Does the Risk of Rerevision Vary Between Porous Tantalum Cups and Other Cementless Designs After Revision Hip Arthroplasty?

Authors: Inari Laaksonen, MD, PhD, Michelle Lorimer, BSc, Kirill Gromov, MD, PhD, Ola Rolfson, MD, PhD, Keijo T. Mäkelä, MD, PhD, Stephen E. Graves, MD, DPhil, Henrik Malchau, MD, PhD, Maziar Mohaddes, MD, PhD

Published in: Clinical Orthopaedics and Related Research® | Issue 12/2017

Login to get access

Abstract

Background

Earlier results with porous tantalum acetabular cups in revision THA generally have been favorable. Recently there has been some evidence presented that porous tantalum cups might decrease the risk of rerevision in the setting of revision hip surgery performed owing to prosthetic joint infection (PJI). As the data supporting this assertion come from a study with a limited study population, examining this issue with a large registry approach may be enlightening.

Questions/purposes

By combining results from two large, national registries, we asked: (1) Do porous tantalum cups show improved survival after revision THA compared with other cementless designs? (2) Does the use of porous tantalum cups influence survivorship when rerevision for PJI is the endpoint?

Methods

A total of 2442 first-time THA revisions with porous tantalum cups and 4401 first-time revisions with other uncemented cups were included in this collaborative study between the Australian and Swedish national joint registries. The mean age of the patients was 69 years (range, 19–97 years), 3754 (55%) of the patients were women, and the mean followup for the porous tantalum and uncemented control groups were 3.0 years (SD, ± 2.1 years) and 3.4 years (SD, ± 2.3 years), respectively. Concomitant stem revision was more common in the porous tantalum group (43% versus 36%). The use of porous tantalum augments also was analyzed as a proxy for more complex acetabular reconstructions. In an attempt to further reduce selection bias, we performed subgroup analysis for primary operations attributable to osteoarthritis and first revision attributable to aseptic loosening.

Results

Kaplan-Meier survivorship with rerevisison for any reason up to 7 years was comparable between the porous tantalum cup group and the uncemented cup control group (86% [95% CI, 85%–89%] and 87% [95% CI, 85%–89%], respectively; p = 0.85) and the overall survivorship up to 7 years with a second revision for PJI as the endpoint (97% [95% CI, 95%–98%] and 97% [95% CI, 96%–98%], respectively; p = 0.64). Excluding procedures where augments had been used or studying primary osteoarthritis and first revision owing to aseptic loosening subgroups did not change this result.

Conclusions

Implant survival for a porous tantalum cup in first-time THA revision was similar to the survival of the uncemented cup control group. With the numbers available, no benefit in survival with rerevision for infection as the endpoint could be ascribed to the porous tantalum cup group, as has been suggested by earlier work. Further studies with acetabular bone deficiency data, greater insight into host comorbidity factors, and a longer followup are needed to corroborate or refute these results.

Level of Evidence

Level III, therapeutic study.
Literature
2.
go back to reference Bobyn JD, Stackpool GJ, Hacking SA, Tanzer M, Krygier JJ. Characteristics of bone ingrowth and interface mechanics of a new porous tantalum biomaterial. J Bone Joint Surg Br. 1999;81:907–914.CrossRefPubMed Bobyn JD, Stackpool GJ, Hacking SA, Tanzer M, Krygier JJ. Characteristics of bone ingrowth and interface mechanics of a new porous tantalum biomaterial. J Bone Joint Surg Br. 1999;81:907–914.CrossRefPubMed
3.
go back to reference Bozic KJ, Kurtz S, Lau E, Ong K, Chiu V, Vail TP, Rubash HE, Berry DJ. The epidemiology of bearing surface usage in total hip arthroplasty in the United States. J Bone Joint Surg Am. 2009;91:1614–1620.CrossRefPubMed Bozic KJ, Kurtz S, Lau E, Ong K, Chiu V, Vail TP, Rubash HE, Berry DJ. The epidemiology of bearing surface usage in total hip arthroplasty in the United States. J Bone Joint Surg Am. 2009;91:1614–1620.CrossRefPubMed
4.
go back to reference Davies JH, Laflamme GY, Delisle J, Fernandes J. Trabecular metal used for major bone loss in acetabular hip revision. J Arthroplasty. 2011;26:1245–1250.CrossRefPubMed Davies JH, Laflamme GY, Delisle J, Fernandes J. Trabecular metal used for major bone loss in acetabular hip revision. J Arthroplasty. 2011;26:1245–1250.CrossRefPubMed
5.
go back to reference Jafari SM, Bender B, Coyle C, Parvizi J, Sharkey PF, Hozack WJ. Do tantalum and titanium cups show similar results in revision hip arthroplasty? Clin Orthop Relat Res. 2010;468:459–465.CrossRefPubMed Jafari SM, Bender B, Coyle C, Parvizi J, Sharkey PF, Hozack WJ. Do tantalum and titanium cups show similar results in revision hip arthroplasty? Clin Orthop Relat Res. 2010;468:459–465.CrossRefPubMed
6.
go back to reference Kamath AF, Ong KL, Lau E, Chan V, Vail TP, Rubash HE, Berry DJ, Bozic KJ. Quantifying the burden of revision total joint arthroplasty for periprosthetic infection. J Arthroplasty. 2015;30:1492–1497.CrossRefPubMed Kamath AF, Ong KL, Lau E, Chan V, Vail TP, Rubash HE, Berry DJ, Bozic KJ. Quantifying the burden of revision total joint arthroplasty for periprosthetic infection. J Arthroplasty. 2015;30:1492–1497.CrossRefPubMed
7.
go back to reference Kosashvili Y, Backstein D, Safir O, Lakstein D, Gross AE. Dislocation and infection after revision total hip arthroplasty: comparison between the first and multiply revised total hip arthroplasty. J Arthroplasty. 2011;26:1170–1175.CrossRefPubMed Kosashvili Y, Backstein D, Safir O, Lakstein D, Gross AE. Dislocation and infection after revision total hip arthroplasty: comparison between the first and multiply revised total hip arthroplasty. J Arthroplasty. 2011;26:1170–1175.CrossRefPubMed
8.
go back to reference Kremers HM, Howard JL, Loechler Y, Schleck CD, Harmsen WS, Berry DJ, Cabanela ME, Hanssen AD, Pagnano MW, Trousdale RT, Lewallen DG. Comparative long-term survivorship of uncemented acetabular components in revision total hip arthroplasty. J Bone Joint Surg Am. 2012;94:e82.CrossRefPubMed Kremers HM, Howard JL, Loechler Y, Schleck CD, Harmsen WS, Berry DJ, Cabanela ME, Hanssen AD, Pagnano MW, Trousdale RT, Lewallen DG. Comparative long-term survivorship of uncemented acetabular components in revision total hip arthroplasty. J Bone Joint Surg Am. 2012;94:e82.CrossRefPubMed
9.
go back to reference Kurtz S, Ong K, Lau E, Mowat F, Halpern M. Projections of primary and revision hip and knee arthroplasty in the United States from 2005 to 2030. J Bone Joint Surg Am. 2007;89:780–785.PubMed Kurtz S, Ong K, Lau E, Mowat F, Halpern M. Projections of primary and revision hip and knee arthroplasty in the United States from 2005 to 2030. J Bone Joint Surg Am. 2007;89:780–785.PubMed
10.
go back to reference Kurtz SM, Lau E, Watson H, Schmier JK, Parvizi J. Economic burden of periprosthetic joint infection in the United States. J Arthroplasty. 2012;27(8 suppl):61–65.e1. Kurtz SM, Lau E, Watson H, Schmier JK, Parvizi J. Economic burden of periprosthetic joint infection in the United States. J Arthroplasty. 2012;27(8 suppl):61–65.e1.
11.
go back to reference Mohaddes M, Garellick G, Kärrholm J. Method of fixation does not influence the overall risk of rerevision in first-time cup revisions. Clin Orthop Relat Res. 2013;471:3922–3931.CrossRefPubMedPubMedCentral Mohaddes M, Garellick G, Kärrholm J. Method of fixation does not influence the overall risk of rerevision in first-time cup revisions. Clin Orthop Relat Res. 2013;471:3922–3931.CrossRefPubMedPubMedCentral
12.
go back to reference Mohaddes M, Rolfson O, Kärrholm J. Short-term survival of the trabecular metal cup is similar to that of standard cups used in acetabular revision surgery. Acta Orthop. 2015;86:26–31.CrossRefPubMedPubMedCentral Mohaddes M, Rolfson O, Kärrholm J. Short-term survival of the trabecular metal cup is similar to that of standard cups used in acetabular revision surgery. Acta Orthop. 2015;86:26–31.CrossRefPubMedPubMedCentral
13.
go back to reference Noiseux NO, Long WJ, Mabry TM, Hanssen AD, Lewallen DG. Uncemented porous tantalum acetabular components: early follow-up and failures in 613 primary total hip arthroplasties. J Arthroplasty. 2014;29:617–620.CrossRefPubMed Noiseux NO, Long WJ, Mabry TM, Hanssen AD, Lewallen DG. Uncemented porous tantalum acetabular components: early follow-up and failures in 613 primary total hip arthroplasties. J Arthroplasty. 2014;29:617–620.CrossRefPubMed
14.
go back to reference Ong KL, Kurtz SM, Lau E, Bozic KJ, Berry DJ, Parvizi J. Prosthetic joint infection risk after total hip arthroplasty in the Medicare population. J Arthroplasty. 2009;24(6 suppl):105–109.CrossRefPubMed Ong KL, Kurtz SM, Lau E, Bozic KJ, Berry DJ, Parvizi J. Prosthetic joint infection risk after total hip arthroplasty in the Medicare population. J Arthroplasty. 2009;24(6 suppl):105–109.CrossRefPubMed
15.
go back to reference Patel A, Pavlou G, Mújica-Mota RE, Toms AD. The epidemiology of revision total knee and hip arthroplasty in England and Wales: a comparative analysis with projections for the United States. A study using the National Joint Registry dataset. Bone Joint J. 2015;97:1076–1081. Patel A, Pavlou G, Mújica-Mota RE, Toms AD. The epidemiology of revision total knee and hip arthroplasty in England and Wales: a comparative analysis with projections for the United States. A study using the National Joint Registry dataset. Bone Joint J. 2015;97:1076–1081.
16.
go back to reference Phillips CB, Barrett JA, Losina E, Mahomed NN, Lingard EA, Guadagnoli E, Baron JA, Harris WH, Poss R, Katz JN. Incidence rates of dislocation, pulmonary embolism, and deep infection during the first six months after elective total hip replacement. J Bone Joint Surg Am. 2003;85:20–26.CrossRefPubMed Phillips CB, Barrett JA, Losina E, Mahomed NN, Lingard EA, Guadagnoli E, Baron JA, Harris WH, Poss R, Katz JN. Incidence rates of dislocation, pulmonary embolism, and deep infection during the first six months after elective total hip replacement. J Bone Joint Surg Am. 2003;85:20–26.CrossRefPubMed
17.
go back to reference Pulido L, Rachala SR, Cabanela ME. Cementless acetabular revision: past, present, and future. Revision total hip arthroplasty: the acetabular side using cementless implants. Int Orthop. 2011;35:289–298.CrossRefPubMedPubMedCentral Pulido L, Rachala SR, Cabanela ME. Cementless acetabular revision: past, present, and future. Revision total hip arthroplasty: the acetabular side using cementless implants. Int Orthop. 2011;35:289–298.CrossRefPubMedPubMedCentral
18.
go back to reference Siegmeth A, Duncan CP, Masri BA, Kim WY, Garbuz DS. Modular tantalum augments for acetabular defects in revision hip arthroplasty. Clin Orthop Relat Res. 2009;467:199–205.CrossRefPubMed Siegmeth A, Duncan CP, Masri BA, Kim WY, Garbuz DS. Modular tantalum augments for acetabular defects in revision hip arthroplasty. Clin Orthop Relat Res. 2009;467:199–205.CrossRefPubMed
19.
go back to reference Skyttä ET, Eskelinen A, Paavolainen PO, Remes VM. Early results of 827 trabecular metal revision shells in acetabular revision. J Arthroplasty. 2011;26:342–345.CrossRefPubMed Skyttä ET, Eskelinen A, Paavolainen PO, Remes VM. Early results of 827 trabecular metal revision shells in acetabular revision. J Arthroplasty. 2011;26:342–345.CrossRefPubMed
20.
go back to reference Sternheim A, Backstein D, Kuzyk PR, Goshua G, Berkovich Y, Safir O, Gross AE. Porous metal revision shells for management of contained acetabular bone defects at a mean follow-up of six years: a comparison between up to 50% bleeding host bone contact and more than 50% contact. J Bone Joint Surg Br. 2012;94:158–162.CrossRefPubMed Sternheim A, Backstein D, Kuzyk PR, Goshua G, Berkovich Y, Safir O, Gross AE. Porous metal revision shells for management of contained acetabular bone defects at a mean follow-up of six years: a comparison between up to 50% bleeding host bone contact and more than 50% contact. J Bone Joint Surg Br. 2012;94:158–162.CrossRefPubMed
22.
go back to reference Tokarski AT, Novack TA, Parvizi J. Is tantalum protective against infection in revision total hip arthroplasty? Bone Joint J. 2015;97:45–49.CrossRefPubMed Tokarski AT, Novack TA, Parvizi J. Is tantalum protective against infection in revision total hip arthroplasty? Bone Joint J. 2015;97:45–49.CrossRefPubMed
23.
go back to reference Weeden SH, Schmidt RH. The use of tantalum porous metal implants for Paprosky 3A and 3B defects. J Arthroplasty. 2007;22(6 suppl 2):151–155.CrossRefPubMed Weeden SH, Schmidt RH. The use of tantalum porous metal implants for Paprosky 3A and 3B defects. J Arthroplasty. 2007;22(6 suppl 2):151–155.CrossRefPubMed
24.
go back to reference Zhan C, Kaczmarek R, Loyo-Berrios N, Sangl J, Bright RA. Incidence and short-term outcomes of primary and revision hip replacement in the United States. J Bone Joint Surg Am. 2007;89:526–533.PubMed Zhan C, Kaczmarek R, Loyo-Berrios N, Sangl J, Bright RA. Incidence and short-term outcomes of primary and revision hip replacement in the United States. J Bone Joint Surg Am. 2007;89:526–533.PubMed
Metadata
Title
Does the Risk of Rerevision Vary Between Porous Tantalum Cups and Other Cementless Designs After Revision Hip Arthroplasty?
Authors
Inari Laaksonen, MD, PhD
Michelle Lorimer, BSc
Kirill Gromov, MD, PhD
Ola Rolfson, MD, PhD
Keijo T. Mäkelä, MD, PhD
Stephen E. Graves, MD, DPhil
Henrik Malchau, MD, PhD
Maziar Mohaddes, MD, PhD
Publication date
01-12-2017
Publisher
Springer US
Published in
Clinical Orthopaedics and Related Research® / Issue 12/2017
Print ISSN: 0009-921X
Electronic ISSN: 1528-1132
DOI
https://doi.org/10.1007/s11999-017-5417-3

Other articles of this Issue 12/2017

Clinical Orthopaedics and Related Research® 12/2017 Go to the issue