01-12-2021 | Original Research

Simulation study on flow rate accuracy of infusion pumps in vibration conditions during emergency patient transport

Authors: Kwan Young Hong, You Yeon Kim, Seung Yeon Yoo, Jong-Hwan Lee, Duk Kyung Kim, Jeong-Jin Min

Published in: Journal of Clinical Monitoring and Computing | Issue 6/2021

Login to get access

Abstract

Infusion pumps are frequently used when transferring critically ill patients via patient transport cart, ambulance, or helicopter. However, the performance of various infusion pumps under these circumstances has not been explored. The aim of this study was to evaluate the flow rate accuracy of infusion pumps under various clinical vibration conditions. Experiments were conducted with four different types of pumps, including two conventional syringe pumps (Injectomat MC Agilia, Fresenius Kabi and TE-331, Terumo), one conventional peristaltic pump (Volumed μVP7000; Arcomed), and one new cylinder pump (H-100, Meinntech). The flow rate was measured using an infusion pump analyzer on a stable table (0 m/s2) for 1 h with 1 ml/h and 5 ml/h. Experiments were repeated in mild vibration (2 m/s2) (representing vibration of patients in a moving stretcher or ambulance), and in moderate vibration (6 m/s2) (representing vibration in helicopter transport). Any accidental bolus occurrence in extreme vibration situations (20 m/s2) was also analyzed. Simulated vibrations were reproduced by a custom-made vibration table. In the resting state without vibration and in mild vibration conditions, all pumps maintained good performance. However, in moderate vibration, flow rates in syringe pumps increased beyond their known error ranges, while flow rates in peristaltic pumps remained stable. In extreme vibration, accidental fluid bolus occurred in syringe pumps but not in peristaltic pumps. The newly developed cylinder pump maintained stable performance and was unaffected by external vibration environments.
Appendix
Available only for authorised users
Literature
1.
go back to reference Fanara B, Manzon C, Barbot O, Desmettre T, Capellier G. Recommendations for the intra-hospital transport of critically ill patients. Crit Care. 2010;14(3):R87.CrossRef Fanara B, Manzon C, Barbot O, Desmettre T, Capellier G. Recommendations for the intra-hospital transport of critically ill patients. Crit Care. 2010;14(3):R87.CrossRef
2.
go back to reference Duke GJ, Green JV. Outcome of critically ill patients undergoing interhospital transfer. Med J Aust. 2001;174(3):122–5.CrossRef Duke GJ, Green JV. Outcome of critically ill patients undergoing interhospital transfer. Med J Aust. 2001;174(3):122–5.CrossRef
3.
go back to reference Droogh JM, Smit M, Absalom AR, Ligtenberg JJ, Zijlstra JG. Transferring the critically ill patient: are we there yet? Crit Care. 2015;19(1):62.CrossRef Droogh JM, Smit M, Absalom AR, Ligtenberg JJ, Zijlstra JG. Transferring the critically ill patient: are we there yet? Crit Care. 2015;19(1):62.CrossRef
4.
go back to reference Fan E, MacDonald RD, Adhikari NK, Scales DC, Wax RS, Stewart TE, Ferguson ND. Outcomes of interfacility critical care adult patient transport: a systematic review. Crit Care. 2005;10(1):R6.CrossRef Fan E, MacDonald RD, Adhikari NK, Scales DC, Wax RS, Stewart TE, Ferguson ND. Outcomes of interfacility critical care adult patient transport: a systematic review. Crit Care. 2005;10(1):R6.CrossRef
5.
go back to reference Taylor J, Chulay J, Landers C, Hood W Jr, Abelmann W. Monitoring high-risk cardiac patients during transportation in hospital. The Lancet. 1970;296(7685):1205–8.CrossRef Taylor J, Chulay J, Landers C, Hood W Jr, Abelmann W. Monitoring high-risk cardiac patients during transportation in hospital. The Lancet. 1970;296(7685):1205–8.CrossRef
6.
go back to reference Kim UR, Peterfreund RA, Lovich MA. Drug infusion systems: technologies, performance, and pitfalls. Anesth Analg. 2017;124(5):1493–505.CrossRef Kim UR, Peterfreund RA, Lovich MA. Drug infusion systems: technologies, performance, and pitfalls. Anesth Analg. 2017;124(5):1493–505.CrossRef
7.
go back to reference Snijder RA, Konings MK, Lucas P, Egberts TC, Timmerman AD. Flow variability and its physical causes in infusion technology: a systematic review of in vitro measurement and modeling studies. Biomed Eng. 2015;60(4):277–300.CrossRef Snijder RA, Konings MK, Lucas P, Egberts TC, Timmerman AD. Flow variability and its physical causes in infusion technology: a systematic review of in vitro measurement and modeling studies. Biomed Eng. 2015;60(4):277–300.CrossRef
8.
go back to reference Husch M, Sullivan C, Rooney D, Barnard C, Fotis M, Clarke J, Noskin G. Insights from the sharp end of intravenous medication errors: implications for infusion pump technology. BMJ Qual Saf. 2005;14(2):80–6.CrossRef Husch M, Sullivan C, Rooney D, Barnard C, Fotis M, Clarke J, Noskin G. Insights from the sharp end of intravenous medication errors: implications for infusion pump technology. BMJ Qual Saf. 2005;14(2):80–6.CrossRef
10.
go back to reference Lee SB, Lee K-h, Lee JJ, Choi JS, Lee SW. Cylinder pump. Google Patents; 2013. Lee SB, Lee K-h, Lee JJ, Choi JS, Lee SW. Cylinder pump. Google Patents; 2013.
11.
go back to reference Yun JG, Jeung KW, Lee BK, Ryu HH, Lee HY, Kim MJ, Heo T, Min YI, You Y. Performance of an automated external defibrillator in a moving ambulance vehicle. Resuscitation. 2010;81(4):457–62.CrossRef Yun JG, Jeung KW, Lee BK, Ryu HH, Lee HY, Kim MJ, Heo T, Min YI, You Y. Performance of an automated external defibrillator in a moving ambulance vehicle. Resuscitation. 2010;81(4):457–62.CrossRef
12.
go back to reference Chung TN, Kim SW, Cho YS, Chung SP, Park I, Kim SH. Effect of vehicle speed on the quality of closed-chest compression during ambulance transport. Resuscitation. 2010;81(7):841–7.CrossRef Chung TN, Kim SW, Cho YS, Chung SP, Park I, Kim SH. Effect of vehicle speed on the quality of closed-chest compression during ambulance transport. Resuscitation. 2010;81(7):841–7.CrossRef
13.
go back to reference Je SM, You JS, Chung TN, Park YS, Chung SP, Park IC. Performance of an automated external defibrillator during simulated rotor-wing critical care transports. Resuscitation. 2011;82(4):454–8.CrossRef Je SM, You JS, Chung TN, Park YS, Chung SP, Park IC. Performance of an automated external defibrillator during simulated rotor-wing critical care transports. Resuscitation. 2011;82(4):454–8.CrossRef
17.
go back to reference Min JJ, Kim DK, Hong KY, Choi JW, Choi KY. Comparison of operator workloads associated with the single-unit Anyfusion® pump and the changeover from a syringe pump to an infusion pump. J Korean Med Sci. 2019;34(49):e314.CrossRef Min JJ, Kim DK, Hong KY, Choi JW, Choi KY. Comparison of operator workloads associated with the single-unit Anyfusion® pump and the changeover from a syringe pump to an infusion pump. J Korean Med Sci. 2019;34(49):e314.CrossRef
19.
go back to reference Batista E, e Sousa JA, Ribeiro Á, Martins L, Pereira M, Navas H Calibration of infusion pumps analyser. In: Journal of Physics: Conference Series. IOP Publishing; 2018. vol 9. p. 092003. Batista E, e Sousa JA, Ribeiro Á, Martins L, Pereira M, Navas H Calibration of infusion pumps analyser. In: Journal of Physics: Conference Series. IOP Publishing; 2018. vol 9. p. 092003.
20.
go back to reference Stokes D, Peacock J, Lewis R, Hutton P. The Ohmeda 9000 syringe pump. The first of a new generation of syringe drivers. Anaesthesia. 1990;45(12):1062–6.CrossRef Stokes D, Peacock J, Lewis R, Hutton P. The Ohmeda 9000 syringe pump. The first of a new generation of syringe drivers. Anaesthesia. 1990;45(12):1062–6.CrossRef
21.
go back to reference Dunster K, Colditz P. Flow continuity of infusion systems at low flow rates. Anaesth Intensive Care. 1995;23(5):605–9.CrossRef Dunster K, Colditz P. Flow continuity of infusion systems at low flow rates. Anaesth Intensive Care. 1995;23(5):605–9.CrossRef
22.
go back to reference Stull JC, Erenberg A, Leff RD. Flow rate variability from electronic infusion devices. Crit Care Med. 1988;16(9):888–91.CrossRef Stull JC, Erenberg A, Leff RD. Flow rate variability from electronic infusion devices. Crit Care Med. 1988;16(9):888–91.CrossRef
23.
go back to reference Connor SB, Quill TJ, Jacobs JR. Accuracy of drug infusion pumps under computer control. IEEE Trans Biomed Eng. 1992;39(9):980–2.CrossRef Connor SB, Quill TJ, Jacobs JR. Accuracy of drug infusion pumps under computer control. IEEE Trans Biomed Eng. 1992;39(9):980–2.CrossRef
24.
go back to reference Linden V, Palmer K, Reinhard J, Westman R, Ehren H, Granholm T, Frenckner B. Inter-hospital transportation of patients with severe acute respiratory failure on extracorporeal membrane oxygenation–national and international experience. Intensive Care Med. 2001;27(10):1643–8.CrossRef Linden V, Palmer K, Reinhard J, Westman R, Ehren H, Granholm T, Frenckner B. Inter-hospital transportation of patients with severe acute respiratory failure on extracorporeal membrane oxygenation–national and international experience. Intensive Care Med. 2001;27(10):1643–8.CrossRef
25.
go back to reference Coppola CP, Tyree M, Larry K, DiGeronimo R. A 22-year experience in global transport extracorporeal membrane oxygenation. J Pediatr Surg. 2008;43(1):46–52.CrossRef Coppola CP, Tyree M, Larry K, DiGeronimo R. A 22-year experience in global transport extracorporeal membrane oxygenation. J Pediatr Surg. 2008;43(1):46–52.CrossRef
26.
go back to reference Bryner B, Cooley E, Copenhaver W, Brierley K, Teman N, Landis D, Rycus P, Hemmila M, Napolitano LM, Haft J. Two decades’ experience with interfacility transport on extracorporeal membrane oxygenation. Ann Thorac Surg. 2014;98(4):1363–70.CrossRef Bryner B, Cooley E, Copenhaver W, Brierley K, Teman N, Landis D, Rycus P, Hemmila M, Napolitano LM, Haft J. Two decades’ experience with interfacility transport on extracorporeal membrane oxygenation. Ann Thorac Surg. 2014;98(4):1363–70.CrossRef
27.
go back to reference Capes DF, Dunster KR, Sunderland VB, McMillan D, Colditz PB, McDonald C. Fluctuations in syringe-pump infusions: association with blood pressure variations in infants. Am J Health Syst Pharm. 1995;52(15):1646–53.CrossRef Capes DF, Dunster KR, Sunderland VB, McMillan D, Colditz PB, McDonald C. Fluctuations in syringe-pump infusions: association with blood pressure variations in infants. Am J Health Syst Pharm. 1995;52(15):1646–53.CrossRef
28.
go back to reference Klem SA, Farrington JM, Leff RD. Influence of infusion pump operation and flow rate on hemodynamic stability during epinephrine infusion. Crit Care Med. 1993;21(8):1213–7.CrossRef Klem SA, Farrington JM, Leff RD. Influence of infusion pump operation and flow rate on hemodynamic stability during epinephrine infusion. Crit Care Med. 1993;21(8):1213–7.CrossRef
29.
go back to reference Pelletier NP, Fisher A. Infusion pump in UH60L/M flight in Afghanistan: why failures occur. Air Med J. 2017;36(5):248–57.CrossRef Pelletier NP, Fisher A. Infusion pump in UH60L/M flight in Afghanistan: why failures occur. Air Med J. 2017;36(5):248–57.CrossRef
30.
go back to reference van der Eijk AC, van Rens RM, Dankelman J, Smit BJ. A literature review on flow-rate variability in neonatal IV therapy. Pediatric Anesth. 2013;23(1):9–21.CrossRef van der Eijk AC, van Rens RM, Dankelman J, Smit BJ. A literature review on flow-rate variability in neonatal IV therapy. Pediatric Anesth. 2013;23(1):9–21.CrossRef
31.
go back to reference Kern H, Kuring A, Redlich U, Döpfmer U, Sims N, Spies C, Kox W. Downward movement of syringe pumps reduces syringe output. Br J Anaesth. 2001;86(6):828–31.CrossRef Kern H, Kuring A, Redlich U, Döpfmer U, Sims N, Spies C, Kox W. Downward movement of syringe pumps reduces syringe output. Br J Anaesth. 2001;86(6):828–31.CrossRef
32.
go back to reference Lönnqvist P, Löfqvist B. Design flaw can convert commercially available continuous syringe pumps to intermittent bolus injectors. Intensive Care Med. 1997;23(9):998–1001.CrossRef Lönnqvist P, Löfqvist B. Design flaw can convert commercially available continuous syringe pumps to intermittent bolus injectors. Intensive Care Med. 1997;23(9):998–1001.CrossRef
33.
go back to reference Weiss M, Bänziger O, Neff T, Fanconi S. Influence of infusion line compliance on drug delivery rate during acute line loop formation. Intensive Care Med. 2000;26(6):776–9.CrossRef Weiss M, Bänziger O, Neff T, Fanconi S. Influence of infusion line compliance on drug delivery rate during acute line loop formation. Intensive Care Med. 2000;26(6):776–9.CrossRef
34.
go back to reference Kim DW, Steward DJ. The effect of syringe size on the performance of an infusion pump. Pediatric Anesth. 1999;9(4):335–7.CrossRef Kim DW, Steward DJ. The effect of syringe size on the performance of an infusion pump. Pediatric Anesth. 1999;9(4):335–7.CrossRef
35.
go back to reference Lin JA. The effect of syringe size on reliability and safety of low-flow infusions. Pediatric Crit Care Med. 2009;10(5):592–6.CrossRef Lin JA. The effect of syringe size on reliability and safety of low-flow infusions. Pediatric Crit Care Med. 2009;10(5):592–6.CrossRef
36.
go back to reference Palmer KT, Griffin MJ, Bendall H, Pannett B, Coggon D. Prevalence and pattern of occupational exposure to whole body vibration in Great Britain: findings from a national survey. Occup Environ Med. 2000;57(4):229–36.CrossRef Palmer KT, Griffin MJ, Bendall H, Pannett B, Coggon D. Prevalence and pattern of occupational exposure to whole body vibration in Great Britain: findings from a national survey. Occup Environ Med. 2000;57(4):229–36.CrossRef
37.
go back to reference Bongers P, Hulshof C, Dljkstra L, Boshuizen H, Groenhout H, Valken E. Back pain and exposure to whole body vibration in helicopter pilots. Ergonomics. 1990;33(8):1007–266.CrossRef Bongers P, Hulshof C, Dljkstra L, Boshuizen H, Groenhout H, Valken E. Back pain and exposure to whole body vibration in helicopter pilots. Ergonomics. 1990;33(8):1007–266.CrossRef
38.
go back to reference Kim JH, Kim TK, In J, Lee DK, Lee S, Kang H. Assessment of risk of bias in quasi-randomized controlled trials and randomized controlled trials reported in the Korean Journal of Anesthesiology between 2010 and 2016. Korean J Anesthesiol. 2017;70(5):511.CrossRef Kim JH, Kim TK, In J, Lee DK, Lee S, Kang H. Assessment of risk of bias in quasi-randomized controlled trials and randomized controlled trials reported in the Korean Journal of Anesthesiology between 2010 and 2016. Korean J Anesthesiol. 2017;70(5):511.CrossRef
Metadata
Title
Simulation study on flow rate accuracy of infusion pumps in vibration conditions during emergency patient transport
Authors
Kwan Young Hong
You Yeon Kim
Seung Yeon Yoo
Jong-Hwan Lee
Duk Kyung Kim
Jeong-Jin Min
Publication date
01-12-2021
Publisher
Springer Netherlands
Published in
Journal of Clinical Monitoring and Computing / Issue 6/2021
Print ISSN: 1387-1307
Electronic ISSN: 1573-2614
DOI
https://doi.org/10.1007/s10877-020-00588-7

Other articles of this Issue 6/2021

Journal of Clinical Monitoring and Computing 6/2021 Go to the issue