Assessment of noninvasive acoustic respiration rate monitoring in patients admitted to an Emergency Department for drug or alcoholic poisoning

Authors: Youcef Guechi, Amélie Pichot, Denis Frasca, Fatima Rayeh-Pelardy, Jean-Yves Lardeur, Olivier Mimoz

Published in: Journal of Clinical Monitoring and Computing | Issue 6/2015

Login to get access

Abstract

To compare respiration rate measurement by an acoustic method and thoracic impedance to capnometry as the reference method, in patients at the Emergency Department after drug or alcoholic poisoning. In this observational study, 30 patients aged 18 or older, hospitalized at the Emergency Department for drug or alcoholic poisoning, without any contraindication to a face mask and/or a cervical acoustic sensor, were included in the study. They benefited from a simultaneous recording of their respiration rate by the acoustic method (RRa®, Masimo Corp., Irvine, CA, USA), by thoracic impedance (Philips Intellivue® MP2, Suresnes, France) and by capnometry (Capnostream® 20, Oridion, Jerusalem, Israël) through a face mask (Capnomask®, Mediplus Ltd, Raleigh, NC, USA) for 40–60 min. Of the 86,578 triplets collected, 77,155 (89.1 %) were exploitable. Median (range) respiration rate measured by capnometry was 18 (7–29) bpm. Compared to capnometry, bias and limits of agreement were 0.1 ± 3.8 bpm for the acoustic method and 0.3 ± 5.5 bpm for thoracic impedance. The proportions of RR values collected by acoustic method or by thoracic impedance which differed over 10 or 20 % during more than 15 s, compared to capnometry, were 8.3 versus 14.3, and 1.5 versus 3.8 %, respectively (p < 0.0001). The acoustic sensor had to be repositioned on three patients. For 11 patients, the Capnomask® was removed several times. In patients with drug or alcoholic poisoning, the acoustic method seems more accurate than thoracic impedance and better tolerated than face mask capnometry.
Literature
1.
2.
go back to reference Bordet R, Gautier S, Le Louet H, et al. Analysis of the direct cost of adverse drug reactions in hospitalised patients. Eur J Clin Pharmacol. 2001;56:935–41.CrossRefPubMed Bordet R, Gautier S, Le Louet H, et al. Analysis of the direct cost of adverse drug reactions in hospitalised patients. Eur J Clin Pharmacol. 2001;56:935–41.CrossRefPubMed
3.
go back to reference Folke M, Cernerud L, Ekström M, et al. Critical review of non-invasive respiratory monitoring in medical care. Med Biol Eng Comput. 2003;41:377–83.CrossRefPubMed Folke M, Cernerud L, Ekström M, et al. Critical review of non-invasive respiratory monitoring in medical care. Med Biol Eng Comput. 2003;41:377–83.CrossRefPubMed
4.
go back to reference Keidan I, Gravenstein D, Berkenstadt H, et al. Supplemental oxygen compromises the use of pulse oximetry for detection of apnea and hypoventilation during sedation in simulated pediatric patients. Pediatrics. 2008;122:293–8.CrossRefPubMed Keidan I, Gravenstein D, Berkenstadt H, et al. Supplemental oxygen compromises the use of pulse oximetry for detection of apnea and hypoventilation during sedation in simulated pediatric patients. Pediatrics. 2008;122:293–8.CrossRefPubMed
5.
go back to reference Fu ES, Downs JB, Schweiger JW, et al. Supplemental oxygen impairs detection of hypoventilation by pulse oximetry. Chest. 2004;126:1552–8.CrossRefPubMed Fu ES, Downs JB, Schweiger JW, et al. Supplemental oxygen impairs detection of hypoventilation by pulse oximetry. Chest. 2004;126:1552–8.CrossRefPubMed
6.
go back to reference Petterson MT, Begnoche VL, Graybeal JM. The effect of motion on pulse oximetry and its clinical significance. Anesth Analg. 2007;105:S78–84.CrossRefPubMed Petterson MT, Begnoche VL, Graybeal JM. The effect of motion on pulse oximetry and its clinical significance. Anesth Analg. 2007;105:S78–84.CrossRefPubMed
7.
go back to reference Wilkinson JN, Thanawala VU. Thoracic impedance monitoring of respiratory rate during sedation—Is it safe? Anaesthesia. 2009;64:455–6.CrossRefPubMed Wilkinson JN, Thanawala VU. Thoracic impedance monitoring of respiratory rate during sedation—Is it safe? Anaesthesia. 2009;64:455–6.CrossRefPubMed
8.
go back to reference Cohen KP, Ladd WM, Beams DM, et al. Comparison of impedance and inductance ventilation sensors on adults during breathing, motion, and simulated airway obstruction. IEEE Trans Biomed Eng. 1997;44:555–66.CrossRefPubMed Cohen KP, Ladd WM, Beams DM, et al. Comparison of impedance and inductance ventilation sensors on adults during breathing, motion, and simulated airway obstruction. IEEE Trans Biomed Eng. 1997;44:555–66.CrossRefPubMed
9.
go back to reference Drummond GB, Nimmo AF, Elton RA. Thoracic impedance used for measuring chest wall movement in postoperative patients. Br J Anaesth. 1996;77:327–32.CrossRefPubMed Drummond GB, Nimmo AF, Elton RA. Thoracic impedance used for measuring chest wall movement in postoperative patients. Br J Anaesth. 1996;77:327–32.CrossRefPubMed
10.
go back to reference Brouillette RT, Morrow AS, Weese-Mayer DE, et al. Comparison of respiratory inductive plethysmography and thoracic impedance for apnea monitoring. J Pediatr. 1987;111:377–83.CrossRefPubMed Brouillette RT, Morrow AS, Weese-Mayer DE, et al. Comparison of respiratory inductive plethysmography and thoracic impedance for apnea monitoring. J Pediatr. 1987;111:377–83.CrossRefPubMed
11.
go back to reference Mimoz O, Benard T, Gaucher A, et al. Accuracy of respiratory rate monitoring using a non-invasive acoustic method after general anaesthesia. Br J Anaesth. 2012;108:872–5.CrossRefPubMed Mimoz O, Benard T, Gaucher A, et al. Accuracy of respiratory rate monitoring using a non-invasive acoustic method after general anaesthesia. Br J Anaesth. 2012;108:872–5.CrossRefPubMed
12.
go back to reference Ramsay MAE, Usman M, Lagow E, et al. The accuracy, precision and reliability of measuring ventilatory rate and detecting ventilatory pause by rainbow acoustic monitoring and capnometry. Anesth Analg. 2013;117:69–75.CrossRefPubMed Ramsay MAE, Usman M, Lagow E, et al. The accuracy, precision and reliability of measuring ventilatory rate and detecting ventilatory pause by rainbow acoustic monitoring and capnometry. Anesth Analg. 2013;117:69–75.CrossRefPubMed
13.
go back to reference Autet L, Frasca D, Pinsard M, et al. Evaluation of acoustic respiration rate monitoring following extubation in ICU patients. Br J Anaesth. 2014;113:195–7.CrossRefPubMed Autet L, Frasca D, Pinsard M, et al. Evaluation of acoustic respiration rate monitoring following extubation in ICU patients. Br J Anaesth. 2014;113:195–7.CrossRefPubMed
14.
go back to reference Bland JM, Altman DG. Agreement between methods of measurement with multiple observations per individual. J Biopharm Stat. 2007;17:571–82.CrossRefPubMed Bland JM, Altman DG. Agreement between methods of measurement with multiple observations per individual. J Biopharm Stat. 2007;17:571–82.CrossRefPubMed
15.
go back to reference Critchley LA, Lee A, Ho AM-H. A critical review of the ability of continuous cardiac output monitors to measure trends in cardiac output. Anesth Analg. 2010;111:1180–92.CrossRefPubMed Critchley LA, Lee A, Ho AM-H. A critical review of the ability of continuous cardiac output monitors to measure trends in cardiac output. Anesth Analg. 2010;111:1180–92.CrossRefPubMed
16.
go back to reference Gaucher A, Frasca D, Mimoz O, et al. Accuracy of respiratory rate monitoring by capnometry using the Capnomask® in extubated patients receiving supplemental oxygen after surgery. Br J Anaesth. 2012;108:316–20.CrossRefPubMed Gaucher A, Frasca D, Mimoz O, et al. Accuracy of respiratory rate monitoring by capnometry using the Capnomask® in extubated patients receiving supplemental oxygen after surgery. Br J Anaesth. 2012;108:316–20.CrossRefPubMed
17.
go back to reference Sesay M, Tauzin-Fin P, Verdonck O, et al. Clinical evaluation of the Capnomask in the supine vs. prone position during monitored anaesthesia care. Eur J Anaesthesiol. 2008;25:769–71.CrossRefPubMed Sesay M, Tauzin-Fin P, Verdonck O, et al. Clinical evaluation of the Capnomask in the supine vs. prone position during monitored anaesthesia care. Eur J Anaesthesiol. 2008;25:769–71.CrossRefPubMed
18.
go back to reference Maddox RR, Williams CK, Oglesby H, et al. Clinical experience with patient-controlled analgesia using continuous respiratory monitoring and a smart infusion system. Am J Health Syst Pharm. 2006;63:157–64.CrossRefPubMed Maddox RR, Williams CK, Oglesby H, et al. Clinical experience with patient-controlled analgesia using continuous respiratory monitoring and a smart infusion system. Am J Health Syst Pharm. 2006;63:157–64.CrossRefPubMed
19.
go back to reference Friesen RH, Alswang M. End-tidal PCO2 monitoring via nasal cannulae in pediatric patients: accuracy and sources of error. J Clin Monit. 1996;12:155–9.CrossRefPubMed Friesen RH, Alswang M. End-tidal PCO2 monitoring via nasal cannulae in pediatric patients: accuracy and sources of error. J Clin Monit. 1996;12:155–9.CrossRefPubMed
20.
go back to reference Goudra B, Penugonda L, Speck R, et al. Comparison of acoustic respiration rate, impedance pneumography and capnometry monitors for respiration rate accuracy and apnea detection during GI endoscopy anesthesia. Open J Anesthesiol. 2013;3:74–9.CrossRef Goudra B, Penugonda L, Speck R, et al. Comparison of acoustic respiration rate, impedance pneumography and capnometry monitors for respiration rate accuracy and apnea detection during GI endoscopy anesthesia. Open J Anesthesiol. 2013;3:74–9.CrossRef
21.
go back to reference Voscopoulos CJ, MacNabb CM, Freeman J, et al. Continuous noninvasive respiratory volume monitoring for the identification of patients at risk for opioid-induced respiratory depression and obstructive breathing patterns. J Trauma Acute Care Surg. 2014;77:S208–15.CrossRefPubMed Voscopoulos CJ, MacNabb CM, Freeman J, et al. Continuous noninvasive respiratory volume monitoring for the identification of patients at risk for opioid-induced respiratory depression and obstructive breathing patterns. J Trauma Acute Care Surg. 2014;77:S208–15.CrossRefPubMed
22.
go back to reference Voscopoulos CJ, MacNabb CM, Brayanov J, et al. The evaluation of a non-invasive respiratory volume monitor in surgical patients undergoing elective surgery with general anesthesia. J Clin Monit Comput. 2014. doi:10.1007/s10877-014-9596-0. Voscopoulos CJ, MacNabb CM, Brayanov J, et al. The evaluation of a non-invasive respiratory volume monitor in surgical patients undergoing elective surgery with general anesthesia. J Clin Monit Comput. 2014. doi:10.​1007/​s10877-014-9596-0.
Metadata
Title
Assessment of noninvasive acoustic respiration rate monitoring in patients admitted to an Emergency Department for drug or alcoholic poisoning
Authors
Youcef Guechi
Amélie Pichot
Denis Frasca
Fatima Rayeh-Pelardy
Jean-Yves Lardeur
Olivier Mimoz
Publication date
01-12-2015
Publisher
Springer Netherlands
Published in
Journal of Clinical Monitoring and Computing / Issue 6/2015
Print ISSN: 1387-1307
Electronic ISSN: 1573-2614
DOI
https://doi.org/10.1007/s10877-015-9658-y

Other articles of this Issue 6/2015

Journal of Clinical Monitoring and Computing 6/2015 Go to the issue