Effects of infusion tubing line lengths and syringe sizes on infusion system compliance: an experimental study using a syringe-type infusion pump at low flow rate

Authors: Kwan Young Hong, Eun Kyung Lee, Youyoung Kim, Dan-Cheong Choi, Jeong-Jin Min

Published in: Journal of Clinical Monitoring and Computing | Issue 5/2023

Login to get access

Abstract

Ideally, the flow delivery of an infusion system is proportional only to the rate of mechanical actuation of the syringe pump plunger. However, in the real world, overall infusion system compliance may be affected by components such as an extension of tubing lines, or different sizes of syringes. With higher compliance, there may be greater chances of flow irregularity. In this experimental study, we investigated the effects of lengths of infusion lines and syringe sizes on the compliance of syringe pumps with low flow rate (2 ml h−1). In the first experiment, infusion system compliance was measured in various settings by occlusion release. As the infusion tubing length and size of the syringe increased, the time to reach each pressure was delayed and the infusion system compliance increased. The contributions to system compliance from syringes were significantly greater compared to those of extended infusion lines. In the occlusion alarm experiment, the occlusion alarm could be delayed by 69.76 ± 3.98 min for the 50-ml syringe with a 560 cm infusion line set-up. In conclusion, the compliance of a syringe pump system increases as the loaded syringe size, or the length of the infusion tubing increases. The occlusion alarm may be much delayed and not useful in highly compliant systems with respect to the potential occlusion of the infusion system, so more attention is required when using a highly compliant infusion system.
Literature
22.
go back to reference Chu DK, Akl EA, Duda S, Solo K, Yaacoub S, Schünemann HJ, Chu DK, Akl EA, El-harakeh A, Bognanni A, Lotfi T, Loeb M, Hajizadeh A, Bak A, Izcovich A, Cuello-Garcia CA, Chen C, Harris DJ, Borowiack E, Chamseddine F, Schünemann F, Morgano GP, Muti Schünemann GEU, Chen G, Zhao H, Neumann I, Chan J, Khabsa J, Hneiny L, Harrison L, Smith M, Rizk N, Giorgi Rossi P, AbiHanna P, El-khoury R, Stalteri R, Baldeh T, Piggott T, Zhang Y, Saad Z, Khamis A, Reinap M, Duda S, Solo K, Yaacoub S, Schünemann HJ. Physical distancing, face masks, and eye protection to prevent person-to-person transmission of SARS-CoV-2 and COVID-19: a systematic review and meta-analysis. Lancet. 2020;395:1973–87. https://doi.org/10.1016/s0140-6736(20)31142-9.CrossRefPubMedPubMedCentral Chu DK, Akl EA, Duda S, Solo K, Yaacoub S, Schünemann HJ, Chu DK, Akl EA, El-harakeh A, Bognanni A, Lotfi T, Loeb M, Hajizadeh A, Bak A, Izcovich A, Cuello-Garcia CA, Chen C, Harris DJ, Borowiack E, Chamseddine F, Schünemann F, Morgano GP, Muti Schünemann GEU, Chen G, Zhao H, Neumann I, Chan J, Khabsa J, Hneiny L, Harrison L, Smith M, Rizk N, Giorgi Rossi P, AbiHanna P, El-khoury R, Stalteri R, Baldeh T, Piggott T, Zhang Y, Saad Z, Khamis A, Reinap M, Duda S, Solo K, Yaacoub S, Schünemann HJ. Physical distancing, face masks, and eye protection to prevent person-to-person transmission of SARS-CoV-2 and COVID-19: a systematic review and meta-analysis. Lancet. 2020;395:1973–87. https://​doi.​org/​10.​1016/​s0140-6736(20)31142-9.CrossRefPubMedPubMedCentral
Metadata
Title
Effects of infusion tubing line lengths and syringe sizes on infusion system compliance: an experimental study using a syringe-type infusion pump at low flow rate
Authors
Kwan Young Hong
Eun Kyung Lee
Youyoung Kim
Dan-Cheong Choi
Jeong-Jin Min
Publication date
17-01-2023
Publisher
Springer Netherlands
Published in
Journal of Clinical Monitoring and Computing / Issue 5/2023
Print ISSN: 1387-1307
Electronic ISSN: 1573-2614
DOI
https://doi.org/10.1007/s10877-023-00973-y

Other articles of this Issue 5/2023

Journal of Clinical Monitoring and Computing 5/2023 Go to the issue