Open Access 01-10-2020 | Atrial Fibrillation

Comparing rates of atrioesophageal fistula with contact force-sensing and non-contact force-sensing catheters: analysis of post-market safety surveillance data

Authors: Hugh Calkins, Andrea Natale, Tara Gomez, Alex Etlin, Moe Bishara

Published in: Journal of Interventional Cardiac Electrophysiology | Issue 1/2020

Login to get access

Abstract

Purpose

There is limited data on the specific incidence of serious adverse events, such as atrioesophageal fistula (AEF), associated with either contact force (CF) or non-CF ablation catheters. Since the actual number of procedures performed with each type of catheter is unknown, making direct comparisons is difficult. The purpose of this study was to assess the incidence of AEF associated with the use of CF and non-CF catheters. Additionally, we aimed to understand the workflow present in confirmed AEF cases voluntarily provided by physicians.

Methods

The number of AEFs for 2014–2017 associated with each type of catheter was extracted from an ablation device manufacturer’s complaint database. Proprietary device sales data, a proxy for the total number of procedures, were used as the denominator to calculate the incidence rates. Additional survey and workflow data were systematically reviewed.

Results

Both CF and non-CF ablation catheters have comparably low incidence of AEF (0.006 ± 0.003% and 0.005 ± 0.003%, respectively, p = 0.69). CF catheters are the catheter of choice for left atrium (LA) procedures which pose the greatest risk for AEF injury. Retrospective analysis of seven AEF cases demonstrated that high power and force and long RF duration were delivered on the posterior wall of the left atrium in all cases.

Conclusions

CF and non-CF ablation catheters were found to have similar AEF incidence, despite CF catheters being the catheter of choice for LA procedures. More investigation is needed to understand the range of parameters which may create risk for AEF.
Literature
1.
go back to reference Reddy VY, Shah D, Kautzner J, Schmidt B, Saoudi N, Herrera C, et al. The relationship between contact force and clinical outcome during radiofrequency catheter ablation of atrial fibrillation in the TOCCATA study. HRTHM. Elsevier Inc.; 2012;9:1789–95. Reddy VY, Shah D, Kautzner J, Schmidt B, Saoudi N, Herrera C, et al. The relationship between contact force and clinical outcome during radiofrequency catheter ablation of atrial fibrillation in the TOCCATA study. HRTHM. Elsevier Inc.; 2012;9:1789–95.
2.
go back to reference Marijon E, Fazaa S, Guy-Moyat B, Bouzeman A, Providencia R, Treguer F, et al. Real-time contact force sensing for pulmonary vein isolation in the setting of paroxysmal atrial fibrillation: Procedural and One-Year Results. J Cardiovasc Electrophysiol. 2013. Marijon E, Fazaa S, Guy-Moyat B, Bouzeman A, Providencia R, Treguer F, et al. Real-time contact force sensing for pulmonary vein isolation in the setting of paroxysmal atrial fibrillation: Procedural and One-Year Results. J Cardiovasc Electrophysiol. 2013.
3.
go back to reference Natale A, Reddy VY, Monir G, Wilber DJ, Lindsay BD, McElderry HT, et al. Paroxysmal AF catheter ablation with a contact force sensing catheter: Results of the prospective, multicenter SMART-AF trial. J Am Coll Cardiol. 2014;64:647–56.CrossRef Natale A, Reddy VY, Monir G, Wilber DJ, Lindsay BD, McElderry HT, et al. Paroxysmal AF catheter ablation with a contact force sensing catheter: Results of the prospective, multicenter SMART-AF trial. J Am Coll Cardiol. 2014;64:647–56.CrossRef
4.
go back to reference Calkins H, Hindricks G, Cappato R, Kim YH, Saad EB, Aguinaga L, et al. 2017 HRS/EHRA/ECAS/APHRS/SOLAECE expert consensus statement on catheter and surgical ablation of atrial fibrillation. Hear Rhythm. Elsevier Inc.; 2017;14:e275–444. Calkins H, Hindricks G, Cappato R, Kim YH, Saad EB, Aguinaga L, et al. 2017 HRS/EHRA/ECAS/APHRS/SOLAECE expert consensus statement on catheter and surgical ablation of atrial fibrillation. Hear Rhythm. Elsevier Inc.; 2017;14:e275–444.
5.
go back to reference Pambrun T, Combes S, Sousa P, Le Bloa M, El Bouazzaoui R, Grand-Larrieu D, et al. Contact-force guided single-catheter approach for pulmonary vein isolation: Feasibility, outcomes, and cost-effectiveness. Hear Rhythm United States. 2017;14:331–8.CrossRef Pambrun T, Combes S, Sousa P, Le Bloa M, El Bouazzaoui R, Grand-Larrieu D, et al. Contact-force guided single-catheter approach for pulmonary vein isolation: Feasibility, outcomes, and cost-effectiveness. Hear Rhythm United States. 2017;14:331–8.CrossRef
6.
go back to reference Martinek M, Lemes C, Sigmund E, Derndorfer M, Aichinger J, Winter S, et al. Clinical impact of an open-irrigated radiofrequency catheter with direct force measurement on atrial fibrillation ablation. PACE - Pacing Clin Electrophysiol. 2012;35:1312–8.CrossRef Martinek M, Lemes C, Sigmund E, Derndorfer M, Aichinger J, Winter S, et al. Clinical impact of an open-irrigated radiofrequency catheter with direct force measurement on atrial fibrillation ablation. PACE - Pacing Clin Electrophysiol. 2012;35:1312–8.CrossRef
7.
go back to reference Liu E, Shehata M, Liu T, Amorn A, Cingolani E, Kannarkat V, et al. Prevention of esophageal thermal injury during radiofrequency ablation for atrial fibrillation. J Interv Card Electrophysiol Netherlands. 2012;35:35–44.CrossRef Liu E, Shehata M, Liu T, Amorn A, Cingolani E, Kannarkat V, et al. Prevention of esophageal thermal injury during radiofrequency ablation for atrial fibrillation. J Interv Card Electrophysiol Netherlands. 2012;35:35–44.CrossRef
8.
go back to reference Black-Maier E, Pokorney SD, Barnett AS, Zeitler EP, Sun AY, Jackson KP, et al. Risk of atrioesophageal fistula formation with contact force–sensing catheters. Hear Rhythm. Elsevier Inc.; 2017;14:1328–33. Black-Maier E, Pokorney SD, Barnett AS, Zeitler EP, Sun AY, Jackson KP, et al. Risk of atrioesophageal fistula formation with contact force–sensing catheters. Hear Rhythm. Elsevier Inc.; 2017;14:1328–33.
9.
go back to reference Deshmukh A, Patel NJ, Pant S, Shah N, Chothani A, Mehta K, et al. In-hospital complications associated with catheter ablation of atrial fibrillation in the United States between 2000 and 2010: analysis of 93 801 procedures. Circulation United States. 2013;128:2104–12.CrossRef Deshmukh A, Patel NJ, Pant S, Shah N, Chothani A, Mehta K, et al. In-hospital complications associated with catheter ablation of atrial fibrillation in the United States between 2000 and 2010: analysis of 93 801 procedures. Circulation United States. 2013;128:2104–12.CrossRef
10.
go back to reference Tripathi B, Arora S, Kumar V, Abdelrahman M, Lahewala S, Dave M, et al. Temporal trends of in-hospital complications associated with catheter ablation of atrial fibrillation in the United States: An update from Nationwide Inpatient Sample database (2011-2014). J Cardiovasc Electrophysiol United States. 2018;29:715–24.CrossRef Tripathi B, Arora S, Kumar V, Abdelrahman M, Lahewala S, Dave M, et al. Temporal trends of in-hospital complications associated with catheter ablation of atrial fibrillation in the United States: An update from Nationwide Inpatient Sample database (2011-2014). J Cardiovasc Electrophysiol United States. 2018;29:715–24.CrossRef
11.
go back to reference Perna F, Heist EK, Danik SB, Barrett CD, Ruskin JN, Mansour M. Assessment of catheter tip contact force resulting in cardiac perforation in swine atria using force sensing technology. Circ Arrhythm Electrophysiol United States. 2011;4:218–24.CrossRef Perna F, Heist EK, Danik SB, Barrett CD, Ruskin JN, Mansour M. Assessment of catheter tip contact force resulting in cardiac perforation in swine atria using force sensing technology. Circ Arrhythm Electrophysiol United States. 2011;4:218–24.CrossRef
12.
go back to reference Kuck K, Reddy VY, Schmidt B, Natale A, Neuzil P, Saoudi N, et al. A novel radiofrequency ablation catheter using contact force sensing : Toccata study. HRTHM. Elsevier Inc.; 2012;9:18–23. Kuck K, Reddy VY, Schmidt B, Natale A, Neuzil P, Saoudi N, et al. A novel radiofrequency ablation catheter using contact force sensing : Toccata study. HRTHM. Elsevier Inc.; 2012;9:18–23.
13.
go back to reference Thiagalingam A, D’Avila A, Foley L, Guerrero JL, Lambert H, Leo G, et al. Importance of catheter contact force during irrigated radiofrequency ablation: evaluation in a porcine ex vivo model using a force-sensing catheter. J Cardiovasc Electrophysiol United States. 2010;21:806–11. Thiagalingam A, D’Avila A, Foley L, Guerrero JL, Lambert H, Leo G, et al. Importance of catheter contact force during irrigated radiofrequency ablation: evaluation in a porcine ex vivo model using a force-sensing catheter. J Cardiovasc Electrophysiol United States. 2010;21:806–11.
14.
go back to reference Hoffmayer KS, Gerstenfeld EP. Contact force-sensing catheters. Curr Opin Cardiol United States. 2015;30:74–80.CrossRef Hoffmayer KS, Gerstenfeld EP. Contact force-sensing catheters. Curr Opin Cardiol United States. 2015;30:74–80.CrossRef
Metadata
Title
Comparing rates of atrioesophageal fistula with contact force-sensing and non-contact force-sensing catheters: analysis of post-market safety surveillance data
Authors
Hugh Calkins
Andrea Natale
Tara Gomez
Alex Etlin
Moe Bishara
Publication date
01-10-2020
Publisher
Springer US
Published in
Journal of Interventional Cardiac Electrophysiology / Issue 1/2020
Print ISSN: 1383-875X
Electronic ISSN: 1572-8595
DOI
https://doi.org/10.1007/s10840-019-00653-5

Other articles of this Issue 1/2020

Journal of Interventional Cardiac Electrophysiology 1/2020 Go to the issue