Claros System: A Rapid Microfluidics-Based Point-of-Care System for Quantitative Prostate Specific Antigen Analysis from Finger-Stick Blood

Authors: Agnieszka Maj-Hes, Sabina Sevcenco, Tibor Szarvas, Gero Kramer

Published in: Advances in Therapy | Issue 4/2019

Login to get access

Abstract

Introduction

Determination of circulating prostate specific antigen (PSA) is commonly used in the diagnosis and treatment monitoring of prostate cancer [1]. Presently, PSA testing is performed in centralized laboratories, which is associated with prolonged time between venipuncture and the PSA value being available. In this prospective study, we present a new and rapid test system for the quantitative determination of PSA levels from finger-stick blood.

Methods

The Claros1® analyzer is a rapid microfluidics-based point-of-care system for quantitative PSA analysis from 10-µl finger-stick blood that requires only 10 min for testing. Total PSA concentrations by the Claros system in 100 consecutive asymptomatic men (median age 57 years, range 44–81 years) were compared with two commercially available, commonly used PSA assays (Abbott and Elecsys by Roche) performed by a reference laboratory.

Results

Eighty-six percent of finger-stick blood-borne probes from 100 men were evaluable for PSA testing by the Claros1® analyzer system. In 13/14 cases the expiry date of the microfluid cassettes of the Claros system was exceeded and one blood puncture was performed inadequately. The correlations between the Claros results and OPKO–Abbott and OPKO–Roche assay results were high, with R2 values of 0.982 and 0.985, respectively. The R2 value for the Roche–Abbott correlation was 0.991 with a slope value of 1.160. Prostate cancer was diagnosed in seven cases, with a median PSA of 1.8 ng/ml in the Claros group compared to 1.75 ng/ml and 2.1 ng/ml in the Abbott and Roche groups, respectively.

Conclusion

The Claros1® PSA assay combines the advantages of rapid, accurate detection with a low required sample volume, allowing the analysis to be performed using finger-stick blood. Provided that further analysis proves the reproducibility of the test, it may help to reduce the number of office visits, thus decreasing costs to the health care system.
Literature
2.
go back to reference Catalona WJ, et al. Comparison of digital rectal examination and serum prostate specific antigen in the early detection of prostate cancer: results of a multicenter clinical trial of 6,630 men. J Urol. 1994;151:1283–90.CrossRefPubMed Catalona WJ, et al. Comparison of digital rectal examination and serum prostate specific antigen in the early detection of prostate cancer: results of a multicenter clinical trial of 6,630 men. J Urol. 1994;151:1283–90.CrossRefPubMed
3.
go back to reference Welch HG, Gorski DH, Albertsen PC. Trends in metastatic breast and prostate cancer–lessons in cancer dynamics. N Engl J Med. 2015;373:1685–7.CrossRefPubMed Welch HG, Gorski DH, Albertsen PC. Trends in metastatic breast and prostate cancer–lessons in cancer dynamics. N Engl J Med. 2015;373:1685–7.CrossRefPubMed
4.
go back to reference Cooperberg MR, Lubeck DP, Meng MV, Mehta SS, Carroll PR. The changing face of low-risk prostate cancer: trends in clinical presentation and primary management. J Clin Oncol. 2004;22:2141–9.CrossRefPubMed Cooperberg MR, Lubeck DP, Meng MV, Mehta SS, Carroll PR. The changing face of low-risk prostate cancer: trends in clinical presentation and primary management. J Clin Oncol. 2004;22:2141–9.CrossRefPubMed
6.
go back to reference Schröder FH, et al. Screening and prostate-cancer mortality in a randomized European study. N Engl J Med. 2009;360:1320–8.CrossRefPubMed Schröder FH, et al. Screening and prostate-cancer mortality in a randomized European study. N Engl J Med. 2009;360:1320–8.CrossRefPubMed
7.
go back to reference Fitzpatrick JM, Banu E, Oudard S. Prostate-specific antigen kinetics in localized and advanced prostate cancer. BJU Int. 2009;103:578–87.CrossRefPubMed Fitzpatrick JM, Banu E, Oudard S. Prostate-specific antigen kinetics in localized and advanced prostate cancer. BJU Int. 2009;103:578–87.CrossRefPubMed
8.
go back to reference Scattoni V, et al. Diagnosis of local recurrence after radical prostatectomy. BJU Int. 2004;93:680–8.CrossRefPubMed Scattoni V, et al. Diagnosis of local recurrence after radical prostatectomy. BJU Int. 2004;93:680–8.CrossRefPubMed
9.
go back to reference Wilson DH, et al. Fifth-generation digital immunoassay for prostate-specific antigen by single molecule array technology. Clin Chem. 2011;57:1712–21.CrossRefPubMedPubMedCentral Wilson DH, et al. Fifth-generation digital immunoassay for prostate-specific antigen by single molecule array technology. Clin Chem. 2011;57:1712–21.CrossRefPubMedPubMedCentral
10.
go back to reference Lepor H, et al. Clinical evaluation of a novel method for the measurement of prostate-specific antigen, AccuPSA(TM), as a predictor of 5-year biochemical recurrence-free survival after radical prostatectomy: results of a pilot study. BJU Int. 2012;109:1770–5.CrossRefPubMed Lepor H, et al. Clinical evaluation of a novel method for the measurement of prostate-specific antigen, AccuPSA(TM), as a predictor of 5-year biochemical recurrence-free survival after radical prostatectomy: results of a pilot study. BJU Int. 2012;109:1770–5.CrossRefPubMed
11.
go back to reference Okada H, Hosokawa K, Maeda M. Power-free microchip immunoassay of PSA in human serum for point-of-care testing. Anal Sci. 2011;27:237–41.CrossRefPubMed Okada H, Hosokawa K, Maeda M. Power-free microchip immunoassay of PSA in human serum for point-of-care testing. Anal Sci. 2011;27:237–41.CrossRefPubMed
12.
go back to reference Lin YY, et al. A nanoparticle label/immunochromatographic electrochemical biosensor for rapid and sensitive detection of prostate-specific antigen. Biosens Bioelectron. 2008;23:1659–65.CrossRefPubMed Lin YY, et al. A nanoparticle label/immunochromatographic electrochemical biosensor for rapid and sensitive detection of prostate-specific antigen. Biosens Bioelectron. 2008;23:1659–65.CrossRefPubMed
13.
go back to reference Kim DJ, et al. Organic electrochemical transistor based immunosensor for prostate specific antigen (PSA) detection using gold nanoparticles for signal amplification. Biosens Bioelectron. 2010;25:2477–82.CrossRefPubMed Kim DJ, et al. Organic electrochemical transistor based immunosensor for prostate specific antigen (PSA) detection using gold nanoparticles for signal amplification. Biosens Bioelectron. 2010;25:2477–82.CrossRefPubMed
14.
go back to reference Adel Ahmed H, Azzazy HME. Power-free chip enzyme immunoassay for detection of prostate specific antigen (PSA) in serum. Biosens Bioelectron. 2013;49:478–84.CrossRefPubMed Adel Ahmed H, Azzazy HME. Power-free chip enzyme immunoassay for detection of prostate specific antigen (PSA) in serum. Biosens Bioelectron. 2013;49:478–84.CrossRefPubMed
15.
go back to reference Karim O, et al. Point-of-care PSA testing: an evaluation of PSAwatch. Prostate Cancer Prostatic Dis. 2007;10:270–3.CrossRefPubMed Karim O, et al. Point-of-care PSA testing: an evaluation of PSAwatch. Prostate Cancer Prostatic Dis. 2007;10:270–3.CrossRefPubMed
16.
go back to reference Dok An C, et al. Evaluation of a rapid qualitative prostate specific antigen assay, the One Step PSA(TM) test. Cancer Lett. 2001;162(2):135–9.CrossRefPubMed Dok An C, et al. Evaluation of a rapid qualitative prostate specific antigen assay, the One Step PSA(TM) test. Cancer Lett. 2001;162(2):135–9.CrossRefPubMed
17.
go back to reference Jung K, et al. Rapid detection of elevated prostate-specific antigen levels in blood: performance of various membrane strip tests compared. Urology 1999;53(1):155–60.CrossRefPubMed Jung K, et al. Rapid detection of elevated prostate-specific antigen levels in blood: performance of various membrane strip tests compared. Urology 1999;53(1):155–60.CrossRefPubMed
18.
go back to reference Lein M, et al. Rapid screening of PSA: evaluation of an immunochemical membrane strip test. Clin Chem. 1995;41(10):1545–7.PubMed Lein M, et al. Rapid screening of PSA: evaluation of an immunochemical membrane strip test. Clin Chem. 1995;41(10):1545–7.PubMed
19.
go back to reference Madersbacher S, et al. Validation of a 10-minute dipstick test for serum prostate-specific antigen. Eur Urol. 1996;30(4):446–50.CrossRefPubMed Madersbacher S, et al. Validation of a 10-minute dipstick test for serum prostate-specific antigen. Eur Urol. 1996;30(4):446–50.CrossRefPubMed
20.
go back to reference Berg W, et al. Possibility of improving the acceptance rateof early detection testing for prostate cancerwith a one-step test for prostate-specific antigen in whole blood. Urol Int. 1999;63(2):102–6.CrossRefPubMed Berg W, et al. Possibility of improving the acceptance rateof early detection testing for prostate cancerwith a one-step test for prostate-specific antigen in whole blood. Urol Int. 1999;63(2):102–6.CrossRefPubMed
21.
go back to reference Fernandez-Sanchez C, et al. One-step immunostrip test for the simultaneous detection of free and total prostate specific antigen in serum. J Immunol Methods 2005;307(1–2):1–12.CrossRefPubMed Fernandez-Sanchez C, et al. One-step immunostrip test for the simultaneous detection of free and total prostate specific antigen in serum. J Immunol Methods 2005;307(1–2):1–12.CrossRefPubMed
22.
go back to reference Miano R, et al. Evaluation of a new, rapid, qualitative, one-step PSA Test for prostate cancer screening: the PSA RapidScreen test. Prostate Cancer Prostatic Dis. 2005;8(3):219–23.CrossRefPubMed Miano R, et al. Evaluation of a new, rapid, qualitative, one-step PSA Test for prostate cancer screening: the PSA RapidScreen test. Prostate Cancer Prostatic Dis. 2005;8(3):219–23.CrossRefPubMed
23.
go back to reference Wu CC, et al. Evaluation of a rapid quantitative determination method of PSA concentration with gold immunochromatographic strips. BMC Urol. 2015;15:109.CrossRefPubMedPubMedCentral Wu CC, et al. Evaluation of a rapid quantitative determination method of PSA concentration with gold immunochromatographic strips. BMC Urol. 2015;15:109.CrossRefPubMedPubMedCentral
Metadata
Title
Claros System: A Rapid Microfluidics-Based Point-of-Care System for Quantitative Prostate Specific Antigen Analysis from Finger-Stick Blood
Authors
Agnieszka Maj-Hes
Sabina Sevcenco
Tibor Szarvas
Gero Kramer
Publication date
01-04-2019
Publisher
Springer Healthcare
Published in
Advances in Therapy / Issue 4/2019
Print ISSN: 0741-238X
Electronic ISSN: 1865-8652
DOI
https://doi.org/10.1007/s12325-019-0888-0

Other articles of this Issue 4/2019

Advances in Therapy 4/2019 Go to the issue