Randomized controlled trial of Amigo® robotically controlled versus manually controlled ablation of the cavo-tricuspid isthmus using a contact force ablation catheter

Authors: Kurt S. Hoffmayer, Felix Krainski, Sanjay Shah, Jessica Hunter, Maylene Alegre, Jonathan C. Hsu, Gregory K. Feld

Published in: Journal of Interventional Cardiac Electrophysiology | Issue 2/2018

Login to get access

Abstract

Background

Radiofrequency catheter ablation (RFCA) of the cavo-tricuspid isthmus (CTI) is a common treatment for atrial flutter (AFL). However, achieving bi-directional CTI conduction block may be difficult, partly due to catheter instability.

Objective

To evaluate the safety and efficacy of the Amigo® Remote Catheter System (RCS) compared to manual catheter manipulation, during CTI ablation for AFL.

Methods

Fifty patients (pts) were prospectively randomized to robotically (25 pts) versus manually (25 pts) controlled catheter manipulation during CTI ablation, using a force-contact sensing, irrigated ablation catheter. The primary outcome was recurrence of CTI conduction after a 30-min waiting period. Secondary outcomes included total ablation, procedure, and fluoroscopy times, contact force measurement, and catheter stability.

Results

Recurrence of CTI conduction 30 min after ablation was less with robotically (0/25) versus manually (6/25) controlled ablation (p = 0.023). Total ablation and procedure times to achieve persistent CTI block (6.7 ± 3 vs. 7.4 ± 2.5 min and 14.9 ± 7.5 vs. 15.2 ± 7 min, respectively) were not significantly different (p = 0.35 and p = 0.91, respectively). There was a non-significant trend toward a greater force time integral (FTI in gm/s) with robotically versus manually controlled CTI ablation (571 ± 278 vs. 471 ± 179, p = 0.13). Fluoroscopy time was longer with robotically versus manually controlled CTI ablation (6.8 ± 4.4 min vs. 3.8 ± 2.3 min, p = 0.0027). There were no complications in either group.

Conclusion

Robotically controlled CTI ablation resulted in fewer acute recurrences of CTI conduction compared to manually controlled CTI ablation, and a trend toward higher FTI. The longer fluoroscopy time during robotically controlled ablation was likely due to a steep learning curve.

Trial registration

Clinicaltrials.​gov Identifier: NCT02467179
Literature
6.
go back to reference Lee KW, Yang Y, Scheinman MM, University of California-San Francisco SF, CA, USA. Atrial flutter: a review of its history, mechanisms, clinical features, and current therapy. Curr Probl Cardiol. 2005;30:121–67.CrossRefPubMed Lee KW, Yang Y, Scheinman MM, University of California-San Francisco SF, CA, USA. Atrial flutter: a review of its history, mechanisms, clinical features, and current therapy. Curr Probl Cardiol. 2005;30:121–67.CrossRefPubMed
7.
go back to reference Saoudi N, Cosio F, Waldo A, et al. Classification of atrial flutter and regular atrial tachycardia according to electrophysiologic mechanism and anatomic bases: a statement from a joint expert group from the Working Group of Arrhythmias of the European Society of Cardiology and the North American Society of Pacing and Electrophysiology. J Cardiovasc Electrophysiol. 2001;12:852–66.CrossRefPubMed Saoudi N, Cosio F, Waldo A, et al. Classification of atrial flutter and regular atrial tachycardia according to electrophysiologic mechanism and anatomic bases: a statement from a joint expert group from the Working Group of Arrhythmias of the European Society of Cardiology and the North American Society of Pacing and Electrophysiology. J Cardiovasc Electrophysiol. 2001;12:852–66.CrossRefPubMed
8.
go back to reference Saoudi N, Nair M, Abdelazziz A, et al. Electrocardiographic patterns and results of radiofrequency catheter ablation of clockwise type I atrial flutter. J Cardiovasc Electrophysiol. 1996;7:931–42.CrossRefPubMed Saoudi N, Nair M, Abdelazziz A, et al. Electrocardiographic patterns and results of radiofrequency catheter ablation of clockwise type I atrial flutter. J Cardiovasc Electrophysiol. 1996;7:931–42.CrossRefPubMed
11.
go back to reference Scheinman MM and Yang Y. Atrial flutter: historical notes—part 1. PACE. 2004:1–3. Scheinman MM and Yang Y. Atrial flutter: historical notes—part 1. PACE. 2004:1–3.
14.
go back to reference Bazan V, Marti-Almor J, Perez-Rodon J, et al. Incremental pacing for the diagnosis of complete cavotricuspid isthmus block during radiofrequency ablation of atrial flutter. J Cardiovasc Electrophysiol. 2010;21:33–9.CrossRefPubMed Bazan V, Marti-Almor J, Perez-Rodon J, et al. Incremental pacing for the diagnosis of complete cavotricuspid isthmus block during radiofrequency ablation of atrial flutter. J Cardiovasc Electrophysiol. 2010;21:33–9.CrossRefPubMed
18.
go back to reference Oral H, Sticherling C, Tada H, et al. Role of transisthmus conduction intervals in predicting bidirectional block after ablation of typical atrial flutter. J Cardiovasc Electrophysiol. 2001;12:169–74.CrossRefPubMed Oral H, Sticherling C, Tada H, et al. Role of transisthmus conduction intervals in predicting bidirectional block after ablation of typical atrial flutter. J Cardiovasc Electrophysiol. 2001;12:169–74.CrossRefPubMed
21.
go back to reference Tada H, Oral H, Sticherling C, et al. Electrogram polarity and cavotricuspid isthmus block during ablation of typical atrial flutter. J Cardiovasc Electrophysiol. 2001;12:393–9.CrossRefPubMed Tada H, Oral H, Sticherling C, et al. Electrogram polarity and cavotricuspid isthmus block during ablation of typical atrial flutter. J Cardiovasc Electrophysiol. 2001;12:393–9.CrossRefPubMed
22.
go back to reference Tada H, Oral H, Sticherling C, et al. Double potentials along the ablation line as a guide to radiofrequency ablation of typical atrial flutter. J Am Coll Cardiol. 2001;38:750–5.CrossRefPubMed Tada H, Oral H, Sticherling C, et al. Double potentials along the ablation line as a guide to radiofrequency ablation of typical atrial flutter. J Am Coll Cardiol. 2001;38:750–5.CrossRefPubMed
27.
go back to reference Montenero AS, Bruno N, Antonelli A, et al. Long-term efficacy of cryo catheter ablation for the treatment of atrial flutter: results from a repeat electrophysiologic study. Am Coll Cardiol. 2005;45:573–80.CrossRef Montenero AS, Bruno N, Antonelli A, et al. Long-term efficacy of cryo catheter ablation for the treatment of atrial flutter: results from a repeat electrophysiologic study. Am Coll Cardiol. 2005;45:573–80.CrossRef
Metadata
Title
Randomized controlled trial of Amigo® robotically controlled versus manually controlled ablation of the cavo-tricuspid isthmus using a contact force ablation catheter
Authors
Kurt S. Hoffmayer
Felix Krainski
Sanjay Shah
Jessica Hunter
Maylene Alegre
Jonathan C. Hsu
Gregory K. Feld
Publication date
01-03-2018
Publisher
Springer US
Published in
Journal of Interventional Cardiac Electrophysiology / Issue 2/2018
Print ISSN: 1383-875X
Electronic ISSN: 1572-8595
DOI
https://doi.org/10.1007/s10840-018-0319-1

Other articles of this Issue 2/2018

Journal of Interventional Cardiac Electrophysiology 2/2018 Go to the issue