01-12-2022 | Computed Tomography | Original Article

Development of an open-source measurement system to assess the areal bone mineral density of the proximal femur from clinical CT images

Authors: Keisuke Uemura, Yoshito Otake, Masaki Takao, Hiroki Makino, Mazen Soufi, Makoto Iwasa, Nobuhiko Sugano, Yoshinobu Sato

Published in: Archives of Osteoporosis | Issue 1/2022

Login to get access

Abstract

Summary

Commercial software is generally needed to measure the areal bone mineral density (aBMD) of the proximal femur from clinical computed tomography (CT) images. This study developed and verified an open-source reproducible system to quantify CT-aBMD to screen osteoporosis using clinical CT images.

Purpose

For existing CT images acquired for various reasons other than osteoporosis, it might be beneficial to estimate areal BMD as assessed by dual-energy X-ray absorptiometry (DXA-based BMD) to ascertain the bone status based on DXA. In this study, we aimed to (1) develop an open-source reproducible measurement system to quantify DXA-based BMD from CT images and (2) validate its accuracy.

Methods

This study analyzed 75 pairs of hip CT and DXA images of women that were acquired for the preoperative assessment of total hip arthroplasty. From the CT images, the femur and a calibration phantom were automatically segmented using pre-trained codes/models available at https://​github.​com/​keisuke-uemura. The proximal femoral region was isolated by manually selected landmarks and was projected onto the coronal plane to measure the areal density (CT-aHU). The calibration phantom was employed to convert the CT-aHU into CT-aBMD. Each parameter was correlated with DXA-based BMD, and the residual errors of CT images to estimate the T-scores in DXA were calculated using the standard error of estimate (SEE).

Results

The correlation coefficients of DXA-based BMD with CT-aHU and CT-aBMD were 0.947 and 0.950, respectively (both p < 0.001). The SEE for quantifying the T-scores in DXA were 0.51 and 0.50 for CT-aHU and CT-aBMD, respectively.

Conclusion

With the method developed herein, CT permits estimation of the DXA-based BMD of the proximal femur within the standard DXA total hip region of interest with an SEE of 0.5 in T-scores. The radiation dose for CT acquisition needs consideration; therefore, our data do not provide a rationale for performing CT for screening osteoporosis. However, on CT images already acquired for clinical indications other than osteoporosis, researchers may use this open-source system to investigate osteoporosis status through the estimated DXA-based BMD of the proximal femur.
Literature
16.
go back to reference Styner M, Lee J, Chin B, et al 3D Segmentation in the clinic: a grand challenge II: MS lesion segmentation. 6 Styner M, Lee J, Chin B, et al 3D Segmentation in the clinic: a grand challenge II: MS lesion segmentation. 6
21.
go back to reference Landis JR, Koch GG (1977) The measurement of observer agreement for categorical data. Biometrics 33:159–174CrossRefPubMed Landis JR, Koch GG (1977) The measurement of observer agreement for categorical data. Biometrics 33:159–174CrossRefPubMed
Metadata
Title
Development of an open-source measurement system to assess the areal bone mineral density of the proximal femur from clinical CT images
Authors
Keisuke Uemura
Yoshito Otake
Masaki Takao
Hiroki Makino
Mazen Soufi
Makoto Iwasa
Nobuhiko Sugano
Yoshinobu Sato
Publication date
01-12-2022
Publisher
Springer London
Published in
Archives of Osteoporosis / Issue 1/2022
Print ISSN: 1862-3522
Electronic ISSN: 1862-3514
DOI
https://doi.org/10.1007/s11657-022-01063-3

Other articles of this Issue 1/2022

Archives of Osteoporosis 1/2022 Go to the issue