantable cardioverter defibrillator patients. Europace. 2011;13:389–94.CrossRefPubMed
15.
go back to reference Moss AJ, Zareba W, Hall WJ, et al. Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction. N Engl J Med. 2002;346:877–83.CrossRefPubMed Moss AJ, Zareba W, Hall WJ, et al. Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction. N Engl J Med. 2002;346:877–83.CrossRefPubMed
16.
go back to reference Bardy GH, Lee KL, Mark DB, et al. Amiodarone or an implantable cardioverter-defibrillator for congestive heart failure. N Engl J Med. 2005;352:225–37.CrossRefPubMed Bardy GH, Lee KL, Mark DB, et al. Amiodarone or an implantable cardioverter-defibrillator for congestive heart failure. N Engl J Med. 2005;352:225–37.CrossRefPubMed
17.
go back to reference Bashir J, Cowan S, Raymakers A, et al. A cost-effectiveness analysis of a proactive management strategy for the Sprint Fidelis recall: a probabilistic decision analysis model. Heart Rhythm. 2013;10:1761–7.CrossRefPubMed Bashir J, Cowan S, Raymakers A, et al. A cost-effectiveness analysis of a proactive management strategy for the Sprint Fidelis recall: a probabilistic decision analysis model. Heart Rhythm. 2013;10:1761–7.CrossRefPubMed
18.
go back to reference Wilkoff BL, Williamson BD, Stern RS, et al. Strategic programming of detection and therapy parameters in implantable cardioverter-defibrillators reduces shocks in primary prevention patients: results from the PREPARE (Primary Prevention Parameters Evaluation) study. J Am Coll Cardiol. 2008;52:541–50.CrossRefPubMed Wilkoff BL, Williamson BD, Stern RS, et al. Strategic programming of detection and therapy parameters in implantable cardioverter-defibrillators reduces shocks in primary prevention patients: results from the PREPARE (Primary Prevention Parameters Evaluation) study. J Am Coll Cardiol. 2008;52:541–50.CrossRefPubMed
19.
go back to reference Moss AJ, Schuger C, Beck CA, et al. Reduction in inappropriate therapy and mortality through ICD programming. N Engl J Med. 2012;367:2275–83.CrossRefPubMed Moss AJ, Schuger C, Beck CA, et al. Reduction in inappropriate therapy and mortality through ICD programming. N Engl J Med. 2012;367:2275–83.CrossRefPubMed
20.
go back to reference Wijers SC, Kolk BYM van der, Tuinenburg AE, Doevendans PAF, Vos MA, Meine M. Implementation of guidelines for implantable cardioverter-defibrillator therapy in clinical practice: which patients do benefit? Neth Heart J 2013;21:274–83.CrossRefPubMedCentralPubMed Wijers SC, Kolk BYM van der, Tuinenburg AE, Doevendans PAF, Vos MA, Meine M. Implementation of guidelines for implantable cardioverter-defibrillator therapy in clinical practice: which patients do benefit? Neth Heart J 2013;21:274–83.CrossRefPubMedCentralPubMed
21.
go back to reference Krahn AD, Champagne J, Healey JS, et al. Outcome of the Fidelis implantable cardioverter-defibrillator lead advisory: a report from the Canadian Heart Rhythm Society Device Advisory Committee. Heart Rhythm. 2008;5:639–42.CrossRefPubMed Krahn AD, Champagne J, Healey JS, et al. Outcome of the Fidelis implantable cardioverter-defibrillator lead advisory: a report from the Canadian Heart Rhythm Society Device Advisory Committee. Heart Rhythm. 2008;5:639–42.CrossRefPubMed
22.
go back to reference Sung RK, Massie BM, Varosy PD, et al. Long-term electrical survival analysis of Riata and Riata ST silicone leads: National Veterans Affairs experience. Heart Rhythm. 2012;9:1954–61.CrossRefPubMed Sung RK, Massie BM, Varosy PD, et al. Long-term electrical survival analysis of Riata and Riata ST silicone leads: National Veterans Affairs experience. Heart Rhythm. 2012;9:1954–61.CrossRefPubMed
23.
go back to reference Poole JE, Johnson GW, Hellkamp AS, et al. Prognostic importance of defibrillator shocks in patients with heart failure. N Engl J Med. 2008;359:1009–17.CrossRefPubMedCentralPubMed Poole JE, Johnson GW, Hellkamp AS, et al. Prognostic importance of defibrillator shocks in patients with heart failure. N Engl J Med. 2008;359:1009–17.CrossRefPubMedCentralPubMed
24.
go back to reference Pedersen SS, Hoogwegt MT, Jordaens L, Theuns DAMJ. Procedure- and device-related complications and psychological morbidity in implantable cardioverter defibrillator patients. Int J Cardiol. 2013;168:606–7.CrossRefPubMed Pedersen SS, Hoogwegt MT, Jordaens L, Theuns DAMJ. Procedure- and device-related complications and psychological morbidity in implantable cardioverter defibrillator patients. Int J Cardiol. 2013;168:606–7.CrossRefPubMed
25.
go back to reference Keren A, Sears SF, Nery P, et al. Psychological adjustment in ICD patients living with advisory fidelis leads. J Cardiovasc Electrophysiol. 2011;22:57–63.CrossRefPubMed Keren A, Sears SF, Nery P, et al. Psychological adjustment in ICD patients living with advisory fidelis leads. J Cardiovasc Electrophysiol. 2011;22:57–63.CrossRefPubMed
26.
go back to reference Magyar-Russell G, Thombs BD, Cai JX, et al. The prevalence of anxiety and depression in adults with implantable cardioverter defibrillators: a systematic review. J Psychosom Res. 2011;71:223–31.CrossRefPubMed Magyar-Russell G, Thombs BD, Cai JX, et al. The prevalence of anxiety and depression in adults with implantable cardioverter defibrillators: a systematic review. J Psychosom Res. 2011;71:223–31.CrossRefPubMed
27.
go back to reference Kallinen LM, Hauser RG, Tang C, et al. Lead integrity alert algorithm decreases inappropriate shocks in patients who have Sprint Fidelis pace-sense conductor fractures. Heart Rhythm. 2010;7:1048–55.CrossRefPubMed Kallinen LM, Hauser RG, Tang C, et al. Lead integrity alert algorithm decreases inappropriate shocks in patients who have Sprint Fidelis pace-sense conductor fractures. Heart Rhythm. 2010;7:1048–55.CrossRefPubMed
28.
go back to reference Swerdlow CD, Gunderson BD, Ousdigian KT, Abeyratne A, Sachanandani H, Ellenbogen KA. Downloadable software algorithm reduces inappropriate shocks caused by implantable cardioverter-defibrillator lead fractures: a prospective study. Circulation. 2010;122:1449–55.CrossRefPubMed Swerdlow CD, Gunderson BD, Ousdigian KT, Abeyratne A, Sachanandani H, Ellenbogen KA. Downloadable software algorithm reduces inappropriate shocks caused by implantable cardioverter-defibrillator lead fractures: a prospective study. Circulation. 2010;122:1449–55.CrossRefPubMed
29.
go back to reference Hauser RG, Hayes DL. Increasing hazard of Sprint Fidelis implantable cardioverter-defibrillator lead failure. Heart Rhythm. 2009;6:605–10.CrossRefPubMed Hauser RG, Hayes DL. Increasing hazard of Sprint Fidelis implantable cardioverter-defibrillator lead failure. Heart Rhythm. 2009;6:605–10.CrossRefPubMed
30.
go back to reference Johansen JB, Jørgensen OD, Møller M, Arnsbo P, Mortensen PT, Nielsen JC. Infection after pacemaker implantation: infection rates and risk factors associated with infection in a population-based cohort study of 46299 consecutive patients. Eur Heart J. 2011;32:991–8.CrossRefPubMedCentralPubMed Johansen JB, Jørgensen OD, Møller M, Arnsbo P, Mortensen PT, Nielsen JC. Infection after pacemaker implantation: infection rates and risk factors associated with infection in a population-based cohort study of 46299 consecutive patients. Eur Heart J. 2011;32:991–8.CrossRefPubMedCentralPubMed
31.
go back to reference Gould PA, Krahn AD, Canadian Heart Rhythm Society Working Group on Device Advisories. Complications associated with implantable cardioverter-defibrillator replacement in response to device advisories. JAMA. 2006;295:1907–11.CrossRefPubMed Gould PA, Krahn AD, Canadian Heart Rhythm Society Working Group on Device Advisories. Complications associated with implantable cardioverter-defibrillator replacement in response to device advisories. JAMA. 2006;295:1907–11.CrossRefPubMed
32.
go back to reference Costea A, Rardon DP, Padanilam BJ, Fogel RI, Prystowsky EN. Complications associated with generator replacement in response to device advisories. J Cardiovasc Electrophysiol. 2008;19:266–9.CrossRefPubMed Costea A, Rardon DP, Padanilam BJ, Fogel RI, Prystowsky EN. Complications associated with generator replacement in response to device advisories. J Cardiovasc Electrophysiol. 2008;19:266–9.CrossRefPubMed
33.
go back to reference Poole JE, Gleva MJ, Mela T, et al. Complication rates associated with pacemaker or implantable cardioverter-defibrillator generator replacements and upgrade procedures: results from the REPLACE registry. Circulation. 2010;122:1553–61.CrossRefPubMed Poole JE, Gleva MJ, Mela T, et al. Complication rates associated with pacemaker or implantable cardioverter-defibrillator generator replacements and upgrade procedures: results from the REPLACE registry. Circulation. 2010;122:1553–61.CrossRefPubMed
34.
go back to reference Eckstein J, Koller MT, Zabel M, et al. Necessity for surgical revision of defibrillator leads implanted long-term: causes and management. Circulation. 2008;117:2727–33.CrossRefPubMed Eckstein J, Koller MT, Zabel M, et al. Necessity for surgical revision of defibrillator leads implanted long-term: causes and management. Circulation. 2008;117:2727–33.CrossRefPubMed
35.
go back to reference Olde Nordkamp LRA, Knops RE, Bardy GH, et al. Rationale and design of the PRAETORIAN trial: a Prospective, RAndomizEd comparison of subcuTaneOus and tRansvenous ImplANtable cardioverter-defibrillator therapy. Am Heart J. 2012;163:753–60.e2.CrossRefPubMed Olde Nordkamp LRA, Knops RE, Bardy GH, et al. Rationale and design of the PRAETORIAN trial: a Prospective, RAndomizEd comparison of subcuTaneOus and tRansvenous ImplANtable cardioverter-defibrillator therapy. Am Heart J. 2012;163:753–60.e2.CrossRefPubMed
36.
go back to reference Wilkoff BL, Love CJ, Byrd CL, et al. Transvenous lead extraction: Heart Rhythm Society expert consensus on facilities, training, indications, and patient management. Heart Rhythm. 2009;6:1085–104.CrossRefPubMed Wilkoff BL, Love CJ, Byrd CL, et al. Transvenous lead extraction: Heart Rhythm Society expert consensus on facilities, training, indications, and patient management. Heart Rhythm. 2009;6:1085–104.CrossRefPubMed
37.
go back to reference Maytin M, Wilkoff BL, Brunner M, et al. Multicenter experience with extraction of the Riata/Riata ST ICD lead. Heart Rhythm. 2014;11:1613–8.CrossRefPubMed Maytin M, Wilkoff BL, Brunner M, et al. Multicenter experience with extraction of the Riata/Riata ST ICD lead. Heart Rhythm. 2014;11:1613–8.CrossRefPubMed
38.
go back to reference Bohm A, Pinter A, Duray G, et al. Complications due to abandoned noninfected pacemaker leads. Pacing Clin Electrophysiol. 2001;24:1721–4.CrossRefPubMed Bohm A, Pinter A, Duray G, et al. Complications due to abandoned noninfected pacemaker leads. Pacing Clin Electrophysiol. 2001;24:1721–4.CrossRefPubMed
39.
go back to reference Suga C, Hayes D, Hyberger L, Lloyd M. Is there an adverse outcome from abandoned pacing leads? J Interv Card Electrophysiol. 2000;4:493–9.CrossRefPubMed Suga C, Hayes D, Hyberger L, Lloyd M. Is there an adverse outcome from abandoned pacing leads? J Interv Card Electrophysiol. 2000;4:493–9.CrossRefPubMed
40.
go back to reference Glikson M, Suleiman M, Luria DM, et al. Do abandoned leads pose risk to implantable cardioverter-defibrillator patients? Heart Rhythm. 2009;6:65–8.CrossRefPubMed Glikson M, Suleiman M, Luria DM, et al. Do abandoned leads pose risk to implantable cardioverter-defibrillator patients? Heart Rhythm. 2009;6:65–8.CrossRefPubMed
41.
go back to reference Bode F, Himmel F, Reppel M, Mortensen K, Schunkert H, Wiegand UKH. Should all dysfunctional high-voltage leads be extracted? Results of a single-centre long-term registry. Europace. 2012;14:1764–70.CrossRefPubMed Bode F, Himmel F, Reppel M, Mortensen K, Schunkert H, Wiegand UKH. Should all dysfunctional high-voltage leads be extracted? Results of a single-centre long-term registry. Europace. 2012;14:1764–70.CrossRefPubMed
42.
go back to reference Amelot M, Foucault A, Scanu P, et al. Comparison of outcomes in patients with abandoned versus extracted implantable cardioverter defibrillator leads. Arch Cardiovasc Dis. 2011;104:572–7.CrossRefPubMed Amelot M, Foucault A, Scanu P, et al. Comparison of outcomes in patients with abandoned versus extracted implantable cardioverter defibrillator leads. Arch Cardiovasc Dis. 2011;104:572–7.CrossRefPubMed
43.
44.
go back to reference Cock C de, Vinkers M, Campe L Van, Verhorst P, Visser C. Long-term outcome of patients with multiple (> or = 3) noninfected transvenous leads: a clinical and echocardiographic study. Pacing Clin Electrophysiol. 2000;23:423–6.CrossRefPubMed Cock C de, Vinkers M, Campe L Van, Verhorst P, Visser C. Long-term outcome of patients with multiple (> or = 3) noninfected transvenous leads: a clinical and echocardiographic study. Pacing Clin Electrophysiol. 2000;23:423–6.CrossRefPubMed
45.
go back to reference Ricciardi D, La Meir M, Asmundis C de, Brugada P. A case of in vivo thrombogenicity of an externalized Riata ST lead. Europace. 2013;15:428.CrossRefPubMed Ricciardi D, La Meir M, Asmundis C de, Brugada P. A case of in vivo thrombogenicity of an externalized Riata ST lead. Europace. 2013;15:428.CrossRefPubMed
46.
go back to reference Goyal SK, Ellis CR, Rottman JN, Whalen SP. Lead thrombi associated with externalized cables on Riata ICD leads: a case series. J Cardiovasc Electrophysiol. 2013;24:1047–50.CrossRefPubMed Goyal SK, Ellis CR, Rottman JN, Whalen SP. Lead thrombi associated with externalized cables on Riata ICD leads: a case series. J Cardiovasc Electrophysiol. 2013;24:1047–50.CrossRefPubMed
47.
go back to reference Pfeiffer D, Jung W, Fehske W, et al. Complications of pacemaker-defibrillator devices: diagnosis and management. Am Heart J. 1994;127:1073–80.CrossRefPubMed Pfeiffer D, Jung W, Fehske W, et al. Complications of pacemaker-defibrillator devices: diagnosis and management. Am Heart J. 1994;127:1073–80.CrossRefPubMed
48.
go back to reference Venkataraman G, Hayes DL, Strickberger SA. Does the risk-benefit analysis favor the extraction of failed, sterile pacemaker and defibrillator leads? J Cardiovasc Electrophysiol. 2009;20:1413–5.CrossRefPubMed Venkataraman G, Hayes DL, Strickberger SA. Does the risk-benefit analysis favor the extraction of failed, sterile pacemaker and defibrillator leads? J Cardiovasc Electrophysiol. 2009;20:1413–5.CrossRefPubMed
49.
go back to reference Mehrotra AK, Knight BP, Smelley MP, et al. Medtronic Sprint Fidelis lead recall: determining the initial 5-year management cost to Medicare. Heart Rhythm. 2011;8:1192–7.CrossRefPubMed Mehrotra AK, Knight BP, Smelley MP, et al. Medtronic Sprint Fidelis lead recall: determining the initial 5-year management cost to Medicare. Heart Rhythm. 2011;8:1192–7.CrossRefPubMed
50.
go back to reference Maytin M, Love CJ, Fischer A, et al. Multicenter experience with extraction of the Sprint Fidelis implantable cardioverter-defibrillator lead. J Am Coll Cardiol. 2010;56:646–50.CrossRefPubMed Maytin M, Love CJ, Fischer A, et al. Multicenter experience with extraction of the Sprint Fidelis implantable cardioverter-defibrillator lead. J Am Coll Cardiol. 2010;56:646–50.CrossRefPubMed
51.
go back to reference Parkash R, Crystal E, Bashir J, et al. Complications associated with revision of Sprint Fidelis leads: report from the Canadian Heart Rhythm Society Device Advisory Committee. Circulation. 2010;121:2384–7.CrossRefPubMed Parkash R, Crystal E, Bashir J, et al. Complications associated with revision of Sprint Fidelis leads: report from the Canadian Heart Rhythm Society Device Advisory Committee. Circulation. 2010;121:2384–7.CrossRefPubMed
52.
go back to reference Patel D, Adelstein E, Nemec J, et al. Extraction of defibrillator leads recalled for cable externalization and failure. J Interv Card Electrophysiol. 2012;34:105–13.CrossRef Patel D, Adelstein E, Nemec J, et al. Extraction of defibrillator leads recalled for cable externalization and failure. J Interv Card Electrophysiol. 2012;34:105–13.CrossRef
53.
go back to reference Brunner MP, Cronin EM, Jacob J, et al. Transvenous extraction of implantable cardioverter-defibrillator leads under advisory–a comparison of Riata, Sprint Fidelis, and non-recalled implantable cardioverter-defibrillator leads. Heart Rhythm. 2013;10:1444–50.CrossRefPubMed Brunner MP, Cronin EM, Jacob J, et al. Transvenous extraction of implantable cardioverter-defibrillator leads under advisory–a comparison of Riata, Sprint Fidelis, and non-recalled implantable cardioverter-defibrillator leads. Heart Rhythm. 2013;10:1444–50.CrossRefPubMed
54.
go back to reference Brunner MP, Cronin EM, Duarte VE, et al. Clinical predictors of adverse patient outcomes in an experience of more than 5000 chronic endovascular pacemaker and defibrillator lead extractions. Heart Rhythm. 2014;11:799–805.CrossRefPubMed Brunner MP, Cronin EM, Duarte VE, et al. Clinical predictors of adverse patient outcomes in an experience of more than 5000 chronic endovascular pacemaker and defibrillator lead extractions. Heart Rhythm. 2014;11:799–805.CrossRefPubMed
Metadata
Title
Managing patients with advisory defibrillator leads: what can we learn from published data?
Authors
F. A. Bracke
B. M. van Gelder
Publication date
01-04-2015
Publisher
Bohn Stafleu van Loghum
Published in
Netherlands Heart Journal / Issue 4/2015
Print ISSN: 1568-5888
Electronic ISSN: 1876-6250
DOI
https://doi.org/10.1007/s12471-015-0669-6

Other articles of this Issue 4/2015

Netherlands Heart Journal 4/2015 Go to the issue