Initial experience with a novel re-sterilisable decapolar electrophysiology catheter

Authors: Anna L. Wilson, Kellie A. Timmins, Rosemary F. Allen, Paula M. Jones, Sacha L. Ware, Dean Boddington, Janice Swampillai, Spencer C. Heald, Martin K. Stiles

Published in: Journal of Interventional Cardiac Electrophysiology | Issue 2/2020

Login to get access

Abstract

Purpose

Diagnostic electrophysiology catheters are single-use and their expense can act as a barrier to their use. A decapolar catheter with a re-sterilisable handle and inner core was developed, along with a disposable outer sheath. This initial experience evaluated the performance of the Khelix partially recyclable steerable decapolar electrophysiology catheter.

Methods

Sequential patients requiring a decapolar steerable catheter were included in the study. Parameters such as time to positioning from right femoral vein, fluoroscopy time and dose, pacing and sensing thresholds, and the requirement for reposition and/or support sheath were recorded. A retrospective analysis from cases using a standard disposable decapolar catheter was used to compare CS cannulation success, support sheath requirement and cost.

Results

11 re-sterilisable catheter handles were used in 64 patients, each time with a disposable sheath, at two New Zealand centres from December 2017 to June 2018. Pre-defined safety measures were met in all patients, with no catheter-related adverse events. The catheter was successfully positioned in the coronary sinus for 59 cases (92%). Based on initial manufacturer recommendations, the inner recyclable portion of the catheter was able to be re-used successfully up to 12 times with re-sterilisation between cases.

Conclusions

This initial study of an innovative, partially reusable, steerable decapolar electrophysiology catheter demonstrates that it may be used safely and successfully. Use of such a catheter may decrease the cost required for electrophysiology studies.
Literature
1.
go back to reference Jackman WM, Friday KJ, Yeung-Lai-Wah JA, Fitzgerald DM, Beck B, Bowman AJ, et al. New catheter technique for recording left free-wall accessory atrioventricular pathway activation. Identification of pathway fiber orientation. Circulation. 1988;78(3):598–611.CrossRef Jackman WM, Friday KJ, Yeung-Lai-Wah JA, Fitzgerald DM, Beck B, Bowman AJ, et al. New catheter technique for recording left free-wall accessory atrioventricular pathway activation. Identification of pathway fiber orientation. Circulation. 1988;78(3):598–611.CrossRef
2.
go back to reference Joseph JP, Rajappan K. Radiofrequency ablation of cardiac arrhythmias: past, present and future. QJM-INT J MED. 2012;105(4):303–14. Joseph JP, Rajappan K. Radiofrequency ablation of cardiac arrhythmias: past, present and future. QJM-INT J MED. 2012;105(4):303–14.
4.
go back to reference Haines DE, Beheiry S, Akar JG, Baker JL, Beinborn D, Beshai JF, et al. Heart Rhythm Society expert Consensus statement on electrophysiology laboratory standards: process, protocols, equipment, personnel, and safety. Heart Rhythm. 2014;11(8):e9–e51.CrossRef Haines DE, Beheiry S, Akar JG, Baker JL, Beinborn D, Beshai JF, et al. Heart Rhythm Society expert Consensus statement on electrophysiology laboratory standards: process, protocols, equipment, personnel, and safety. Heart Rhythm. 2014;11(8):e9–e51.CrossRef
5.
go back to reference Mont L, Brugada J. Electrophysiology: it is time to simplify! Europace. 2009;11(8):985–6.CrossRef Mont L, Brugada J. Electrophysiology: it is time to simplify! Europace. 2009;11(8):985–6.CrossRef
6.
go back to reference Liew R, Baker V, Richmond L, Rajappan K, Gupta D, Finlay M, et al. A randomized-controlled trial comparing conventional with minimal catheter approaches for the mapping and ablation of regular supraventricular tachycardias. Europace. 2009;11:1057–64.CrossRef Liew R, Baker V, Richmond L, Rajappan K, Gupta D, Finlay M, et al. A randomized-controlled trial comparing conventional with minimal catheter approaches for the mapping and ablation of regular supraventricular tachycardias. Europace. 2009;11:1057–64.CrossRef
7.
go back to reference Avitall B, Khan M, Krum D, Jazayeri M, Hare J. Repeated use of ablation catheters: a prospective study. J Am Coll Cardiol. 1993;22:1367–72.CrossRef Avitall B, Khan M, Krum D, Jazayeri M, Hare J. Repeated use of ablation catheters: a prospective study. J Am Coll Cardiol. 1993;22:1367–72.CrossRef
8.
go back to reference Shandling AH, Rieders D, Edwards M. Use of a unique long pre-curved sheath to facilitate femoral placement of coronary sinus catheters. IPEJ. 2009;9(5):247–50. Shandling AH, Rieders D, Edwards M. Use of a unique long pre-curved sheath to facilitate femoral placement of coronary sinus catheters. IPEJ. 2009;9(5):247–50.
9.
go back to reference Skala T, Taborsky M. Electromechanical mapping in electrophysiology and beyond. Czech Society of Cardiology. 2015;57(6):e470–82. Skala T, Taborsky M. Electromechanical mapping in electrophysiology and beyond. Czech Society of Cardiology. 2015;57(6):e470–82.
10.
go back to reference Kuck KH, Schluter M. Single-catheter approach to radiofrequency current ablation of left-sided accessory pathways in patients with Wolff–Parkinson–White syndrome. Circulation. 1991;84:2366–75.CrossRef Kuck KH, Schluter M. Single-catheter approach to radiofrequency current ablation of left-sided accessory pathways in patients with Wolff–Parkinson–White syndrome. Circulation. 1991;84:2366–75.CrossRef
11.
go back to reference Kleman JM, Pinski SL, Sterba R, et al. A technique for rapid transfemoral catheterization of the coronary sinus with multielectrode catheters. Pacing Clin Electrophysiol. 1994;17(11 Pt1):1737–40.CrossRef Kleman JM, Pinski SL, Sterba R, et al. A technique for rapid transfemoral catheterization of the coronary sinus with multielectrode catheters. Pacing Clin Electrophysiol. 1994;17(11 Pt1):1737–40.CrossRef
12.
go back to reference Kuniyoshi RR, Sternick EB, Nadalin E, Hachul DT. Reprocessing of medical products in electrophysiology. Arq Bras Cardiol. 2017;108(2):169–72.PubMedPubMedCentral Kuniyoshi RR, Sternick EB, Nadalin E, Hachul DT. Reprocessing of medical products in electrophysiology. Arq Bras Cardiol. 2017;108(2):169–72.PubMedPubMedCentral
13.
go back to reference Bracklo M, Marczak U. Reprocessing cardiology devices - reviewing 15 years of experience in Germany. ECR. 2012;8(2):82–3. Bracklo M, Marczak U. Reprocessing cardiology devices - reviewing 15 years of experience in Germany. ECR. 2012;8(2):82–3.
Metadata
Title
Initial experience with a novel re-sterilisable decapolar electrophysiology catheter
Authors
Anna L. Wilson
Kellie A. Timmins
Rosemary F. Allen
Paula M. Jones
Sacha L. Ware
Dean Boddington
Janice Swampillai
Spencer C. Heald
Martin K. Stiles
Publication date
01-08-2020
Publisher
Springer US
Published in
Journal of Interventional Cardiac Electrophysiology / Issue 2/2020
Print ISSN: 1383-875X
Electronic ISSN: 1572-8595
DOI
https://doi.org/10.1007/s10840-019-00583-2

Other articles of this Issue 2/2020

Journal of Interventional Cardiac Electrophysiology 2/2020 Go to the issue