Published in: Journal of Nuclear Cardiology 5/2016

01-10-2016 | Original Article

Simultaneous reconstruction of attenuation and activity in cardiac PET can remove CT misalignment artifacts

Authors: L. Presotto, PhD, E. Busnardo, MD, D. Perani, MD, L. Gianolli, MD, M. C. Gilardi, MSc, V. Bettinardi, MSc

Published in: Journal of Nuclear Cardiology | Issue 5/2016

Login to get access

Abstract

Background

Misalignment between positron emission tomography (PET) and computed tomography (CT) data is known to generate artifactual defects in cardiac PET images due to imprecise attenuation correction (AC). In this work, the use of a maximum likelihood attenuation and activity (MLAA) algorithm is proposed to avoid such artifacts in time-of-flight (TOF) PET.

Methods

MLAA was implemented and tested using a thorax/heart phantom and retrospectively on fourteen 13N-ammonia PET/CT perfusion studies. Global and local misalignments between PET and CT data were generated by shifting matched CT images or using CT data representative of the end-inspiration phase. PET images were reconstructed with MLAA and a 3D-ordered-subsets-expectation-maximization (OSEM)-TOF algorithm. Images obtained with 3D-OSEM-TOF and matched CT were used as references. These images were compared (qualitatively and semi-quantitatively) with those reconstructed with 3D-OSEM-TOF and MLAA for which a misaligned CT was used, respectively, for AC and initialization.

Results

Phantom experiment proved the capability of MLAA to converge toward the correct emission and attenuation distributions using, as input, only PET emission data, but convergence was very slow. Initializing MLAA with phantom CT images markedly improved convergence speed. In patient studies, when shifted or end-inspiration CT images were used for AC, 3D-OSEM-TOF reconstructions showed artifacts of increasing severity, size, and frequency with increasing mismatch. Such artifacts were absent in the corresponding MLAA images.

Conclusion

The proposed implementation of the MLAA algorithm is a feasible and robust technique to avoid AC mismatch artifacts in cardiac PET studies provided that a CT of the source is available, even if poorly aligned.
Appendix
Available only for authorised users
Literature
2.
go back to reference Ghosh N, Rimoldi OE, Beanlands RS, Camici PG. Assessment of myocardial ischaemia and viability: Role of positron emission tomography. Eur Heart J 2010;31:2984-95.CrossRefPubMed Ghosh N, Rimoldi OE, Beanlands RS, Camici PG. Assessment of myocardial ischaemia and viability: Role of positron emission tomography. Eur Heart J 2010;31:2984-95.CrossRefPubMed
3.
go back to reference Martinez-Möller A, Souvatzoglou M, Navab N, Schwaiger M, Nekolla SG. Artifacts from misaligned CT in cardiac perfusion PET/CT studies: Frequency, effects, and potential solutions. J Nucl Med 2007;48:188-93.PubMed Martinez-Möller A, Souvatzoglou M, Navab N, Schwaiger M, Nekolla SG. Artifacts from misaligned CT in cardiac perfusion PET/CT studies: Frequency, effects, and potential solutions. J Nucl Med 2007;48:188-93.PubMed
4.
go back to reference Gould KL, Pan T, Loghin C, Johnson NP, Guha A, Sdringola S. Frequent diagnostic errors in cardiac PET/CT due to misregistration of CT attenuation and emission PET images: A definitive analysis of causes, consequences, and corrections. J Nucl Med 2007;48:1112-21.CrossRefPubMed Gould KL, Pan T, Loghin C, Johnson NP, Guha A, Sdringola S. Frequent diagnostic errors in cardiac PET/CT due to misregistration of CT attenuation and emission PET images: A definitive analysis of causes, consequences, and corrections. J Nucl Med 2007;48:1112-21.CrossRefPubMed
5.
go back to reference Le Meunier L, Maass-Moreno R, Carrasquillo JA, Dieckmann W, Bacharach SL. PET/CT imaging: Effect of respiratory motion on apparent myocardial uptake. J Nucl Cardiol 2006;13:821-30.CrossRefPubMed Le Meunier L, Maass-Moreno R, Carrasquillo JA, Dieckmann W, Bacharach SL. PET/CT imaging: Effect of respiratory motion on apparent myocardial uptake. J Nucl Cardiol 2006;13:821-30.CrossRefPubMed
6.
go back to reference Slomka PJ, Diaz-Zamudio M, Dey D, Motwani M, Brodov Y, Choi D, et al. Automatic registration of misaligned CT attenuation correction maps in Rb-82 PET/CT improves detection of angiographically significant coronary artery disease. J Nucl Cardiol 2015. doi:10.1007/s12350-014-0060-9. Slomka PJ, Diaz-Zamudio M, Dey D, Motwani M, Brodov Y, Choi D, et al. Automatic registration of misaligned CT attenuation correction maps in Rb-82 PET/CT improves detection of angiographically significant coronary artery disease. J Nucl Cardiol 2015. doi:10.​1007/​s12350-014-0060-9.
7.
go back to reference Pourmoghaddas A, Klein R, deKemp RA, Wells RG. Respiratory phase alignment improves blood-flow quantification in Rb82 PET myocardial perfusion imaging. Med Phys 2013;40:022503.CrossRefPubMed Pourmoghaddas A, Klein R, deKemp RA, Wells RG. Respiratory phase alignment improves blood-flow quantification in Rb82 PET myocardial perfusion imaging. Med Phys 2013;40:022503.CrossRefPubMed
9.
go back to reference Alessio AM, Kinahan PE, Champley KM, Caldwell JH. Attenuation-emission alignment in cardiac PET/CT based on consistency conditions. Med Phys 2010;37:1191-200.CrossRefPubMedPubMedCentral Alessio AM, Kinahan PE, Champley KM, Caldwell JH. Attenuation-emission alignment in cardiac PET/CT based on consistency conditions. Med Phys 2010;37:1191-200.CrossRefPubMedPubMedCentral
10.
go back to reference Bond S, Kadir T, Hamil J, Casey M, Platsch G, Burckhardt D, et al. Automatic registration of cardiac PET/CT for attenuation correction. In: Nuclear Science Symposium Conference Record, 2008. NSS’08. IEEE (p. 5512-7). Bond S, Kadir T, Hamil J, Casey M, Platsch G, Burckhardt D, et al. Automatic registration of cardiac PET/CT for attenuation correction. In: Nuclear Science Symposium Conference Record, 2008. NSS’08. IEEE (p. 5512-7).
11.
go back to reference Khurshid K, McGough RJ, Berger K. Automated cardiac motion compensation in PET/CT for accurate reconstruction of PET myocardial perfusion images. Phys Med Biol 2008;53:5705-18.CrossRefPubMed Khurshid K, McGough RJ, Berger K. Automated cardiac motion compensation in PET/CT for accurate reconstruction of PET myocardial perfusion images. Phys Med Biol 2008;53:5705-18.CrossRefPubMed
12.
go back to reference Nye JA, Hamill J, Tudorascu D, Carew J, Esteves F, Votaw JR. Comparison of low-pitch and respiratory-averaged CT protocols for attenuation correction of cardiac PET studies. Med Phys 2009;36:1618-23.CrossRefPubMed Nye JA, Hamill J, Tudorascu D, Carew J, Esteves F, Votaw JR. Comparison of low-pitch and respiratory-averaged CT protocols for attenuation correction of cardiac PET studies. Med Phys 2009;36:1618-23.CrossRefPubMed
13.
go back to reference Souvatzoglou M, Bengel F, Busch R, Kruschke C, Fernolendt H, Lee D, et al. Attenuation correction in cardiac PET/CT with three different CT protocols: A comparison with conventional PET. Eur J Nucl Med Mol Imaging 2007;34:1991-2000.CrossRefPubMed Souvatzoglou M, Bengel F, Busch R, Kruschke C, Fernolendt H, Lee D, et al. Attenuation correction in cardiac PET/CT with three different CT protocols: A comparison with conventional PET. Eur J Nucl Med Mol Imaging 2007;34:1991-2000.CrossRefPubMed
14.
go back to reference Nye JA, Esteves F, Votaw JR. Minimizing artifacts resulting from respiratory and cardiac motion by optimization of the transmission scan in cardiac PET/CT. Med Phys 2007;34:1901-6.CrossRefPubMed Nye JA, Esteves F, Votaw JR. Minimizing artifacts resulting from respiratory and cardiac motion by optimization of the transmission scan in cardiac PET/CT. Med Phys 2007;34:1901-6.CrossRefPubMed
15.
16.
go back to reference Dorbala S, Di Carli MF, Delbeke D, Abbara S, DePuey EG, Dilsizian V, et al. SNMMI/ASNC/SCCT guideline for cardiac SPECT/CT and PET/CT 1.0. J Nucl Med 2013;54:1485-507.CrossRefPubMed Dorbala S, Di Carli MF, Delbeke D, Abbara S, DePuey EG, Dilsizian V, et al. SNMMI/ASNC/SCCT guideline for cardiac SPECT/CT and PET/CT 1.0. J Nucl Med 2013;54:1485-507.CrossRefPubMed
17.
go back to reference Censor Y, Gustafson DE, Lent A, Tuy H. A new approach to the emission computerized tomography problem: Simultaneous calculation of attenuation and activity coefficients. IEEE Trans Nucl Sci 1979;26:2775-9.CrossRef Censor Y, Gustafson DE, Lent A, Tuy H. A new approach to the emission computerized tomography problem: Simultaneous calculation of attenuation and activity coefficients. IEEE Trans Nucl Sci 1979;26:2775-9.CrossRef
18.
go back to reference Nuyts J, Dupont P, Stroobants S, Benninck R, Mortelmans L, Suetens P. Simultaneous maximum a posteriori reconstruction of attenuation and activity distributions from emission sinograms. IEEE Trans Med Imaging 1999;18:393-403.CrossRefPubMed Nuyts J, Dupont P, Stroobants S, Benninck R, Mortelmans L, Suetens P. Simultaneous maximum a posteriori reconstruction of attenuation and activity distributions from emission sinograms. IEEE Trans Med Imaging 1999;18:393-403.CrossRefPubMed
19.
go back to reference Rezaei A, Defrise M, Bal G, Michel C, Conti M, Watson C, Nuyts J. Simultaneous reconstruction of activity and attenuation in time-of-flight PET. IEEE Trans Med Imaging 2012;31:2224-33.CrossRefPubMed Rezaei A, Defrise M, Bal G, Michel C, Conti M, Watson C, Nuyts J. Simultaneous reconstruction of activity and attenuation in time-of-flight PET. IEEE Trans Med Imaging 2012;31:2224-33.CrossRefPubMed
20.
go back to reference Defrise M, Rezaei A, Nuyts J. Time-of-flight PET data determine the attenuation sinogram up to a constant. Phys Med Biol 2012;57:885.CrossRefPubMed Defrise M, Rezaei A, Nuyts J. Time-of-flight PET data determine the attenuation sinogram up to a constant. Phys Med Biol 2012;57:885.CrossRefPubMed
21.
go back to reference Presotto L, Gianolli L, Gilardi MC, Bettinardi V. Evaluation of image reconstruction algorithms encompassing time-of-flight and point spread function modelling for quantitative cardiac PET: Phantom studies. J Nucl Cardiol 2015;22:351-63.CrossRefPubMed Presotto L, Gianolli L, Gilardi MC, Bettinardi V. Evaluation of image reconstruction algorithms encompassing time-of-flight and point spread function modelling for quantitative cardiac PET: Phantom studies. J Nucl Cardiol 2015;22:351-63.CrossRefPubMed
22.
go back to reference Bettinardi V, Presotto L, Rapisarda E, Picchio M, Gianolli L, Gilardi MC. Physical performance of the new hybrid PET/CT discovery-690. Med Phys 2011;38:5394-411.CrossRefPubMed Bettinardi V, Presotto L, Rapisarda E, Picchio M, Gianolli L, Gilardi MC. Physical performance of the new hybrid PET/CT discovery-690. Med Phys 2011;38:5394-411.CrossRefPubMed
23.
go back to reference Bettinardi V, De Bernardi E, Presotto L, Gilardi MC. Motion-tracking hardware and advanced applications in PET and PET/CT. PET Clin 2013;8:11-28.CrossRefPubMed Bettinardi V, De Bernardi E, Presotto L, Gilardi MC. Motion-tracking hardware and advanced applications in PET and PET/CT. PET Clin 2013;8:11-28.CrossRefPubMed
24.
go back to reference Cerqueira MD, Weissman NJ, Dilsizian V, Jacobs AK, Kaul S, Laskey WK, et al. Standardized myocardial segmentation and nomenclature for tomographic imaging of the heart. Circulation 2002;105:539-42.CrossRefPubMed Cerqueira MD, Weissman NJ, Dilsizian V, Jacobs AK, Kaul S, Laskey WK, et al. Standardized myocardial segmentation and nomenclature for tomographic imaging of the heart. Circulation 2002;105:539-42.CrossRefPubMed
25.
go back to reference Porenta G, Kuhle W, Czernin J, Ratib O, Brunken RC, Phelps ME, Schelbert HR. Semiquantitative assessment of myocardial blood flow and viability using polar map displays of cardiac PET images. J Nucl Med 1992;33:1628-36.PubMed Porenta G, Kuhle W, Czernin J, Ratib O, Brunken RC, Phelps ME, Schelbert HR. Semiquantitative assessment of myocardial blood flow and viability using polar map displays of cardiac PET images. J Nucl Med 1992;33:1628-36.PubMed
26.
go back to reference Beanlands RS, Muzik O, Hutchins GD, Wolfe ER, Schwaiger M. Heterogeneity of regional nitrogen 13-labeled ammonia tracer distribution in the normal human heart: Comparison with rubidium 82 and copper 62-labeled PTSM. J Nucl Cardiol 1994;1:225-35.CrossRefPubMed Beanlands RS, Muzik O, Hutchins GD, Wolfe ER, Schwaiger M. Heterogeneity of regional nitrogen 13-labeled ammonia tracer distribution in the normal human heart: Comparison with rubidium 82 and copper 62-labeled PTSM. J Nucl Cardiol 1994;1:225-35.CrossRefPubMed
27.
go back to reference Nuyts J, De Man B, Dupont P, Defrise M, Suetens P, Mortelmans L. Iterative reconstruction for helical CT: A simulation study. Phys Med Biol 1998;43:729.CrossRefPubMed Nuyts J, De Man B, Dupont P, Defrise M, Suetens P, Mortelmans L. Iterative reconstruction for helical CT: A simulation study. Phys Med Biol 1998;43:729.CrossRefPubMed
Metadata
Title
Simultaneous reconstruction of attenuation and activity in cardiac PET can remove CT misalignment artifacts
Authors
L. Presotto, PhD
E. Busnardo, MD
D. Perani, MD
L. Gianolli, MD
M. C. Gilardi, MSc
V. Bettinardi, MSc
Publication date
01-10-2016
Publisher
Springer US
Published in
Journal of Nuclear Cardiology / Issue 5/2016
Print ISSN: 1071-3581
Electronic ISSN: 1532-6551
DOI
https://doi.org/10.1007/s12350-015-0239-8