Efficacy of a Novel Prefilled, Single-Use, Needle-Free Device (Zeneo®) in Achieving Intramuscular Agent Delivery: An Observational Study

Authors: Marc Bardou, Maxime Luu, Paul Walker, Christophe Auriel, Xavière Castano

Published in: Advances in Therapy | Issue 1/2017

Login to get access

Abstract

Background

It is recognized that, as a result of variation in tissue anatomy, current auto-injectors may have insufficient needle length to achieve successful intramuscular agent delivery in a number of patients. The Zeneo® auto-injector is a novel prefilled, single-use, needle-free device currently in development for intradermal, subcutaneous, and intramuscular agent delivery across a variety of clinical indications. We aimed to evaluate delivery depth of the device calibrated at pressure appropriate for intramuscular (IM) administration.

Methods

This was a prospective single-center study in healthy adult volunteers, in whom each received a single injection of saline into the anterolateral thigh. Using sequential MRI scans, we measured skin-to-muscle distance (STMD) agent delivery depth, and the success of IM agent penetration. Device dynamic pressure measurements were also recorded.

Results

Results are reported for 37 subjects with evaluable MRI scans; 19 men, 18 women; mean age 38 years (range 20–58); mean BMI 27.0 kg/m2 (range 21.2–30.8 kg/m2). Mean STMD values were 18.6 mm (range 13.4–23.6 mm) in women and 10.0 mm (range 5.0–21.7 mm) in men, with gender differences due primarily to greater subcutaneous thickness in women. A trend for greater STMD in subjects with BMI greater than 25 kg/m2 was seen. Mean injectate penetration depths of 30.1 mm (range 20.2–45.6 mm) were observed with values similar in male and female subjects. Successful IM delivery was reported in 95% of subjects. When failure occurred, this was not due to inadequate injection depth. Device pressure (P max) had the greatest influence on injectate muscle penetration.

Conclusion

Use of the Zeneo® auto-injector achieves delivery depth that ensures intramuscular delivery in both men and women, regardless of BMI. Consistent with other reported data, STMD is greater in women.

Funding

Crossject.
Literature
1.
go back to reference Lieberman P, Nicklas RA, Randolph C, et al. Anaphylaxis–a practice parameter update 2015. Ann Allergy Asthma Immunol. 2015;115(5):341–84.CrossRefPubMed Lieberman P, Nicklas RA, Randolph C, et al. Anaphylaxis–a practice parameter update 2015. Ann Allergy Asthma Immunol. 2015;115(5):341–84.CrossRefPubMed
2.
go back to reference Simons FE, Ebisawa M, Sanchez-Borges M, et al. 2015 update of the evidence base: World Allergy Organization anaphylaxis guidelines. World Allergy Organ J. 2015;8(1):32.CrossRefPubMedPubMedCentral Simons FE, Ebisawa M, Sanchez-Borges M, et al. 2015 update of the evidence base: World Allergy Organization anaphylaxis guidelines. World Allergy Organ J. 2015;8(1):32.CrossRefPubMedPubMedCentral
4.
go back to reference Simons FER, Gu X, Simons KJ. Epinephrine absorption in adults: intramuscular versus subcutaneous injection. J Allergy Clin Immunol. 2001;108(5):871–3.CrossRefPubMed Simons FER, Gu X, Simons KJ. Epinephrine absorption in adults: intramuscular versus subcutaneous injection. J Allergy Clin Immunol. 2001;108(5):871–3.CrossRefPubMed
5.
go back to reference Song TT, Nelson MR, Chang JH, Engler RJ, Chowdhury BA. Adequacy of the epinephrine autoinjector needle length in delivering epinephrine to the intramuscular tissues. Ann Allergy Asthma Immunol. 2005;94(5):539–42.CrossRefPubMed Song TT, Nelson MR, Chang JH, Engler RJ, Chowdhury BA. Adequacy of the epinephrine autoinjector needle length in delivering epinephrine to the intramuscular tissues. Ann Allergy Asthma Immunol. 2005;94(5):539–42.CrossRefPubMed
6.
go back to reference Tsai G, Kim L, Nevis IF, et al. Auto-injector needle length may be inadequate to deliver epinephrine intramuscularly in women with confirmed food allergy. Allergy Asthma Clin Immunol. 2014;10(1):39.CrossRefPubMedPubMedCentral Tsai G, Kim L, Nevis IF, et al. Auto-injector needle length may be inadequate to deliver epinephrine intramuscularly in women with confirmed food allergy. Allergy Asthma Clin Immunol. 2014;10(1):39.CrossRefPubMedPubMedCentral
7.
go back to reference Bhalla MC, Gable BD, Frey JA, Reichenbach MR, Wilber ST. Predictors of epinephrine autoinjector needle length inadequacy. Am J Emerg Med. 2013;31(12):1671–6.CrossRefPubMed Bhalla MC, Gable BD, Frey JA, Reichenbach MR, Wilber ST. Predictors of epinephrine autoinjector needle length inadequacy. Am J Emerg Med. 2013;31(12):1671–6.CrossRefPubMed
8.
go back to reference Stecher D, Bulloch B, Sales J, Schaefer C, Keahey L. Epinephrine auto-injectors: is needle length adequate for delivery of epinephrine intramuscularly? Pediatrics. 2009;124(1):65–70.CrossRefPubMed Stecher D, Bulloch B, Sales J, Schaefer C, Keahey L. Epinephrine auto-injectors: is needle length adequate for delivery of epinephrine intramuscularly? Pediatrics. 2009;124(1):65–70.CrossRefPubMed
9.
go back to reference Dreborg S, Wen X, Kim L, et al. Do epinephrine auto-injectors have an unsuitable needle length in children and adolescents at risk for anaphylaxis from food allergy? Allergy Asthma Clin Immunol. 2016;12:11.CrossRefPubMedPubMedCentral Dreborg S, Wen X, Kim L, et al. Do epinephrine auto-injectors have an unsuitable needle length in children and adolescents at risk for anaphylaxis from food allergy? Allergy Asthma Clin Immunol. 2016;12:11.CrossRefPubMedPubMedCentral
10.
go back to reference Money AG, Barnett J, Kuljis J, Lucas J. Patient perceptions of epinephrine auto-injectors: exploring barriers to use. Scand J Caring Sci. 2013;27(2):335–44.CrossRefPubMed Money AG, Barnett J, Kuljis J, Lucas J. Patient perceptions of epinephrine auto-injectors: exploring barriers to use. Scand J Caring Sci. 2013;27(2):335–44.CrossRefPubMed
11.
go back to reference Simons FE, Lieberman PL, Read Jr EJ, Edwards ES. Hazards of unintentional injection of epinephrine from autoinjectors: a systematic review. Ann Allergy Asthma Immunol. 2009;102(4):282–7.CrossRefPubMed Simons FE, Lieberman PL, Read Jr EJ, Edwards ES. Hazards of unintentional injection of epinephrine from autoinjectors: a systematic review. Ann Allergy Asthma Immunol. 2009;102(4):282–7.CrossRefPubMed
12.
go back to reference World Medical Association. World Medical Association Declaration of Helsinki: ethical principles for medical research involving human subjects. JAMA. 2013;310(20):2191–4.CrossRef World Medical Association. World Medical Association Declaration of Helsinki: ethical principles for medical research involving human subjects. JAMA. 2013;310(20):2191–4.CrossRef
13.
go back to reference Gibney M, Arce CH, Byron KJ, Hirsch LJ. Skin and subcutaneous adipose layer thickness in adults with diabetes at sites used for insulin injections: implications for needle length recommendations. Curr Med Res Opin. 2010;26(6):1519–30.CrossRefPubMed Gibney M, Arce CH, Byron KJ, Hirsch LJ. Skin and subcutaneous adipose layer thickness in adults with diabetes at sites used for insulin injections: implications for needle length recommendations. Curr Med Res Opin. 2010;26(6):1519–30.CrossRefPubMed
14.
go back to reference Hirsch L, Byron K, Gibney M. Intramuscular risk at insulin injection sites–measurement of the distance from skin to muscle and rationale for shorter-length needles for subcutaneous insulin therapy. Diabet Technol Ther. 2014;16(12):867–73.CrossRef Hirsch L, Byron K, Gibney M. Intramuscular risk at insulin injection sites–measurement of the distance from skin to muscle and rationale for shorter-length needles for subcutaneous insulin therapy. Diabet Technol Ther. 2014;16(12):867–73.CrossRef
15.
go back to reference Johnstone J, Hobbins S, Parekh D, O’Hickey S. Excess subcutaneous tissue may preclude intramuscular delivery when using adrenaline autoinjectors in patients with anaphylaxis. Allergy. 2015;70(6):703–6.CrossRefPubMed Johnstone J, Hobbins S, Parekh D, O’Hickey S. Excess subcutaneous tissue may preclude intramuscular delivery when using adrenaline autoinjectors in patients with anaphylaxis. Allergy. 2015;70(6):703–6.CrossRefPubMed
16.
go back to reference Lippert WC, Wall EJ. Optimal intramuscular needle-penetration depth. Pediatrics. 2008;122(3):e556–63.CrossRefPubMed Lippert WC, Wall EJ. Optimal intramuscular needle-penetration depth. Pediatrics. 2008;122(3):e556–63.CrossRefPubMed
17.
go back to reference Song TT, Lieberman P. Epinephrine auto-injector needle length: what is the ideal length? Curr Opin Allergy Clin Immunol. 2016;16(4):361–5.CrossRefPubMed Song TT, Lieberman P. Epinephrine auto-injector needle length: what is the ideal length? Curr Opin Allergy Clin Immunol. 2016;16(4):361–5.CrossRefPubMed
18.
go back to reference Song TT, Merrill NL, Cole JW. Delivery depth of epinephrine by auto-injector into the subcutaneous tissue of pig. Ann Allergy Asthma Immunol. 2013;111(2):143–5.CrossRefPubMed Song TT, Merrill NL, Cole JW. Delivery depth of epinephrine by auto-injector into the subcutaneous tissue of pig. Ann Allergy Asthma Immunol. 2013;111(2):143–5.CrossRefPubMed
19.
go back to reference Resuscitation Council (UK). Emergency treatment of anaphylactic reactions: guidelines for healthcare providers. London: Resuscitation Council (UK); 2008. Resuscitation Council (UK). Emergency treatment of anaphylactic reactions: guidelines for healthcare providers. London: Resuscitation Council (UK); 2008.
20.
go back to reference Centers for Disease Control and Prevention. Epidemiology and prevention of vaccine-preventable diseases. In: Hamborsky J, Kroger A, Wolfe S, editors. 13th ed. Washington: Public Health Foundation; 2015. Centers for Disease Control and Prevention. Epidemiology and prevention of vaccine-preventable diseases. In: Hamborsky J, Kroger A, Wolfe S, editors. 13th ed. Washington: Public Health Foundation; 2015.
Metadata
Title
Efficacy of a Novel Prefilled, Single-Use, Needle-Free Device (Zeneo®) in Achieving Intramuscular Agent Delivery: An Observational Study
Authors
Marc Bardou
Maxime Luu
Paul Walker
Christophe Auriel
Xavière Castano
Publication date
01-01-2017
Publisher
Springer Healthcare
Published in
Advances in Therapy / Issue 1/2017
Print ISSN: 0741-238X
Electronic ISSN: 1865-8652
DOI
https://doi.org/10.1007/s12325-016-0452-0

Other articles of this Issue 1/2017

Advances in Therapy 1/2017 Go to the issue
Live Webinar | 27-06-2024 | 18:00 (CEST)

Keynote webinar | Spotlight on medication adherence

Live: Thursday 27th June 2024, 18:00-19:30 (CEST)

WHO estimates that half of all patients worldwide are non-adherent to their prescribed medication. The consequences of poor adherence can be catastrophic, on both the individual and population level.

Join our expert panel to discover why you need to understand the drivers of non-adherence in your patients, and how you can optimize medication adherence in your clinics to drastically improve patient outcomes.

Prof. Kevin Dolgin
Prof. Florian Limbourg
Prof. Anoop Chauhan
Developed by: Springer Medicine
Obesity Clinical Trial Summary

At a glance: The STEP trials

A round-up of the STEP phase 3 clinical trials evaluating semaglutide for weight loss in people with overweight or obesity.

Developed by: Springer Medicine

Highlights from the ACC 2024 Congress

Year in Review: Pediatric cardiology

Watch Dr. Anne Marie Valente present the last year's highlights in pediatric and congenital heart disease in the official ACC.24 Year in Review session.

Year in Review: Pulmonary vascular disease

The last year's highlights in pulmonary vascular disease are presented by Dr. Jane Leopold in this official video from ACC.24.

Year in Review: Valvular heart disease

Watch Prof. William Zoghbi present the last year's highlights in valvular heart disease from the official ACC.24 Year in Review session.

Year in Review: Heart failure and cardiomyopathies

Watch this official video from ACC.24. Dr. Biykem Bozkurt discusses last year's major advances in heart failure and cardiomyopathies.