01-12-2016 | Original Research

Comparison of the accuracy of hemoglobin point of care testing using HemoCue and GEM Premier 3000 with automated hematology analyzer in emergency room

Authors: Jan Zatloukal, Jiri Pouska, Jakub Kletecka, Richard Pradl, Jan Benes

Published in: Journal of Clinical Monitoring and Computing | Issue 6/2016

Login to get access

Abstract

The laboratory analysis provides accurate, but time consuming hemoglobin level estimation especially in the emergency setting. The reliability of time-sparing point of care devices (POCT) remains uncertain. We tested two POCT devices accuracy (HemoCue®201+ and Gem®Premier™3000) in routine emergency department workflow. Blood samples taken from patients admitted to the emergency department were analyzed for hemoglobin concentration using a laboratory reference Beckman Coulter LH 750 (HBLAB), the HemoCue (HBHC) and the Gem Premier 3000 (HBGEM). Pairwise comparison for each device and HbLAB was performed using correlation and the Bland–Altman methods. The reliability of transfusion decision was assessed using three-zone error grid. A total of 292 measurements were performed in 99 patients. Mean hemoglobin level were 115 ± 33, 110 ± 28 and 111 ± 30 g/l for HbHC, HbGEM and HbLAB respectively. A significant correlation was observed for both devices: HbHC versus HbLAB (r2 = 0.93, p < 0.001) and HBGEM versus HBLAB (r2 = 0.86, p < 0.001). The Bland–Altman method revealed bias of −3.7 g/l (limits of agreement −20.9 to 13.5) for HBHC and HBLAB and 2.5 g/l (−18.6 to 23.5) for HBGEM and HBLAB, which significantly differed between POCT devices (p < 0.001). Using the error grid methodology: 94 or 91 % of values (HbHC and HbGEM) fell in the zone of acceptable difference (A), whereas 0 and 1 % (HbHC and HbGEM) were unacceptable (zone C). The absolute accuracy of tested POCT devices was low though reaching a high level of correlation with laboratory measurement. The results of the Morey´s error grid were unfavorable for both POCT devices.
Literature
1.
go back to reference Valeri CR, Cassidy G, Pivacek LE, Ragno G, Lieberthal W, Crowley JP, Khuri SF, Loscalzo J. Anemia-induced increase in the bleeding time: implications for treatment of nonsurgical blood loss. Transfusion. 2001;41(8):977–83.CrossRefPubMed Valeri CR, Cassidy G, Pivacek LE, Ragno G, Lieberthal W, Crowley JP, Khuri SF, Loscalzo J. Anemia-induced increase in the bleeding time: implications for treatment of nonsurgical blood loss. Transfusion. 2001;41(8):977–83.CrossRefPubMed
2.
go back to reference Patel AJ, Wesley R, Leitman SF, Bryant BJ. Capillary versus venous haemoglobin determination in the assessment of healthy blood donors. Vox Sang. 2013;104(4):317–23.CrossRefPubMedPubMedCentral Patel AJ, Wesley R, Leitman SF, Bryant BJ. Capillary versus venous haemoglobin determination in the assessment of healthy blood donors. Vox Sang. 2013;104(4):317–23.CrossRefPubMedPubMedCentral
3.
go back to reference Skelton VA, Wijayasinghe N, Sharafudeen S, Sange A, Parry NS, Junghans C. Evaluation of point-of-care haemoglobin measuring devices: a comparison of Radical-7™ pulse co-oximetry, HemoCue(®) and laboratory haemoglobin measurements in obstetric patients*. Anaesthesia. 2013;68(1):40–5.CrossRefPubMed Skelton VA, Wijayasinghe N, Sharafudeen S, Sange A, Parry NS, Junghans C. Evaluation of point-of-care haemoglobin measuring devices: a comparison of Radical-7™ pulse co-oximetry, HemoCue(®) and laboratory haemoglobin measurements in obstetric patients*. Anaesthesia. 2013;68(1):40–5.CrossRefPubMed
4.
go back to reference Mimoz O, Frasca D, Médard A, Soubiron L, Debaene B, Dahyot-Fizelier C. Reliability of the HemoCue® hemoglobinometer in critically ill patients: a prospective observational study. Min Anestesiol. 2011;77(10):979–85. Mimoz O, Frasca D, Médard A, Soubiron L, Debaene B, Dahyot-Fizelier C. Reliability of the HemoCue® hemoglobinometer in critically ill patients: a prospective observational study. Min Anestesiol. 2011;77(10):979–85.
5.
go back to reference Adam I, Ahmed S, Mahmoud MH, Yassin MI. Comparison of HemoCue® hemoglobin-meter and automated hematology analyzer in measurement of hemoglobin levels in pregnant women at Khartoum hospital, Sudan. Diagn Pathol. 2012;7:30.CrossRefPubMedPubMedCentral Adam I, Ahmed S, Mahmoud MH, Yassin MI. Comparison of HemoCue® hemoglobin-meter and automated hematology analyzer in measurement of hemoglobin levels in pregnant women at Khartoum hospital, Sudan. Diagn Pathol. 2012;7:30.CrossRefPubMedPubMedCentral
6.
go back to reference Paiva ADA, Rondó PHC, Silva SSDB, Latorre MDRDO. Comparison between the HemoCue and an automated counter for measuring hemoglobin. Rev Saude Publica. 2004;38(4):585–7.CrossRef Paiva ADA, Rondó PHC, Silva SSDB, Latorre MDRDO. Comparison between the HemoCue and an automated counter for measuring hemoglobin. Rev Saude Publica. 2004;38(4):585–7.CrossRef
7.
go back to reference Giraud B, Frasca D, Debaene B, Mimoz O. Comparison of haemoglobin measurement methods in the operating theatre. Br J Anaesth. 2013;111(6):946–54.CrossRefPubMed Giraud B, Frasca D, Debaene B, Mimoz O. Comparison of haemoglobin measurement methods in the operating theatre. Br J Anaesth. 2013;111(6):946–54.CrossRefPubMed
8.
go back to reference Seguin P, Kleiber A, Chanavaz C, Morcet J, Mallédant Y. Determination of capillary hemoglobin levels using the HemoCue system in intensive care patients. J Crit Care. 2011;26(4):423–7.CrossRefPubMed Seguin P, Kleiber A, Chanavaz C, Morcet J, Mallédant Y. Determination of capillary hemoglobin levels using the HemoCue system in intensive care patients. J Crit Care. 2011;26(4):423–7.CrossRefPubMed
9.
go back to reference Ray JG, Post JR, Hamielec C. Use of a rapid arterial blood gas analyzer to estimate blood hemoglobin concentration among critically ill adults. Crit Care. 2002;6(1):72–5.CrossRefPubMed Ray JG, Post JR, Hamielec C. Use of a rapid arterial blood gas analyzer to estimate blood hemoglobin concentration among critically ill adults. Crit Care. 2002;6(1):72–5.CrossRefPubMed
10.
go back to reference Shah N, Osea EA, Martinez GJ. Accuracy of noninvasive hemoglobin and invasive point-of-care hemoglobin testing compared with a laboratory analyzer. Int J Lab Hematol. 2014;36(1):56–61.CrossRefPubMed Shah N, Osea EA, Martinez GJ. Accuracy of noninvasive hemoglobin and invasive point-of-care hemoglobin testing compared with a laboratory analyzer. Int J Lab Hematol. 2014;36(1):56–61.CrossRefPubMed
11.
go back to reference Richards N, Boyce H, Yentis S. Estimation of blood haemoglobin concentration using the HemoCue during caesarean section: the effect of sampling site. Int J Obstet Anesth. 2010;19(1):67–70.CrossRefPubMed Richards N, Boyce H, Yentis S. Estimation of blood haemoglobin concentration using the HemoCue during caesarean section: the effect of sampling site. Int J Obstet Anesth. 2010;19(1):67–70.CrossRefPubMed
12.
go back to reference Patel KP, Hay GW, Cheteri MK, Holt DW. Hemoglobin test result variability and cost analysis of eight different analyzers during open heart surgery. J Extra Corpor Technol. 2007;39(1):10–7.PubMedPubMedCentral Patel KP, Hay GW, Cheteri MK, Holt DW. Hemoglobin test result variability and cost analysis of eight different analyzers during open heart surgery. J Extra Corpor Technol. 2007;39(1):10–7.PubMedPubMedCentral
13.
go back to reference Spielmann N, Mauch J, Madjdpour C, Schmugge M, Weiss M, Haas T. Accuracy and precision of hemoglobin point-of-care testing during major pediatric surgery. Int J Lab Hematol. 2012;34(1):86–90.CrossRefPubMed Spielmann N, Mauch J, Madjdpour C, Schmugge M, Weiss M, Haas T. Accuracy and precision of hemoglobin point-of-care testing during major pediatric surgery. Int J Lab Hematol. 2012;34(1):86–90.CrossRefPubMed
14.
go back to reference Lardi AM, Hirst C, Mortimer AJ, McCollum CN. Evaluation of the HemoCue for measuring intra-operative haemoglobin concentrations: a comparison with the Coulter Max-M. Anaesthesia. 1998;53(4):349–52.CrossRefPubMed Lardi AM, Hirst C, Mortimer AJ, McCollum CN. Evaluation of the HemoCue for measuring intra-operative haemoglobin concentrations: a comparison with the Coulter Max-M. Anaesthesia. 1998;53(4):349–52.CrossRefPubMed
15.
go back to reference Neville RG. Evaluation of portable haemoglobinometer in general practice. Br Med J (Clin Res Ed). 1987;294(6582):1263–5.CrossRef Neville RG. Evaluation of portable haemoglobinometer in general practice. Br Med J (Clin Res Ed). 1987;294(6582):1263–5.CrossRef
16.
go back to reference Louw A, Lasserre N, Drouhin F, Thierry S, Lecuyer L, Caen D, Tenaillon A. Reliability of HemoCue in patients with gastrointestinal bleeding. Intensive Care Med. 2007;33(2):355–8.CrossRefPubMed Louw A, Lasserre N, Drouhin F, Thierry S, Lecuyer L, Caen D, Tenaillon A. Reliability of HemoCue in patients with gastrointestinal bleeding. Intensive Care Med. 2007;33(2):355–8.CrossRefPubMed
17.
go back to reference McNulty SE, Torjman M, Grodecki W, Marr A, Schieren H. A comparison of four bedside methods of hemoglobin assessment during cardiac surgery. Anesth Analg. 1995;81(6):1197–202.PubMed McNulty SE, Torjman M, Grodecki W, Marr A, Schieren H. A comparison of four bedside methods of hemoglobin assessment during cardiac surgery. Anesth Analg. 1995;81(6):1197–202.PubMed
18.
go back to reference Lamhaut L, Apriotesei R, Combes X, Lejay M, Carli P, Vivien B. Comparison of the accuracy of noninvasive hemoglobin monitoring by spectrophotometry (SpHb) and HemoCue® with automated laboratory hemoglobin measurement. Anesthesiology. 2011;115(3):548–54.CrossRefPubMed Lamhaut L, Apriotesei R, Combes X, Lejay M, Carli P, Vivien B. Comparison of the accuracy of noninvasive hemoglobin monitoring by spectrophotometry (SpHb) and HemoCue® with automated laboratory hemoglobin measurement. Anesthesiology. 2011;115(3):548–54.CrossRefPubMed
19.
go back to reference Gehring H, Hornberger C, Dibbelt L, Rothsigkeit A, Gerlach K, Schumacher J, Schmucker P. Accuracy of point-of-care-testing (POCT) for determining hemoglobin concentrations. Acta Anaesthesiol Scand. 2002;46(8):980–6.CrossRefPubMed Gehring H, Hornberger C, Dibbelt L, Rothsigkeit A, Gerlach K, Schumacher J, Schmucker P. Accuracy of point-of-care-testing (POCT) for determining hemoglobin concentrations. Acta Anaesthesiol Scand. 2002;46(8):980–6.CrossRefPubMed
20.
go back to reference Muñoz Gómez M, Naveira Abeigón E, Romero Ruiz A, Ramírez Ramírez G. Precision and accuracy of the immediate determination of hemoglobin using HemoCueB Hemoglobin in urgent, surgical, and critical patients. Rev Esp Anestesiol Reanim. 2003;50(7):332–9.PubMed Muñoz Gómez M, Naveira Abeigón E, Romero Ruiz A, Ramírez Ramírez G. Precision and accuracy of the immediate determination of hemoglobin using HemoCueB Hemoglobin in urgent, surgical, and critical patients. Rev Esp Anestesiol Reanim. 2003;50(7):332–9.PubMed
21.
go back to reference Morey TE, Gravenstein N, Rice MJ. Let’s think clinically instead of mathematically about device accuracy. Anesth Analg. 2011;113(1):89–91.CrossRefPubMed Morey TE, Gravenstein N, Rice MJ. Let’s think clinically instead of mathematically about device accuracy. Anesth Analg. 2011;113(1):89–91.CrossRefPubMed
22.
go back to reference American Society of Anesthesiologists Task Force on Perioperative Blood Transfusion and Adjuvant Therapies. Practice guidelines for perioperative blood transfusion and adjuvant therapies: an updated report by the American Society of Anesthesiologists Task Force on Perioperative Blood Transfusion and Adjuvant Therapies. Anesthesiology. 2006;105(1):198–208.CrossRef American Society of Anesthesiologists Task Force on Perioperative Blood Transfusion and Adjuvant Therapies. Practice guidelines for perioperative blood transfusion and adjuvant therapies: an updated report by the American Society of Anesthesiologists Task Force on Perioperative Blood Transfusion and Adjuvant Therapies. Anesthesiology. 2006;105(1):198–208.CrossRef
23.
go back to reference Spahn DR, Bouillon B, Cerny V, Coats TJ, Duranteau J, Fernández-Mondéjar E, Filipescu D, Hunt BJ, Komadina R, Nardi G, Neugebauer E, Ozier Y, Riddez L, Schultz A, Vincent JL, Rossaint R. Management of bleeding and coagulopathy following major trauma: an updated European guideline. Crit Care. 2013;17(2):R76.CrossRefPubMedPubMedCentral Spahn DR, Bouillon B, Cerny V, Coats TJ, Duranteau J, Fernández-Mondéjar E, Filipescu D, Hunt BJ, Komadina R, Nardi G, Neugebauer E, Ozier Y, Riddez L, Schultz A, Vincent JL, Rossaint R. Management of bleeding and coagulopathy following major trauma: an updated European guideline. Crit Care. 2013;17(2):R76.CrossRefPubMedPubMedCentral
24.
go back to reference Muñoz M, Romero A, Gómez JF, Manteca A, Naveira E, Ramírez G. Utility of point-of-care haemoglobin measurement in the HemoCue-B haemoglobin for the initial diagnosis of anaemia. Clin Lab Haematol. 2005;27(2):99–104.CrossRefPubMed Muñoz M, Romero A, Gómez JF, Manteca A, Naveira E, Ramírez G. Utility of point-of-care haemoglobin measurement in the HemoCue-B haemoglobin for the initial diagnosis of anaemia. Clin Lab Haematol. 2005;27(2):99–104.CrossRefPubMed
25.
go back to reference Gayat E, Bodin A, Sportiello C, Boisson M, Dreyfus JF, Mathieu E, Fischler M. Performance evaluation of a noninvasive hemoglobin monitoring device. Ann Emerg Med. 2011;57(4):330–3.CrossRefPubMed Gayat E, Bodin A, Sportiello C, Boisson M, Dreyfus JF, Mathieu E, Fischler M. Performance evaluation of a noninvasive hemoglobin monitoring device. Ann Emerg Med. 2011;57(4):330–3.CrossRefPubMed
26.
go back to reference Conway AM, Hinchliffe RF, Earland J, Anderson LM. Measurement of haemoglobin using single drops of skin puncture blood: is precision acceptable? J Clin Pathol. 1998;51(3):248–50.CrossRefPubMedPubMedCentral Conway AM, Hinchliffe RF, Earland J, Anderson LM. Measurement of haemoglobin using single drops of skin puncture blood: is precision acceptable? J Clin Pathol. 1998;51(3):248–50.CrossRefPubMedPubMedCentral
27.
go back to reference Jaeger M, Ashbury T, Adams M, Duncan P. Perioperative on-site haemoglobin determination: as accurate as laboratory values? Can J Anaesth. 1996;43(8):795–8.CrossRefPubMed Jaeger M, Ashbury T, Adams M, Duncan P. Perioperative on-site haemoglobin determination: as accurate as laboratory values? Can J Anaesth. 1996;43(8):795–8.CrossRefPubMed
Metadata
Title
Comparison of the accuracy of hemoglobin point of care testing using HemoCue and GEM Premier 3000 with automated hematology analyzer in emergency room
Authors
Jan Zatloukal
Jiri Pouska
Jakub Kletecka
Richard Pradl
Jan Benes
Publication date
01-12-2016
Publisher
Springer Netherlands
Published in
Journal of Clinical Monitoring and Computing / Issue 6/2016
Print ISSN: 1387-1307
Electronic ISSN: 1573-2614
DOI
https://doi.org/10.1007/s10877-015-9799-z

Other articles of this Issue 6/2016

Journal of Clinical Monitoring and Computing 6/2016 Go to the issue